K063252 · Cook, Inc. · DQY · Nov 14, 2006 · Cardiovascular
Device Facts
Record ID
K063252
Device Name
ATB ADVANCED PTA DILATATION CATHETER
Applicant
Cook, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Nov 14, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Intended Use
For percutaneous transluminal angioplasty of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and ilio femoral and are also intended to treat obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The ATB Advance PTA Dilatation Catheter is also intended for post-dilatation of balloonexpandable peripheral vascular stents.
Device Story
ATB Advance PTA Dilatation Catheter is an over-the-wire balloon catheter used for peripheral artery angioplasty and post-dilatation of balloon-expandable stents. Device utilizes 5.0 or 6.0 French nylon tubing compatible with 0.035-inch guidewires. Operated by physicians in clinical settings to dilate narrowed or obstructed vessels or dialysis fistulae. Mechanical inflation of the balloon exerts radial force on the lesion to restore vessel patency. Supplied sterile for single-use. Benefits include improved blood flow and vessel patency.
Clinical Evidence
Bench testing only. Testing included tensile strength, balloon inflation/deflation, balloon burst, balloon compliance, balloon fatigue, balloon profile, and balloon burst/fatigue within a stent.
Technological Characteristics
Over-the-wire balloon catheter; 5.0 or 6.0 French nylon tubing; 0.035-inch guidewire compatibility; sterile; single-use; mechanical inflation/deflation mechanism.
Indications for Use
Indicated for patients requiring percutaneous transluminal angioplasty of peripheral artery lesions (iliac, renal, popliteal, infrapopliteal, femoral, iliofemoral) or treatment of obstructive lesions in native/synthetic arteriovenous dialysis fistulae, and post-dilatation of balloon-expandable peripheral vascular stents.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Related Devices
K052036 — ATB ADVANCE PTA DILATATION CATHETER · Cook, Inc. · Aug 25, 2005
K032931 — PTA BALLOON CATHETER · Cook, Inc. · Oct 22, 2003
K981906 — MODIFICATION FOR PTA BALLOON CATHETER · Cook, Inc. · Jun 29, 1998
K023504 — MODIFICATION TO ATB ALL-TERRAIN BALLOON PTA DILATATION CATHETER · Cook, Inc. · Nov 13, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(k) Summary
# Submitted By:
Karen Bradburn, RAC Senior Requlatory Affairs Specialist Cook Incorporated 750 Daniels Way, PO Box 489 Bloomington, IN 47402 812-339-2235 October 26, 2006
# NOV 1 4 2006
K063252
2. Lot 2
#### Device:
Trade Name: Proposed Classification: ATB™ Advance PTA Dilatation Catheter Catheter, Angioplasty, Peripheral, Transluminal (74 DQY)
## Predicate Devices:
The ATB™ Advance PTA Dilatation Catheter is similar in terms of intended use, materials of construction and technological characteristics to predicate devices reviewed as devices for transluminal percutaneous angioplasty of vessel lumens which are narrowed or obstructed.
#### Device Description:
The ATB™ Advance PTA Dilatation Catheter is an over-the wire balloon catheter indicated for percutaneous transluminal angioplasty of lesions in peripheral arteries including illac, renal, popliteal, infrapopliteal, femoral and ilio femoral and are also intended to treat obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The device will be made with 5.0 or 6.0 French nylon tubing compatible with an 0.035-inch guidewire. It will be supplied sterile, intended for one-time use.
## Substantial Equivalence:
This device will be manufactured according to specified process controls and a Quality Assurance Program. This device will undergo packaging similar to the devices currently marketed and distributed by Cook Incorporated. This device will undergo sterilization similar to the devices currently marketed and distributed. Being similar with respect to indications for use, materials and physical construction to predicate devices, this device meets the requirements for section 510(k) substantial equivalence.
{1}------------------------------------------------
## Test Data:
The ATB™ Advance PTA Dilatation Catheter was subjected to the following tests to assure The ATD … Auvance TTA Blatation Outhoter was catholor was cathers. These tests were comprised of:
- Tensile tests 1.
- Balloon inflation/deflation tests 2.
- Balloon burst tests 3.
- Balloon compliance 4.
- Balloon fatigue tests 5.
- Balloon profile റ്.
- Balloon burst within a stent 7.
- Balloon fatigue within a stent 8.
The results of these tests provide reasonable assurance that the device has been designed i The results of these tests provide roubonable addrance and a PTA dilatation balloon catheter.
KO63252 p. 20f2
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 4 2006
Cook Incorporated c/o Ms. Karen Bradburn Senior Regulatory Affairs Specialist P.O. Box 489 Bloomington, IN 47402-0489
Re: K063252 ATB Advanced PTA Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (Two) Product Code: DQY Dated: October 26, 2006 Received: October 27, 2006
Dear Ms. Bradburn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/9 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
{3}------------------------------------------------
Page 2 - Ms. Karen Bradburn
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements af the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Duma R. balmer
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Special 510(k) Premarket Notification PTA Balloon Catheter COOK INCORPORATED
510(k) Number (if known): K063252
PTA Balloon Catheter Device Name:
Indications for Use:
For percutaneous transluminal angioplasty of lesions in peripheral arteries including iliac, renal, popliteal, infrapopliteal, femoral and ilio femoral and are also intended to treat obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The ATB Advance PTA Dilatation Catheter is also intended for post-dilatation of balloonexpandable peripheral vascular stents.
Prescription Use __ × (Per 21 CFR 801 Subpart D)
ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
pring R. Jackner
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_K 0 63252
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.