K063427 · Lynton Lasers Limited · GEX · Jul 25, 2007 · General, Plastic Surgery
Device Facts
Record ID
K063427
Device Name
LUMINA INTENSE PULSED LIGHT (& LASER) SYSTEM
Applicant
Lynton Lasers Limited
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 25, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LUMINA Intense Pulsed Light (& Laser) System is intended for use in applications requiring the selective ablation, vaporization and photothermolysis (photocoagulation or coagulation) of soft tissue in the medical specialities of general & plastic surgery and dermatology.
Device Story
LUMINA Intense Pulsed Light (& Laser) System is a floor-standing platform for dermatological and surgical procedures. It comprises a main cabinet with central control electronics, a touchscreen interface, and interchangeable IPL handpieces or an internal Nd:YAG laser accessory. The system utilizes selective photothermolysis to target soft tissue, melanin, or vascular structures. The device is operated by clinicians in medical settings. The control unit manages timing and dosing parameters and includes a water-cooling system for handpieces. Clinicians use the touchscreen to configure treatment parameters based on the specific lesion or hair removal application. The output is delivered via light pulses to the patient's skin, affecting clinical decision-making by enabling non-invasive or minimally invasive tissue treatment. Benefits include the clearance of vascular/pigmented lesions, tattoo removal, and hair reduction.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, compliance with international standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1), and comparison of technical specifications to the predicate device.
Technological Characteristics
Floor-standing cabinet with touchscreen control. Energy sources: Intense Pulsed Light (IPL) and Nd:YAG laser (1064nm and 532nm). Features water-cooling system for handpieces. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser safety), and IEC 60825-1 (laser radiation safety).
Indications for Use
Indicated for patients with Fitzpatrick skin types I-IV. IPL handpieces: treatment of benign pigmented epidermal lesions (dyschromia, hyperpigmentation, ephelides); benign cutaneous vascular lesions (port wine stains, hemangiomas, telangiectasias, rosacea, angiomas, poikiloderma of Civatte, leg veins, venous malformations); and stable long-term/permanent hair removal. Nd:YAG 1064nm: treatment of benign vascular lesions (port wine stains, hemangiomas, telangiectasias, reticular leg veins, rosacea, venous lake, leg/spider veins, poikiloderma of Civatte, angiomas); removal of blue/black tattoos; non-ablative facial wrinkle treatment; hair removal/reduction; and pseudofolliculitis barbae. Nd:YAG 532nm: removal of light blue, yellow, and red tattoos; treatment of benign pigmented lesions (café au lait macules, lentigines, freckles, chloasma, nevi, nevus spillus, nevus of Ota, Becker's nevi).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Lovely System (Lovely II, or "Harmony") (K033946, K042000)
Submission Summary (Full Text)
{1}------------------------------------------------
K063427
## Appendix A11 The 510(k) Summary
. .
| Applicant & Submitter : | Lynton Lasers Limited | | JUL 25 2007 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-------------|
| Address : | Lynton House, Manor Lane,<br>Holmes Chapel, Cheshire, CW4 8AF, UK | | |
| Telephone : | +44 (0)1477 536977 | | |
| Fax : | +44 (0)1477 536978 | | |
| Contact Person : | Dr. Andrew J Berry<br>AJBerry@lynton.co.uk | | |
| Preparation Date : | 9th November 2006 | | |
| Device Submitted : | LUMINA Intense Pulsed Light (& Laser)<br>System | | |
| Common Name : | Light Therapy Device | | |
| Classification Name : | Laser surgical instrument for use in general<br>and plastic surgery and in dermatology. | | |
| Product Code : | GEX | | |
| Predicate Device : | Lovely System (Lovely II, or "Harmony"),<br>manufactured by MSq(M2) Ltd.<br>7 Haeshel St. P.O.B 3021<br>Caesarea Industrial Park, 38900 Israel<br>(K033946, K042000) | | |
| Device Description : | The LUMINA Intense Pulsed Light (& Laser)<br>System is a platform for a range of Intense<br>Pulse Light (IPL) Handpieces and an internal<br>Nd:YAG Laser Accessory for a number of<br>applications in general and plastic surgery and<br>dermatology, including :<br>Hair RemovalVascular Lesion treatmentsPigmented Lesion treatmentsTattoo removalFacial wrinkle reductionThe System includes : | | |
{2}------------------------------------------------
- . A floor-standing main cabinet unit (including the central control electronics) that controls timing and dosing parameters. This unit also contains a water-cooling system used to remove heat from the IPL Handpieces and the internal Nd:YAG Laser Accessory (if fitted).
