DIMENSION VISTA; CER FLEX REAGENT CARTRIDGE, PROTEIN 1 CALIBRATOR, PROTEIN 1 CONTROL MEDIUM, PROTEIN 1 CONTROL HIGH
K063655 · Dade Behring, Inc. · CHN · Feb 15, 2007 · Immunology
Device Facts
Record ID
K063655
Device Name
DIMENSION VISTA; CER FLEX REAGENT CARTRIDGE, PROTEIN 1 CALIBRATOR, PROTEIN 1 CONTROL MEDIUM, PROTEIN 1 CONTROL HIGH
Applicant
Dade Behring, Inc.
Product Code
CHN · Immunology
Decision Date
Feb 15, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5210
Device Class
Class 2
Indications for Use
The CER method is an in vitro diagnostic test for the quantitative determination of Ceruloplasmin in human serum, heparinized plasma or EDTA plasma on the Dimension Vista System. Measurements of ceruloplasmin aid in the diagnosis of copper metabolism disorders. PROT1 CAL is an in vitro diagnostic product for the calibration of the C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Immunoglobulin A (IGA), Immunoglobulin G (IGG), Immunoglobulin M (IGM) and Prealbumin (PREALB) methods on the Dimension Vista® System. PROT1 CON L, M and H are assayed intralaboratory quality controls for assessment of precision and analytical bias in the determination of C3 Complement (C3), C4 Complement (C4), ceruloplasmin (CER), immunoglobulin A (IGA), immunoglobulin G (IGG), Immunoglobulin M (IGM) and prealbumin (PREALB) on the Dimension Vista® System.
Device Story
The Dimension Vista CER Flex reagent cartridge is an in vitro diagnostic assay for quantitative ceruloplasmin measurement in human serum/plasma. It utilizes an immunochemical reaction where proteins in the sample form immune complexes with specific antibodies; these complexes scatter light. The intensity of scattered light is proportional to protein concentration. The system evaluates results by comparing them against a known standard. The device is used in clinical laboratory settings by trained personnel. Results assist clinicians in diagnosing copper metabolism disorders. The system includes associated liquid, human serum-based calibrators and assayed quality controls to ensure analytical precision and bias assessment.
Clinical Evidence
Bench testing only. Precision study (20 days, duplicate analysis) showed within-lab CVs ranging from 1.78% to 5.22%. Linearity confirmed across 0.07 to 3.40 g/L range (slope 1.018, r=1.000). Method comparison against predicate (n=130) yielded slope 1.009, intercept 0.008, r=0.994. Interference testing confirmed no significant bias (±10%) for common endogenous substances, RF up to 685 IU/mL, and total protein up to 127 g/L. Matrix comparison (serum vs. plasma) showed high agreement (r ≥ 0.998).
Technological Characteristics
Reagent cartridge contains reaction buffer (phosphate buffer, polyethylene) and rabbit antiserum to human ceruloplasmin. Quantitative nephelometric sensing principle. Energy source: system-integrated light source for scattering. Dimensions/form factor: multi-well reagent cartridge. Connectivity: system-specific (Dimension Vista®). Sterilization: not specified. Software: automated system-controlled analysis.
Indications for Use
Indicated for the quantitative determination of ceruloplasmin in human serum, heparinized plasma, or EDTA plasma to aid in the diagnosis of copper metabolism disorders. For prescription use only.
Regulatory Classification
Identification
A ceruloplasmin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the ceruloplasmin (copper-transporting serum protein) in serum, other body fluids, or tissues. Measurements of ceruloplasmin aid in the diagnosis of copper metabolism disorders.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9.
Predicate Devices
Dade Behring N Antisera to Human Ceruloplasmin (k053074)
Dade Behring N Protein Standard SL (k012470)
Dade Behring N/T Protein Control SL (k012468)
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k063655
B. Purpose for Submission:
New Device
C. Measurand:
Ceruloplasmin
D. Type of Test:
Quantitative, nephelometric.
