AXSYM ANTI-CCP REAGENT, STANDARD CALIBRATOR AND CONTROL KITS, MODELS 3L91-20, 3L91-01 AND 3L91-10
Applicant
Axis-Shield Diagnostics, Ltd.
Product Code
NHX · Immunology
Decision Date
Mar 20, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5775
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K063347 · Mar 20, 2007
AXSYM ANTI-CCP REAGENT, STANDARD CALIBRATOR AND CONTROL KITS, MODELS 3L91-20, 3L91-01 AND 3L91-10
Axis-Shield Diagnostics, Ltd.
Clinical serum/plasma specimens from five sites representing various disease states
The applicant used a retrospective cohort of 931 clinical specimens to compare the performance of the AxSYM Anti-CCP assay against the predicate device (DIASTAT) and to establish clinical sensitivity and specificity across various disease groups.
931 clinical specimens covering various disease states (RA, SLE, Sjögren’s, Scleroderma, Polymyositis/Dermatomyositis, Osteoarthritis, Psoriatic Arthritis, Inflammatory Polyarthritis, Ankylosing Spondylitis, EBV, Infectious disease, and Asymptomatic Healthy); Sample Size: 931; Number of Sites: 5
DIASTAT™ Anti-CCP assay
Percent concordance, total agreement, positive agreement, negative agreement, clinical sensitivity, and clinical specificity
Indications for Use
AxSYM® Anti-CCP is a Microparticle Enzyme Immunoassay (MEIA) for the semiquantitative determination of the IgG class of autoantibodies specific to cyclic citrullinated peptide (CCP) in human serum and plasma on the AxSYM System. Detection of anti-CCP antibodies is used as an aid in the diagnosis of Rheumatoid Arthritis (RA) and should be used in conjunction with other clinical information. Autoantibody levels represent one parameter in a multicriterion diagnostic process, encompassing both clinical and laboratory-based assessments. The AxSYM® Anti-CCP Standard Calibrators are for the standard calibration of the AxSYM System when used for the semi-quantitative determination the IgG class of autoantibodies specific to cyclic citrullinated peptide (CCP) in human serum and plasma. The AxSYM® Anti-CCP Controls are for the use in quality control to monitor the accuracy and precision of the AxSYM Anti-CCP assay when used for the semiquantitative determination of the IgG class of autoantibodies specific to cyclic citrullinated peptide (CCP) in human serum and plasma on the AxSYM System.
Device Story
AxSYM Anti-CCP is an in vitro diagnostic assay for Rheumatoid Arthritis (RA) diagnosis. It utilizes Microparticle Enzyme Immunoassay (MEIA) technology on the AxSYM System. The device processes human serum or plasma samples; reagents are pipetted into reaction vessels where anti-CCP antibodies bind to CCP-coated microparticles. These complexes are transferred to a matrix cell, washed, and reacted with an alkaline phosphatase-labeled conjugate. A substrate (4-Methylumbelliferyl Phosphate) is added; the enzyme catalyzes the production of a fluorescent product (4-Methylumbelliferone), which is measured by the MEIA optical assembly. The system is operated by laboratory personnel in a clinical setting. Output is a semiquantitative measurement of IgG anti-CCP levels, which clinicians use as one parameter within a broader diagnostic process to support RA diagnosis.
Clinical Evidence
Method comparison study of 931 specimens (529 RA, 402 non-RA) compared AxSYM Anti-CCP to DIASTAT Anti-CCP. Total agreement was 97.3% (95% CI: 96.1-98.3%). Sensitivity in RA patients was 72.2%; specificity in non-RA samples was 94.5%. Precision studies (n=80) showed total %CV between 7.7% and 11.1%. Linearity demonstrated from 0 to 200 U/mL.
Technological Characteristics
Microparticle Enzyme Immunoassay (MEIA) technology. Reagents include CCP-coated microparticles, mouse anti-human IgG:alkaline phosphatase conjugate, and 4-methylumbelliferyl phosphate substrate. Automated system (AxSYM) performs pipetting, incubation, washing, and fluorescence measurement. Connectivity: AxSYM System Software version 5.0. Storage: 2-8°C.
