EZMANAGER MAX DIABETES MANAGEMENT SOFTWARE

K063674 · Animas Corp. · LZG · Mar 30, 2007 · General Hospital

Device Facts

Record IDK063674
Device NameEZMANAGER MAX DIABETES MANAGEMENT SOFTWARE
ApplicantAnimas Corp.
Product CodeLZG · General Hospital
Decision DateMar 30, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesSoftware as a Medical Device, Therapeutic

Indications for Use

The Animas ezManager Plus diabetes management software is indicated for use as an accessory to Animas insulin pumps and specified commercially available blood glucose meters. The software supports diabetes management by the patient and/or health care professional by allowing for the review, analysis and evaluation of insulin delivery and blood glucose history information. The Animas ezManager Plus will also allow download of information from Animas pumps and commercially available blood glucose meters to a hand-held electronic device for display of the information. The Animas ezManager Plus is also indicated for the management of diabetes by calculating an insulin or carbohydrate dose based on health care professional prescribed settings and user entered data.

Device Story

Software application for PC and handheld devices; interfaces with Animas insulin pumps and blood glucose meters via IR dongle or connection cables. Enables data download/upload of pump settings (basal rates, ratios, sensitivity factors, target ranges, food database). Allows users to view/analyze insulin delivery and glucose history in various formats for trending. Facilitates dose calculation based on HCP-prescribed settings and user-entered data. Used by patients and HCPs for diabetes management; supports therapy optimization through data visualization and automated dose recommendations. Requires user verification of pump settings after upload.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Software-based diabetes management system; operates on personal computers and handheld electronic devices. Connectivity via IR dongle (for Animas pumps) and manufacturer-specified cables (for glucose meters).

Indications for Use

Indicated for patients with diabetes and healthcare professionals to manage diabetes via review of insulin delivery/blood glucose history and calculation of insulin or carbohydrate doses based on prescribed settings.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAR 3 0 2007 ## 510(k) Summary #### SUBMITTER: Animas Corporation 200 Lawrence Drive West Chester, PA 19380 Contact: Amy Smith; Manager, Regulatory & Clinical Affairs Animas ezManager Plus Diabetes Management Software COMMON OR USUAL NAME: PREDICATE DEVICE: DEVICE CLASSIFICATION: DESCRIPTION: DEVICE NAME: Class II per CFR 21 §880.5725, Infusion Pump, product code LZG Animas ezManager Plus, K022394 Deltec, CoZmanager, K062323 Insulin infusion pump accessory The Animas ezManager Plus diabetes management system is software that is loaded onto a patient's or health care professional's personal computer to allow for downloading information from Animas insulin pumps and specified commercially available blood glucose meters. The system allows for users to view data stored in the pump and meter history such as basal deliveries, bolus deliveries and blood glucose measurements. The data can be displayed in numerous formats and configurations to facilitate trending and better diabetes management. The Animas ezManager Plus diabetes management system also allows for Animas pump users to upload personal setting information to their pump. Personal setting information includes basal programs and rates, insulin to carbohydrate ratios, insulin sensitivity factors, target blood glucose ranges, and when available pump food database information (food, serving size and carbohydrate information). This feature allows users the convenience of programming and saving all of this information on their personal computers. Once the information is uploaded to the pump, the user is prompted to verify the active basal program before initiating therapy. {1}------------------------------------------------ Kc63674 1 of 2 For Animas pumps the download and upload is completed using an IR dongle and the IR window on the back of the For the specified meters, the download is pump. accomplished by using the manufacturer specified connection cables. ### INTENDED USE: The Animas ezManager Plus diabetes management software is indicated for use as an accessory to Animas insulin pumps and specified commercially available blood glucose meters. I m The software supports diabetes management by the patient and/or health care professional by allowing for the review, analysis and evaluation of insulin delivery and blood glucose history information. The Animas ezManager Plus will also allow download of information from Animas pumps and commercially available blood glucose meters to a hand-held electronic device for display of the information. The Animas ezManager Plus is also indicated for the management of diabetes by calculating an insulin or carbohydrate dose based on health care professional prescribed settings and user entered data. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" around the perimeter. Inside the circle is a stylized symbol featuring three human profiles in a row, with a flowing design element above them. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Amy Smith Manager, Regulatory & Clinical Affairs Animas Corporation 200 Lawrence Drive West Chester, Pennsylvania 19380 MAR 3 0 2007 Re: K063674 Trade/Device Name: ezManager Plus Diabetes Management Software Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: March 8, 2007 Received: March 9, 2007 Dear Ms. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ## Page 2 - Ms. Smith Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Snette y. Michael Oms. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K063674 Device Name: ezManager Plus Diabetes Management Software Indications for Use: The Animas ezManager Plus diabetes management software is indicated for use as an accessory to Animas insulin pumps and specified commercially available blood glucose meters. The software supports diabetes management by the patient and/or health care professional by allowing for the review, analysis and evaluation of insulin delivery and blood glucose history information. The Animas ezManager Plus will also allow download of information from Animas pumps and commercially available blood glucose meters to a hand-held electronic device for display of the information. The Animas ezManager Plus is also indicated for the management of diabetes by calculating an insulin or carbohydrate dose based on health care professional prescribed settings and user entered data. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Anten Vm f Anesthesiology, Regional Hospital Memorial Hospital Suction Control, Control Devices 1) Number: K463674 Page 1 of 1 i K063674 - Animas ezManager Plus Diabetes Management Software
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...