IMPELLA RECOVER LP 2.5 PERCUTANEOUS CARDIAC SUPPORT SYSTEM
K063723 · Abiomed, Inc. · KFM · May 30, 2008 · Cardiovascular
Device Facts
Record ID
K063723
Device Name
IMPELLA RECOVER LP 2.5 PERCUTANEOUS CARDIAC SUPPORT SYSTEM
Applicant
Abiomed, Inc.
Product Code
KFM · Cardiovascular
Decision Date
May 30, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IMPELLA RECOVER® LP 2.5 Percutaneous Cardiac Support System is intended for partial circulatory support using an extracorporeal bypass control unit, for periods up to 6 hours. It is also intended to be used to provide partial circulatory support (for periods up to 6 hours) during procedures not requiring cardiopulmonary bypass. The IMPELLA RECOVER® LP 2.5 Percutaneous Cardiac Support System also provides pressure measurements which are useful in determining intravascular pressure.
Device Story
System provides partial circulatory support and intravascular pressure monitoring. Components include a catheter with integrated pump motor, infusate lumen, pressure lumen, and cannula; an extracorporeal controller/console; and an infusion system. Used in clinical settings for procedures requiring temporary circulatory assistance. Physician operates system to deliver support and monitor pressure. Output provides hemodynamic assistance and pressure data to aid clinical decision-making during short-term procedures. Benefits include temporary circulatory stabilization.
Clinical Evidence
Clinical safety analysis provided based on data from 129 patients (109 OUS, 20 US). Bench testing included sterilization validation (EN 550, ISO 10993-7), biocompatibility (ISO 10993), electrical safety/EMC (EN 60601, EN 61000), and in vitro performance testing.
Technological Characteristics
System includes catheter with integrated pump, controller, and infusion system. Sterilized via EtO gas (SAL 10^-6) per EN 550. Patient-contacting materials validated per ISO 10993. Electrical safety/EMC compliance per EN 60601 and EN 61000.
Indications for Use
Indicated for patients requiring partial circulatory support for up to 6 hours, including during procedures not requiring cardiopulmonary bypass, and for intravascular pressure monitoring.
Regulatory Classification
Identification
A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability;
(ii) The patient-contacting components of the device must be demonstrated to be biocompatible;
(iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and
(iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(b)
*Nonroller-type temporary ventricular support blood pump* —(1)*Identification.* A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2)
*Classification.* Class III (premarket approval).(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper half of the circle, with the words separated by dots.
MAY 19 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Abiomed, Inc. c/o Mr. Robert T. Kung 22 Cherry Hill Drive Danvers, MA 01923
Re: K063723 Impella® Recover® LP 2.5 Percutaneous Cardiac Suppport System Regulation Number: 21 CFR 870.4360 Regulation Name: Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type Catheter, Cannula and Tubing Vascular, Cardiopulmonary Bypass Catheter, Intravascular, Diagnostic Regulatory Class: Class III (three) Product Code: KFM, DWF, DQO Dated: May 15, 2008 Received: May 19, 2008
Dear Mr. Kung:
This letter corrects our substantially equivalent letter of May 30, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 -- Mr. Robert T. Kung
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
R. Vahner
A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosur
{2}------------------------------------------------
ABIOMED, Inc.
IMPELLA RECOVER® LP 2.5 Perculaneous Cardiac Support System Traditional 510(k)
## Indications for Use
510{k) Number (if known): K063723
Device Name: IMPELLA RECOVER® LP 2.5 Percutaneous Cardiac Support System
#### Indications for Use:
The IMPELLA RECOVER® LP 2.5 Percutaneous Cardiac Support System is intended for partial circulatory support using an extracorporeal bypass control unit, for periods up to 6 hours. It is also intended to be used to provide partial circulatory support (for periods up to 6 hours) during procedures not requiring cardiopulmonary bypass.
The IMPELLA RECOVER® LP 2.5 Percutaneous Cardiac Support System also provides pressure measurements which are useful in determining intravascular pressure.
