← Product Code [MMI](/productcode/MMI) · K063243

# VIDAS TROPONIN I ULTRA (TNIU) ASSAY, MODEL 30 448 (K063243)

_bioMerieux, Inc. · MMI · Dec 14, 2007 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K063243

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [MMI](/productcode/MMI.md)
- **Decision Date:** Dec 14, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1215
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

VIDAS® Troponin I Ultra is an automated quantitative test for use on the VIDAS instruments for the determination of human cardiac troponin I in human serum or plasma (lithium heparin) using the ELFA (Enzyme-Linked Fluorescent Assay) technique. VIDAS Troponin I Ultra is intended to be used as an aid in diagnosis of myocardial infarction.

## Device Story

VIDAS Troponin I Ultra is an automated, quantitative, two-site sandwich ELFA immunoassay for cardiac troponin I (cTnI) in human serum or plasma. The device utilizes a disposable Solid Phase Receptacle (SPR) coated with mouse monoclonal anti-cTnI antibodies. The assay is performed on VIDAS or miniVIDAS instruments, which control all steps: sample incubation with alkaline phosphatase-labeled conjugate, wash cycles, and substrate (4-Methyl-umbelliferyl phosphate) addition. The enzyme catalyzes substrate hydrolysis into a fluorescent product, measured at 450nm. Fluorescence intensity is directly proportional to cTnI concentration. Used in clinical laboratories, the test provides quantitative results to assist physicians in diagnosing myocardial infarction. The system requires prescription use and is intended for serial sampling to monitor cardiac markers over time.

## Clinical Evidence

Clinical performance was evaluated at two sites (US and France) with patients presenting with chest pain and suspected myocardial infarction. Total sample size was 602 (US) and 396 (France). Sensitivity and specificity were calculated at 0-6 hours and 4-12 hours post-admission using a 0.11 ng/mL cutoff. US site sensitivity was 69.44% (0-6h) and 97.22% (4-12h); specificity was 96.24% (0-6h) and 96.21% (4-12h). French site sensitivity was 78.45% (0-6h) and 98.70% (4-12h); specificity was 90.55% (0-6h) and 92.11% (4-12h).

## Technological Characteristics

Two-site sandwich ELFA immunoassay. Solid phase: SPR coated with mouse monoclonal anti-cTnI antibody. Conjugate: alkaline phosphatase-labeled mouse monoclonal anti-cTnI antibody. Substrate: 4-Methyl-umbelliferyl phosphate. Energy source: automated VIDAS/miniVIDAS instrument optical scanner (450nm). Connectivity: standalone instrument. Calibration: traceable to NIST SRM 2921. Reportable range: 0.01-30 ng/mL.

## Regulatory Identification

A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

## Predicate Devices

- Dimension RxL® Cardiac Troponin I (CTNI) Assay

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k063243

B. Purpose for Submission:
New device

C. Measurand:
Troponin I

D. Type of Test:
Two-site sandwich immunoassay, quantitative

E. Applicant:
bioMerieux, Inc.

F. Proprietary and Established Names:
VIDAS® Troponin I Ultra

G. Regulatory Information:

1. Regulation section:
21 CFR 862.1215, Creatine phosphokinase/creatine kinase or isoenzymes test system

2. Classification:
Class II

3. Product code:
MMI (Immunoassay method, troponin subunit)

4. Panel:
Chemistry (75)

H. Intended Use:

1. Intended use(s):
See indications for use below.

{1}

2. **Indication(s) for use:**
VIDAS® Troponin I Ultra is an automated quantitative test for use on the VIDAS instruments for the determination of human cardiac troponin I in human serum or plasma (lithium heparin) using the ELFA (Enzyme-Linked Fluorescent Assay) technique. VIDAS Troponin I Ultra is intended to be used as an aid in diagnosis of myocardial infarction.

3. **Special conditions for use statement(s):**
For prescription use

4. **Special instrument requirements:**
VIDAS (k891385) and miniVIDAS (k923579)

I. **Device Description:**
Each VIDAS® Troponin I Ultra kit contains 60 tests. The kit is comprised of: 60 TNIU Reagent Strips, 60 Solid Phase Receptacles (SPR), TNIU controls (C1 and C2), TNIU calibrators (S1 and S2), TNIU Diluent, and one Master Lot Entry (MLE) Card.

The TNIU Reagent Strips consist of 10 wells covered with a labeled foil seal. Five of the wells contain either conjugate (alkaline phosphatase-labeled mouse monoclonal anti-cardiac troponin I antibody and preservative), wash buffer, or a cuvette with substrate (4-Methyl-umberlliferyl phosphate, dietholamine, and preservative). One well is designated for the sample and the remaining wells are empty.

The interior of the Solid Phase Receptacles (SPR) are coated with mouse monoclonal anti-cardiac troponin I antibody.

The TNIU controls (C1 and C2) are supplied with the kit as four, 2 mL vials of lyophilized human serum, troponin I, and preservative; 2 vials of C1 and 2 vials of C2.

The TNIU calibrators (S1 and S2) are supplied with the kit as four, 2 mL vials of lyophilized human serum, troponin I, and preservative; 2 vials of C1 and 2 vials of C2.

The TNIU diluent is read-to-use. It is supplied as one 2 mL vial and contains human serum with preservatives.

Human source material was tested and found negative for HIV-1/2, HBsAg, and HCV by FDA or European Union approved methods.

J. **Substantial Equivalence Information:**

1. **Predicate device name(s):**
Dade Behring Dimension RxL CTNI immunoassay
Dade Behring Dimension CTNI calibrator

{2}

2. Predicate 510(k) number(s): k010313
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended use | For the quantitative measurement of cardiac troponin I in human serum or plasma. Measurements of cardiac troponin I are used to aid in the diagnosis of acute myocardial infarction (AMI). | For the quantitative measurement of cardiac troponin I in human serum and plasma. Measurements of cardiac troponin I are used to aid in the diagnosis of acute myocardial infarction (AMI).  |
|  Sample type | Serum or plasma | Serum or plasma  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Additional indication | None | Risk stratification of patients with acute coronary syndromes with respect to their relative risk of mortality  |
|  Reportable range | 0.01 to 30 ng/mL | 0.04 to 40 ng/mL  |
|  Sample volume | 200 μL | 50 μL  |

K. Standard/Guidance Document Referenced (if applicable):
- Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline- Second Edition (CLSI EP5-A2)
- Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (CLSI EP6-A)
- Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (CLSI EP6-A)

L. Test Principle:
The VIDAS Troponin I Ultra (TNIU) Assay is one-step, sandwich enzyme-linked fluorescent immunoassay (ELFA) performed with an automated VIDAS or a miniVIDAS instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR), serves as the solid phase as well as a pipettor for the assay. Reagents for the assay are pre-dispensed in the sealed TNIU Reagent Strips.

All assay steps and assay temperatures are controlled by the instrument. The sample is transferred into the wells containing the conjugate (alkaline phosphatase-labeled

{3}

mouse monoclonal anti-cardiac troponin I antibody). The sample and conjugate mixture is cycled in and out of the SPR several times. Unbound sample is removed from the SPR during the wash step. During the detection step, the fluorescent substrate (4-Methyl-umberlliferyl phosphate) is cycled in and out of the SPR. The conjugate enzyme remaining in the SPR catalyzes the hydrolysis of the substrate into a fluorescent product 4-methylumbellferone. Fluorescence is measured at  $450\mathrm{nm}$  wavelength by the optical scanner in the instrument. The intensity of the fluorescence is directly proportional to the concentration of analyte present in the sample.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

The precision study was performed over a period of 10 days, 2 runs per day with 2 reagent lots and 3 test sites  $(n = 240)$ . The repeatability, inter-lot reproducibility, and inter-site reproducibility were calculated based on CLSI EP5-A2. Samples with TnI concentration at 4 different levels were tested (0.58, 3.55, 6.66, and  $17.05\mathrm{ng / mL}$ ). Repeatability ranged from  $1.28 - 2.7\%$  coefficient of variation (CV), inter-lot reproducibility ranged from  $3.4 - 6.61\%$  and inter-site reproducibility ranged from  $3.74 - 7.86\%$  CV.

A Limit of Quantitation (LoQ) study was performed as part of the same study to determine the lowest concentration of TnI that corresponded to a  $10\%$  CV. Four samples were tested in duplicate, 2 runs per day, for 10 days with 2 kit lots and across 3 testing sites  $(n = 240)$ . The LoQ was determined to be 0.11 ng/mL and corresponds to a coefficient of variation (CV) of approximately  $10\%$ . Three samples with values ranging from 0.05-0.08 ng/mL had CVs ranging from  $13 - 18\%$ . The results are shown in the table below.

