OK METER BLOOD GLUCOSE MONITORING SYSTEM

K063026 · Ok Biotech Co., Ltd. · NBW · Feb 9, 2007 · Clinical Chemistry

Device Facts

Record IDK063026
Device NameOK METER BLOOD GLUCOSE MONITORING SYSTEM
ApplicantOk Biotech Co., Ltd.
Product CodeNBW · Clinical Chemistry
Decision DateFeb 9, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The OK Meter Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger. Testing is done outside the body (in vitro diagnostic use). It is indicated for both lay uses by people with diabetes, and in a clinical setting by health care professionals, as an aid to monitoring levels in Diabetes Mellitus. Not for use on neonates.

Device Story

OK Meter Blood Glucose Monitoring System; electrochemical biosensor; capillary action. Input: 0.7 μL fresh capillary whole blood applied to test strip. Operation: glucose oxidase enzyme reaction on strip electrodes; current measured by meter proportional to glucose concentration. Output: blood glucose result displayed in 6 seconds. Used by patients (home) or healthcare professionals (clinic). User enters batch-specific code from strip vial into meter for calibration. Output used to monitor glucose levels in Diabetes Mellitus; aids clinical decision-making for glycemic control.

Clinical Evidence

Pre-clinical and clinical data were submitted in accordance with FDA Guidance for In Vitro Diagnostic Test Systems. No specific performance metrics (e.g., sensitivity, specificity) are provided in the summary document.

Technological Characteristics

Electrochemical biosensor; glucose oxidase enzyme; amperometric detection. Sample volume: 0.7 μL. Measurement range: 20-600 mg/dL. Calibration: batch-specific code entry. Connectivity: none. Sterilization: not specified. Software: embedded firmware for signal conversion and display.

Indications for Use

Indicated for quantitative glucose measurement in fresh capillary whole blood from the finger. For use by people with diabetes (lay users) and healthcare professionals in clinical settings to monitor Diabetes Mellitus. Not for use on neonates.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo with the word "Biotech" in a stylized font. Above the word "Biotech" is a circular design with lines and curves. Below the word "Biotech" are Chinese characters. The logo appears to be for a company or organization related to biotechnology. TEL 886-3-5160258 FAX 886-3-5160028 TEL 86-791-3899362 FAX : 86-791-3880131 http www.okbrotech.com E-mail service dokbictech com K063026 ### 510(K) Summary of Safety and Effectiveness II. (Per 21 CFR 807.92) 2.1. General Information Establishment FEB 9 2007 트 Manufacturer: OK Biotech Co., Ltd. 1F, No.87, Sec. 2, Gongdao 5th Road. Hsinchu City, Taiwan. 30070 트 Address: 트 Registration Number: 3005862821 를 Contact Person: Dr. Jen, Ke-Min Official Correspondent 886-3-5208829 (Tel) 886-3-5209783 (Fax) . Date Prepared: September 25, 2006 Device . Proprietary Name: OK Meter Blood Glucose Monitoring System ● Common Name: Blood Glucose Monitoring System ● Classification Name: SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER, Class II, ## 2.2. Safety and Effectiveness Information - . Predicate Device: Claim of Substantial Equivalence (SE) is made to ROCHE DIAGNOSTICS CORP. -- Accu-Chek Active Test System (K012324) - . Device Description: The OK Meter Blood Glucose Monitoring System based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. The OK Meter Blood Glucose Monitoring System only needs a 0.7 ul small amount of blood. Capillary action both side near the end of the test strip that draws the blood into the action chamber and the blood glucose result is precisely and displayed in 6 seconds. 3 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a logo with the letters "KR" stacked on top of each other. To the right of the letters is the word "Biotech" in a stylized font. Below the logo is a line of text in Chinese characters. The logo appears to be for a company or organization in the biotechnology field. TEL : 886-3-5160258 FAX : 886-3-5160028 TEL 86-791-3899362 FAX : 86-791-3880131 #### Intended Use: . The OK Meter Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the Testing is done outside the body ( in vitro diagnostic use). It is finger. indicated for both lay uses by people with diabetes and in a clinical setting by health care professionals, as an aid to monitoring levels in Diabetes Mellitus. #### . Synopsis of Test Methods and Results Pre-clinical and clinical data are employed upon submission of this 510(K) premarket notification according to the Guidance Document for In Vitro Diagnostic Test System; Guidance for Industry and FDA document provided by CDRH/ FDA. #### . Substantial Equivalence (SE) A claim of substantial equivalence is made to ROCHE DIAGNOSTICS CORP. -- Accu-Chek Active Test System (K012324). Both of them have the same working principle and technologies. The differences are electric voltage, dimensions of the unit and strip, weight. There are no safety and effectiveness aspects arising from the subject device. They are substantially equivalent. Ke-Miz-Ter Ke-Min Jen, Dr. Official Correspondent for OK Biotech Co., Ltd. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an emblem that features a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus is depicted with a staff entwined by two snakes and topped with wings. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ke-Min Jen OK Biotech Co., Ltd. No. 58, Fu Chiun Street Hsin Chu City, 30067 Taiwan, R.O.C. 9 2007 FF k063026 Re: Trade/Device Name: OK Meter Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, CGA, JJX Dated: January 27, 2007 Received: January 30, 2007 Dear Dr. Jen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jean M. Cooper, M.S., D.V.M. Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use K063026 510 (K) Number ( If Known ):__________________________________________________________________________________________________________________________________________________ Device Name: OK Meter Blood Glucose Monitoring System Indications for Use: The OK Meter Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger. Testing is done outside the body ( in vitro diagnostic use). It is indicated for both lay uses by people with diabetes and in a clinical setting by health care professionals, as an aid to monitoring levels in Diabetes Mellitus. Not for use on neonates. Prescription Use ## AND/OR Over-The-Counter Use √ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) AC H Division Sign-Off Page 1 of 1 Office of In Vitro Diagnostic Device Evaluation and Safety 510(KK) 06 3026
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...