MEDRAD STELLANT CT INJECTOR SYSTEM WITH XDS ACCESSORY

K063090 · Medrad, Inc. · IZQ · Mar 27, 2007 · Cardiovascular

Device Facts

Record IDK063090
Device NameMEDRAD STELLANT CT INJECTOR SYSTEM WITH XDS ACCESSORY
ApplicantMedrad, Inc.
Product CodeIZQ · Cardiovascular
Decision DateMar 27, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1650
Device ClassClass 2

Indications for Use

The MEDRAD Stellant CT Injector System is intended for the specific purpose of injecting intravenous contrast media into humans for diagnostic studies in computed tomography (CT) applications. The MEDRAD Extravasation Detection System (XDS) accessory is designed to aid in the detection of intravenous contrast media from the injection site into the surrounding tissue during diagnostic studies in computed tomography (CT) applications.

Device Story

MEDRAD Stellant CT Injector System delivers intravenous contrast media for CT scans; XDS accessory provides supplemental monitoring for extravasation. XDS uses gold-plated antenna sensor at injection site; detects fluid leakage into surrounding tissue. If extravasation detected (up to 20ml), system triggers audible/visual alerts and halts injector. Operated by clinicians in CT suites; interfaces exclusively with Stellant models Sx and Dx. Provides safety monitoring to supplement standard clinical diligence; reduces potential patient harm from contrast extravasation.

Clinical Evidence

Bench testing only. No clinical data provided. Performance validated via technical specifications and functional testing of extravasation detection capabilities.

Technological Characteristics

Syringe-based fluid delivery; PET barrel; polycarbonate plunger; silicone coating. XDS sensor: gold-plated antenna, ferro-plastic absorber, silicone overmold. Connectivity: proprietary interface to Stellant Sx/Dx injectors. Sterilization: EtO. Materials: Medical grade PVC, ABS, polycarbonate, polyester/acrylate tapes. Compliant with ISO 594-1/2 for Luer fittings.

Indications for Use

Indicated for adults undergoing CT diagnostic studies requiring intravenous contrast media injection via antecubital site. Contraindicated for pediatric patients, back-of-hand injections, portable use, and MR environments.

