K063047 · Medhot Thermal Imaging, Inc. · LHQ · Oct 19, 2006 · Obstetrics/Gynecology
Device Facts
Record ID
K063047
Device Name
MEDHOT MTI 2000 THERMAL IMAGING SYSTEM
Applicant
Medhot Thermal Imaging, Inc.
Product Code
LHQ · Obstetrics/Gynecology
Decision Date
Oct 19, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2980
Device Class
Class 1
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Intended for qualified healthcare personnel who are trained in its use, the MedHot MTI 2000 is a non-contact, non-invasive, non-radiating, thermographic imaging device for use as an adjunct to other clinical diagnostic procedures in the capture, quantification, and screening, of differences in skin surface temperature changes. The device can digitally render quantitative images of thermal data via the capture and algorithmic interpretation of relative infrared information from the imaging camera in collaborative use with a compatible computer system which provides a user interface and image storage. Patient populations and environmental use includes all age groups from adult to pediatric and neonatal in hospital, sub acute and public areas i.e. airports.
Device Story
MedHot MTI 2000 is a non-invasive telethermographic system; captures infrared radiation via FPA microbolometer camera; converts data array to temperature data using algorithmic derivative; renders quantitative thermal images on computer screen. Operated by trained healthcare personnel in hospitals, sub-acute, or public areas (e.g., airports). System includes camera and computer with onboard CPUs; provides GUI for image visualization, quantification, and storage. Used as adjunct to clinical diagnostic procedures; assists providers in screening and quantifying skin surface temperature differences; aids clinical decision-making by providing visual thermal data.
Clinical Evidence
No clinical data. Bench testing only; device successfully tested for EMC and EMI compliance per IEC 60601-1-2.
Technological Characteristics
FPA UC microbolometer detector; amorphous silicon material; spectral response 7-14 µm; thermal sensitivity < 100 mK; 160x120 array resolution; manual focus; analog NTSC video output; 20 fps refresh rate; emissivity correction 0.1-1.0; connectivity via computer interface; power 100 VAC.
Indications for Use
Indicated for use by trained healthcare personnel for non-contact, non-invasive, non-radiating thermographic imaging as an adjunct to clinical diagnostic procedures. Used for capture, quantification, and screening of skin surface temperature changes in all age groups, including adults, pediatric, and neonatal patients in hospital, sub-acute, and public settings.
Regulatory Classification
Identification
A telethermographic system for adjunctive diagnostic screening for detection of breast cancer or other uses is an electrically powered device with a detector that is intended to measure, without touching the patient's skin, the self-emanating infrared radiation that reveals the temperature variations of the surface of the body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Telethermographic camera Series A, E, P, and S (k033967)
Submission Summary (Full Text)
{0}------------------------------------------------
# KO63047
## Exhibit C
510(k) Summary
0C1 1 9 2006
| Name and address of sponsor of the<br>510(k) submission: | MedHot Thermal Imaging, Inc. |
|------------------------------------------------------------|------------------------------------------------|
| | 315 Doris Drive, Lakeland. Florida - 33813 |
| Official contact person for all<br>correspondence: | Carol Chandler (address as above) |
| | Phone: 863 709 9565, Fax: 863 619 7525 |
| | E-mail: infrared@tampabay.rr.com |
| Date Prepared: | Sept 27, 2006 |
| Device Name: | MedHot MTI 2000 Thermal Imaging System |
| Generic name of the device: | Telethermographic System (Adjunctive) |
| Classification, Product Code and<br>CFR Regulation Number: | Class I, LHQ and 21 CFR 884.2980 |
| Classification Panel: | Obstetrics/Gynecology |
| Predicate Device Name: | Telethermographic camera Series A, E, P, and S |
| Predicate 510(k) Number: | k033967 |
#### Device Description:
The MedHot MTI 2000 thermal imaging system is a non-invasive system that provides the user a visual representation of thermal temperature data in the form of a quantitative image on the users computer screen. The process begins by the cameras' capture of a data array representative of infrared radiation naturally emitted by the object being imaged; the data array is converted to a temperature data array using an algorithmic derivative. Once the infrared data is in the form of temperature data, software on the user's computer is used to generate an image from the rendered temperature data. The user can use this image in a number of ways to quantify the temperature of objects imaged.