- A control and display panel including a ● touchscreen.
- A range of optional IPL handpieces that . include the flashlamp light source, electrical and cooling water connections.
- The LUMINA Intense Pulsed Light (& Laser) Intended Use : System is intended for use in applications requiring the selective ablation, vaporization and photothermolysis (photocoagulation or coagulation) of soft tissue in the medical specialities of general & plastic surgery and dermatology.
- The LUMINA Intense Pulsed Light (& Laser) Indications for Use : stem, when using the IPL Handpieces, is indicated for :
- The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation and ephelides (freckles).
- The treatment of benign cutaneous . vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations.
- The removal of unwanted hair and to effect . stable long-term or permanent hair removal.
- Indicated for use on all skin types (Fitzpatrick I-IV)
The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 1064nm (Long Pulsed and Q-Switched), is indicated for treatment and clearance of :
{3}------------------------------------------------
- . Benign vascular lesions such as, but not limited to treatment of :
- > Port wine stains
- > Hemangiomas
> Superficial and deep telangiectasias (venulectasias)
> Reticular veins (0.1 - 4.0mm diameter) of the leg
- > Rosacea
- > Venus Lake
- > Leg veins
- > Spider veins
- > Poikiloderma of Civatte
- > Angiomas
- . The removal of blue or black tattoos (significant reduction in the intensity of black and/or blue/black tattoos)
- The non-ablative treatment of facial ● wrinkles, such as, but not limited to: > Periocular wrinkles
- > Perioral wrinkles
- The removal of unwanted hair, for stable ● long-term or permanent hair reduction through selective targeting of melanin in
- hair follicles. Removal or lightening of unwanted hair . (with and without adjuvant preparation).
- Pseudofolliculitis barbae (PFB) .
- Indicated for use on all skin types . (Fitzpatrick I-IV)
The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 532nm (Long Pulsed and Q-Switched), is indicated for :
- Tattoo removal :
- > Light blue
- > Yellow
- > Red
- Benign pigmented lesions such as, but not . limited to :
- > Café au lait macules
- > Lentigines (senile and solar)
- > Freckles (ephelides)
{4}------------------------------------------------
- > Chloasma
- > Nevi
- > Nevus spillus
- > Nevus of Ota
- > Becker's nevi
- The LUMINA Intense Pulsed Light (& Laser) Performance Data : System complies with the European Medical Devices Directive 93/42/EEC (Annex II) and the standards IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22 and IEC 60825-1
- Substantial Equivalence : The LUMINA Intense Pulsed Light (& Laser) System is substantially equivalent in terms of technological characteristics, performance, intended use, indications for use and operator interface to the previously cleared Lovely System (Lovely II, or "Harmony" System), cleared under K033946, K042000. There are no unique applications, indications, materials or specifications presented. Evidence of equivalence has been demonstrated through :
- · The LUMINA Intense Pulsed Light (& Laser) System's intended use and indications for use were previously cleared by the FDA for the predicate device.
- The technical characteristics of The LUMINA Intense Pulsed Light (& Laser) System are similar to those of the predicate device.
- Laser output values of the LUMINA Intense ● Pulsed Light (& Laser) System are similar to those of the predicate device (and other FDA cleared devices on the market)
- The predicate device, and other previously ● cleared devices with similar output specifications, have a proven safety and effectiveness in the treatment of the claimed indications.