Indications for Use
Indicated for the semiquantitative determination of IgG class autoantibodies to cyclic citrullinated peptide (CCP) in human serum and plasma as an aid in the diagnosis of Rheumatoid Arthritis (RA). For use in conjunction with other clinical and laboratory assessments.
Regulatory Classification
Identification
A rheumatoid factor immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the rheumatoid factor (antibodies to immunoglobulins) in serum, other body fluids, and tissues. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k063347
B. Purpose for Submission:
New application on an approved system
C. Measurand:
Anti-cyclic citrullinated peptide (CCP)
D. Type of Test:
Microparticle Enzyme Immunoassay (MEIA) Semi-quantitative
E. Applicant:
AXIS-SHIELD DIAGNOSTICS, LTD.
F. Proprietary and Established Names:
AxSYM® Anti-CCP Reagent kit
AxSYM® Anti-CCP Standard Calibrator kit
AxSYM® Anti-CCP Control kit
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| Antibodies, Anti-Cyclic Citrullinated Peptide (CCP) (NHX) | Class II | 21 CFR § 866.5775, Rheumatoid factor immunological test system. | 82 Immunology |
| Calibrator, Multi-Analyte Mixture (JIX) | Class II | 21 CFR § 862.1150 | 75 Chemistry |
| Quality Control Material (Assayed and Unassayed) (JJY) | Class I | 21 CFR § 862.1660 | 75 Chemistry |
H. Intended Use:
1. Intended use(s):
AxSYM® Anti-CCP is a Microparticle Enzyme Immunoassay (MEIA) for the semi-quantitative determination of the IgG class of autoantibodies specific to cyclic citrullinated peptide (CCP) in human serum and plasma on the AxSYM System. Detection of anti-CCP antibodies is used as an aid in the diagnosis of Rheumatoid Arthritis (RA) and should be used in conjunction with other clinical information. Autoantibody levels represent one parameter in a multicriterion diagnostic process, encompassing both clinical and laboratory-based assessments.
The AxSYM® Anti-CCP Standard Calibrators are for the standard calibration of the AxSYM System when used for the semi-quantitative determination the IgG
{1}
class of autoantibodies specific to cyclic citrullinated peptide (CCP) in human serum and plasma.
The AxSYM® Anti-CCP Controls are for the use in quality control to monitor the accuracy and precision of the AxSYM® Anti-CCP assay when used for the semiquantitative determination of the IgG class of autoantibodies specific to cyclic citrullinated peptide (CCP) in human serum and plasma on the AxSYM System.
2. Indication(s) for use:
Same as above
3. Special conditions for use statement(s):
This device is for prescription use only.
4. Special instrument requirements:
For use with AxSYM System Software version 5.0 cleared under k974651
# I. Device Description:
Each AxSYM® Anti-CCP Reagent kit contains 1 bottle of each component of Conjugate, Microparticles, Sample Diluent and Matrix Cell Blocker.
- Conjugate: 1 Bottle (12.2 mL) Mouse Anti-Human IgG:Alkaline Phosphatase Conjugate in TRIS buffer containing detergent with protein (Bovine) stabilizers. Minimum concentration $0.1\mu \mathrm{g} / \mathrm{mL}$ . Preservative: Sodium Azide. (Reagent Bottle 1)
- Microparticles: 1 Bottle (4.8 mL) CCP-Coated Microparticles in phosphate buffer with protein (Bovine) stabilizers. Minimum concentration: 0.1% solids (w/v). Preservative: Sodium Azide. (Reagent Bottle 2)
- Sample Diluent: 1 Bottle (24.7 mL) Sample Diluent containing detergent with protein (Bovine) stabilizers. Preservative: Sodium Azide. (Reagent Bottle 3)
- Matrix Cell Blocker: 1 Bottle (50.2 mL) Matrix Cell Blocker containing detergent. Preservative: Sodium Azide. (Reagent Bottle 4)
Each Anti-CCP Standard Calibrator kit contains six bottles of AxSYM® Anti-CCP Standard Calibrators (4.3 mL each). Calibrator A is a phosphate buffer with protein (Bovine) stabilizers. Calibrators B-F contains anti-CCP positive human plasma in phosphate buffer with protein (Bovine) stabilizers to yield the concentrations (U/mL) shown in the following table:
| Standard Calibrator | Anti-CCP Concentration (U/mL) |
| --- | --- |
| Cal A | 0 |
| Cal B | 5 |
| Cal C | 25 |
| Cal D | 50 |
| Cal E | 100 |
| Cal F | 200 |
{2}
Each AxSYM® Anti-CCP Control kit contains two bottles (7.0 mL each) of controls with one positive and the other negative human plasma in phosphate buffer to yield the following concentrations (U/mL):
| Bottle | Anti-CCP Concentration (U/mL) | Range (U/mL) |
| --- | --- | --- |
| Positive Control | 24.0 | 14.0 – 34.0 |
| Negative Control | 0.0 | ≤ 3.0 |
Calibrators and controls are sold separately.