Prescription Use × (Part 21 CFR 801-Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Carmovascular Devices
510(k) Number_Ko63723
Page 1 of 1
{3}------------------------------------------------
K063723
MAY 30 2008
# 510(k) Summary (Prepared in accordance with 21 CFR Part 807.92)
### a. Submitted
| Applicant Name: | ABIOMED, Inc. |
|---------------------------|--------------------------------------------------------------------------------------------------|
| | 22 Cherry Hill Drive, Danvers, MA 01923 |
| Contact Person: | William J. Bolt |
| Date Summary<br>Prepared: | May 15, 2008 |
| b. Device Information | |
| Trade Name: | IMPELLA RECOVER® LP 2.5 Percutaneous<br>Cardiac Support System |
| Common Name: | Percutaneous Cardiac Support System |
| Classification<br>Name: | Pump, Blood, Cardiopulmonary Bypass, non rolle<br>type (classified under 870.4360) |
| | Catheter, Cannula and Tubing, Vascular,<br>Cardiopulmonary Bypass (classified under<br>870.4210) |
| | Diagnostic intravascular catheter (classified under<br>870.1200) |
| Product Code: | 74KFM, 74DWF, 74DQO |
#### c. Legally Marketed Predicate Devices
- TandemHeart PTVA System K991783, K052570 K924642 , K924643 .
- Vascular Solutions Langston Dual Lumen Catheter K050168 .
### d. Device Description:
The IMPELLA RECOVER® LP 2.5 Percutaneous Cardiac Support System (IMPELLA RECOVER LP 2.5 System) provides circulatory support with the ability to deliver anticoagulant through an infusion system. The System is comprised of: 1) a catheter which contains an integrated pump motor/infusate lumen, integrated intravascular pressure lumen and integral cannula, 2) a controller/console and 3) infusion system designed to work together, and 4) accessories.
# e. Intended Use:
The IMPELLA RECOVER® LP 2.5 Percutaneous Cardiac Support System is intended for partial circulatory support using an extracorporeal bypass control unit, for periods up to 6 hours. It is also intended to be used to provide partial circulatory support (for periods up to 6 hours) during procedures not requiring cardiopulmonary bypass.
The IMPELLA RECOVER® LP 2.5 Percutaneous Cardiac Support System also provides pressure measurements which are useful in determining intravascular pressure.
{4}------------------------------------------------
# f. Technological Characteristics and Comparison to Predicate Device(s):
The technological characteristics of the IMPELLA RECOVER® LP 2,5 are the same as the TANDEM HEART systems and the Vascular solutions Langston catheter with the exception of the following differences:
- · pump location
- certain materials of construction .
- . pump speed
#### g. Test Results:
Pre - Clinical:
To validate the device design of the IMPELLA RECOVER® Percutaneous Cardiac Support System, ABIOMED performed the following in vitro testing:
With regard to sterilization, packaging, and shelf-life, the IMPELLA RECOVER® LP 2.5 Percutaneous Cardiac Support System is sterilized using EtO gas with a SAL of 10-6. The sterilization method/cycle was validated using EN 550 "Sterilization of Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization." The EtO sterilization residual values for EO and ECH and are within the allowable limits of ISO 10993-7. The LAL test was used to ensure a pyrogen free determination. The packaging material has been validated to ensure its integrity.
Biocompatibility testing of all patient contacting materials was conducted on the finished sterilized devices in accordance with ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". All testing results are acceptable.
The design and testing validation of the software contained in the IMPELLA MCS was conducted in compliance with the FDA 2005 guidance document entitled "Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
With regard to electromagnetic compatibility and electrical safety, the IMPELLA RECOVER® LP 2.5 Percutaneous Cardiac Support System was tested in accordance with EN 60601 and EN 61000 and their applicable subparts. The testing results demonstrate that the device is in conformance with these FDA recognized standards.
With regard to In vitro performance testing, ABIOMED conducted a full range of testing demonstrating that the entire IMPELLA RECOVER® LP 2.5 System operates as intended. All tests were acceptable.
#### Clinical :
Abiomed provided a detailed analysis based on a clinical data collected from a combination of 109 OUS and 20 US patients used to address patient safety.
{5}------------------------------------------------
# h. Conclusion:
。
The ABIOMED IMPELLA RECOVER® LP 2.5 Percutaneous Cardiac Support System is substantially equivalent to the TandemHeart PTVA System and Control system (K991783), the TandemHeart Transseptal Cannula (K052570), the Medtronic Biomedicus 15F arterial cannula and introducer (K924642, K924643) and the Vascular Solutions Langston Dual Lumen Catheter – K050168
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.