The LADARWave® CustomCornea® Wavefront System is used for measuring, recording, and analyzing visual aberrations (such as myopia, hyperopia, astigmatism, coma and spherical aberration) and for displaying refractive error maps of the eye to assist in prescribing refractive corrections. This device is enabled to export wavefront data and associated anatomical registration information to a compatible treatment laser where a wavefront-guided treatment is indicated. This device can also export centration data and associated registration information for conventional phoropter-based refractive surgery with the LADAR6000™ System.
Device Story
LADARWave® CustomCornea® Wavefront System is an ophthalmic aberrometer using Hartmann-Shack wavefront sensing. Input: infrared probe beam reflected from the eye. Subsystems: fixation (adjusts for spherocylindrical error), video (operator eye positioning/registration), probe beam (infrared), and wavefront detection (imaging re-emitted wavefront). Operation: operator uses video imagery to align eye and record wavefront geometry relative to anatomical features. Output: refractive error maps and electronic files containing wavefront data, patient ID, and registration information. Used in clinical settings by eye care professionals. Data exported to compatible lasers (LADARVision®4000 or LADAR6000™) for customized ablative surgery. Assisted registration (overlay method) automates alignment, improving translational and cyclotorsional registration accuracy compared to manual methods. Benefits: enables precise wavefront-guided or conventional refractive surgery planning.
Clinical Evidence
Bench testing only. Equivalence study compared automated 'overlay' registration method against manual 'Sputnik' method. Sample: eye images collected for sessions with and without eye marks. Primary endpoints: accuracy of translational (x,y) and cyclotorsional registration. Results: 'overlay' method demonstrated accuracy as good as or better than 'Sputnik' method across all registration error values. Software verification and validation performed per FDA guidance.
Technological Characteristics
Hartmann-Shack wavefront sensing aberrometer. Infrared probe beam. Four optical subsystems: fixation, video, probe beam, wavefront detection. Proprietary software for data analysis and registration. Connectivity: export to removable media for compatible laser systems. Encrypted proprietary file format. Class I device.
Indications for Use
Indicated for patients requiring measurement and analysis of visual aberrations (myopia, hyperopia, astigmatism, coma, spherical aberration) to assist in prescribing refractive corrections and planning refractive surgery.
Regulatory Classification
Identification
An ophthalmic refractometer is an automatic AC-powered device that consists of a fixation system, a measurement and recording system, and an alignment system intended to measure the refractive power of the eye by measuring light reflexes from the retina.
Predicate Devices
LADARWave® CustomCornea® Wavefront System (K023249)
Reference Devices
LADARVision®4000
LADAR6000™ System
Submission Summary (Full Text)
{0}------------------------------------------------
K062930
# 510(k) Summary LADARWave® CustomCornea® Wavefront System
Alcon Research Ltd., on behalf of Alcon RefractiveHorizons 2501 Discovery Drive, Suite 500 Orlando, FL 32826
## A. Contact Person:
NOV - 8 2006
Janet G. Johnson Senior Manager Regulatory Affairs Telephone (407) 513-8954 Fax (407) 513-7855 Date Summary was Prepared: September 27, 2006
## B. Device Information:
| Proprietary Name: | LADARWave® CustomCornea® Wavefront System |
|-----------------------|-------------------------------------------|
| Common Name: | Abberrometer, Ophthalmic |
| Classification Name | Ophthalmic Refractometer |
| and Regulatory Class: | 21 CFR 886.7160 Class I |
| Product Code: | NCF |
## C. Substantial Equivalence Device (Predicate):
LADARWave® CustomCornea® Wavefront System (K023249)
### D. Indications for Use:
The LADARWave® CustomCornead Wavefront System is used for measuring, recording, and analyzing visual aberrations (such as myopia, astigmatism, coma and spherical aberration) and for displaying refractive error maps of the eye to assist in prescribing refractive corrections.
This device is enabled to export wavefront data and associated anatomical registration information to a compatible treatment laser where a wavefront-guided treatment is indicated.
This device can also export centration data and associated registration information for conventional phoropter-based refractive surgery with the LADAR6000" System.
## E. Device Description:
The LADARWave® CustomCornea® Wavefront System is an aberrometer, utilizing Hartmann-shack wavefront sensing to measure the aberrations in the human eye. The device contains four major optical subsystems used in the clinical wavefront examination.
- A fixation subsystem provides the patient with an unambiguous point of fixation. . Optics in this path automatically adjusts to correct for the patient's spherocylindrical error so that the target is clearly observed.
{1}------------------------------------------------
- A video subsystem provides the device operator with a view of the eye at the . measurement plane. The operator uses the video imagery to position the eye for the measurement and to record the geometry of the wavefront relative to anatomical features.
- A probe beam subsystem directs a narrow beam of eye-safe infrared radiation . into the eye to generate the re-emitted wavefront.
- A wavefront detection subsystem images the re-emitted wavefront onto the . entrance face of the Hartmann-shack wavefront sensor.
These subsystems are all under control of the device software.
Once the wavefront examination is complete, the operator may export the exam data to removable media. The exported electronic file contains all information necessary to perform customized ablative surgery using either the LADARVision®4000 or LADAR6000" System. The information includes the wavefront measurement, essential patient identification information, and geometric registration data. The electronic file is in a proprietary format and is encrypted so that wavefront data cannot be exported for use by an incompatible treatment device.
#### F. Substantial Equivalence Comparison:
The LADARWave® CustomCornea® Wavefront System with Assisted Registration has the same indications for use, intended use, and same technological characteristics (i.e., design, material, chemical composition, energy source) as the current LADARWave® CustomCornea® Wavefront System.
The LADARWave® CustomCornea® Wavefront System with Assisted Registration software has automated the registration process, allows the option for LADARWave wavefront measurements (pre-op) to be performed on an alternate day than the surgery, and allows centration data and associated registration data, for conventional surgery, to be performed on the LADARWave and exported to the LADAR6000™ System.
## G. Testing:
Eye images were collected to perform equivalence testing of the new automated registration process. For each eye, one LADARWave wavefront session (centration photo and five wavefront measurements with associated photos) with eye marks and a session without eye marks were taken for the testing. These images were compared using the current manual "Sputnik" method and the "Overlay" assisted registration method. The primary test was to establish equivalence of the two methods of registration by testing that the accuracy in translational (x,y) and cyclotorsional registration for the "overlay" method are at least as good as with the current "sputnik" method.
The results demonstrate that the "overlay" method is as good or better than the "sputnik" method in offset registration and rotational registration at all values of registration error (i.e. at all x-axis values).
Additionally, system level software verification and validation were successfully completed in accordance with General Principles of Software Validation; Final Guidance for Industry and FDA Staff dated January 11, 2002.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" surrounding it.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Alcon Research, Ltd. c/o Ms. Janet G Johnson Senior Manager, Regulatory Affairs 2501 Discovery Drive, Suite 500 Orlando. FL 32826
NOV - 8 2006
Re: K062930
Trade/Device Name: LADARWave® CustomCornea® Wavefront System Regulation Number: 21 CFR 886.1760 Regulation Name: Ophthalmic Refractometer Regulatory Class: I Product Code: NCF Dated: September 27, 2006 Received: September 28, 2006
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbrandijng and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your devire can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Janet G. Johnson
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Egelmisi MD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nosc and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### KO62930 510(k) Number (if known): Alcon LADARWave® CustomCornea® Wavefront System Device Name:
## Indications for Use Statement
The LADARWave® CustomCornea® Wavefront System is used for measuring, recording, and analyzing visual aberrations (such as myopia, hyperopia, astigmatism, coma and spherical aberration) and for displaying refractive error maps of the eye to assist in prescribing refractive corrections.
This device is enabled to export wavefront data and associated anatomical registration information to a compatible treatment laser where a wavefront-guided treatment is indicated.
This device can also export centration data and associated registration information for conventional phoropter-based refractive surgery with the LADAR6000™ System.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Arnl L. Kouchi
(Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises
510(k) Number k062930
Prescription Use: X (Per 21 CFR 801.109)
Over the Counter Use:
Or
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.