← Product Code [MNS](/productcode/MNS) · K063533

# BIPAP SYNCHRONY (K063533)

_Respironics, Inc. · MNS · Feb 20, 2007 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K063533

## Device Facts

- **Applicant:** Respironics, Inc.
- **Product Code:** [MNS](/productcode/MNS.md)
- **Decision Date:** Feb 20, 2007
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 868.5895
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Respironics BiPAP Synchrony 2 is intended to provide non-invasive ventilation for pediatric patients 7 years or older (> 40 lbs) and adult patients (> 66 Ibs) with respiratory insufficiency or obstructive sleep apnea, in the hospital or home.

## Device Story

Microprocessor-controlled, blower-based bi-level positive pressure system; delivers IPAP/EPAP levels to provide pressure support. Inputs: flow and pressure data from patient airway sensors. Processing: microprocessor regulates blower assembly based on sensor feedback; supports CPAP and four bilevel modes (Spontaneous, Spontaneous/Timed, Timed, Pressure Control). Outputs: therapeutic airflow, clinical data display, and alarm notifications. Used in hospital or home; operated by clinicians or patients. Healthcare providers use displayed data to adjust clinical parameters. Benefits: improved patient comfort via dual pressure levels; non-invasive respiratory support.

## Clinical Evidence

Bench testing only. No clinical data. Testing verified pressure performance, waveform performance, triggering, cycling, and alarm functionality against predicate VPAP III ST-A (K060105) to ensure safety and efficacy for pediatric patients.

## Technological Characteristics

Microprocessor-controlled blower system; dual-level (IPAP/EPAP) positive pressure; flow and redundant pressure sensors; 22mm smooth lumen tubing; user interface for parameter adjustment; fixed and adjustable alarms. Non-invasive ventilation.

## Regulatory Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

## Predicate Devices

- Respironics BiPAP Synchrony HC ([K992530](/device/K992530.md))
- Respironics BiPAP Synchrony S/T ([K012323](/device/K012323.md))
- Respironics BiPAP Harmony ([K031656](/device/K031656.md))
- Resmed VPAP III ST-A / Kidsta system ([K060105](/device/K060105.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Abbreviated 510(k) Tab 5 – 510(k) summary

4063533

BiPAP Synchrony 2

TAB 5

# 510(K) SU

| Date of Submission       | 20 November 2006                                                                                                                                                                                        |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact         | Zita A. Yurko<br>Manager, Regulatory Affairs<br>Respironics, Inc.<br>1001 Murry Ridge Lane<br>Murrysville, PA 15668<br>Zita.yurko@respironics.com<br>724-387-4120 t<br>724-882-4120 c<br>724-387-4216 f |
| Classification Reference | 21 CFR 868.5895                                                                                                                                                                                         |
| Product Code             | MNS - Non-Continuous ventilator                                                                                                                                                                         |
| Common/Usual Name        | Ventilator, continuous, non-life supporting                                                                                                                                                             |
| Proprietary Name         | Respironics BiPAP Synchrony 2 Ventilatory Support System                                                                                                                                                |
| Predicate Device(s)      | Respironics BiPAP Synchrony HC (K992530)<br>Respironics BiPAP Synchrony S/T (K012323)<br>Respironics BiPAP Harmony (K031656)<br>Resmed VPAP III ST-A / Kidsta system (K060105)                          |
| Reason for submission    | Include pediatric (7 years and older > 40 lbs) indication                                                                                                                                               |

## Substantial Equivalence

The BiPAP Synchrony 2/ Kidsta system has the following similarities to the previously cleared predicate device:

- ロ Same intended use.
- Same operating principle.
- Same technology.

{1}------------------------------------------------

Tab 5 – 510(k) summary

D Same manufacturing process.