E. Applicant:
Dade Behring, Inc.
F. Proprietary and Established Names:
Dimension Vista™ Ceruloplasmin Flex® reagent cartridge (CER)
Dimension Vista™ Protein 1 Calibrator
Dimension Vista™ Protein 1 Control L
Dimension Vista™ Protein 1 Control M
Dimension Vista™ Protein 1 Control H
G. Regulatory Information:
1. Regulation section:
21 CFR 866.5210; Ceruloplasmin Immunological Test System
21 CFR 862.1150; Calibrator
21 CFR 862.1660; Quality Control Material (assayed and unassayed)
2. Classification:
Class II; Ceruloplasmin test system and Calibrator
Class I; Control
3. Product code:
CHN; Immunochemical, Ceruloplasmin
JIX; Calibrator, Multi-Analyte Mixture
JJY; Multi Analyte Controls
4. Panel:
Immunology 82; Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
Dimension Vista™ CER Flex® reagent cartridge:
The CER method is an *in vitro* diagnostic test for the quantitative determination of ceruloplasmin in human serum, heparinized plasma or EDTA plasma on the Dimension Vista® System. Measurements of ceruloplasmin aid in the diagnosis of copper metabolism disorders.
Dimension Vista™ Protein 1 Calibrator:
PROT1 CAL is an *in vitro* diagnostic product for the calibration of the C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Immunoglobulin A (IGA), Immunoglobulin G (IGG), Immunoglobulin M (IGM) and Prealbumin (PREALB) methods on the Dimension Vista® System.
{2}
2
# Dimension Vista™ Protein 1 Control L, M and H:
PROT1 CON L, M and H are assayed intralaboratory quality controls for assessment of precision and analytical bias in the determination of C3 complement (C3), C4 complement (C4), ceruloplasmin (CER), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB) on the Dimension Vista® System.
2. Indication(s) for use:
Same as intended use
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
For use on the Dade Behring Dimension Vista® System
# I. Device Description:
# Dimension Vista™ CER Flex® reagent cartridge
Dimension Vista™ CER Flex® reagent cartridge consists of 12 wells containing reaction buffer: phosphate buffer and polyethylene and rabbit antiserum to human ceruloplasmin.
# Dimension Vista™ Protein 1 Calibrator
Protein 1 Calibrator is a multi-analyte, liquid, human serum based product containing C3 complement, C4 complement, ceruloplasmin (CER), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB).
# Dimension Vista™ Protein 1 Control L, M and H
Protein 1 Control L, M and H are multi-analyte, liquid, human serum based products containing C3 complement, C4 complement, ceruloplasmin (CER), immunoglobulin A (IGA), immunoglobulin G (IGG), immunoglobulin M (IGM) and prealbumin (PREALB).
Calibrators and Controls are sold separately.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Dade Behring N Antisera to Human Ceruloplasmin
Dade Behring N Protein Standard SL
Dade Behring N/T Protein Control SL
2. Predicate 510(k) number(s):
k053074; k012470; k012468
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | For the quantitative determination of ceruloplasmin | Same |
| Antibody | Rabbit polyclonal | Same |
| Reportable Range | 0.07 to 2.2 g/L | Same |
| Controls | L, M and H | Same |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Traceability | Calibrators traceable to ERM® DA470 reference standard | Same |
| Differences | | |
| Item | Device | Predicate |
| Sample matrix | Serum, heparinized plasma or EDTA plasma | Serum or heparinized plasma |
| Instrumentation | Dimension Vista® System | BN™ Systems |
| Calibrator | Dimension Vista™ Protein 1 Calibrator | N Protein Standard SL |
| Calibrator Constituents | C3, C4, ceruloplasmin, IgA, IgG, IgM, and prealbumin | IgG, IgG1, IgG2, IgG3, IgG4, IgA, IgM, IgE, C3c, C4, transferrin, albumin, α1-antitrypsin, α2-macroglobulin, haptoglobin, α1-acid glycoprotein, prealbumin, hemopexin, ceruloplasmin, retinol binding protein, Ig Light-chain kappa, Ig Light-chain lambda, soluble transferrin receptor, ferritin, β2-microglobulin and total protein |
