The Fugo Blade system is intended for incision, excision, ablation, vaporization and hemostasis of oral soft tissue e.g. excisional and incisional biopsies; exposure of unerupted teeth; fibroma removal; frenectomy and frenotomy; gingival troughing for crown impressions; gingiyoplasty: gingivectomy; gingival incision and excision; hemostasis; implant recovery; incision and drainage of abscess; leukoplakia; operculectomy; oral papillectomies; pulpotomy; pulpotomy as an adjunct to root canal therapy; reduction of gingival hypertrophy; soft tissue crown lengthening; secular debridement; treatment of aphthous ulcers; vestibuloplasty; biopsy incision and excision; and lesion (tumor) removal.
Device Story
Electrosurgical device powered by C-cell batteries; generates electromagnetic (EM) energy (9.8 +/- 2.0 x 10^6 Hz). Energy conditioned/tuned by proprietary network; focused on blunt cutting filament. Interaction with tissue creates plasma field; ionizes molecular bonds; causes micro-fragmentation of tissue lattice. Enables incision, excision, ablation, vaporization, and hemostasis. Used in dental clinics by surgeons. Surgeon activates via on/off switch. Output visualized as plasma cloud. Benefits include precise tissue cutting with minimal trauma and effective hemostasis.
Clinical Evidence
No new clinical data for dental application; relies on clinical evidence from ophthalmic predicates (K001498, K041019, K050933). Studies at University of South Carolina (lens capsule) and Louisiana State University (scleral tissue) demonstrated structural integrity, biomechanical strength, and resistance-free incisions with minimal trauma. Bench testing confirmed compliance with ANSI/AAMI HF18:2001 for thermal safety and high-frequency leakage current.
Technological Characteristics
Electrosurgical unit; 7.2V battery-powered; 9.8 +/- 2.0 MHz frequency. Proprietary electronic network conditions/tunes EM energy to blunt filament. Plasma formation mechanism. Complies with ANSI/AAMI HF18:2001 (thermal safety, leakage current). UL approved and CSA compliant.
Indications for Use
Indicated for oral soft tissue procedures including biopsies, gingivectomy, gingivoplasty, frenectomy, pulpotomy, lesion/tumor removal, and hemostasis in patients requiring dental surgery.
Regulatory Classification
Identification
A dental electrosurgical unit and accessories is an AC-powered device consisting of a controlled power source and a set of cutting and coagulating electrodes. This device is intended to cut or remove soft tissue or to control bleeding during surgical procedures in the oral cavity. An electrical current passes through the tip of the electrode into the tissue and, depending upon the operating mode selected, cuts through soft tissue or coagulates the tissue.
{0}------------------------------------------------
K063468
APR 1 & 2007
10001
Medisurg Ltd. 100 West Fornance St. Norristown, PA 19401 Phone (610) 277-3937 Fax (610) 277-7256
# Pre-market Notification [510(k)] Summary
Submitters name: Richard J Fugo M.D., Ph.D. Contact person: Richard J Fugo M.D., Ph.D.
Date: October 30, 2006
Names: (807.92(a)(2))
- B) Classification name: Unit, electrosurgical, and accessories, dental
- C) Common/ usual name: The Fugo Blade for Dentistry
- D) Proprietary name: The Fugo Blade
Equivalence predicate device: (807.92(a)(3))
- A) Equivalence from an intended use standpoint: FDA cleared
Wallach Quantum 500 electrosurgical generator for surgery -K000768
- B) Equivalence from a technologic standpoint:
- a. Fugo Blade for Capsulotomy FDA # K001498
- b. Fugo Blade for Glaucoma FDA # K041019
- c. Fugo Blade for Peripheral Iridotomy FDA # K050933
Description of the device (807.92(a)(4))
The Fugo Blade for dentistry is an electrosurgical device that is powered by the electromagnetic (EM) energy from flashlight size batteries. This energy is conditioned, tuned (9.8+/- 2.0 x 106 Hz) and focused on a thin blunt cutting filament. Moreover, the EM energy from flashlight size (C cell) batteries is fed into a proprietary electronic network system which is activated by the surgeon with an on/off switch. This electronic system is fed EM energy from the battery energy supply source, then conditions it in a proprietary fashion and transfers it to the surgeon cutting filament. This
{1}------------------------------------------------
column of EM energy is capable of reacting with tissue. This column of EM energy causes the molecular bonds in the tissue which come in contact with the EM field to ionize and thereby break apart, a condition that plasma physicists refer to as "plasma formation". In this way, the molecular lattice of the tissue is broken into micro-fragments, thereby creating an incision, excision, ablation, vaporization, and hemostasis of oral soft tissue. This process was evaluated extensively in the FDA 510K submission K050933: The Fugo Blade for Peripheral Iridotomy.
### Intended use of device (807.92(a)(5))
The Fugo Blade system is intended for incision, excision, ablation, vaporization and hemostasis of oral soft tissue e.g. excisional and incisional biopsies; exposure of unerupted teeth; fibroma removal; frenectomy and frenotomy; gingival troughing for crown impressions; gingiyoplasty: gingivectomy; gingival incision and excision; hemostasis; implant recovery; incision and drainage of abscess; leukoplakia; operculectomy; oral papillectomies; pulpotomy; pulpotomy as an adjunct to root canal therapy; reduction of gingival hypertrophy; soft tissue crown lengthening; secular debridement; treatment of aphthous ulcers; vestibuloplasty; biopsy incision and excision; and lesion (tumor) removal.
### Summary of technological characteristics (807.92(a)(6))
The Fugo Blade for Dentistry: Electromagnetic (EM) energy from flashlight size ("C" cell ) batteries is fed into a proprietary network system. This electronic system takes EM energy from the batteries then conditions, tunes (9.8 +/- 2.0 x 106 Hz) and focuses this EM energy around an incising probe at the end of the Fugo Blade electrosurgical hand piece. This column of EM energy is capable of reacting with tissue in an intended path of incision. This reaction causes the molecular bonds in the tissues which come in contact with the Fugo Blade to absorb the Fugo Blade EM energy and thereby have the molecular bonds shatter. Upon shattering, the molecules ionize and create a substance that physicists refer to as a "plasma field". In this way, the molecular lattice of the tissue is broken down, thereby creating an incision, excision, ablation, vaporization as well as hemostasis of oral soft tissue. The plasma field generation can easily be visualized under a microscope. Such images have been published by Medisurg Ltd. in scientific journals and advertisements. This Fugo Blade plasma cloud has been studied and analyzed by the Plasma Lab at Swarthmore College, Swarthmore, PA. The issue of Fugo Blade field generation was addressed and evaluated in the
{2}------------------------------------------------
FDA 510K submission # K050933 "The Fugo Blade for Peripheral Iridotomy".
### Non-clinical performance data (807.92(b)(1))
The Fugo Blade uses 7.2 V batteries to focus an electromagnetic field into tissue in a path of incision, thereby ionizing molecules in the tissue lattice thus creating a plasma field as discussed in FDA 510 submission K050933. This was substantiated by proprietary studies under the supervision of Dr Michael Brown, director of the Swarthmore college plasma lab, Swarthmore, PA. It is well know in the electrosurgical literature that all electrosurgical incision units including a Wallach Quantum 500 generate "arcs" which are plasma clouds at the incising tip.
The same electronic system as the Fugo Blade for Dentistry has previously been FDA cleared in FDA 510 K submissions K001498. K041019 and K050933. The Fugo Blade system has also received UL approval and is CSA compliant. The Fugo Blade system complies with ANSI/AAMI HF18:2001, clause 4.2.3.1 for thermal safety. The Fugo Blade system complies with high frequency leakage current requirements specified by ANSI/AAMI HF18:2001, clause 4.2.5.2. These findings support substantial equivalency with the Wallach Quantum 500 system.