|  Sample conc. | SD - inter-site | % CV - inter-site  |
| --- | --- | --- |
|  0.05 μg/L | 0.006 | 13.61  |
|  0.06 | 0.011 | 18.06  |
|  0.08 | 0.013 | 16.97  |
|  0.11 | 0.011 | 10.68  |
|  0.58 | 0.022 | 3.74  |
|  3.55 | 0.149 | 4.19  |
|  6.66 | 0.281 | 4.22  |
|  17.05 | 1.340 | 7.86  |

# b. Linearity/assay reportable range:

The reportable range of the assay is 0.01 to  $30\mathrm{ng / mL}$

{4}

The recovery across the range was evaluated by comparing the observed versus expected values obtained with the VIDAS TnI Ultra method. Two studies were performed, one evaluating the range from 0.02 to 32.4 ng/mL and the other evaluating the range from 0.03 to 0.68 ng/mL. In each case, a high plasma pool was created by spiking natural troponin antigen into normal plasma. A very low plasma pool was created by spiking a very low amount of the same antigen into normal plasma. The two pools were mixed in different proportions across the range of the assay. The observed results recovered in the range of 91-110% of the expected TnI values for test samples in the range of 3.66 to 32.4 ng/mL and 90-124% for test samples in the range of 0.06 to 0.68 ng/mL.

High dose hook effect was evaluated by testing normal human sera spiked with troponin I at high concentrations up to 1,000 ng/mL. No hook effect was observed with samples up to this level.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability and Value Assignment:

The kit calibrators are traceable to reference calibrators that are prepared gravimetrically from NIST SRM 2921.

Stability: The shelf life of the product is stated to be 12 months when stored refrigerated at 2-8°C. Stability studies were performed to support these claims.

The calibrators and controls are lyophilized. After reconstitution, they are stable for up to 8 hours at 2-8°C or up to the expiration date of the kit when stored at -25±6°C. Calibrators and controls may be frozen and thawed up to 4 times.

The stability protocols and acceptance criteria were reviewed and found to be acceptable.

d. Detection limit:

Studies were performed following a protocol similar to CLSI EP17-A. A sample containing no troponin I was tested in duplicate using 2 kit lots over 15 assay runs. In all, 30 measurements (n=30) of the zero standard were performed. The Limit of Blank (LoB) was defined as the concentration corresponding to two standard deviations above the mean signal of the zero standard. The value is then converted to TnI concentration. Based on the sponsor's analysis limit of blank (LoB) was determined to be 0.0013 ng/mL.

The sponsor defined Limit of Detection (LoD) as the lowest signal that can be detected reliably. This value is then converted into TnI concentration. Two different samples with very low TnI concentration of approximately 0.006 0.003 ng/mL were tested with 2 kit lots. The limits of detection (LoD) were

5

{5}

determined to be 0.002 ng/mL and 0.008 ng/mL respectively. The sponsor chose 0.01 ng/mL as the LoD. The device reports results greater or equal to 0.01 ng/mL.

The Limit of Quantitation when defined as where the device achieves a coefficient of variation (CV) of approximately 10% is 0.11 ng/mL (see precision section 1. a. above).

e. Analytical specificity:

Evaluation of analytical specificity (cross-reactivity) was done by spiking each cross-reactant to target concentration into a troponin I negative serum. A control for each cross-reactant was prepared by spiking the samples with the same volume of the solvent used for reconstituting the cross-reactant. The cross-reactant test samples and the control samples were measured on the VIDAS TNIU and the cross-reactivity was calculated.

|  Cross-reactant | Concentration | % Cross-reactivity  |
| --- | --- | --- |
|  Troponin-C (cardiac) | 1000 ng/mL | <0.001  |
|  Troponin-T (cardiac human) | 1000 ng/mL | 0.2  |
|  Troponin-I (skeletal human) | 1000 ng/mL | <0.001  |
|  Troponin-I (skeletal human) | 1000 ng/mL | <0.001  |

The effect of potentially interfering endogenous substances (hemoglobin, bilirubin and triglycerides) was tested by spiking a samples containing approximately 0.4, 1, 2 and 4 ng/mL TnI with the appropriate concentration of the test substance and comparing the TnI recovery of the sample to that of a control sample. Hemoglobin up to 332 uM (2,178 mg/dL), bilirubin up to 510 uM (29 mg/dL), and triglycerides up to 30 g/L did not interfere with the test. Additionally, the device package insert recommends that samples that appear hemolyzed, icteric, or lipemic not use used and that, if possible, a new sample be collected.