## J. Substantial Equivalence Information:
1. Predicate device name(s): DIASTAT™ Anti-CCP
2. Predicate 510(k) number(s): k023285
3. Comparison with predicate:
Similarities
| Parameter | Submission Device
Axis-Shield AxSYM Anti-CCP | Predicate Device
Axis-Shield DIASTAT Anti-CCP |
| --- | --- | --- |
| Intended Use | Aid in the diagnosis of Rheumatoid Arthritis (RA) | Same |
| Specimen Type | Human serum or plasma (sodium citrate, lithium heparin) | Same |
| Capture Antigen | Cyclic citrullinated peptide (CCP), second generation | Same |
| Storage Conditions | The AxSYM Anti-CCP Reagent Pack, Standard Calibrator Pack and Control Pack must be stored at 2°-8°C | Same |
| Suggested Cut-off | 5.0 U/mL | Same |
{3}
# Differences
| Parameter | Submission Device Axis-Shield AxSYM Anti-CCP | Predicate Device Axis-Shield DIASTAT Anti-CCP |
| --- | --- | --- |
| Technology Format | Automated, sandwich format, Microparticle Enzyme Immunoassay (MEIA) | Manual, microtitre plate format, Enzyme-linked Immunosorbent Assay (ELISA) |
| Conjugate Antibody | Mouse anti-human IgG: alkaline phosphatase | Alkaline phosphatase-labelled murine monoclonal antibody to human IgG |
| Substrate | 4-Methylumbelliferyl phosphate | Phenolphthalein monophosphate |
| Assay End-Point | Fluorescence | Color, read at 540-565nm |
| Quantitation | Results are determined from a standard calibration curve (0, 5, 25, 50, 100, 200 U/mL) generated and stored on the instrument | Results are determined from a standard calibration curve (0, 2, 8, 30, 100 U/mL) generated on each microtiter plate |
| Calibrator Range | 0-200 U/mL Auto-dilute (1/10) up to 2000 U/mL | 0-100 U/mL |
| Calibration | Semi-quantitative assay | Semi-quantitative/qualitative assay |
| Calibrators and Controls | Sold separately | Kit components |
| Expected Values in Asymptomatic Population | < 1.0 to 2.9 U/mL in a representative study | 0.05 – 3.8 U/mL from a reference population |
# K. Standard/Guidance Document Referenced (if applicable):
STANDARDS
Title and Reference Number
Other Standards
GUIDANCE
| Document Title | Office | Division | Web Page |
| --- | --- | --- | --- |
| | | | |
# L. Test Principle:
The AxSYM® Anti-CCP is based on Microparticle Enzyme Immunoassay (MEIA) technology. The AxSYM® Anti-CCP reagents and sample are pipetted in the following sequence:
SAMPLING CENTER
{4}
- Sample and all AxSYM Anti-CCP reagents required for one test are pipetted by the Sampling Probe into various wells of a Reaction Vessel (RV).
- AxSYM Line Diluent and sample are pipetted into the Incubation Well of the RV.
- The AxSYM Anti-CCP Sample Diluent and diluted sample are pipetted into the Sample Well of the RV.
- The AxSYM Anti-CCP Matrix Cell Blocker is pipetted into the Buffer Well of the RV.
