← Product Code [HKI](/productcode/HKI) · K063717

# CF-1 (K063717)

_Canon, Inc. · HKI · Jan 23, 2007 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K063717

## Device Facts

- **Applicant:** Canon, Inc.
- **Product Code:** [HKI](/productcode/HKI.md)
- **Decision Date:** Jan 23, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1120
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

CF-1 is intended to be used for taking digital images of retina of human eye with a mydriatic.

## Device Story

The Canon CF-1 is an ophthalmic camera designed to capture digital images of the human retina. It functions as an improved model of the CF-60DSi. The device requires the use of a mydriatic agent to dilate the patient's pupil. The operator, typically an ophthalmologist, aligns the camera using split lines and a 45mm working distance to capture images in COLOR, RED FREE, or FLOU modes. The system utilizes a mounted Canon EOS Digital Camera to record the retinal images, which are then available for immediate viewing, facilitating clinical applications such as telemedicine and electronic filing. The device provides variable power settings (50/43 degrees) and includes an external eye fixation lamp. By providing high-quality digital retinal documentation, the device assists clinicians in diagnosing and monitoring ocular conditions, improving procedural efficiency compared to traditional film-based or older digital systems.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and comparison of technical specifications, optical design, and physical safety compliance (UL60601-1).

## Technological Characteristics

Ophthalmic camera; 50/43 degree field of view; 45mm working distance; 300WS max photography light source; EOS Digital Camera image unit; manual/automatic filter set; UL60601-1 compliant; 350VA power consumption; 27kg main unit weight.

## Regulatory Identification

An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

## Predicate Devices

- Canon CF-60DSi ([K041546](/device/K041546.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Appendix G: 510(k) Summary 510(k) Summary

October 11, 200

# Prepared:

October 11, 2006

### Submitter:

Company Name: Company Address:

Contact Person: Phone Number Fax Number:

#### Proposed Device:

Reason For 510(k): Manufacturer: Trade Name: Model Name: Classification Name: Regulatory Class: FDA 510(k) #:

Canon USA, Inc. (U.S. agent for Canon Inc.) One Canon Plaza Lake Success, NY 11042 Ms. Sheila Driscoll (516) 328-5602 (516) 328-5169

- New Model Canon Inc. Canon CF-1 86HKI, Ophthalmic cameras Class II 063717 To be assigned
Predicate Device:

Manufacturer: Trade Name: Model Name: Classification Name Regulatory Class: FDA 510(k) #

Canon Inc. Canon CF-60DS! 86HKI, Ophthalmic cameras Class II K041546

Intended Use:

CF-1 is intended to be used for taking digital images of retina of human eye with a mydriatic.

Description Of Device: CF 1 is an improved model of CF-60DSi. Canon EOS Digital Camera is mounted with CF-1, can be viewed immediately, making procedures more efficient vaniera hand different applications, such as telemedicine and electronic filing.

CF-1's intended use is the same as that of CF-60DSi. However, the differences in design are as follows:

- CF-1 can take images in three modes, such as COLOR, RED FREE, and FLOU .
- CF-60DSi can take images in four modes, such as COLOR, RED FREE, FLOU and ICG. t
- ICG digital camera cannot be attached to CF-1. .
- CF-1 has 2 variable powers (50//43 degree), where 43 degree is digital magnification. . CF-60DSi has 2 optical variable powers (60//40 degree).

CF-1 is equivalent to CF-60DSi in the following respect:

- The optical components and alignment and the mechanical structures of the CF-1 arc . almost same as the CF 60DSi. Picase refer to the CF 1 comparison table provided in this section.
JAN 2 3 2007

{1}------------------------------------------------

## Appendix G: 510(k) Summary

・・ ".

. .

:

.

.

