← Product Code [DPT](/productcode/DPT) · K063586

# MOORFLPI FULL-FIELD LASER PERFUSION IMAGER (K063586)

_Moor Instruments, Ltd. · DPT · Jan 25, 2007 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K063586

## Device Facts

- **Applicant:** Moor Instruments, Ltd.
- **Product Code:** [DPT](/productcode/DPT.md)
- **Decision Date:** Jan 25, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2120
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The moorFLPI Full-Field Laser Perfusion Imager is intended for blood flow measurements in the microcirculation. This device is intended for clinical research use.

## Device Story

The moorFLPI performs non-contact imaging of tissue blood perfusion in the microcirculation. It illuminates tissue with a diverging infrared laser beam, creating a laser speckle pattern captured by a CCD camera. The device uses speckle contrast analysis to process video data, where high perfusion causes rapid speckle variation (low contrast/blurring) and low perfusion causes stable speckle (high contrast). The system generates color-coded perfusion maps at video rates (25 frames/second). Used in clinical research settings, the device is operated by researchers to assess superficial blood flow. By providing real-time, full-field imaging, it reduces the impact of temporal perfusion variations compared to scanning-based systems. The device is a Class 1 laser product, enhancing safety for patients and operators.

## Clinical Evidence

Bench testing only. Performance evaluated using a simple flow model and comparative image scans of skin blood flow against the predicate device (moorLDI2-IR). Results indicate substantially equivalent performance for superficial blood flow measurements within the specified range and imaging area. Device also tested for compliance with electrical safety, laser radiation safety (IEC 60825-1:2001), electromagnetic compatibility, and programmable medical device standards.

## Technological Characteristics

Infrared laser source (Class 1, IEC 60825-1:2001); CCD camera imaging; PC-based image processing. Principle: Speckle contrast analysis (quantifies flux based on speckle pattern blurring). Full-field illumination via diverging beam. Connectivity: PC-based video data acquisition. Software: Perfusion mapping algorithm.

## Regulatory Identification

An extravascular blood flow probe is an extravascular ultrasonic or electromagnetic probe used in conjunction with a blood flowmeter to measure blood flow in a chamber or vessel.

## Predicate Devices

- moorLDI2-IR Infrared Laser Doppler Imager ([K032841](/device/K032841.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Moor Instruments. The logo consists of a stylized sunburst graphic on the left, followed by the words "moor instruments" in a sans-serif font. Below the company name is the tagline "laser Doppler blood flow assessment" in a smaller, lighter font.

K063586

## 7 510(k) Summary

:

| Submitter:             | Moor Instruments Ltd                                                                                                                                                 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:               | Millwey<br>Axminster, Devon<br>EX13 5HU, United Kingdom                                                                                                              |
|                        | JAN 2 5 2007                                                                                                                                                         |
| Telephone:             | (+44) 1297 35715                                                                                                                                                     |
| Fax:                   | (+44) 1297 35716                                                                                                                                                     |
| Contact:               | Xiabing Huang                                                                                                                                                        |
| Contact title:         | Technical Manager                                                                                                                                                    |
| E-mail:                | xhuang@moor.co.uk                                                                                                                                                    |
| Date:                  | November 27, 2006                                                                                                                                                    |
| Model Name:            | moorFLPI Full-Field Laser Perfusion Imager                                                                                                                           |
| Model Number:          | moorFLPI                                                                                                                                                             |
| Common Name:           | Full-Field Laser Perfusion Imager                                                                                                                                    |
| Classification Name:   | Extravascular blood flow probe, DPT, 21 CFR 870.2120<br>Laser surgical instrument for use in general and plastic<br>surgery and in dermatology, GEX, 21 CFR 878.4810 |
| Regulatory Status:     | Class II                                                                                                                                                             |
| Establishment Reg No:  | 8043564                                                                                                                                                              |
| Type of 510(k):        | Traditional                                                                                                                                                          |
| Reason for submission: | New device                                                                                                                                                           |
| Predicate Device:      | moorLDI2-IR Infrared Laser Doppler Imager<br>510(k) Number K032841                                                                                                   |

Moor Instruments L.td Millwey Axminster Devon EX13 5HU UK 1el +44 (0)1297 35715 fax +44 (0)1297 35716 emall sales@moor.co.uk website www.moor.co.uk Company Registered in England No. 2205367 VAT Registration No. GB490867906

Image /page/0/Picture/5 description: The image contains two certification logos. The logo on the left has a check mark inside a circle with the text "ISO 9001:2015 SGS". The logo on the right has a check mark with a crown above it and the text "UKAS QUALITY MANAGEMENT 005" below it.

{1}------------------------------------------------

#### Description of the Device 7.1

The moorFLPI is a device to perform non-contact imaging of tissue blood perfusion in the microcirculation, for example skin, using speckle contrast analysis. The tissue surface is illuminated with a diverging infra-red laser beam resulting in a laser speckle pattern. The pattern is imaged by a CCD camera and image processing of the speckle contrast is used to generate colour code images of the tissue blood perfusion in the microcirculation.

