MICROSCAN SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANEL WITH SYNERCID (0.12-2UG/ML 5DILUTION BREAKPOINT SEQUENCE
K063366 · Dade Behring, Inc. · LON · Dec 8, 2006 · Microbiology
Device Facts
Record ID
K063366
Device Name
MICROSCAN SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANEL WITH SYNERCID (0.12-2UG/ML 5DILUTION BREAKPOINT SEQUENCE
Applicant
Dade Behring, Inc.
Product Code
LON · Microbiology
Decision Date
Dec 8, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Intended Use
The MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive enterococci and staphylococci. After inoculation, panels are incubated for 4.5 -- 18 hours at 35℃ +/- 1℃, in a WalkAway® SI, or equivalent, and read by the MicroScan® Instrumentation. Additionally, the panels may be incubated in a non-CO2 incubator and the Antimicrobial Susceptibility Testing (AST) portions can be read visually, according to the Package Insert.
Device Story
MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels utilize growth-based detection to determine antimicrobial susceptibility. Input: isolated colonies from solid media cultures, prepared via turbidity method in saline/Pluronic® broth. Panels contain dehydrated antibiotic (Synercid) and control wells. Operation: panels incubated in WalkAway®-SI system; colorimetric optics system (lamp/photosensor) monitors growth periodically from 4.5 to 16 hours (rapid) or 16-24 hours (overnight). Output: MIC values and qualitative susceptibility interpretations. Used in clinical microbiology laboratories by technicians. Results assist clinicians in selecting appropriate antibiotic therapy for Gram-positive infections.
Clinical Evidence
Clinical study evaluated 350 Gram-positive isolates (308 fresh, 42 stock) across 3 sites. Performance compared rapid read (<16h) and overnight methods against CLSI reference broth dilution. Rapid read method achieved 97.1% Essential Agreement (EA) and 97.7% Category Agreement (CA) for Long Dilution, and 98.0% CA for 5-Dilution Breakpoint. Reproducibility >95%. No very major errors in challenge set. Bench testing confirmed growth rates >95%.
Technological Characteristics
Miniaturized broth dilution susceptibility test; dehydrated antimicrobial agents in panel wells. Rehydration with Synergies plus™ Pos Broth. Incubation 4.5–18 hours at 35°C. Automated reading via MicroScan® instrumentation (WalkAway® SI) or visual reading. Connectivity: WalkAway® SI system integration.
Indications for Use
Indicated for quantitative and/or qualitative antimicrobial susceptibility testing of rapidly growing aerobic and facultative anaerobic Gram-positive enterococci (vancomycin-resistant and multi-drug resistant E. faecium) and staphylococci (methicillin-susceptible S. aureus). Testing performed on colonies grown on solid media. Concentrations: 0.12–8 µg/mL (Long Dilution) and 0.12–2 µg/mL (5-Dilution Breakpoint). Incubation: 4.5–16 hours (rapid) or 16–20 hours (overnight).
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
MicroScan® Dried Gram-Positive Panels
Related Devices
K062705 — SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH LINEZOLID · Dade Behring, Inc. · Oct 20, 2006
K063564 — MICROSCAN SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH CHLORAMPHENICOL (4-16 UG/ML) · Dade Behring, Inc. · Feb 8, 2007
K092919 — MICROSCAN SYNERGIES PLUS MIC/COMBO PANELS, MODEL B1025 · Siemens Healthcare Diagnostics · Oct 23, 2009
K070860 — MICROSCAN SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH CLINDAMYCIN · Dade Behring, Inc. · May 11, 2007
K063365 — SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH GENTAMICIN (0.06-32MG/ML LONG DILUTION SEQUENCE · Dade Behring, Inc. · Dec 8, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Ka63366
## 510(k) Summary
510(k) Submission Information: Device Manufacturer: Dade Behring Inc. DEC = 8 2006 Contact name: Robert Eusebio, Regulatory Affairs Manager Fax: 916-374-3144 Date prepared: October 31, 2006 Microdilution Minimum Inhibitory Concentration (MIC) Panels Product Name: Trade Name: MicroScan® Synergies plus" Gram-Positive MIC/Combo Panels Intended Use: To determine antimicrobial agent susceptibility 510(k) Notification: New antimicrobial - Synercid MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels and Predicate device: MicroScan® Dried Gram-Positive Panels
510(k) Summary:
MicroScan® Synergies plus" Gram-Positive MIC/Combo Panels, utilizing both the MicroScan® Rapid Fluorogenic Identification and Dried Overnight Antimicrobial Susceptibility Testing (AST) technologies, are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic Gram-positive enterococci and staphylococci.