Clinical performance data 807.92(b)(2)
FDA submission K001498 provides statistical significance for safety and efficacy for the Fugo Blade system in the incision/ ablation/ vaporization of highly delicate tissue, namely anterior human lens capsule. Lens capsule is less than 50 microns thick and is extremely fragile tissue. Yet, studies at the University of South Carolina show that the Fugo Blade incision wall of lens capsule had a structural integrity which was untraumatized and a biomechanical strength characteristic equal to or better than other electrosurgical units. (Appendix #8)
FDA submission 510K K041019 provides statistical significance for safety and efficacy in the incision/ablation/vaporization of extremely dense, strong tissue, namely scleral tissue of the eye. Also, studies at Louisiana State University demonstrated that the Fugo Blade can provide resistance free incisions with minimal trauma to the tissue adjacent to the incision. (Appendix #7)
FDA submission K 050933 provides statistical significance for safety and efficacy for incision/ablation/vaporization with hemostasis in highly vascular tissue, namely the vascular uveal layer of the eye.
{3}------------------------------------------------
K063468
4074
These studies support substantial equivalency in clinical performance of the Fugo Blade system to the Wallach Quantum 500 system for Dentistry (FDA 510K K000768) from an intended use perspective (intended use predicate). From a technologic perspective (technologic predicate), the Fugo Blade for Dentistry has been shown to be equivalent to the Fugo Blade for Capsulotomy (FDA 510K K001498), the Fugo Blade for Glaucoma (FDA 510K K041019), and the Fugo Blade for Peripheral Iridotomy (FDA 510K K050933).
#### Summary 807.92(b)(3)
C
The Fugo Blade for Dentistry is as safe and as effective as the predicate devices. From a technologic perspective (technologic predicate) the Fugo Blade for Dentistry is the same electronic system as FDA cleared systems K001498, K041019, K050933. Therefore, the Fugo Blade for Dentistry has been shown to be equivalent to the technologic predicates.
From an intended use perspective, the Fugo Blade for Dentistry has been shown to be equivalent in terms of safety and effectiveness to the Wallach Quantum 500 system for dentistry (FDA # K000768). These findings are based on preclinical data and clinical data discussed in this application. This data has also previously been described in FDA 510K application K001498, K041019, K050933.
Therefore, the Fugo Blade for Dentistry has been demonstrated to have equivalent safety and efficacy as its technologic predicate and its intended use predicate based on non-clinical and clinical tests.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal for the Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Richard J. Fugo M.D., Ph.D. MediSurg Limited 100 West Fornance Street Norristown, Pennsylvania 19401
APR 1 & 2007
Re: K063468
Trade/Device Name: The Fugo Blade for Dentistry Regulation Number: 21 CFR 872.4920 Regulation Name: Dental Electrosurgical Unit and Accessories Regulatory Class: II Product Code: EKZ Dated: March 19, 2007 Received: April 9, 2007
Dear Dr. Fugo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Dr. Fugo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Surfie Y. Michael Davis
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K063468
## Indications for Use
#### 510(k) Number (if known): K063468
The Fugo Blade for Dentistry Device Name:
Indications For Use:
- Excisional and incisional biopsy .
- Exposure of unerupted teeth .
- Fibroma removal
- Frenectomy and frenotomy .
- Gingival troughing for crown impressions .
- Leukoplakia .
- Operculectomy t
- Gingivoplasty (contouring of gingival tissue) .
- Gingivectomy (excision of unsupported gingival tissue) .
- Gingival incision and excision .
- Hemostatis ●
- Implant recovery .
- Incision and drainage of abscess .
- Oral papillectomies .
- Pulpotomy .
- Pulpotomy as an adjunct to root canal therapy .
- Reduction of gingival hypertrophy ●
- Soft tissue crown lengthening .
- Treatment of aphthous ulcers .
- Vestibuloplasty .
- Biopsy incision and excision .
- Lesion removal .
- Tumor removal �
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

| of Anesthesiology, General Hospital, |
|--------------------------------------|
| Infection Control, Dental Devices |
| 510(k) Number: | K063468 |
|----------------|---------|
|----------------|---------|
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.