K063448 · Emg Technology Co., Ltd. · JCX · Nov 30, 2006 · General, Plastic Surgery
Device Facts
Record ID
K063448
Device Name
EMG AC/DC SUCTION UNIT
Applicant
Emg Technology Co., Ltd.
Product Code
JCX · General, Plastic Surgery
Decision Date
Nov 30, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The device is a vacuum suction device. It is to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory. It is for use on the order of a physician only.
Device Story
Compact medical suction unit; powered by AC adapter, internal rechargeable battery, or automobile 14V DC. Vacuum generated by internal pump (DC motor, cylinder, piston, valves). User interface includes power switch, vacuum adjusting knob, and vacuum meter. Device connects via PVC tubing to bacteria filter, collection container, and patient-applied drainage tubing. Used to suction sputum or body fluids. Operated by healthcare professionals or patients under physician order in home or transport settings. Provides continuous vacuum pressure; vacuum level adjustable via knob with +/-5% accuracy. Benefits include effective fluid/infectious material removal from wounds or airways.
Clinical Evidence
Bench testing only. Device compliance verified against ISO10079-1, IEC60601-1-1, and IEC60601-1-2 standards. No clinical data presented.
Technological Characteristics
Powered suction pump; DC motor-driven piston pump. Materials: PVC compound (ISO10993 compliant). Dimensions: compact. Connectivity: AC/DC power. Standards: UL60601-1, CSA C22.2 No.601.1-M90, IEC60601-1-2, ISO10079-1:1991. Control: Manual vacuum adjustment knob with analog meter display. Software: None (mechanical control).
Indications for Use
Indicated for removal of infectious materials from wounds or fluids from airway/respiratory system. Prescription use only.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Kob3448
Page 1 of 3
# NOV 3 U 2006
#### Exihibit #2
#### 510(K)Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of CFR 807.92 1
The assigned 510(k) number is :_
### 1. Submitter's identifications :
EMG Technology Co. , LTD 4F-2, No.210 , 38RD., Taichung Industrial Park, Shituen Chiu, Taichung, Taiwan , R.O.C. TEL : 886-4-23596033
FAX : 886-4-23596031
#### 510(k) owner's name :
Mr. Roger Huang General Manager
### Contact person :
Sally Chuang Email: Sally@emgtech.com.tw
#### 2. Name of the Device :
Trade Name Including Model Number of Device : EMG AC/DC Suction Unit , Model SUF01-XXX Series
Common Name or Classification Name : Powered Suction Pump
## 3. Information of the 510(k) Cleared Device(Predicate Device) : DeVilbiss Suction Unit , Model 7305P-I 510(k) Number : K#982304
### 4. Device description :
The EMG AC/DC Suction unit is a compact medical suctioning device which can be powered by the AC current through a attached AC Adapter unit connecting to the wall electric outlet at home or DC power source through a set of built-in
Page 1 of 3
{1}------------------------------------------------
Ko63448
rechargeable battery or automobile DC 14V electric power through power cord connecting to the receptacle during transportation.
The vacuum pressure of EMG AC/DC Suction unit is to be produced through a built-in vacuum pump which is consisted of DC motor, cylinder, piston connecting rod, cup seal, eccentric crank, inlet valve, outlet valve---etc parts. When turn on the suction unit, motor start to run, its shaft drive the eccentric crank which actuate the piston connecting rod moving up and down in the cylinder. When the piston is being in the down stroke, a vacuum pressure is produced in the cylinder, then air will be suctioned into the cylinder through an one way inlet port, when the piston is being in the up stroke, it pressed out these suctioned air through another one way outlet port to the atmosphere, these two one way port avoid air being suctioned from the atmosphere as well as air in cylinder being pressed back to the inlet port. When motor keep running, then a continuous vacuum pressure source through inlet port is produced.
One set of plastic cover enclose the vacuum pump, control p. c. board, battery and wire harness to protect user from electrical shock and mechanical hurt hazard. Following components on the cover are to be the operating interface to the use: a power switch, a receptacle, a vacuum adjusting set, vacuum meter and device vacuum inlet port. The vacuum adjusting set through inner tubing connecting to the inlet port of vacuum pump, the vacuum meter and the device vacuum inlet port are connecting to the vacuum adjusting set through a three way connector and inner tubing. Device's vacuum pressure can be adjusted by tuning the knob of vacuum adjusting set, accuracy within +/-5%. The vacuum meter display the vacuum pressure data during operation, the device vacuum inlet port provide the vacuum pressure source.
Connect the device vacuum inlet port, bacteria filter (Bacteria filter pore size: 0.2microns,Model:3928) and collection container with two PVC tubing respectively. then connect one end of a biocompatible patient tubing (comply with ISO10993 / material: PVC compound, Type: 3MSA060p3X) to the collection container inlet port, the other end of the patient tubing connected to the applied part (like drainage tubing). Put the applied part to the adequate location of patient body, then this device can start to carry out the suctioning sputum and other body liquid's operation.
{2}------------------------------------------------
K063448
Page 3 of 3
The device is designed and manufactured to comply with UL60601-1 , CSA C22.2 No.601.1-M90 and IEC60601-1-2 EMC safety standards and meet the performance standard per ISO10079-1:1991
#### 5. Intended Use :
The device is a vacuum suction device. It is to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory. lt is for use on the order of a physician only, and has the same intended use, with the predicate Device DeVilbiss Suction Unit, Model 7305P-1.
## 6. The technological characteristics comparison to Predicate Devices :
The fundamental technological characteristics of the EMG AC/DC Suction Unit same as the predicate device DeVilbiss Suction Unit , Model 7305P-1 For a complete technological characteristics comparison please refer Exhibit #5 .
7. Discussion of Non-Clinical Tests performed for Determination of Substantial Equivalence are follows :
The EMG AC/DC Suction Unit Compliance to the applicable voluntary standards include ISO10079-1 , IEC60601-1-1 and IEC60601-1-2 requirements .
It is same as the predicate device 510(k) summary statement .
#### 8. Conclusion :
In terms of The EMG AC/DC Suction Unit have the same construction , function , safety and effectiveness and intended use as the 510(k) predicate device . The EMG AC/DC Suction Unit is substantially equivalent to other legally market suction pumps used for this application .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and head. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
EMG Technology Co., Ltd. % Underwriters Laboratories, Inc. Mr. Marc M. Mouser Senior Project Engineer, Reviewer 2600 NW Lake Road Camas, Washington 98607
NOV 3 U 2006
Re: K063448
Trade/Device Name: EMG AD/DC Suction Unit Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: II Product Code: JCX Dated: October 26, 2006 Received: November 15, 2006
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or re and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, TDA movie can publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Marc M. Mouser
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to 1egally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You mon ohtain other general information on your responsibilities under the Act from the Division of Soull Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/suport/index20tml.
Sincerely vor
Por Plothn
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
063448
## Indications for Use
510(k) Number (if known):
Device Name: EMG AC/DC Suction Unit
Indications For Use:
The device is a vacuum suction device. It is to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory. It is for use on the order of a physician only.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Psts mo
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page 1 of 1
**510(k) Number** K063448
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.