- The AxSYM Anti-CCP Mouse Anti-Human IgG:Alkaline Phosphatase Conjugate is pipetted into Reagent Well 3 of the RV.
- AxSYM Line Diluent and CCP-Coated Microparticles are pipetted into Reagent Well 2 of the RV.
- A reaction mixture is formed by combining diluted sample and diluted microparticles coated with CCP in Reagent Well 1 of the RV.
- When anti-CCP antibody is present in the sample, it binds to the CCP-Coated Microparticles, forming antigen-antibody complexes on the microparticles.
- The RV is immediately transferred into the Processing Center. Further pipetting is done in the Processing Center by the Processing Probe.
## PROCESSING CENTER
- An aliquot of Matrix Cell Blocker is transferred to the Matrix Cell.
- An aliquot of the reaction mixture, containing microparticles and bound antigen-antibody complex, is transferred to the Matrix Cell. The microparticles bind irreversibly to the glass fiber matrix.
- The Matrix Cell is washed to remove materials not bound to the microparticles.
- The AxSYM Anti-CCP Mouse Anti-Human IgG:Alkaline Phosphatase Conjugate is dispensed onto the Matrix Cell and it binds with the antigen-antibody complexes.
- The Matrix Cell is washed to remove conjugate not bound to the microparticles.
- The substrate, 4-Methylumbelliferyl Phosphate, is added to the Matrix Cell. The alkaline phosphatase-labeled conjugate catalyzes the removal of a phosphate group from the substrate, yielding the fluorescent product, 4-Methylumbelliferone. This fluorescent product is measured by the MEIA optical assembly.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Four samples covering the measuring range of the device were assayed in replicates of two at two separate times of the day for twenty days (n=80). Testing was performed on two AxSYM Systems, using one reagent lot. The precision acceptance criteria for total %CV and within-run %CV was <15%. Data from the studies are summarized below:
{5}
| N | Mean (mg/dL) | Within run %CV | Total %CV |
| --- | --- | --- | --- |
| 80 | 5.2 | 6.9 | 9.3 |
| 80 | 4.3 | 10.7 | 11.1 |
| 80 | 13.8 | 7.1 | 8.4 |
| 80 | 15.0 | 5.7 | 7.7 |
| 80 | 24.4 | 7.0 | 8.6 |
| 80 | 25.6 | 6.6 | 8.3 |
| 80 | 139.8 | 6.8 | 10.0 |
| 80 | 149.7 | 7.1 | 8.5 |
b. Linearity/assay reportable range:
Linearity testing was demonstrated by testing five high patient sample pools, with anti-CCP concentrations between 100 and $200\mathrm{U / mL}$ . These samples were serially diluted with AxSYM Anti-CCP Negative control at the following dilutions: 1:2, 1:4, 1:6, 1:8, and 1:10. Each dilution was assayed in triplicate and a median is calculated. Data was analyzed using linear regression analysis (x-axis: expected concentration and y-axis: actual concentration). The data showed the following regression equation:
Sample Pool 1 $\mathrm{y} = 1.0842\mathrm{x} + 2.3294,\mathrm{r}^2 = 0.991$
Sample Pool 2 $\mathrm{y} = 1.0025\mathrm{x} + 4.6563,\mathrm{r}^2 = 0.9873$
Sample Pool 3 $\mathrm{y} = 0.9507\mathrm{x} + 6.6989,\mathrm{r}^2 = 0.9869$
Sample Pool 4 $\mathrm{y} = 1.0288\mathrm{x} + 3.1923,\mathrm{r}^2 = 0.9813$
Sample Pool 5 $\mathrm{y} = 1.000\mathrm{x} + 1.9658,\mathrm{r}^2 = 0.9903$
The AxSYM Anti-CCP assay showed linearity from 0 to $200\mathrm{U / mL}$
In addition, an automated dilution protocol was designed to assist in quantitation of test results using the AxSYM system. The AxSYM system performs a 1:10 dilution of the specimen prior to analysis. The system automatically calculates the concentration of the diluted specimen and reports the result. Five high sample pools were used in this study. Each sample was diluted manually and auto-dilute using Anti-CCP Negative Control. The assay was designed to have a mean auto-dilute recovery of $100 \pm 20\%$ . The $\%$ recoveries for the five sample pools were 107.3, 90.0, 86.3, 91.3, 100.2, and 95.0. The results support the use of both automated and manual 1:10 dilution for the AxSYM Anti-CCP assay.
