K063422 · Innovative Bioceramix, Inc. · KIF · Nov 21, 2006 · Dental
Device Facts
Record ID
K063422
Device Name
BIOAGGREGATE
Applicant
Innovative Bioceramix, Inc.
Product Code
KIF · Dental
Decision Date
Nov 21, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
• Repair of Root Perforations • Repair of Root Resorption • Root End Filling • Apexification • Pulp capping
Device Story
BioAggregate is a hydraulic, ceramic-based powder mixed with a liquid to form a cement for root canal repair. Used by dental professionals in clinical settings to create a hermetic seal within the root canal. The material promotes cementogenesis. It serves as an aluminum-free alternative to traditional mineral trioxide aggregate (MTA) materials, incorporating calcium phosphate to enhance safety and effectiveness. The device is applied directly by the clinician to the site of perforation, resorption, or root end, or used for apexification and pulp capping. It benefits patients by providing a biocompatible, non-cytotoxic, and non-mutagenic material for endodontic repair, ensuring tissue tolerance and effective sealing of the root canal system.
Clinical Evidence
Bench testing only. Tests included working time, setting time, solubility, and radiopacity. Biocompatibility testing confirmed the material is non-mutagenic, non-cytotoxic, lacks allergenic potential, and is well-tolerated by subcutaneous tissues.
Technological Characteristics
Hydraulic cement-like powder composed of biocompatible ceramic particles; aluminum-free; contains calcium phosphate. Physical state: powder-liquid mix. Properties: radiopaque, non-mutagenic, non-cytotoxic. No specific ASTM standards cited.
Indications for Use
Indicated for dental patients requiring root canal repair procedures, including root perforation repair, root resorption repair, root end filling, apexification, and pulp capping.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
Predicate Devices
Dentsply International WHITE MTA MATERIAL (K011009)
NovaBone Products, LLC PerioGlas - Bioglass Bone Graft Particulate (K040278)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo and contact information for Innovative BioCeramix Inc. The logo includes the company name in bold, uppercase letters, followed by the letters "IBC" inside of a circle. The address is listed as 3650 Wesbrook Mall, Vancouver, BC V6S 2L2 Canada. The phone number is (604) 221 6800, and the fax number is (604) 677 6129.
# 510(k) SUMMARY
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92
| SUBMITTER: | Innovative BioCeramix Inc.<br>3650 Wesbrook Mall<br>Vancouver, BC V6S 2L2 Canada<br>Tel: 604-221-6800<br>Fax: 604-677-6129 | NOV 2 1 2006 |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| CONTACT: | Dr. Quanzu Yang | |
| SUMMARY PREPARED: | Jul 25, 2006 | |
| TRADE NAME: | BioAggregate | |
| COMMON NAME: | Root Canal Repair Filling Material | |
| CLASSIFICATION NAME: | Root Canal Filling Resin (21 CFR 872.3820, Product Code: KIF) | |
| PREDICATE DEVICE: | Dentsply International WHITE MTA MATERIAL (K011009) | |
| | For specific chemical compositions:<br>• Biomet, Inc. Tantalum Beads - Radiographic Marker (K010348)<br>• Synthes (USA) chronOSTM(K041350)<br>• NovaBone Products, LLC PerioGlas - Bioglass Bone Graft Particulate<br>(K040278) | |
| DEVICE DESCRIPTION: | BioAggregate is a white hydraulic cement-like powder composed of<br>biocompatible ceramic particles. The BioAggregate Powder promotes<br>cementogenesis upon mixing with BioA Liquid and forms a hermetic seal<br>inside the root canal. | |
| INTENDED USE: | • Repair of Root Perforations<br>• Repair of Root Resorption<br>• Root End Filling<br>• Apexification<br>• Pulp capping | |
| TECHNOLOGICAL<br>CHARACTERISTICS: | BioAggregate is equivalent to WHITE MTA MATERIAL,<br>in terms of the physical state (powder-liquid mix), working time, setting<br>time, solubility and radiopacity. Similar major chemical compounds are<br>utilized in the production of both BioAggregate and WHITE MTA<br>MATERIAL. Both materials are non-mutagenic and non-cytotoxic, and are<br>designated for the same indications for use. The primary difference between | |
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Innovative BioCeramix Inc. The logo includes the company name in bold, uppercase letters. Below the company name is the address, 3650 Wesbrook Mall.
Vancouver BC Tel. (604) 221 6800 Fax: (604) 677 6129
the two materials is that BioAggregate has an aluminum-free composition, which contains chemical compounds that improve BioAggregate's properties in terms of safety and effectiveness, and includes calcium phosphate.
The additional predicate devices include: Tantalum Beads - Radiographic Marker, Synthes (USA) chronOS™ and PerioGlas - Bioglass Bone Graft Particulate each containing specific chemical compounds that are equivalent to BioAggregate, providing evidence that these chemical compounds are safe and effective for medical devices use.
NON-CLINICAL TESTS BioAggregate has undergone extensive bench and biocompatibility testing PERFORMED: to provide evidences that BioAggregate's physico-chemical properties are substantially equivalent to WHITE MTA MATERIAL. Bench tests included Working Time, Setting Time, Solubility and Radiopacity. Biocompatibility tests results determined that BioAggregate is nonmutagenic, non-cytotoxic, does not cause allergenic potential to tissues after multiple uses and has good tolerance by subcutaneous tissues.
CONCLUSIONS: BioAggregate has the same indications for use, includes similar materials and comparable technological characteristics as WHITE MTA MATERIAL. In addition, the bench and biocompatibility testing performed demonstrates that any differences in their technological characteristics do not raise any new questions as to safety and effectiveness. The chemical compounds of BioAggregate that were not equivalent to MTA were found to be safe and effective in Tantalum Beads - Radiographic Marker, Synthes (USA) chronOSTM and PerioGlas - Bioglass Bone Graft Particulate. Therefore, it is concluded that BioAggregate is safe, effective and substantially equivalent to the predicate devices.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 1 2006
Innovative Bioceramix Incorporated C/O Mr. Neil E. Devine, Jr. Responsible Third Party Official Intertek Testing Services NA, Incorporated 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087
Re: K063422
Trade/Device Name: BioAggregate Regulation Number: 21 CFR 872.3820 Regulation Name: Regulatory Class: II Product Code: KIF Dated: November 10, 2006 Received: November 13, 2006
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Mr. Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Senitte y. Michieu Davis
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
Koballing 510(k) Number (if Known): Device Name: BioAggregate Indications for Use: Repair of Root Perforation . Repair of Root Resorption . Root End Filling . Apexification . Pulp Capping .
ん Prescription Use _ AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Rump
Page 1 of 1
3.01-4.10) n of Anesthesiology, General Hospital, con Control, Dental Devices
0634122
5-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.