X-FORCE N30 NEPHROSTOMY BALLOON DILATION CATHETER
K063632 · C.R. Bard, Inc. · LJE · Jan 5, 2007 · GU
Device Facts
| Record ID | K063632 |
| Device Name | X-FORCE N30 NEPHROSTOMY BALLOON DILATION CATHETER |
| Applicant | C.R. Bard, Inc. |
| Product Code | LJE · GU |
| Decision Date | Jan 5, 2007 |
| Decision | SESE |
| Submission Type | Special |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
The Bard X-Force™ Nephrostomy Balloon Dilation Catheter is recommended for use in the dilation of the nephrostomy tract and for placement of the working sheath.
Device Story
The X-Force™ N30 Nephrostomy Balloon Dilation Catheter is a medical device used by clinicians to dilate the nephrostomy tract and facilitate the placement of a working sheath. The device consists of a dual-lumen shaft featuring an inflation lumen and a guidewire lumen, with a dilation balloon located at the distal end. A radiopaque band is positioned at the distal end of the balloon's working length to assist with radiographic visualization and accurate placement during the procedure. A high-pressure stopcock is attached to the inflation lumen to maintain balloon pressure after the pressurization apparatus is removed. The device is used in a clinical setting to assist in urological procedures, providing a mechanical means to expand the tract for subsequent instrumentation.
Clinical Evidence
Bench testing only. The device was subjected to design verification and validation activities to confirm it meets the same performance specifications as the predicate device.
Technological Characteristics
Dual-lumen catheter shaft; inflation and guidewire lumens; distal dilation balloon; radiopaque band for visualization; high-pressure stopcock for pressure maintenance. Mechanical device; no energy source or software.
Indications for Use
Indicated for dilation of the nephrostomy tract and placement of a working sheath in patients requiring nephrostomy procedures.
Predicate Devices
- X-Force™ N30 Nephrostomy Balloon Dilation Catheter (K051316)
Submission Summary (Full Text)
{0}------------------------------------------------
K063632
PG. i uf l
## 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
- A. Submitter's Information:
| Submitter's Name: | C. R. Bard, Inc., Urological Division |
|-------------------------|-----------------------------------------------|
| Address: | 13183 Harland Dr.<br>Covington, Georgia 30014 |
| Contact Person: | Terri Morris |
| Contact Person's Phone: | (678) 342-4922 |
| Contact Person's Fax: | (678) 342-4992 |
| Date of Preparation: | November 28, 2006 |
| JAN - 5 2007 | |
- B. Device Name:
| Trade Name: | X-Force™ N30 Nephrostomy Balloon Dilation Catheter |
|----------------------|----------------------------------------------------|
| Common / Usual Name: | Nephrostomy Balloon Dilation Catheter |
| Classification Name: | 78 LJE - Catheter, Nephrostomy |
| CFR Reference: | N/A - Unclassified |
- C. Predicate Device Name:
X-Force™ N30 Nephrostomy Balloon Dilation Catheter # K051316
- D. Device Description:
The Bard X-Force™ Nephrostomy catheter is composed of a dual lumen shaft with inflation and guidewire lumens and a dilation balloon on the distal end. The catheter has a band which corresponds with the distal end of the balloon's working length to radiopaque facilitate radiographic visualization and placement. A high pressure stopcock is used on the inflation lumen to maintain pressure after removal of the pressurization apparatus,
- E. Intended Use:
The Bard X-Force™ Nephrostomy Balloon Dilation Catheter is recommended for use in the dilation of the nephrostomy tract and for placement of the working sheath.
- F. Technological Characteristics Summary:
The modified device has the same intended use, general design and fundamental scientific technology as the predicate device.
- G. Performance Data Summary:
The X-Force™ N30 Nephrostomy Balloon Dilation Catheter referenced in this submission is held to the same design, manufacture, and performance specifications as those catheters currently manufactured by Bard. The appropriate design verification and validation activities for the modifications to the device were conducted.
Trade Name:
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Ms. Terri Morris Regulatory Affairs Specialist II Bard Urological Division C. R. Bard, Inc. 13183 Harland Drive COVINGTON GA 30014
## JAN - 5 2007
Re: K063632
Trade/Device Name: X-Force™ N30 Nephrostomy Balloon Dilation Catheter Regulatory Class: Unclassified Regulation Number: None Product Code: LJE Dated: November 30, 2006 Received: December 8, 2006
Dear Ms. Morris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/1/Picture/10 description: The image shows a circular logo with the letters FDA in the center. The logo also includes the words "Centennial" and "1906-2006". There are three stars below the word Centennial. The logo is surrounded by text that reads "A Century of Consumer Protection".
ling Public
{2}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancv C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Bard Urological Division, C.R. Bard, Inc. X-Force™ N30 Nephrostomy Balloon Dilation Catheter Premarket Notification [510(k)]
1.3 Indications for Use Statement
K-063632 510(k) Number (if known):
Device Name:
Indications for Use:
The Bard X-Force™ Nephrostomy Balloon Dilation Catheter is recommended for use in the dilation of the nephrostomy tract and for placement of the working sheath.
X Prescription Use (Part 21 CFR 801 Subpart D) ANDIOR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
David W. Wyman
(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number __
(Recommended Format 11/13/2003)