| Control | Dimension Vista™ Protein 1 Control L, M and H | N/T Protein Control SL L, M and H |
| Control Constituents | C3, C4, ceruloplasmin, IgA, IgG, IgM, and prealbumin | IgG, IgG1, IgG2, IgG3, IgG4, IgA, IgM, IgE, C3c, C4, transferrin, albumin, α1-antitrypsin, α2-macroglobulin, haptoglobin, α1-acid glycoprotein, prealbumin, hemopexin, ceruloplasmin, retinol binding protein, Ig Light-chain kappa, Ig Light-chain lambda, soluble transferrin receptor, ferritin, β2-microglobulin and total protein. |
# K. Standard/Guidance Document Referenced (if applicable):
None
# L. Test Principle:
Proteins contained in human body fluids form immunochemical reaction with specific antibodies. These complexes scatter a beam o flight passed through the sample. The
{4}
intensity of the scattered light is proportional to the concentration of the relevant protein in the sample. The result is evaluated by comparison with a standard of known concentration.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
**Precision Study**
Specimens at various levels (Protein 1 Control L, M, H; Serum High, Serum Pool High and Plasma High) were analyzed in duplicate, twice a day, for 20 days. The repeatability and within-lab standard deviations (SD) and percent coefficient of variation (%CV) were calculated and a summary of the precision data is presented in the Table below.
| Material | Mean Value (g/L) | Within-run SD (%CV) | Within-Lab SD (%CV) |
| --- | --- | --- | --- |
| PROT1 CON L | 0.22 | 0.01 (3.81) | 0.01 (4.53) |
| PROT1 CON M | 0.26 | 0.01 (3.02) | 0.01 (3.42) |
| PROT1 CON H | 0.41 | 0.02 (3.94) | 0.02 (4.01) |
| Serum High | 1.63 | 0.04 (2.54) | 0.05 (3.08) |
| Serum Pool High | 2.13 | 0.03 (1.54) | 0.04 (1.78) |
| Plasma Pool | 0.51 | 0.01 (2.65) | 0.03 (5.22) |
#### b. Linearity/assay reportable range:
Linearity across the assay range was confirmed by testing a calibrator with a high concentration of ceruloplasmin. This calibrator was serially diluted with System Diluent (3.40 to 0.07 g/L). Each dilution was tested in replicates of three. Data were analyzed using linear regression analysis. The acceptance criteria of slope between 0.9 and 1.1 and correlation coefficient ≥0.95 were met. A summary of the linearity data is presented in the table below.
**Linearity Data Summary**
| Dimension Vista™ | n | Slope | Intercept | Correlation Coefficient |
| --- | --- | --- | --- | --- |
| CER | 6 | 1.018 | -0.015 | 1.000 |
### High Dose Hook Effect
The possibility of hook effect occurring when using the Dimension Vista™ CER assay was evaluated with a serum sample above the assay range. The sample was analyzed on both the BN ProSpec® System and the Dimension Vista® System, indicating no hook effect up to 4.441 g/L. A summary of the hook effect data are presented below.
| BN ProSpec® System Ceruloplasmin Concentration (g/L) | Dimension Vista® Test Dilution | Dimension Vista™ CER Concentration (g/L) |
| --- | --- | --- |
| 4.13 | 1:10 | >2.2 (exceeds assay range) |
| | 1:100 | 4.441 |
{5}
c. Traceability, Stability, Expected values (controls, calibrators, or methods): The analyte in the calibrator and controls is traceable to ERM® DA470 reference standard.
d. Detection limit:
Analytical sensitivity is defined as the minimal detectable level of analyte which can be distinguished from zero. The value was calculated as the mean value of twenty replicates of the System Diluent plus 2SD and was determined to be 0.14 g/L.
e. Analytical specificity:
Potentially interfering endogenous substances
Interference testing was performed to determine the effect of various endogenous and exogenous substances on the Dimension Vista™ CER assay. For all interferents except Triglycerides, RF, and Total Protein, the percent bias was determined by testing a control sample without the interferent and comparing it to the value obtained from a test sample to which the potential interferent had been added. A bias exceeding ±10% was considered a significant interference.
