← Product Code [MMK](/productcode/MMK) · K063722

# SEKURIT SHARPS COLLECTION CONTAINER (K063722)

_Techen Safety, Inc. · MMK · Mar 5, 2007 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K063722

## Device Facts

- **Applicant:** Techen Safety, Inc.
- **Product Code:** [MMK](/productcode/MMK.md)
- **Decision Date:** Mar 5, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

Techen Safety Sharps Disposal Containers with Screw Top Caps are intended for single use disposal of used or contaminated medical sharps, including but not limited to, hypodermic needles, syringes, lancets, and Blood Needles. The containers can be used in a variety of healthcare settings. They are suitable for physicians offices, dental offices, laboratories, home health, patient room, and other generation of medical waste.

## Device Story

Sekurit Sharps Collection Containers are single-use, puncture-resistant vessels for disposal of contaminated medical sharps; hypodermic needles, syringes, lancets, blood needles. Containers are blow-molded high-density polyethylene (HDPE) with puncture-resistant polypropylene screw caps. Available in 1.4 quart, 1 gallon, 2 gallon, and 3 gallon capacities; translucent or opaque red/yellow colors. Used in clinical, laboratory, and home health settings for safe containment of medical waste. Prevents accidental needle sticks and exposure to contaminated sharps.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Materials: High Density Polyethylene (HDPE) container, Polypropylene (PP) screw cap. Design: Blow-molded, single-piece, threaded screw neck, puncture-resistant. Capacities: 1.4 quart, 1, 2, and 3 gallons. Colors: Translucent/opaque red and yellow. Non-reusable.

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- B-D Guardian One Piece Sharps Collectors ([K943139](/device/K943139.md))
- Sage Products, Sharps Disposal Containers with Screw Top Caps ([K980490](/device/K980490.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Exhibit C

K063722

# 510(k) Summary

| Name and address of sponsor of the<br>510(k) submission:   | Techen Safety, Inc.<br>1901 Powis Ct. West Chicago, IL 60185                                                                         |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Official contact person for all<br>correspondence:         | John Allegretti, (address as above)                                                                                                  |
|                                                            | Phone: 800-832-4361, Fax: 630-797-7305                                                                                               |
|                                                            | E-mail: cja@techensafety.com                                                                                                         |
| Date Prepared:                                             | November 28, 2006                                                                                                                    |
| Device Name:                                               | Sekurit Sharps Collection Containers                                                                                                 |
| Generic name of the device:                                | Sharps Container                                                                                                                     |
| Classification, Product Code and<br>CFR Regulation Number: | Class II, MMK and 21 CFR 880.5570                                                                                                    |
| Classification Panel:                                      | General Hospital                                                                                                                     |
| Predicate Device Name and 510(k)<br>Number :               | 1)B-D@ Guardian One Piece Sharps Collectors, K943139<br>2) Sage Products, Sharps Disposal Containers with Screw<br>Top Caps, K980490 |

### Device Description:

The Sekurit Sharps Collection Container is a single piece container which is blow molded of High Density Polyethylene plastic and is available in 1 Gallon, 3 Gallon, 3 Gallon and 1.4 quart sizes. Each container is of the same design type and altered only in its height to achieve the needed capacities. The containers are available in translucent red and yellow colors as well as . opaque red and yellow colors. Sekurit Sharps Collection Containers have a threaded screw neck and utilize a puncture resistant screw cap molded from Polypropylene. These containers are NOT reusable.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Techen Safety, Incorporated C/O Mr. Neil E. Devine Responsible Third Party Official Intertek Testing Services 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087

MAR 0 5 2007

Re: K063722

Trade/Device Name: 1GL01-001 -1 Gallon Red Transluscent Sekurit Sharps Collection Container; 1GL01-002-1 Gallon Red Opaque Sekurit Sharps Collection Container; 1GL01-003-1 Gallon Yellow Transluscent Sekurit Sharps Collection Container; 1GL01-004-1 Gallon Yellow Opaque Sekurit Sharps Collection Container; 2GL01-001-2 Gallon Red Transluscent Sekurit Sharps Collection Container; 2GL01-002-2 Gallon Red Opaque Sekurit Sharps Collection Container; 2GL01-003-2 Gallon Yellow Transluscent Sekurit Sharps Collection Container; 2GL01-004-2 Gallon Yellow Opaque Sekurit Sharps Collection Container; 3GL01-001-3 Gallon Red Transluscent Sekurit Sharps Collection Container; 3GL01-002-3 Gallon Red Opaque Sekurit Sharps Collection Container; 3GL01-003-3 Gallon Yellow Transluscent Sekurit Sharps Collection Container; 3GL01-004-3 Gallon Yellow Opaque Sekurit Sharps Collection Container; 1QT01-001-1.4 Quart Red Transluscent Sekurit Sharps Collection Container; 1QT01-002-1.4 Quart Red Opaque Sekurit Sharps Collection Container; 1QT01-003-1.4 Ouart Yellow Transluscent Sekurit Sharps Collection Container; 1QT01-004-1.4 Quart Yellow Opaque Sekurit Sharps Collection Container Regulation Number: 880.5570

Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: MMK Dated: February 16, 2007 Received: February 20, 2007

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general

{2}------------------------------------------------

Page 2 – Mr. Devine

controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Snyde y. Michael Davis.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## INDICATIONS FOR USE

Page 1

#### 510(k) Number (if known): Not Assigned as of this time

#### Device Name: Sekurit Sharps Collection Container

Model numbers: 1GL01-001 - 1 Gallon Red Transluscent Sekurit Sharps Collection Container 1GL01-002 - 1 Gallon Red Opaque Sekurit Sharps Collection Container 1GL01-003 - 1 Gallon Yellow Transluscent Sekurit Sharps Collection Container 1GL01-004 - 1 Gallon Yellow Opaque Sekurit Sharps Collection Container 2GL01-001 - 2 Gallon Red Transluscent Sekurit Sharps Collection Container 2GL01-002 - 2 Gallon Red Opaque Sekurit Sharps Collection Container 2GL01-003 - 2 Gallon Yellow Transluscent Sekurit Sharps Collection Container 2GL01-004 - 2 Gallon Yellow Opaque Sekurit Sharps Collection Container 3GL01-001 - 3 Gallon Red Transluscent Sekurit Sharps Collection Container 3GL01-002 - 3 Gallon Red Opaque Sekurit Sharps Collection Container 3GL01-003 - 3 Gallon Yellow Transluscent Sekurit Sharps Collection Container 3GL01-004 - 3 Gallon Yellow Opaque Sekurit Sharps Collection Container 1QT01-001 - 1.4 Quart Red Transluscent Sekurit Sharps Collection Container 1QT01-002 - 1.4 Quart Red Opaque Sekurit Sharps Collection Container 1QT01-003 - 1.4 Quart Yellow Transluscent Sekurit Sharps Collection Container 1QT01-004 - 1.4 Quart Yellow Opaque Sekurit Sharps Collection Container

#### Indications for Use:

Techen Safety Sharps Disposal Containers with Screw Top Caps are intended for single use disposal of used or contaminated medical sharps, including but not limited to, hypodermic needles, syringes, lancets, and Blood Needles. The containers can be used in a variety of healthcare settings. They are suitable for physicians offices, dental offices, laboratories, home health, patient room, and other generation of medical waste.

Prescription Use (Per 21 CFR 801 Subpart D) Over-The Counter Use X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices

K063722

---

**Source:** [https://fda.innolitics.com/device/K063722](https://fda.innolitics.com/device/K063722)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