High dose hook effect
The possibility of high dose hook effect occurring when using the device was evaluated with two serum samples with concentrations above the assay range (1200U/mL and 2400 U/mL). The samples were analyzed on two AxSYM instruments in triplicate with two reagent lots. No hook effect was observed at concentrations 12 times that of Calibrator F (200 U/mL).
{6}
c. Traceability, Stability, Expected values (controls, calibrators, or methods): An international reference material for anti-CCP antibodies is not available. The calibrators and controls are assigned relative arbitrary Axis-Shield units (U/mL).
The AxSYM Anti-CCP Reference Calibrators were manufactured using a pool of high titer anti-CCP positive plasmas that has been assigned a value in U/mL. The AxSYM Anti-CCP Standard Calibrators were prepared from an anti-CCP stock (positive plasma) diluted in Calibrator diluent. This is tested by rate matching to the Reference Calibrator and should be within the target range of ± 2.5%. If acceptable, this becomes Calibrator F and is then further diluted to produce calibrators B-E.
Storage conditions
The AxSYM Anti-CCP Reagent Pack must be stored at 2-8°C. When stored and handled as directed, reagents are stable until the expiration date.
d. Detection limit:
The AxSYM assay is designed to have a mean analytical sensitivity/limit of blank (LoB) of ≤ 1.0 U/mL. The LoB was calculated as the concentration at the 95th percentile above the Standard Calibrator A (0.0 U/mL) and was found to be 0.067 U/mL. The design specification (≤ 1.0 U/mL) was met.
e. Analytical specificity:
Interference: Interference testing was performed using guidance supplied by CLSI EP7-A, "Interference testing in Clinical Chemistry". No significant interference was observed in:
- Hemoglobin up to 500 mg/dL
- Bilirubin up to 20 mg/dL
- Triglycerides up to 1500 mg/dL
- Total Protein up to 12 g/dL
- Rheumatoid Factor up to 200 IU/mL
- RBC up to 0.4%
Crossreactivity
To assess the potential cross-reactivity of the CCP antigen used in the AxSYM Anti-CCP assay with other autoantibodies, the assay was evaluated with 16 samples, all with high levels of various autoantibodies and negative for CCP. The following autoantibodies (2 samples each) were tested: SSA, SSB, RNP, dsDNA, Jo-1, and Ribosomal-P. The study showed no significant cross-reactivity of the CCP antigen with any of these autoantibodies.
f. Assay cut-off:
See Expected Value.
2. Comparison studies:
{7}
a. Method comparison with predicate device:
Nine hundred thirty one (931) specimens were tested at Axis-Shield using the AxSYM® Anti-CCP assay and its predicate device, DIASTAT™ Anti-CCP assay. Specimens were received from five sites covering a range of relevant disease states. Sample distribution and percent concordance between the two devices for each disease group are shown below:
| Disease Group | Total N | %American | %European | %Concordance |
| --- | --- | --- | --- | --- |
| Asymptomatic Healthy | 100 | 100 | 0 | 100 |
| EBV | 25 | 100 | 0 | 100 |
| Infectious disease | 25 | 100 | 0 | 100 |
| SLE | 76 | 14 | 86 | 99 |
| Sjögren’s Syndrome | 8 | 50 | 50 | 100 |
| Scleroderma | 9 | 0 | 100 | 100 |
| Polymyositis/Dermatomyositis | 6 | 33 | 67 | 100 |
| Osteoarthritis | 22 | 0 | 100 | 100 |
| Psoriatic Arthritis | 31 | 0 | 100 | 100 |
| Inflammatory Polyarthritis | 80 | 0 | 100 | 100 |
| Ankylosing Spondylitis | 20 | 25 | 75 | 100 |
