The EKOS Micro-Infusion System is intended for regional infusion of contrast materials into selected vessels in the neurovasculature. The EKOS Micro- Infusion System may be used for controlled, regional infusion into selected vessels. The EKOS Micro-Infusion system is intended to deliver physician specified fluids to the coronary vasculature. The EKOS Micro-Infusion System is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.
Device Story
Infusion catheter system designed for controlled, selective delivery of physician-specified fluids (e.g., contrast, thrombolytics) into peripheral, neurovascular, and coronary vessels. Device features distal catheter tip with integrated transducer element delivering ultrasound energy. Operated by physicians in clinical settings. Ultrasound energy is non-therapeutic; intended solely for infusion support. System provides localized fluid delivery to target vasculature; assists in clinical procedures requiring precise fluid administration. Safety/effectiveness for intracoronary thrombolytic therapy not established.
Clinical Evidence
Bench testing and animal studies conducted. Results demonstrate device meets design specifications and is safe/effective for its intended use. No clinical trial data provided.
Technological Characteristics
Infusion catheter system; distal tip transducer element for ultrasound energy delivery; catheter end-hole for fluid infusion. Class II device. Standalone operation.
Indications for Use
Indicated for regional infusion of contrast materials into neurovasculature, controlled regional infusion into selected vessels, delivery of physician-specified fluids to coronary vasculature, and controlled/selective infusion of physician-specified fluids (including thrombolytics) into peripheral vasculature. Contraindicated for intracoronary thrombolytic therapy administration; ultrasound energy is not intended to be therapeutic for coronary thrombolysis or thrombectomy.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font below.
January 7, 2022
EKOS Corporation Jocelyn Kersten Vice President, Regulatory & Clinical Affairs 11911 N Creek Parkway South Bothell, Washington 98011
Re: K063620
Trade/Device Name: EKOS Micro-Infusion System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEY, KRA
Dear Jocelyn Kersten:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 6, 2007. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory.Oconnell(@FDA.HHS.gov.
Sincerely,
Digitally signed by Gregory W. Gregory W. O'connell -O'connell -S Date: 2022.01.07 13:51:23 -05'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human services.
**AUG 6 2007**
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
EKOS Corporation c/o Ms. Jocelyn Kersten Vice President, Regulatory and Clinical Affairs 11911 North Creek Parkway South Bothell, WA 98011
Re: K063620
> Trade/Device Name: EKOS Micro Infusion System Regulation Number: 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: II Product Code: KRA Dated: April 24, 2007 Received: April 25, 2007
Dear Ms. Kersten:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:
The safety and effectiveness of the EKOS Micro-Infusion system used for intracoronary thrombolytic therapy administration have not been established. In particular, the ultrasound energy delivered by the EKOS micro-infusion system is not intended to be therapeutic, and the safety and effectiveness of the EKOS system for coronary thrombolysis or thrombectomy (i.e., clot disruption) have not been established.
{2}------------------------------------------------
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as prooced in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting (1 Mr.), N. May 6 can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Donna-Bea Tillman, Ph.D., M.P.A. Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### 1.10 Indications for Use Statement
## EKOS Corporation 11911 North Creek Parkway South Bothell, WA 98011
### Indications for Use Statement
510(k) Number (if known): K063620
EKOS Micro-Infusion System Device Name:
Indications for Use:
The EKOS Micro-Infusion System is intended for regional infusion of contrast materials into selected vessels in the neurovasculature. The EKOS Micro- Infusion System may be used for controlled, regional infusion into selected vessels.
The EKOS Micro-Infusion system is intended to deliver physician specified fluids to the coronary vasculature.
WARNING: The safety and effectiveness of the EKOS Micro-Infusion system used for intracoronary thrombolytic therapy administration have not been established. In particular, the ultrasound energy delivered by the EKOS Micro-Infusion system is intended to be therapeutic, and the safety and effectiveness of the EKOS system for coronary thrombolysis or thrombectomy (i.e., clot disruption) have not been established.
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | X |
|----------------------|----------------------|
| (Per 21 CFR 801.109) | |
| | OR |
| | Over-The-Counter Use |

Division Sign-OffCONFIDENTIAL
{4}------------------------------------------------
K063620
# 8.0 510(k) Summary
| Applicant Name: | EKOS Corporation | AUG - 6 2007 |
|-----------------|------------------------------------------------------|--------------|
| Address: | 11911 North Creek Parkway South<br>Bothell, WA 98011 | |
| Contact Person: | Jocelyn Kersten<br>Director, Regulatory Affairs | |
| Telephone: | (425) 415-3132 | |
| Fax: | (425) 415-3102 | |
| Device: | EKOS Micro-Infusion System | |
| Classification: | CFR 870.1210 - Continuous Flush Catheter | |
| Panel: | Cardiovascular | |
| Product Code: | KRA | |
| Intended Use: | | |
The EKOS Micro-Infusion System is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.
The EKOS Micro-Infusion System is intended for regional infusion of contrast materials into selected vessels in the neurovasculature. The EKOS Micro- Infusion System may be used for controlled, regional infusion into selected vessels.
The EKOS Micro-Infusion system is intended to deliver physician specified fluids to the coronary vasculature.
| Device Description: | The EKOS Micro-Infusion System is an infusion catheter system<br>designed to deliver fluids through the catheter end-hole while<br>simultaneously delivering ultrasound energy via a transducer<br>element at the distal catheter tip. |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Basis: | The EKOS Micro-Infusion System is substantially equivalent to<br>other legally marketed devices. These devices include:<br><br>EKOS Micro-Infusion System, (EKOS Corporation, K053437,<br>K053432, K062507, K062508)<br><br>Endeavor Infusion Catheter, (Cordis Corporation, K972110) |
CONFIDENTIAL
{5}------------------------------------------------
:
| PV 0.018 F/X, (JoMed, K944004) | |
|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Renegade Hi-Flo Microcatheter, (Boston Scientific<br>Corporation, K000177) | |
| TurboTracker 18 Infusion Catheter, (Boston Scientific<br>Corporation, K960806) | |
| FasTracker 10 Infusion Catheter, (Boston Scientific<br>Corporation, K926243) | |
| Performance: | EKOS has conducted preclinical bench and animal studies with the<br>Micro-Infusion System. These studies demonstrate that the<br>performance of the Micro-Infusion System meets its design<br>specifications and is safe and effective for its intended use. |
:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.