← Product Code FRO · K063063

# SILVACIDE SKIN & WOUND MOISTURIZING SOLUTION (K063063)

_Medical Molecular Therapuetics, LLC · FRO · Jul 3, 2007 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K063063

## Device Facts

- **Applicant:** Medical Molecular Therapuetics, LLC
- **Product Code:** FRO
- **Decision Date:** Jul 3, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Intended Use

Over-The-Counter Indications: For minor burns, abraded skin, irritated areas and minor wounds. Professional Prescription Indications: Wound moistening solution for dermal lesions such as Stage I-IV pressure ulcers, stasis ulcers, foot ulcers, diabetic ulcers, post surgical wounds, first and second degree burns, abrasions and skin irritations.

## Device Story

Silvion is a buffered ionic silver aqueous solution used as a wet dressing for external wound management. It provides a moist environment to aid healing and uses ionic silver as an antimicrobial agent to protect the solution from microbial contamination. The product is clear, odorless, and colorless, supplied in spray bottles (2 oz to 16 oz). It is applied topically to skin, mucous membranes, and acute or chronic wounds by clinicians or patients. The solution maintains a moist wound environment, which is the primary mode of action, while the silver component provides antimicrobial protection. It is intended for use in clinical settings or at home for wound care management.

## Clinical Evidence

No clinical trials were performed. Evidence consists of bench testing including biocompatibility (ISO Modified Intracutaneous Study, USP/ISO Modified Systemic Toxicity, ISO Maximization Sensitization) and cytotoxicity (Agarose Overlay Method). Antimicrobial effectiveness was validated via MIC/MBC dilution methods, Zone of Inhibition, USP Antimicrobial Effectiveness Test <51>, USP Microbial Limit Test <61>, Bioburden Aerobic Total Count, and Microbial Challenge Test.

## Technological Characteristics

Buffered ionic silver aqueous solution. Form factor: liquid spray. Materials: ionic silver, aqueous buffer. Sterilization: not specified. Connectivity: none. Software: none. Sensing/actuation: chemical/physical moisture provision and antimicrobial action.

## Predicate Devices

- Dermacyn Wound Cleanser ([K042729](/device/K042729.md))
- Saline Solution (Wound Dressing) ([K040683](/device/K040683.md))
- Restore Wound Cleanser ([K022670](/device/K022670.md))
- Dermacyn Wound Irrigation ([K042729](/device/K042729.md))
- X-Static Silverseal Hydrogel Wound Dressing ([K040019](/device/K040019.md))
- Acryderm Silver Antimicrobial Wound Gel ([K011994](/device/K011994.md))
- Antimicrobial Barrier Wound Contact Dressing ([K023612](/device/K023612.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Medical Molecular Therapeutics LLC 510(k) PreMarket Notification K063063 Silvion Antibacterial Silver Skin & Wound Moisturizing Solution Rev 15 May 07

# 7. 510(k) Summary Information

JUL - 3 2007 In accordance with 21 CFR 807.92, the 510(k) Summary is presented to provide an understanding for the basis of substantial equivalence.

## 7.1 Submitter's & Owners Name and Address

Arthur B. Flick MD Medical Molecular Therapeutics LLC 36 Lake Rabun Road Lakemont, Georgia 30552

Contact Person: Arthur B. Flick MD Telephone: 706 782 5064 Email: abflick@alltel,net

## 7.2 Date Prepared

20 September 2006

#### 7.3 Device Name

| Trade Name:          | Silvion Antibacterial Silver Skin & Wound<br>Moisturizing Solution |  |  |  |  |  |
|----------------------|--------------------------------------------------------------------|--|--|--|--|--|
| Common Name:         | Skin & Wound Moisturizing Solution                                 |  |  |  |  |  |
| Classification Name: | Liquid Bandage KMF                                                 |  |  |  |  |  |

## 7.4 Predicate Devices 807.92(a)(3)

| Dermacyn Wound Cleanser          | K 042729 |
|----------------------------------|----------|
| Saline Solution (Wound Dressing) | K040683  |
| Restore Wound Cleanser           | K022670  |
| Dermacyn Wound Irrigation        | K042729  |

{1}------------------------------------------------

Medical Molecular Therapeutics LLC 510(k) PreMarket Notification K063063 Rev 15 May 07 Silvion Antibacterial Silver Skin & Wound Moisturizing Solution

| X-Static Silverseal Hydrogel Wound Dressing  | K040019 |
|----------------------------------------------|---------|
| Acryderm Silver Antimicrobial Wound Gel      | K011994 |
| Antimicrobial Barrier Wound Contact Dressing | K023612 |

#### 7.5 Device Description

A buffered ionic silver aqueous wet dressing for external wound management. The product provides a moist environment that aids the body in the healing process. The non-irritating components of Silvion Antibacterial Silver Skin & Wound Moisturizing Solution allow the product to be used on mucous membranes, the skin as well as acute and chronic wounds. The silver provides effective protection of the solution against microbial contamination. The Silvion Antibacterial Silver Skin & Wound Moisturizing Solution is clear, odorless and colorless. The product is available in 2 oz, 4 oz, 8 oz and 16 oz spray bottles.

