PHILIPS XD-S DIRECT RADIOGRAPHY WORKSTATION/PACKAGE
K063781 · Philips Medical Systems North America Co. · LLZ · Jan 5, 2007 · Radiology
Device Facts
Record ID
K063781
Device Name
PHILIPS XD-S DIRECT RADIOGRAPHY WORKSTATION/PACKAGE
Applicant
Philips Medical Systems North America Co.
Product Code
LLZ · Radiology
Decision Date
Jan 5, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
As part of a radiographic system, the Philips XD-S is intended to acquire, process, store, display, and export digital radiographic images. The Philips XD-S is suitable for all routine radiography exams, including specialists areas like intensive care, trauma, or pediatric work, excluding mammography
Device Story
Philips XD-S is a digital radiography workstation comprising a computer, keyboard, display, and mouse, integrated with a flat solid-state X-ray detector. Used by clinicians in clinical settings to preset examination data, acquire, process, store, display, and export digital X-ray images. Operates as part of a larger radiographic system including X-ray generators, tubes, and tables. Supports optional features like image stitching, touch-screen interface, and connectivity to PCR image plate readers or printers. Image processing is fully reversible and non-compressed for storage. Provides healthcare providers with digital images for diagnostic review; facilitates clinical decision-making by enabling efficient image handling and system control integration.
Clinical Evidence
Bench testing only. Device complies with federal X-Ray performance standards (CFR 1020.30, .31), electrical safety standards (UL/IEC 60950-1, 60601-1), and electromagnetic compatibility standards (IEC-601-1-2, CISPR-11).
Technological Characteristics
Workstation includes computer, keyboard, display, mouse, and flat solid-state X-ray detector. Complies with DICOM standards. Software level of concern is Minor. Risk management per ISO 14971. Connectivity includes DICOM, direct printer, and PCR plate reader interfaces.
Indications for Use
Indicated for routine radiographic examinations, including intensive care, trauma, and pediatric applications. Excludes mammography. Prescription use only.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K063781
## 510(k) Summary of Safety and Effectiveness
JAN -5 2007
In accordance with the requirements of the Safe Medical Device Act, Philips Medical Systems North America Company herewith submits a Summary of Safety and Effectiveness.
| MANUFACTURER: | Philips Medical Systems DMC GmBH<br>Roentgenstrasse 24<br>Hamburg, Germany D-22335 |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration No.: | 3003768251 |
| SUBMITTER: | Philips Medical Systems<br>22100 Bothell Everett Highway<br>Bothell, WA 98021-8431 |
| Establishment Registration No.:<br>Contact: | 1217116<br>Lynn Harmer<br>425-487-7312 |
| DATE PREPARED: | 11 November 2006 |
| CLASSIFICATION NAME: | System, image processing, radiological, Class I<br>(LLZ)<br>Solid State X-Ray Imager (Plat Panel/Digital<br>Imager), Class II (MQB) |
| COMMON/USUAL NAME: | Digital image acquisition workstation |
| TRADE/PROPRIETARY NAME: | PHILIPS XD-S Direct Radiography<br>Workstation/Package |
#### PERFORMANCE STANDARDS:
This device complies with the federal X-Ray performance standards (CFR I 020.30, .31) as well as with relevant national and international standards for Electrical Safety (UL 60950-1, IEC 60950-1, UL 60601-1, IEC 60601-1) as well as international standards for Electromagnetic Compatibility (IEC-601-1-2, CISPR-11) and the ACR/NEMA DICOM digital imaging communication standard.
### SYSTEM DESCRIPTION:
The Philips XD-S is a workstation (computer, keyboard, display, mouse), combined with a flat solid state X-ray detector. It is used by the operator to preset examination data, and to generate, process and handle digital X-ray images.
{1}------------------------------------------------
As a part of a radiographic system, the Philips XD-S is intended to acquire, process, store, display, and export digital radiographic images. The Philips XD-S is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding mammography.
The complete X-ray system would further include other subsystems and components, like patient table, X-ray control(s), X-ray high voltage generator, X-ray tube(s), collimator(s), accessories, etc.
There is a standalone version with minimal integration into the X-ray system. With the fully integrated version, the workstation screen also provides displays area and controls for X-ray generator control. The workstation computer can also host parts of the system control software.
Available options are:
- touch-screen monitor .
- . image stitching
- . PCR image plate reader connection
- second flat detector .
- . X-ray generator user interface integration
- direct printer connection .
#### INTENDED USE:
As part of a radiographic system, the Philips XD-S is intended to acquire, process, store, display, and export digital radiographic images. The Philips XD-S is suitable for all routine radiography exams, including specialists areas like intensive care, trauma, or pediatric work, excluding mammography
#### EQUIVALENCE INFORMATION:
The PHILIPS XD-S Direct Radiography Workstation/Package is considered substantially equivalent to the Philips Digital Diagnost, which received FDA marketing on November 25, 1998, under the name Philips Bucky Vision in 510(k) Number K982795.
In relation to the image plate reader connection, the PHILIPS XD-S is substantially equivalent to the Philips Computed Radiography, which received FDA marketing on December 18, 1996, under 510(k) Number K964124.
{2}------------------------------------------------
### SAFETY INFORMATION:
The Philips XD-S Direct Radiography Workstation/Package uses mature technology. It is designed to be in compliance with National and International safety standards well as the DICOM communication standard.
lmage data are not compressed for storage and the applied image processing is fully reversible.
The software used in the Wokstation is equivalent to the software used in the predicate devices. The Level of Software concern is MINOR as determined according to the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May/11/2005.
A product risk mangement is cxecuted according to ISO 14971 and all risks are reduced to an acceptable level by implementation and verification of appropriate measures.
Philips Medical Systems North America Company feels that sufficient information and data are contained in this submission to enable CDRH to reach a determination of substantial equivalence,
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Philips Medical Systems North America Company % Mr. Marc M. Mouser Program Reviewer Underwriters Laboratories, Inc. 2600 NW Lake Road CAMAS WA 98607
# JAN - 5 2007
Re: K063781
Trade/Device Name: PHILIPS XD-S Direct Radiography Workstation/Package Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 20, 2006 Received: December 21, 2006
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the udicec use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act, The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulterstion.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/9 description: The image shows a circular logo with the text "1906-2006" at the top. The letters "FDA" are prominently displayed in the center of the logo. Below the letters, the word "Centennial" is written in a cursive font. Three stars are arranged in a horizontal line beneath the word "Centennial".
Protecting and Promoting Public Health
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Kuc 378i
# Indications for Use
Kor 3781 510(k) Number (if known):
Device Name: PHILIPS XD-S Direct Radiography Workstation/Package
Indications For Use:
As a part of a radiographic system, the Philips XD-S is intended to acquire, process, store, display, and export digital radiographic images. The Philips XD-S is suitable for all routine radiography exams, including specialist areas like intensive care, trauma, or pediatric work, excluding mammography.
Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brogdon
(Division Sign-Off Division of Reproductive, Abdo and Radiological Devices 510(k) Number __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.