AKREOS SINGLE USE INSERTION DEVICE, MODEL AI-28

K063694 · Bausch & Lomb, Inc. · KYB · Mar 7, 2007 · Ophthalmic

Device Facts

Record IDK063694
Device NameAKREOS SINGLE USE INSERTION DEVICE, MODEL AI-28
ApplicantBausch & Lomb, Inc.
Product CodeKYB · Ophthalmic
Decision DateMar 7, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 886.4300
Device ClassClass 1

Indications for Use

The intended use for the Akreos Single Use Insertion Device, Model AI-28 is to fold and deliver a Bausch & Lomb IOL into the eye during normal small incision cataract surgery. The Akreos Single Use Insertion Device should only be used to fold and deliver those IOL models that allow use of this injector in their approved labeling.

Device Story

Sterile, single-use, disposable plastic insertion device; consists of syringe-shaped tube with lens loading deck, lens cap, and plunger; includes transition cell. Used by surgeons during cataract surgery to fold and deliver Akreos Advanced Optics Aspheric IOLs into the eye via a small tubular pathway using one continuous forward motion. Device ensures proper lens folding and recovery; facilitates small incision implantation. Benefits include controlled, consistent lens delivery. Operates mechanically; no electronic or software components.

Clinical Evidence

Bench testing only. Evidence includes fold and recovery testing of the Akreos Aspheric Optic IOL, stability testing of plunger material (tensile and flexural properties), and biocompatibility testing of polypropylene and polyphthalamide materials.

Technological Characteristics

Materials: Polypropylene and polyphthalamide. Principle: Mechanical folding and delivery via tubular pathway. Energy: Manual. Form factor: Syringe-shaped tube with plunger and transition cell. Sterilization: Ethylene Oxide (EtO).

Indications for Use

Indicated for folding and delivery of specific Bausch & Lomb intraocular lens (IOL) models into the eye during small incision cataract surgery.

Regulatory Classification

Identification

An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K063694 # Bausch & Lomb ... #### 510(k) Summary MAR 0 7 2007 ## Akreos Single Use Insertion Device, Model AI-28 | Submitter<br>Information | Owner: | Bausch & Lomb, Incorporated<br>180 Via Verde<br>San Dimas, CA 92886 | |--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------| | | Telephone:<br>Fax: | 909 971 5394<br>909 971 5148 | | | Contact: | Nanette Canepa<br>Associate Manager, Global Regulatory Affairs | | Date Prepared | December 12, 2006 | | | Subject device<br>name | Trade name: Akreos Single Use Insertion Device, Model AI-28<br>Common name: Lens insertion guide<br>Classification name: Intraocular lens guide (21CFR 886.4300,<br>Product Code KYB) | | | Predicate<br>device | Passport foldable lens placement system | | Continued on next page : {2}------------------------------------------------ | Subject device description | Akreos Single Use Insertion Device, Model AI-28, previously<br>known as Hydroport SI, is validated for folding and delivering the<br>Bausch & Lomb one-piece Akreos Advanced Optics Aspheric<br>Intraocular Lens (IOL) into the eye during cataract surgery. Future<br>IOL models may also by qualified with the Akreos Single Use<br>Insertion Device following fold and recovery validation studies. | |-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The device provides a small tubular pathway in which the Akreos<br>Advanced Optics Aspheric Lens can be placed into the eye with one<br>continuous forward motion. The Akreos Single Use Insertion Device<br>consists of two parts: a syringe shaped tube (includes lens loading<br>deck and lens cap) with a plunger, and a transition cell. | | | The Akreos Single Use Insertion Device is a sterile (EtO),<br>disposable plastic device, designed for single use only. | | Technological characteristics | The similarities between the Akreos Single Use Insertion Device,<br>Model AI-28 and the Passport foldable lens placement system are<br>the overall design, components, intended use, operating principle,<br>folding direction of the lens, cartridge design, sterilization methods<br>and component materials with the exception of the plunger material. | | | The change in plunger material of the subject device has not altered<br>the fundamental scientific technology of the predicate device. | | Intended use | The intended use for the Akreos Single Use Insertion Device, Model<br>AI-28 is to fold and deliver a Bausch & Lomb IOL into the eye<br>during normal small incision cataract surgery. | Continued on next page {3}------------------------------------------------ Labeling The labeling for the Akreos Single Use Insertion Device, Model AI-28 and the Passport foldable lens placement system differ in the type of intraocular lens specified for use. These differences in the labeling that specify use, do not affect safety and effectiveness of the device when used as labeled because each insertion device is specifically made to fit the size and shape of the indicated intraocular lens. Substantial Substantial equivalence is based on non-clinical performance data Equivalence that show that the subject device met the acceptance criteria for fold discussion and recovery testing of the Akreos Aspheric Optic IOL using the same principle of operation as the predicate device. Stability testing for the plunger material met the acceptance criteria showing no significant shifts in tensile and flexural properties following sterilization. Therefore, the effectiveness is demonstrated for the indicated use of the subject device. -The Akreos Single Use Insertion Device, Model AI-28 was shown to be safe using the results of the biocompatibility testing for the polypropylene material as well as polyphthalamide, the new plunger material. Sterilization and Shelf-life test results confirm that the subject device is substantially equivalent to the marketed predicate device. Conclusion Based on the 510(k) summary and the information provided herein, we conclude that the modified device is substantially equivalent to the predicate device, is safe and effective under the Federal Food, Drug and Cosmetic Act. {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged around the circumference of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Bausch & Lomb c/o Ms. Nanette Canepa Associate Manager, Global Regulatory Affairs 180 E. Via Verde Drive San Dimas, CA 91773 MAR 0 7 2007 Re: k063694 Trade/Device Name: Akreos Single Use Insertion Device Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I Product Code: KYB Dated: February 7, 2007 Received: February 8, 2007 Dear Ms. Canepa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ #### Page 2 – Ms. Nanette Canepa This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, M.B. Eichelman SimD Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### Indications for Use 510(k) Number (if known): K063694 Device Name: Indications For Use: Akreos Single Use Insertion Device The Akreos Single Use Insertion Device should only be used to fold and deliver those IOL models that allow use of this injector in their approved labeling. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) . (Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises Page 1 of *_*_ **510(k) Number** k063694
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...