#### 7.2 Intended Use

The moorFLPI Full-Field Laser Perfusion Imager is intended for blood flow measurements in the microcirculation. This device is intended for clinical research use.

#### 7.3 Technological Characteristics

Tissue is illuminated by floodlighting with a diverging infra-red laser beam. The tissue is imaged with a CCD camera and video data is acquired by a PC. Image processing software on the PC is used to generate colour-coded maps of tissue perfusion. The image processing software uses the fact that high perfusion produces rapid variation in the laser speckle pattern, which is integrated by the CCD to produce an area of low contrast (seen as blurring of the speckle pattern in the video image). Conversely, low perfusion causes little variation in the speckle pattern and as a result a high contrast area of well-defined speckle is produced in the video image. Contrast is quantified and the resulting flux is colour-coded to produce a perfusion image.

The moorFLPI uses a large diameter diverging laser beam to illuminate an area of tissue and measures tissue perfusion at all points in the image simultaneously, hence the "Full-Field" designation in the device name. The predicate device uses a collimated laser beam to sequentially illuminate very small areas of tissue and generates images by scanning the beam in a raster pattern. These differences in the means to capture a single frame image mean that the moorFLPI is able to generate images more rapidly than the predicate device. The moorFLPI records images at video rates (25 frames/second) and the moorLDI2-IR takes approximately 2 minutes to record a single 'frame' 100x150 pixel blood flow image. The effect of possible temporal variations in tissue perfusion recorded with the moorFLPI are therefore greatly reduced.

The moorFLPI uses a diverging infrared laser beam and is classified as a Class 1 laser product (IEC 60825-1:2001). Class 1 laser products are generally regarded as nonhazardous to the eye and skin and protective eyewear is not required to be worn by the operator or patient. The predicate device uses a collimated laser beam of the same wavelength and is classified as a Class 3R laser product (IEC60825-1:2001) and as such is potentially harmful to the eye. The lower laser classification of the moorFLPI provides an increased level of safety compared with the predicate device.

The moorFLPI and the moorLDI2-IR both measure tissue perfusion by making use of the Doppler frequency shifts that result from coherent laser light being scattered

{2}------------------------------------------------

from moving red blood cells. The moorFLPI measurement of blood perfusion depends on analysing the speckle pattern contrast that results from interference between the Doppler shifted laser light and the laser light scattered from the tissue in which the blood is moving, whereas the moorLDI2-IR measurement of blood perfusion depends on direct measurements of the Doppler frequency shifts. The speckle pattern contrast is generated mainly by laser light scattered from the superficial layers of tissue. Detected light coming from deeper regions e.g. from underlying veins, is of low intensity and contributes little to the recorded light intensity. Flow in veins that are more than a mm or so below the surface are not imaged by the moorFLPI; however because the moorLDI2-IR is measuring the Doppler frequency shifts and calculates blood flow from the first moment of the integrated power spectral density of the detected photocurrent, (i.e. the frequency weighted power spectral density), it records not only the blood flow in the superficial tissue layers but also the faster flow in underlying veins.

Both the moorFLPI and the moorLDI2-IR can be used to image flow in small blood vessels exposed during open surgery.

#### 7.4 Performance Data

In order to evaluate the performance of the moorFLPI Full-Field Laser Perfusion Imager, and determine its substantial equivalence to the predicate device (moorLDI2-IR), a set of comparison tests has been carried out. These include measurements from a simple flow model and single point and image scans from blood flow in skin, using both devices. The results show that the moorFLPI, within its specified range for flow measurements and area of tissue that can be imaged, has substantially the same performance as the predicate device for measurements of superficial blood flow.

The moorFLPI Full-Field Laser Perfusion Imager has been designed and tested for compliance with the standards for electrical safety, laser radiation safety, electromagnetic compatibility and programmable medical device.

# 7.5 Conclusions

From the description of the technological characteristics and the performance data, it can be concluded that the moorFLPI Full-Field Laser Perfusion Imager is substantial equivalence to the predicate device moorLDI2-IR laser Doppler imager in terms of effectiveness and safety.

{3}------------------------------------------------

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white and has a simple, clean design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Moor Instruments, Ltd. % Xiabing Huang Technical Manager Millwey Axminster, Devon EX13 5HU, United Kingdom

JAN 2 5 2007

Re: K063586

Trade/Device Name: moorFLPI Full-Field Laser Perfusion Imager Regulation Number: 21 CFR 870.2120 Regulation Name: Extravascular blood flow probe Regulatory Class: II Product Code: DPT, GEX Dated: November 27, 2006 Received: December 1, 2006

Dear Xiabing Huang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 -- Xiabing Huang

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally . marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours

Sincerely yours,

Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

### Indications for Use Statement 5

510(k) Number: K063586

Device name: moorFLP1 Full-Field Laser Perfusion Imager

Indications for use:

The moorFLPI Full-Field Laser Perfusion Imager is intended for blood flow measurements in the microcirculation. This device is intended for clinical research use.

Prescription Use: Yes
(Part 21 CFR 801 SubpartD)

AND/OR

Over-The-Counter Use: No (21 CFR 807 Subpart C)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

K063586

510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K063586](https://fda.innolitics.com/device/K063586)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