For each spiked sample, the % recovery was determined [% Recovery = (Test result/Baseline) x 100]. The acceptance criterion of ±10% relative deviation from the base pool was met for all interferents tested. A summary of the results is presented below.
| Exogenous Substance | Concentration in sample | 95% Sample + 5% Stock solution | | |
| --- | --- | --- | --- | --- |
| | | Test sample Mean g/L | Control g/L | Recovery (%) |
| Bilirubin (conjugated) | 60 mg/dL | 0.46 | 0.44 | 103% |
| Bilirubin (unconjugated) | 60 mg/dL | 0.45 | 0.44 | 100% |
| Creatinine | 30 mg/dL | 0.44 | 0.44 | 100% |
| Hemoglobin | 1000 mg/dL | 0.44 | 0.44 | 98% |
| Immunoglobulin G (IgG) | 5000 mg/dL | 0.34 | 0.32 | 106% |
| Protein: Albumin | 6000 mg/dL | 0.39 | 0.39 | 101% |
| Urea | 500 mg/dL | 0.45 | 0.44 | 102% |
| Cholesterol | 500 mg/dL | 0.43 | 0.42 | 102% |
| Uric Acid | 20 mg/dL | 0.44 | 0.44 | 100% |
Triglyceride interference was evaluated by testing five samples containing a known amount of triglycerides and compared to results from the same sample after centrifugation. Results showed no significant change in CER recovery after centrifugation. This study however, only showed effect of centrifugation but did not address the triglyceride interference since the actual CER concentration in these samples without triglycerides were not known. The package insert was revised to state that lipemic samples should be avoided in the specimen handling section.
To evaluate interference from total protein (TP) and rheumatoid factor (RF), samples which had RF concentrations > 500 IU/mL and TP concentrations > 120 g/L and samples with no detectable RF concentration or low TP concentrations were used to prepare samples for the study. A (1+1) mixture of
{6}
samples with high concentrations of RF and TP were prepared and the ceruloplasmin concentrations determined in replicates of five on the Dimension® Vista System. For each sample preparation, the $\%$ recovery was calculated [% Recovery = (measured result/calculated result) x 100]. The acceptance criterion of $\pm 10\%$ relative deviation from the calculated result was met. No significant interference was found with the Dimension Vista® CER method from the presence of RF up to $685~\mathrm{IU / mL}$ or from elevated Total Protein at $127~\mathrm{g / L}$ at CER concentrations ranging from 0.19 to $0.754~\mathrm{g / L}$ . Results for TP and RF interference are presented in the following tables:
| | TP (g/L) | CER (g/L) |
| --- | --- | --- |
| Sample A | 213 | 0.173 |
| Sample B | 217 | 0.419 |
| Sample C | 209 | 0.707 |
| Sample D | 131 | 0.191 |
| Sample E | 128 | 0.447 |
| Sample F | 132 | 0.755 |
| Sample G | 40.9 | 0.206 |
| Sample H | 33.1 | 0.327 |
| Sample I | 35.3 | 0.666 |
| Sample ID | Dilution 1+1 | TP (g/L) | CER calculated (g/L) | CER measured (g/L) | Recovery (%) |
| --- | --- | --- | --- | --- | --- |
| 1 | Sample A + G | 127 | 0.190 | 0.196 | 103 |
| 2 | Sample B + H | 125 | 0.373 | 0.374 | 100 |
| 3 | Sample C + I | 122 | 0.687 | 0.693 | 101 |