The BiPAP Synchrony was cleared in K992530/K012323. The BIPAP Harmony was cleared in K031656 using the BiPAP Synchrony as the device predicate. The Mirage Kidsta was cleared in K060105. Respironics has performed a risk analysis to identify the additional considerations of using the BiPAP Synchrony 2 / Kidsta system on pediatric patients (> 7 years and > 40 lbs). Additional testing based on the BiPAP Synchrony 2 / Kidsta risk analysis was performed to ensure the safety and efficacy of the BiPAP Synchrony 2/ Kidsta system when used on the pediatric patient (> 7 years and > 40 lbs). Bench testing has confirmed that the BiPAP Synchrony 2 performs equivalently to the device predicate VPAP III ST-A (K060105). Additionally bench testing was performed to demonstrate compatibility of the Kidsta mask (K060105) with the BiPAP Synchrony. This testing included, pressure performance, waveform performance, triggering, cycling and alarm functionality testing. All tests were verified to meet the required acceptance criteria.

Respironics has followed the FDA's Guidance for Industry and FDA Staff document "pre-market assessment of pediatric medical devices" and applied the principle of FDA's Least Burdensome Approach to demonstrate the Substantial Equivalence of the BiPAP Synchrony 2/ Kidsta mask system to its predicate devices (Respironics BiPAP Synchrony, BiPAP Harmony and the VPAP ST-A/Kidsta mask system). We conclude that the existing and cleared adult indications for use can be safety and effectively applied to pediatric patients (> 7 years and > 40 lbs).

#### Intended Use

The Respironics BiPAP Synchrony 2 is intended to provide non-invasive ventilation for pediatric patients 7 years or older (> 40 lbs) with respiratory insufficiency or obstructive sleep apnea, in the hospital or home.

### Device Description

The Respironics BiPAP Synchrony 2 Ventilatory Support System is a microprocessor controlled blower based Bi-level positive pressure system that delivers two positive pressure levels (IPAP/EPAP). The dual pressure levels provide a more natural means of delivering pressure support therapy to the patient resulting in improved patient comfort. A flow sensor and redundant pressure sensors in the patient airway feed data on measured flow and pressure into a microprocessor controller, which in turn regulates the blower assembly. A user interface displays clinical data and enables the operator to set and adjust certain clinical parameters.

{2}------------------------------------------------

The BiPAP Synchrony 2 also has a CPAP mode in which a fixed pressure is delivered and four bilevel operating modes which determine how the changes between IPAP and EPAP pressures are made, Spontaneous, Spontaneous/Timed, Timed and Pressure Control.

The BiPAP Synchrony 2 is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.

The BiPAP Synchrony 2 Ventilatory Support System is intended for use with a patient circuit that is used to connect the device to the patient interface device (mask). A typical patient circuit consists of a six-foot disposable or reusable smooth lumen 22mm tubing, a method of venting exhaled gases, and the Mirage Kidsta mask (K060105).

(End of Tab.)

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Respironics, Incorporated Ms. Zita A. Yurko Manager, Regulatory Affairs Sleep & Home Respiratory Group 1001 Murry Ridge Lane Murrysville, Pennsylvania 15668

FEB 2 0 2007

Re: K063533

Trade/Device Name: BiPAP Synchrony 2 Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: MNS Dated: November 20, 2006 Received: November 22, 2006

Dear Ms. Yurko:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

#### Page 2 - Ms. Yurko

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Chris

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Indications for Use

510(k) Number (if known): __ ______

Device Name: _________________________________________________________________________________________________________________________________________________________________

The Respironics BiPAP Synchrony 2 is intended to provide non-invasive ventilation for pediatric patients 7 years or older (> 40 lbs) and adult patients (> 66 Ibs) with respiratory insufficiency or obstructive sleep apnea, in the hospital or home.

 Prescription Use ____________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mh. A. Anker

sthesiology, General Hos

S0001 Number K063533

040019

---

**Source:** [https://fda.innolitics.com/device/K063533](https://fda.innolitics.com/device/K063533)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