- Safety and performance testing. .
Therefore, the LUMINA Intense Pulsed Light (& Laser) System is substantially equivalent to the predicate device cited above and raises no new safety and/or effectiveness issues.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three abstract, overlapping shapes that resemble human profiles or waves, creating a sense of depth and movement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 5 2007
Lynton Lasers Ltd. % Dr. Andrew J. Berry Technical Director Lynton House, Manor Lane Holmes Chapel, Cheshire CW4 8AF United Kingdom
Re: K063427
Trade/Device Name: LUMINA Intense Pulsed Light (& Laser) System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgerv and in dermatology Regulatory Class: II Product Code: GEX Dated: June 21, 2007 Received: June 25, 2007
Dear Dr. Berry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Dr. Andrew J. Berry
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permitte your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 6338-241 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/suport/index20th11.
Sincerely yours,
Mark A. Millikusen
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known) : (063427
LUMINA Intense Pulsed Light (& Laser) System Device Name :
Indications for Use :
The LUMINA Intense Pulsed Light (& Laser) System is intended for use in applications requiring the selective ablation. vaporization and photothermolysis (photocoagulation or coagulation) of soft tissue in the medical specialities of general & plastic surgery and dermatology, as follows :
The LUMINA Intense Pulsed Light (& Laser) System, when using the IPL Handpieces, is indicated for :
- . The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation and ephelides (freckles).
- The treatment of benign cutaneous vascular lesions including port wine . stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations.
- The removal of unwanted hair and to effect stable long-term or . permanent hair removal.
- Indicated for use on all skin types (Fitzpatrick I-IV) .
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Mark N. Milliken
Division of General, Restorative, and Neurological Devices
510(k) Number K064327
{8}------------------------------------------------
510(k) Number (if known) :K(063427
Device Name :
Indications for Use : - Continued from previous page -
The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 1064nm (Long Pulsed and Q-Switched), is indicated for treatment and clearance of :
- Benign vascular lesions such as, but not limited to treatment of :
- > Port wine stains
- > Hemangiomas
- > Superficial and deep telangiectasias (venulectasias)
- > Reticular veins (0.1 4.0mm diameter) of the leg
- > Rosacea
- > Venus Lake
- > Leg veins
- > Spider veins
- > Poikiloderma of Civatte
- > Angiomas
- The removal of blue or black tattoos (significant reduction in the . intensity of black and/or blue/black tattoos)
Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{9}------------------------------------------------
510(k) Number (if known) : (063427
Device Name :
Indications for Use : - Continued from previous page -
The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 1064nm (Long Pulsed and Q-Switched), is indicated for treatment and clearance of - continued - :
- . The non-ablative treatment of facial wrinkles, such as, but not limited to:
- > Periocular wrinkles
- > Perioral wrinkles
- The removal of unwanted hair, for stable long-term or permanent hair . reduction through selective targeting of melanin in hair follicles.
- Removal or lightening of unwanted hair (with and without adjuvant preparation).
- Pseudofolliculitis barbae (PFB) ◆
- Indicated for use on all skin types (Fitzpatrick I-IV) ●
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{10}------------------------------------------------
510(k) Number (if known) :)(063427
Device Name :
Indications for Use : - Continued from previous page -
The LUMINA Intense Pulsed Light (& Laser) System, when using the internal Nd:YAG Laser Accessory at 532nm (Long Pulsed and Q-Switched), is indicated for :
- Tattoo removal : ●
- > Light blue = = =
- > Yellow
- > Red
- Benign pigmented lesions such as, but not limited to : �
- > Café au lait macules
- > Lentigines (senile and solar)
- > Freckles (ephelides)
- > Chloasma
- > Nevi
- > Nevus spillus
- > Nevus of Ota
- > Becker's nevi
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.