| 4 | Sample D + G | 86.0 | 0.198 | 0.203 | 102 |
| 5 | Sample E + H | 80.6 | 0.387 | 0.392 | 101 |
| 6 | Sample F + I | 83.7 | 0.711 | 0.791 | 111 |
| | RF (IU/mL) | CER (g/L) |
| --- | --- | --- |
| Sample A | 922 | 0.248 |
| Sample B | 1370 | 0.443 |
| Sample C | 1340 | 0.762 |
| Sample D | 504 | 0.207 |
| Sample E | 774 | 0.470 |
| Sample F | 423 | 0.628 |
| Sample G | < 9.6 | 0.185 |
| Sample H | < 9.6 | 0.430 |
| Sample I | < 9.6 | 0.746 |
| Sample ID | Dilution 1+1 | RF (IU/mL) | CER calculated (g/L) | CER measured (g/L) | Recover (%) |
| --- | --- | --- | --- | --- | --- |
| 1 | Sample A + G | 461 | 0.216 | 0.219 | 101 |
| 2 | Sample B + H | 685 | 0.415 | 0.415 | 100 |
| 3 | Sample C + I | 122 | 0.687 | 0.693 | 101 |
| 4 | Sample D + G | 86.0 | 0.198 | 0.203 | 102 |
| 5 | Sample E + H | 80.6 | 0.387 | 0.392 | 101 |
| 6 | Sample I + I | 83.7 | 0.711 | 0.791 | 111 |
{7}
7
| Sample ID | Dilution 1+1 | RF (IU/mL) | CER calculated (g/L) | CER measured (g/L) | Recover (%) |
| --- | --- | --- | --- | --- | --- |
| 3 | Sample C + I | 670 | 0.754 | 0.733 | 97 |
| 4 | Sample D + G | 252 | 0.196 | 0.194 | 99 |
| 5 | Sample E + H | 387 | 0.450 | 0.461 | 102 |
| 6 | Sample F + I | 212 | 0.507 | 0.506 | 100 |
f. Assay cut-off:
No change.
2. Comparison studies:
a. Method comparison with predicate device:
The Dimension Vista™ CER assay was compared to the Dade Behring N Antisera to Human Ceruloplasmin assay on the BN ProSpec® System by evaluating 60 serum samples, 32 Li plasma and 38 NH-plasma samples with concentrations ranging from 0.09 g/L to 2.20 g/L. Regression analysis of these results yielded the following results:
Method Comparison Study
| Comparative Method | n | Slope | Intercept (g/L) | Correlation Coefficient |
| --- | --- | --- | --- | --- |
| N Antisera to Human Ceruloplasmin on BN ProSpec® (95% CI) | 130 | 1.009 (0.986-1.036) | 0.008 (0.00-0.02) | 0.994 |
b. Matrix comparison:
Serum versus Plasma Comparison:
A comparison was performed with matched specimens of serum, and EDTA, lithium heparin, and sodium heparin plasma (10 for each serum/plasma pairs) assayed on the Dimension Vista® System. The samples covered the reportable range. Linear regression analysis showed agreement between the serum and the plasma samples tested.
| Linear Regression vs. Serum | | | |
| --- | --- | --- | --- |
| | Li Hep | Na Hep | EDTA |
| Slope: | 0.99 | 1.02 | 0.97 |
| Y-int (g/L): | 0.00 | -0.02 | 0.01 |
| r: | 0.999 | 0.998 | 1.000 |
| Syx: | 0.03 | 0.05 | 0.02 |
| Slope 95% CI | 0.971-1.01 | 0.986-1.06 | 0.955-0.99 |
| % Recovery Statistics | | | |
| Mean: | -0.8% | -1.0% | -0.4% |
| Min: | -4.9% | -5.1% | -5.8% |
| Max: | 5.3% | 4.2% | 5.7% |
{8}
3. Clinical studies:
a. Clinical Sensitivity: Not applicable.
b. Clinical specificity: Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.
4. Clinical cut-off: Not applicable.
5. Expected values/Reference range: Expected Values: 0.2 – 0.6 g/L. This reference interval applies to serum samples from healthy adults
N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.