Regulatory Classification

Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ## 510(K) SUMMARY K063090 | OFFICIAL CONTACT: | Troy A. Jack<br>Sr. Regulatory Affairs Specialist<br>MEDRAD, Inc.<br>One MEDRAD Drive<br>Indianola, PA 15051<br>(412) 767-2400 ext. 3305 | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------| | CLASSIFICATION NAME: | Injector with Syringe, Angiographic | | COMMON NAME(S): | Powered Injector with Syringe | | PROPRIETARY NAME: | MEDRAD Stellant CT Injector System with XDS Accessory | | PREDICATE DEVICE: | MEDRAD Stellant CT Injector System with Imaging System<br>Interface Module (K033881) | MAR 27 2007INTENDED USE: The MEDRAD Stellant CT Injector System is intended for the specific purpose of injecting intravenous contrast media into humans for diagnostic studies in computed tomography (CT) applications. The MEDRAD Extravasation Detection System (XDS) accessory is designed to aid in the detection of intravenous contrast media from the injection site into the surrounding tissue during diagnostic studies in computed tomography (CT) applications. CONTRAINDICATIONS: The XDS System is not intended for portable use or for use in an MR environment. The XDS System is designed for use on adults in antecubital injections only. The system is not validated for use on the back of the hand or in pediatrics. The Control Room Module/Stellant Interface signal output is not intended to be used with any equipment other than with Medrad Stellant CT Injector models Sx and Dx. DEVICE DESCRIPTION AND COMPARISON TO PREDICATE: The XDS Accessory may be added to the MEDRAD Stellant CT Injector System as an optional accessory. The injector system, when used with the XDS accessory, maintains the same intended use, same operational parameters, same labeling (with the addition of an XDS-specific operations manual). This device is also used in the same manner as the predicate device. The Stellant CT Injector System with XDS accessory is a syringe-based fluid delivery system indicated for delivery of contrast media during computed tomography procedures. The MEDRAD Stellant CT Injector System is intended for the specific purpose of injecting intravenous contrast media into humans for diagnostic studies in computed tomography (CT) applications. The MEDRAD Extravasation Detection System (XDS) accessory is designed to aid in the detection of intravenous contrast media from the injection site into the surrounding tissue during diagnostic studies in computed tomography (CT) applications. The XDS accessory is designed to detect and notify the user before 20ml of fluid is extravasated within the sensors' field of view. {2}------------------------------------------------ The XDS accessory does not replace or change the existing duties of the healthcare professionals performing a CT scan. This accessory provides a supplemental monitoring function to normal extravasation diligence and therefore does not impose additional risk to the patient. The device will exclusively interface with the Stellant CT Injector System, halting it in the event of an extravasation. This accessory does not otherwise alter the injector system's primary functions or performance during a CT procedure. Differences between the predicate device and the Stellant CT Injector System with XDS functionality are detailed in the Tables 1 and 2: {3}------------------------------------------------ # Table 1 - COMPARISON OF STELLANT CT INJECTOR SYSTEM w/ XDS ACCESSORY and STELLANT CT INJECTOR SYSTEM w/ ISI ... | Feature | Predicate Device:<br>Stellant CT Injector System with ISI<br>Accessory (K033881) | Proposed Device:<br>Stellant CT Injector System with<br>XDS Accessory | | |-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------| | Intended Use | The MEDRAD Stellant CT Injector<br>System with ISI Module is intended<br>for the specific purpose of injecting<br>intravenous contrast media into<br>humans for diagnostic studies in<br>computed tomography (CT)<br>applications. | The MEDRAD Stellant CT Injector<br>System with ISI Module is intended<br>for the specific purpose of injecting<br>intravenous contrast media into<br>humans for diagnostic studies in<br>computed tomography (CT)<br>applications.<br>The XDS accessory is designed<br>specifically to aid in the detection of<br>leakage of intravenous contrast<br>media from the injection site into<br>the surrounding tissue during<br>diagnostic studies in computed<br>tomography (CT) applications. | | | | Single or Dual Syringe<br>System | Single and dual syringe models | Same | | | Information Display | Color LCD | Same | | | Programming Keys | Non-dedicated keys - software<br>determined | Same | | | Touch screen | Yes | Same | | | Multi-Phase | 1 - 6 phases per injection | Same | | | Arming Modes | Single | Same | | Protocol Storage<br>Capability | 32 protocols | Same | | | Hold Capability | 20 minutes max. | Same | | | Scan Delay | 1 – 300 seconds | Same | | | Safety Stop Mechanism | Multi-layered software stops with<br>backup monitoring | Same | | | Syringe System | Single syringe model: one 200 ml<br>syringe<br>Dual syringe model: two 200 ml<br>syringes | Same | | | | Programmed Volume | 1 to 200 ml | Same | | | Volume Remaining<br>Readout | LED on injector head; graphical and<br>numeric on LCD | Same | | | Fill Rate | Variable up to 10 mL/sec | Same | | Flow Rate | 0.1mL/sec to 10.0 mL/sec | Same | | | Programmable<br>Pressure Limit | 325 PSI default; user settable 50 to 325<br>PSI | Same | | | Pause | Programmable - 1 sec to 900 sec in 