← Product Code [HFX](/productcode/HFX) · K063429

# GMD UNIVERSAL CIRCUMCISION CLAMP, MODEL 200 (K063429)

_Generic Medical Device, Inc. · HFX · Jan 17, 2007 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K063429

## Device Facts

- **Applicant:** Generic Medical Device, Inc.
- **Product Code:** [HFX](/productcode/HFX.md)
- **Decision Date:** Jan 17, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.4530
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The GMD Circumcision Clamp is intended to be used in the medical procedure of circumferential excision of the foreskin or prepuce skin at or near the level of coronal sulcus, with minimal amount of preputial skin remaining.

## Device Story

GMD Universal Circumcision Clamp is a manual, reusable stainless steel surgical instrument. Used by clinicians during circumcision procedures to perform circumferential excision of the foreskin or prepuce at or near the coronal sulcus. Operates by crushing the prepuce parallel to the penile axis to facilitate excision. Device is sold non-sterile and requires sterilization before use. Provides a mechanical means to assist the surgeon in achieving circumcision.

## Clinical Evidence

No clinical data or non-clinical laboratory studies were conducted; substantial equivalence is based on the long history of clinical use for this type of technology.

## Technological Characteristics

Manual, reusable surgical instrument constructed of stainless steel. Operates via mechanical crushing of tissue. No energy source, software, or connectivity. Sold non-sterile.

## Regulatory Identification

An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.

## Special Controls

*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.

## Predicate Devices

- GOMCO Circumcision Clamp ([K894201](/device/K894201.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(K) Summary

### Introduction

This 510(k) summary documentation is intended to comply with requirements of 21 CFR § 807.92. FDA may make this summary available to the public within 30 days following a finding of substantial equivalence.

#### 510K Submitted By

Generic Medical Device, Inc. 2201-34th Ave NW Gig Harbor, WA 98335

#### USA Contact Person

Monica R. Montanez, RAC, CQA Consultant Tel: 713-523-9630 / Fax: 713-521-2770

#### Date Prepared

November 5, 2006

#### Trade Name of Device

GMD Universal Circumcision Clamp

#### Proprietary Name:

GMD Universal Circumcision Clamp

#### Common Name:

Clamp, Circumcision

#### Classification Name:

Clamps, Circumcision Class II: Obstetric-gynecologic specialized manual instrument 21 CFR 884.4530, Product Code HFX.

#### Predicate Devices:

K894201 - GOMCO Circumcision Clamp Manufactured by: Zinnanti Surgical Instruments 21540-B Prairie Street Chatsworth, California 91311

510(k) Classification Class II

Device Description

JAN / 7 2007

{1}------------------------------------------------

The GMD Circumcision Clamp is a stainless steel reusable medical device intended to be used in circumcision procedures. The GMD Circumcision clamp will be sold non-Sterile.

#### Intended Use

The GMD Circumcision Clamp is intended to be used in the medical procedure of circumferential excision of the foreskin or prepuce skin at or near the level of coronal sulcus, with minimal amount of preputial skin remaining.

### Comparison to Predicate Devices

GMD Universal Circumcision Clamp has the same intended use, same technological characteristics as the predicate device (GOMCO Circumcision Clamp- K894201). The GMD Universal Circumcision Clamp and the GOMCO device are substantially equivalent in their method of circumcision as they both crush the prepuce parallel to the axis of the penis. They are also equivalent in intended use as they are both are used for circumferencial crushing of the prepuce to achieve circumcision.

#### Clinical/Non-Clinical Studies

The company did not conduct, nor depend on, clinical studies or non-clinical laboratory studies in order to establish substantial equivalence as this type of technology and procedure has a long history of clinical use.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular fashion around the caduceus. The logo is black and white.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Generic Medical Device Company c/o Ms. Monica Montanez Consultant Montanez and Associates 3906 Roseland St. HOUSTON TX 77006

## JAN 1 7 2007

Re: K063429

Trade/Device Name: GMD Universal Circumcision Clamp Regulation Number: 21 CFR §884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: HFX Dated: November 11, 2006 Received: November 13, 2006

#### Dear Ms. Montanez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image is a circular logo for the FDA Centennial, celebrating 100 years from 1906-2006. The logo features the letters "FDA" in a bold, stylized font above the word "Centennial" in a cursive script. Three stars are placed below the word "Centennial". The logo is surrounded by text that follows the curve of the circle.

Protesting and Promoting Public Health.

{3}------------------------------------------------

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I least be advised that I Dris lessant that your device complies with other requirements of the Act that I Dri has made a and regulations administered by other Federal agencies. You must or any I odolar battler and streegirements, including, but not limited to: registration and listing (21 Comply with an the Hot STEQUES Part 801); good manufacturing practice requirements as set CI K Part 6077, adoling (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and w you to substantial equivalence of your device of your device to a legally premarker notinodion: "The states in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use GMD UNIVERSAL CIRCUMCISION CLAMP

K063429 510(k) Number (if known):

Device Name: GMD Universal Circumcision Clamp

Indications For Use: The GMD Circumcision Clamp is intended to be used in the medical procedure of circumferential excision of the foreskin or prepuce skin at or near the level of coronal sulcus, with minimal amount of preputial skin remaining.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C Brogdon

(Division Sign Off)
Division
and
510(k) Number K063429

---

**Source:** [https://fda.innolitics.com/device/K063429](https://fda.innolitics.com/device/K063429)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