| Rheumatoid arthritis | 529 | 18 | 82 | 95 |
| Total | 931 | 37 | 63 | 96 |
The following table summarized the % concordance between the two assays:
RA (n = 529)
| | DIASTAT | |
| --- | --- | --- |
| AxSYM | + | - |
| + | 384 | 0 |
| - | 23 | 122 |
Total Agreement (95% CI) = 95.7% (93.5-97.2%)
Positive Agreement (95% CI) = 94.3% (91.6-96.4%)
Negative Agreement (95% CI) = 100% (97.6-100%)
Non-RA (n = 402)
| | DIASTAT | |
| --- | --- | --- |
| AxSYM | + | - |
| + | 21 | 1 |
| - | 1 | 379 |
Total Agreement (95% CI) = 99.5% (98.2-99.9%)
Positive Agreement (95% CI) = 95.5% (77.2-99.9%)
Negative Agreement (95% CI) = 99.7% (98.5-100%)
{8}
All Samples (n = 931)
Total Agreement (95% CI) = 97.3% (96.1-98.3%)
Positive Agreement (95% CI) = 94.4% (91.8-96.4%)
Negative Agreement (95% CI) = 99.8% (98.9-100%)
b. Matrix comparison:
Matrix comparison studies using matched serum (clot), serum separator, EDTA, lithium and sodium citrate plasma samples covering the dynamic range were performed on the AxSYM instrument using AxSYM® Anti-CCP assay. The results of the linear regression analyses are seen on the table below.
| | N | Slope (95% CI) | Intercept (95% CI) | Correlation Coefficient |
| --- | --- | --- | --- | --- |
| Serum (clot vs. Serum separator) | 10 | 1.0618 (0.9943-1.1294) | -1.2470 (-4.8136-2.3197) | 1.00 |
| Serum vs. Lithium Heparin | 10 | 0.9531 (0.9246-9816) | 0.3433 (-11598-1.8465) | 1.00 |
| Serum vs. Sodium Citrate | 10 | 0.9880 (0.9617-1.0142) | 0.3122 (-1.0726-1.6970) | 1.00 |
| Serum vs. EDTA | 10 | 0.9452 (0.9211-0.9693) | 0.4536 (-0.8189-1.7261) | 1.00 |
3. Clinical studies:
a. Clinical Sensitivity and Specificity
Using the same samples from the method comparison study, % sensitivity and specificity were calculated. The following results were obtained:
| Disease Group | N | Anti-CCP positive | %Sensitivity | % Specificity |
| --- | --- | --- | --- | --- |
| All RA samples | 529 | 382 | 72.2 | |
| All Non-RA samples | 402 | 22 | | 94.5 |
The lower sensitivity of 72.2% was accepted for the following reasons:
1. Anti-CCP assays from different manufacturers showed variable ranges from 69% to 87.8%.
2. Literature showed sensitivity ranging from 65% to 85% in established RA depending on cohort tested.
References:
{9}
Autoantibodies to citrullinated antigens in (early) rheumatoid arthritis. W.J. van Venrooij, A.J.W. Zendman, GJN Pruijn. Autoantibody Reviews: 6 (2006); 37-41.
Anti-cyclic citrullinated peptide antibodies: diagnostic, predictive and monitoring value in RA. PH Schur. Int. J. Advances in Rheumatology: 3 (2005); 77-83.
Diagnostic and predictive value of anti-CCP antibodies in rheumatoid arthritis: a systematic literature review. J Avouac, L Gossec, M Dougados. Ann. Rheum. Dis. (2006). Online: doi:10.1136/ard.2006.051391.
b. Other clinical supportive data (when a. is not applicable): None
4. Clinical cut-off: Not applicable
5. Expected values/Reference range: In a representative study, 100 serum specimens from asymptomatic healthy donors with an age range of 19-73 years, comprised of males (n=51) and females (n=49), were tested in the AxSYM Anti-CCP assay. No differences attributable to gender or age were observed. Specimen values range from <1.0 to 2.9 U/mL. On the basis of this reference population data, the suggested cut-off is 5.0 U/mL. Values ≥5.0 U/mL is considered positive and ≤ 5.0 U/mL is considered negative.
N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.