1<br>sec increments | Same | | | Autofill | Fill rate 4 mL/sec | Same | | | Retract Control | Yes (Automatic) | Same | | | Remote Start Switch | Yes | Same | | | Feature | Predicate Device:<br>Stellant CT Injector System with ISI<br>Accessory (K033881) | Proposed Device:<br>Stellant CT Injector System with<br>XDS Accessory | | | Pressure Graph | Yes | Same | | | Syringe Sensing | Yes | Same | | | Autoload | Yes | Same | | | Auto<br>Dock/Retract/Advance | Yes; user-selectable autodock and<br>advance; user-selectable auto-retract | Same | | | Protocol Lock / Remote<br>Arming | Yes | Same | | | Remote Check for Air<br>(from Head) | Yes | Same | | | Scan Delay | 1 sec to 300 sec in 1 sec increments | Same | | | Store/Recall | 32 protocols | Same | | | Test Inject | Yes | Same | | | Syringe Heat<br>Maintainer | Yes | Same | | | Flow Profile Display | Yes | Same | | | Imaging System<br>Interface Functionality | Yes | Same | | | Patient Sensor | N/A | Gold-plated antenna shrouded in<br>ferro-plastic absorber covered by a<br>silicone overmold | | | Extravasation Detector<br>Performance | N/A | Audible and visual alert if an<br>extravasation of up to 20ml is<br>detected under sensor region. Will<br>halt the Stellant injector system in<br>the event of an extravasation event. | | | Feature | Predicate Device:<br>Stellant CT Injector System<br>(K033881) | Proposed Device:<br>Stellant CT Injector System<br>with XDS Accessory | | | Administration Set Packaging | Tyvek lid covering polystyrene<br>tray | Same | | | Sterilization | EtO sterilization | Same | | | Method of determining<br>Pyrogen-Freeness | Limulus Amebocyte Lysate test to<br>USP requirements | Same | | | Luer Fittings | ISO 594-1 & ISO 594-2 compliant<br>design | Same | | | Syringe: | | | | | Barrel Material Composition | Polyethylene Terephthalate | Same | | | Barrel Length | 7.504" | Same | | | Barrel OD | 2.002" | Same | | | Barrel ID | 1.844" | Same | | | Plunger Material Composition | Polycarbonate | Same | | | Plunger Cover | Polyisoprene | Same | | | Plunger Silicone Coating | Polydimethylsiloxane (Silicone) | Same | | | Barrel Flange | Easy-engage design for non-<br>rotational orientation (no<br>alignment necessary) | Same | | | Syringe sensing feature | Grooves at bottom of barrel to be<br>optically identified | Same | | | Connector Tubing<br>Components: | | | | | Maximum Pressure | 400 psi | Same | | | Tube Material | Medical grade PVC | Same | | | Tube Length | 60" | Same | | | Bonding Agent | Cyclohexanone | Same | | | T-connector | Medical grade polycarbonate | Same | | | Contrast Media/Saline<br>Container Spike | ABS | Same | | | Priming Tube | Low density polyethylene | Same | | | XDS Disposable Components: | | | | | Sensor Interface Disposable<br>(SID) | N/A | Medical Grade Tape<br>(Polyester) with Acrylate<br>Adhesive | | | SID Packaging | N/A | Polyester + Polyethylene with<br>Paper Backing | | | SID Sterilization | N/A | EtO sterilization | | | Cable Management Tape (Quick<br>Strips) | N/A | Medical Grade Tape<br>(Polyethylene) with Acrylate<br>Adhesive | | | Cable Management Tape<br>(Quick Strips) Packaging | N/A | Polyester + Polyethylene with<br>Paper Backing | | | Cable Management Tape<br>(Quick Strips) Sterilization | N/A | EtO sterilization | | {4}------------------------------------------------ and the count {5}------------------------------------------------ # Table 2 - COMPARISON OF STELLANT CT INJECTOR SYSTEM AND XDS ACCESSORY DISPOSABLES ... {6}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms. #### Public Health Service Food and Drug Administration 200 Corporate Boulevard Rockville MD 20850 MAR 2 7 2007 Medrad, Inc. c/o Mr. Troy A. Jack Senior Regulatory Affairs Specialist One Medrad Drive Indianola, PA 15051 Re: K063090 Trade/Device Name: Stellant CT Injector System with Extravasation Detector Accessory Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector and Syringe Regulatory Class: Class II Product Code: IZQ Dated: March 21, 2007 Received: March 23, 2007 Dear Mr. Jack: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {7}------------------------------------------------ Page 2 - Mr. Troy A. Jack Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 /CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Blommor for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K063090 Device Name: MEDRAD Stellant CT Injector System with Extravasation Detection Accessory Indications for Use: The MEDRAD Stellant CT Injector System is intended for the specific purpose of injecting intravenous contrast media into humans for diagnostic studies in computed tomography (CT) applications. The MEDRAD Extravasation Detection System (XDS) accessory is designed to aid in the detection of intravenous contrast media from the injection site into the surrounding tissue during diagnostic studies in computed tomography (CT) applications. Contraindications: The XDS System is not intended for portable use or for use in an MR environment. The XDS System is designed for use on adults with antecubital injections only. The system is not validated for use on the back of the hand or in pediatrics. The Control Room Module / Stellant Interface signal output is not intended to be used with any equipment other than with MEDRAD Stellant CT Injectors models SX and DX. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) B concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Off Division of Cardiovascular Devices 510(k) Number K063090
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...