#### 7.6 Assessment of Performance Data

Silvion Antibacterial Silver Skin & Wound Moisturizing Solution has been subject to in vitro and in vivo biocompatibility (ISO Modified Intracutaneous Study, the USP and ISO Modified Systemic Toxicity Study and the ISO Maximization Sensitization) and cytotoxicity testing (Agarose Overlay Method). These tests support the safe use of Silvion Antibacterial Silver Skin & Wound Moisturizing Solution in contact with breached or compromised skin. In vitro antimicrobial testing was assessed by the standard the MIC and MBC Dilution Methods, Zone of Inhibition, USP Antimicrobial Effectiveness Test <51>. USP Microbial Limit Test <61>, Bioburden Aerobic Total Count and Microbial Challenge Test.

#### 7.7 Statement of Intended Use

OTC: Minor burns, abraded skin, irritated areas minor wounds.

Professional: Stage I-IV pressure ulcers, stasis ulcers, diabetic ulcers, post surgical wounds, first and second degree burns, abrasions and skin irritations.

б

{2}------------------------------------------------

510(k) PreMarket Notification K063063 Medical Molecular Therapeutics LLC Rev 15 May 07 Silvion Antibacterial Silver Skin & Wound Moisturizing Solution

### 7.8 Technological Characteristics and Substantial Equivalence

Silvion Antibacterial Silver Skin & Wound Moisturizing Solution is a combination product within the meaning of section 503(g) of the Federal Food, Drug, and Cosmetic Act (Act) and Title 21 of the Code of Federal Regulations (CFR) section 3.2(e)(1) and (2). In accordance with section 503(g)(1) of the Act and 21 CFR section 3.4, Silvion Antibacterial Silver Skin & Wound Moisturizing Solution had two modes of action. One action of the product is the device components' action to provide moisture to dermal wounds and inflamed skin. Another mode of action of the product is that of the drug component (ionic silver) to act as an antimicrobial agent. The primary mode of action of the combination product is attributable to the device components' action to provide moisture to dermal wounds and inflamed skin.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, superimposed over three horizontal lines that represent waves. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Medical Molecular Therapeutics LLC % Arthur B. Flick, MD 36 Lake Rabun Road Lakemont, Georgia 30552

JUL - 3 2007

Re: K063063

Trade/Device Name: Silvion™ Antibacterial Silver Skin & Wound Moisturizing Solution Regulatory Class: Unclassified Product Code: FRO Dated: May 22, 2007 Received: May 24, 2007

Dear Dr. Flick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally 

{4}------------------------------------------------

Page 2 - Arthur B. Flick, MD

marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

for
Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Medical Molecular Therapeutics LLC 510(k) PreMarket Notification K063063 Silvion Antibacterial Silver Skin & Wound Moisturizing Solution Rev 15 May 07

# 5. INDICATIONS FOR USE

510(k) Number: K063063/S1

Device Name: Silvion™ Antibacterial Silver Skin & Wound Moisturizing Solution

Indications For Use:

Over-The-Counter Indications:

For minor burns, abraded skin, irritated areas and minor wounds.

Professional Prescription Indications:

Wound moistening solution for dermal lesions such as Stage I-IV pressure ulcers, stasis ulcers, foot ulcers, diabetic ulcers, post surgical wounds, first and second degree burns, abrasions and skin irritations.

| Prescription Use            | X | AND/OR | Over-The-Counter Use   | X |
|-----------------------------|---|--------|------------------------|---|
| (Part 21 CFR 801 Subpart D) |   |        | (21 CFR 801 Subpart C) |   |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IS NEEDED)

(Division Sign)

Concurrence of CDRH Division of General and Neurologic Office of Device Evaluation (ODE)

12063062

510

---

**Source:** [https://fda.innolitics.com/device/K063063](https://fda.innolitics.com/device/K063063)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