ﺳ

### Table of Comparison

|                                                     |                                                        |                                                                                           | CF-60DSi                                                                                                                          | CF-1                                                             |
|-----------------------------------------------------|--------------------------------------------------------|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| FDA 510K #                                          |                                                        |                                                                                           | K041546                                                                                                                           | Proposed Device                                                  |
| P<br>R<br>O<br>F<br>O<br>R<br>M<br>A<br>N<br>C<br>E | Angular field of view                                  |                                                                                           | 60/40 degree                                                                                                                      | 50/43 degree (digital magnification)                             |
|                                                     | Actual image size                                      |                                                                                           |                                                                                                                                   | φ29×22 mm (on sensor array)<br>φ7.5×5.7 (Type 1:on sensor array) |
|                                                     | Min. diameter of pupil<br>required                     |                                                                                           |                                                                                                                                   | φ5.5mm                                                           |
|                                                     | Working distance<br>(WD)                               |                                                                                           |                                                                                                                                   | 45mm                                                             |
|                                                     | Focusing                                               |                                                                                           |                                                                                                                                   | By aligning the split lines                                      |
|                                                     | Eye fixation lamp                                      |                                                                                           |                                                                                                                                   | External                                                         |
|                                                     | Filter set                                             |                                                                                           |                                                                                                                                   | Automatic/Manual                                                 |
|                                                     | Light source for<br>photography                        |                                                                                           |                                                                                                                                   | Max. 300WS                                                       |
|                                                     | Image unit                                             |                                                                                           |                                                                                                                                   | EOS Digital Camera<br>ICG digital camera (with Adapter)          |
|                                                     |                                                        | Working range<br>Vertical<br>Forward and back<br>Right and left<br>Chin rest (vertically) |                                                                                                                                   | 38mm<br>70mm<br>120mm<br>65mm                                    |
|                                                     | Panning                                                |                                                                                           |                                                                                                                                   | 30 degree, right or left                                         |
|                                                     | External dimensions<br>Main unit<br>Power control unit |                                                                                           |                                                                                                                                   | W320×D560×H565mm<br>W225×D380×H485mm                             |
|                                                     | Weight                                                 | Main unit<br>Power control unit                                                           |                                                                                                                                   | Approx.27kg<br>Approx.27kg                                       |
| Intended use                                        |                                                        |                                                                                           | Taking digital images of retina of<br>human eye with a mydriatic                                                                  | Same                                                             |
| Energy                                              |                                                        | used                                                                                      | 1000VA                                                                                                                            | 350VA                                                            |
| delivered                                           |                                                        |                                                                                           | Not Applicable                                                                                                                    | Same                                                             |
| Target population                                   |                                                        |                                                                                           | Ophthalmologist                                                                                                                   | Same                                                             |
| Physical safety                                     |                                                        |                                                                                           | UL60601-1                                                                                                                         | Same                                                             |
| Compliance with standards                           |                                                        |                                                                                           | UL60601-1                                                                                                                         | Same                                                             |
| Biocompatibility                                    |                                                        |                                                                                           | Not Applicable                                                                                                                    | Same                                                             |
| Labeling                                            |                                                        |                                                                                           | Printed model name and electrical ratings are changed.<br>"Rx Only" added<br>to and EMC reference removed from CF-1 rating label. |                                                                  |

-

:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle. The logo is simple, using a single color, and is designed to be easily recognizable.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 3 2007

Canon USA, Inc. c/o Jeffrey D. Rongero Underwriters Laboratories, Inc. 1285 Walt Whitman Rd. Melville, NY 11747

Re: K063717

Trade/Device Name: Canon Digital Retinal Camera CF-1 Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: January 9, 2007 Received: January 10, 2007

Dear Mr. Rongero:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

M.B. Egelman Si MD

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications Statement

063717

510(K)Number(if known): -

Device Name: CF-1

Indications for Use:

The device is intended to be used for taking digital images of retina of human eye with a mydriatic.

. .

Prescription Use (Part 21 CFR 801 Subpart D) OR

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHERT PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation(ODE)

Dexider

(Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises

t Devises

1/17/2007

Page 1

of

510(k) Number ________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/device/K063717](https://fda.innolitics.com/device/K063717)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