The effect of potentially interfering common and prescription drugs was tested at concentrations in excess of 100 times the expected dose level in plasma. A list of these substances can be found in the package insert.

f. Assay cut-off:

The sponsor determined the cutoff to be 0.11 ng/mL based on the LoQ. (See precision section 1. 1. above)

2. Comparison studies:

a. Method comparison with predicate device:

Lithium heparin plasma samples (n=534) were assayed on the VIDAS TNIU and another commercially available TnI method. Samples ranged in value

{6}

from  $&lt; 0.01$  to  $19.02\mathrm{ng / mL}$  on the VIDAS TNIU method. The data were evaluated using Passing-Bablok regression (slope  $= 0.42$ , confidence interval 0.38-0.44) and the Pearson's correlation coefficient was calculated (0.97).

For the samples obtained from the US and French study sites, the concordance between the VIDAS TNIU method and a commercially available method is shown in the table below:

|  Times | 0-6 hours after admission | 4-12 hours after first collection  |
| --- | --- | --- |
|  US Site Samples (n) | 302 | 300  |
|  Concordance % | 94.4% | 95.3%  |
|  French Site Samples (n) | 243 | 153  |
|  Concordance | 87.2% | 94.8%  |

# b. Matrix comparison:

In order to demonstrate equivalence between serum and lithium heparin plasma TnI results, thirty-seven (37) sample pairs were evaluated. Both spiked and native samples were included in the study and sample values ranged in value from 0.04 to  $38\mathrm{ng / mL}$ , which is slightly higher than the reportable range. Passing and Bablock regression analysis yielded the following equation with  $95\%$  confidence intervals:

Serum (y) = 0.95 [0.91; 0.97] Plasma (x) - 0.01 [-0.09; 0.11]

The labeling recommends that the same collection tube type, serum or lithium heparinized plasma, should be used when performing serial sampling.

# 3. Clinical studies:

# a. Clinical Sensitivity:

The clinical sensitivity and specificity studies were performed at two clinical sites, one in the United States (US) at a community hospital and one in France at a specialized cardiac hospital. For 16 weeks, assays were performed using samples from patients with chest pain and suspected myocardial infarction. The samples were collected from the patients within 6 hours following their admission, and then between 4 and 12 hours after the first sample collection.

At the US site a total of 602 samples from patients admitted to the emergency department (ED) were collected:

- Within 0-6 hours after admission, out of the 302 patients concerned, 36 were diagnosed as having myocardial infarction.

{7}

- Within 4-12 hours after the first sample collection, out of the 300 patients concerned who met the study inclusion criteria, 36 were diagnosed as having myocardial infarction.

The clinical sensitivity and specificity of VIDAS Troponin I Ultra was determined for a sample collected at the time of admission and the one subsequently collected, using the assay cutoff 0.11 ng/mL. The data are presented in the table below.

US Site

|  Times | 0-6 hours after admission | 4-12 hours after first collection  |
| --- | --- | --- |
|  Samples (n) | 302 | 300  |
|  Sensitivity % | 69.44 | 97.22  |
|  95% Confidence interval | 52.80 - 82.20 | 85.48-99.52  |
|  Specificity % | 96.24 | 96.21  |
|  95% Confidence interval | 93.14 - 97.97 | 93.09-97.95  |

At the French site a total of 396 samples from patients admitted to the Cardiology Intensive Care Unit were collected:

- Within 0-6 hours after admission, out of the 243 patients concerned, 116 were diagnosed as having myocardial infarction.
- Within 4-12 hours after the first sample collection, out of the 153 patients concerned who met the study inclusion criteria, 77 were diagnosed as having myocardial infarction.

The clinical sensitivity and specificity of VIDAS Troponin I Ultra was determined for a sample collected at the time of admission and the one subsequently collected using the assay cutoff of 0.11 ng/mL. The data are presented in the table below.

French site

|  Times | 0-6 hours after admission | 4-12 hours after first collection  |
| --- | --- | --- |
|  Samples (n) | 243 | 153  |
|  Sensitivity % | 78.45 | 98.70  |
|  95% Confidence interval | 69.93 – 85.07 | 92.82 – 99.78  |
|  Specificity % | 90.55 | 92.11  |
|  95% Confidence interval | 84.05 – 94.57 | 83.61 – 96.39  |

{8}

b. Clinical specificity:
See Clinical Sensitivity section above.

c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.

4. Clinical cut-off:
See section 1.f. assay cutoff above.

5. Expected values/Reference range:
In a study of 399 heparinized plasma samples from apparently healthy individuals in the United States, the upper limit of the 99th percentile for the method was determined to be 0.01 ng/mL.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K063243](https://fda.innolitics.com/device/K063243)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