014
{1}------------------------------------------------
| | MedHot MTI 2000 | A20M |
|---------------------------|------------------------------------------------|------------------------------------------------|
| Technology / Detector | FPA UC Microbolometer | FPA UC Microbolometer |
| Material | Amorphous Silicon | Amorphous Silicon |
| Spectral Response | 7-14 µm | 7.5-13 µm |
| Thermal Sensitivity | < 100 mK | < 80 mK |
| Contrast / Brightness | Software Controlled | Software Controlled |
| Data Output | Analog NTSC Video | IEEE-1394 usb firewire |
| Refresh Rate | 20 fps Refresh | 60 fps Refresh |
| Optics | 8.5mm with 50°x 35° | 24° x 18°/ 0.3 m |
| Array Resolution | 160x120 array | 320x240 array |
| Focus | manual | manual |
| Power Supply | 2.0 amps @100 VAC | AC adapter 110/220 VAC,<br>50/60Hz |
| Weight | 2.5 pounds | 1.7 pounds |
| Size | 7.6" / 5.0" / 2.2" | 6.2" / 2.9" / 3.1" |
| Operating Temperatures | "-20°C to +60°C (-4°F<br>to<br>+140°F)" | "-40°C to +70°C (-40°F to<br>+158°F)" |
| Emissivity Correction | Variable from 0.1 to 1.0 | Variable from 0.1 to 1.0 |
| Method of Data Collection | collection instrumentation | |
| data processing | camera and computer each<br>have onboard cpu's | camera and computer each<br>have onboard cpu's |
| Measurement Parameters | | |
| storage | data storage on computer | data storage on computer |
| Camera Output | | |
| display | GUI on computer | GUI on computer |
| user interface | Software on computer | Software on computer |
## Comparison of Technological Characteristics to Predicate Device:
#### Intended Use:
Intended for qualified healthcare personnel who are trained in its use; The MedHot MTI 2000 is a non-contact, non-invasive, non-radiating, thermographic imaging device for use as an adjunct to other clinical diagnostic procedures in the capture, quantification, and screening, of differences in skin surface temperature changes.
The device can digitally render quantitative images of thermal data via the capture and algorithmic interpretation of relative infrared information from the imaging camera in collaborative use with a compatible computer system which provides a user interface and image storage.
Patient populations and environmental use includes all age groups from adult to pediatric and neonatal in hospital, sub acute and public areas i.e. airports.
{2}------------------------------------------------
## Non-Clinical Testing:
Camera unit was successfully tested for EMC and EMI compliance as per IEC 60601-1-2 requirements. Power supply also conforms to appropriate medical grade stines on
#### Clinical Testing:
Not applicable
#### Conclusion:
MedHot MTI 2000 is substantially equivalent to the predicate device and arguments given in the submission show that the differences don't bring up any new questions regarding the safety and effectiveness of the subject device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird-like figure. The logo is black and white.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
MedHot Thermal Imaging, Inc. c/o Mr. Robert Mosenkis President CITECH 5200 Butler Pike PLYMOUTH MEETING PA 19462-1298
## OCT 1 9 2006
Re: K063047
Trade/Device Name: MedHot MTI 2000 Thermal Imaging System Regulation Number: 21 CFR §884.2980 Regulation Name: Telethermographic system Regulatory Class: I Product Code: LHQ Dated: September 27, 2006 Received: October 4, 2006
Dear Mr. Mosenkis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image shows a circular logo with the letters "FDA" in the center. Above the letters, the numbers "1906 - 2006" are displayed. Below the letters, the word "Centennial" is written in cursive. The logo is surrounded by a dotted circle.
coling Public .
{4}------------------------------------------------
Page 2 -
Please be advised that I'DA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
Page _ 1 of 1
K063047
## 510(k) Number (if known): No
#### Device Name: MedHot MTI 2000 Thermal Imaging System
#### Indications for Use:
Intended for qualified healthcare personnel who are trained in its use, the MedHot MTI 2000 is a non-contact, non-invasive, non-radiating, thermographic imaging device for use as an adjunct to other clinical diagnostic procedures in the capture, quantification, and screening, of differences in skin surface temperature changes.
The device can digitally render quantitative images of thermal data via the capture and algorithmic interpretation of relative infrared information from the imaging camera in collaborative use with a compatible computer system which provides a user interface and image storage.
Patient populations and environmental use includes all age groups from adult to pediatric and neonatal in hospital, sub acute and public areas i.e. airports.
Prescription Use (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David h. Seymon
(Division Sign-Off)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K063
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.