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in water and dehydrated. Various antimicrobial agents are diluted in water, buffer, or minute concentrations of broth, to concentrations bridging the range of clinical interest. Panels are rehydrated with Synergies plus™ Pos Broth, after inoculation with a standardized suspension of the organism. After incubation in the WalkAway® SI, or equivalent, for 4.5 - 18 hours, the minimum inhibitory concentration (MIC) for the test organism is read by determining the lowest antimicrobial concentration showing inhibition of growth.
The proposed MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel demonstrated substantially equivalent performance when compared with a frozen Reference Panel, as defined in the FDA document "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", dated February 5, 2003. The Premarket Notification (510[k]) presents data in support of the MicroScan® Synergies plus" Cram-Positive MIC/Combo Panel with Synercid.
The external validation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external validations were designed to confirm the acceptability of the proposed Synergies plus " Gram-Positive Panel by comparing its performance with a frozen Reference panel. Challenge strains were compared to Expected Results determined prior to the evaluation. The Synergies plus" Gram-Positive Panel demonstrated acceptable performance with an overall Essential Agreement of 97.4%, for the Long Dilution Sequence when compared with the frozen Reference panel.
{1}------------------------------------------------
Instrument reproducibility testing demonstrated acceptable reproducibility and precision for Synercid with Turbidity inoculum preparation method and the WalkAway® SI System or equivalent.
Quality Control testing demonstrated acceptable results for the MicroScan® Synergies plus"" Gram-Positive MIC/Combo Panel with Synercid.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC - 8 2006
Mr. Robert Eusebio Regulatory Affairs Manager Dade Behring, Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691-9972
Re: k063366 Trade/Device Name: MicroScan® Synergies plus"M Gram-Positive MIC/Combo Panels Synercid (0.12-8 µg/ml Long Dilution Sequence and 0.12 - 2 ug/ml 5 - Dilution Breakpoint Sequence) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully automated short-term incubation cycle antimicrobial susceptibility system. Regulatory Class: Class II Product Code: LON, LRG, JWY, LTT Dated: November 6, 2006 Received: November 15, 2006
Dear Mr. Eusebio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Sally, artopm
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
K063366 510(k) Number (if known):
Device Name: MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels with Synercid (0.12 - 8 µg/ml Long Dilution Sequence and 0.12 - 2 µg/ml 5-Dilution Breakpoint Sequence)
## Indications For Use:
The MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive enterococci and staphylococci. After inoculation, panels are incubated for 4.5 -- 18 hours at 35℃ +/- 1℃, in a WalkAway® SI, or equivalent, and read by the MicroScan® Instrumentation. Additionally, the panels may be incubated in a non-CO2 incubator and the Antimicrobial Susceptibility Testing (AST) portions can be read visually, according to the Package Insert.
This particular submission is for the addition of the antimicrobial Synercid, at concentrations of 0.12 to 8 µg/ml Long Dilution Sequence and 0.12 – 2 µg/ml 5-Dilution Breakpoint Sequence, for Enterococcus faecium and Staphylococcus spp., to the test panel.
The Gram-positive organisms which may be used for Synercid susceptibility testing in this panel are:
- Enterococcus faecium (vancomycin-resistant and multi-drug resistant strains only) .
- Staphylococcus aureus (methicillin-susceptible strains only)
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Freddie M. Poole
Division Sign-Off
Office of In Vito Diagnostic Device Evaluation and Safeta
Page I of
IX
Syn 510k.doc 12/4/2006
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches