POWERED EZ-IO BONE MARROW ASPIRATION SYSTEM

K062833 · Vidacare Corporation · FCF · Nov 30, 2006 · Gastroenterology, Urology

Device Facts

Record IDK062833
Device NamePOWERED EZ-IO BONE MARROW ASPIRATION SYSTEM
ApplicantVidacare Corporation
Product CodeFCF · Gastroenterology, Urology
Decision DateNov 30, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1075
Device ClassClass 2

Indications for Use

For Bone Marrow Aspiration of the Iliac Crest.

Device Story

Powered EZ-IO Bone Marrow Aspiration System consists of a reusable battery-powered driver and a disposable 15-gauge, 60mm intraosseous (IO) needle assembly. The driver drills the needle through the bone cortex to a target depth in the marrow. Once seated, the driver is detached, the trocar/stylet is removed, and a standard syringe is attached to the Luer lock hub for aspiration. Used by clinicians to access bone marrow; facilitates sample collection for diagnostic purposes. Benefits include efficient, powered access to the iliac crest.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and design comparisons.

Technological Characteristics

Battery-powered driver; disposable 15-gauge, 60mm IO needle assembly; Luer lock connection for syringe aspiration; manual drill-based penetration mechanism.

Indications for Use

Indicated for bone marrow aspiration of the iliac crest.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ## 5. 510(k) Summary ## CARE NOV 30 2006 722 Isom Road San Antonio, TX 78216 210-375-8500 ## SUMMARY Submitter's name: Address: Phone: Fax number: VidaCare Corporation 722 Isom Road San Antonio, TX 78216 210-375-8500 210-375-8537 Name of contact person: Grace Holland Regulatory Specialists, Inc 3722 Ave. Sausalito Irvine, CA 92606 Phone: 949-262-0411 Fax: 949-552-2821 Date the summary was prepared: September 7, 2006 Name of the device: Powered EZ-IO Bone Marrow Aspiration System Trade or proprietary name: Bone Marrow Aspiration System Common or usual name: Aspiration Needle Classification name: Gastroenterology-urology biopsy instrument. The legally marketed devices to which we are claiming equivalence [807.92(a)(3)]: | 510(k)<br>Number | Trade or Proprietary or Model<br>Name | Manufacturer | |------------------|----------------------------------------------|--------------------------------------| | 1 K032885 | VidaPort Intraosseous Infusion<br>System | VidaCare Corp. | | 2 K980196 | Manan Biopsy Set for Bone<br>and bone Marrow | Medical Device<br>Technologies, Inc. | {2}------------------------------------------------ Description of the device: The Powered EZ-IO Bone Marrow Aspiration System consists of a reusable battery powered driver connected to a disposable intraosseous (IO) needle assembly. Upon activation, the drill penetrates through the cortex of the bone to a desired depth within the bone marrow. The driver then is separated from the hub of the IO needle assembly, leaving the cannula securely seated in the bone. The trocar/stylet containing the drill bit is then removed. A standard Luer lock (part of the needle assembly) then permits attachment of standard syringe for aspiration. The needle is 15 gauge and 60mm long. Indications: For Bone Marrow Aspiration of the Iliac Crest. Summary of the technological characteristics of our device compared to the predicate device: The predicates and the Powered EZ-IO Bone Marrow Aspiration System were compared in the following areas and found to have similar technological characteristics and to be equivalent. Indications for Use Target Population Driver Design Features Needle Design Technique Sterility Biocompatibility Anatomical Sites Where Used {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure with three overlapping, curved lines forming its body and wings. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Vidacare Corporation % Regulatory Specialists, Inc. Ms. Grace Holland Regulatory Specialist 3722 Avenue Sausalito Irvine, California 92606 NOV 3 U 2006 Re: K062833 Trade/Device Name: Powered EZ-IO Bone Marrow Aspiration System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II Product Code: FCF Dated: October 20, 2006 Received: October 26, 2006 Dear Ms. Holland: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not mean or any Federal statutes and regulations administered by other Federal agencies. Your must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 -- Ms. Grace Holland forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) I ms letter will anow you to begin manusing of substantial equivalence of your device to a legally premarket notification: "The Privation for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific as no any liance at (240) 276-0115. Also, please note the regulation entitled, Connect the Office or Comment notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours Rok Pelloh Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## 4. Indications for Use Statement Indications for Use 510(k) Number (if known): _ Device Name: Powered EZ-IO Bone Marrow Aspiration System Indications for Use: For Bone Marrow Aspiration of the Iliac Crest. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Pel (Division Sign-O Division of General, Restorative, and Neurological Devices Page 1 of 1 **510(k) Number** L062833
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...