K062805 · Spirus Medical, Inc. · FED · Oct 13, 2006 · Gastroenterology, Urology
Device Facts
Record ID
K062805
Device Name
MODIFICATION TO ENDO-EASE
Applicant
Spirus Medical, Inc.
Product Code
FED · Gastroenterology, Urology
Decision Date
Oct 13, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The Spirus Medical, Inc. Endo-Ease Endoscopic Overtube is indicated for use with a flexible endoscope to aid in endoscopic insertion and advancement during diagnostic and therapeutic lower GI flexible endoscopy.
Device Story
Endo-Ease is a sterile, single-use, flexible overtube used in conjunction with flexible endoscopes during lower GI procedures. It functions as an accessory to facilitate endoscope insertion and advancement. The device is intended for use by clinicians in a clinical setting. It is a modified version of the predicate device, with safety and performance verified through design controls and risk analysis procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, verification, and validation of device modifications under 21 CFR 820.30 design controls.
Technological Characteristics
Sterile, single-use, flexible overtube. Designed for use with flexible endoscopes in the lower GI tract. No specific materials or connectivity features described.
Indications for Use
Indicated for patients undergoing diagnostic or therapeutic lower GI flexible endoscopy to assist in the insertion and advancement of a flexible endoscope.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}------------------------------------------------
# 510(k) Summary
K062805
| Trade Name: | Endo-Ease™ |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Spirus Medical, Inc.<br>1063 Turnpike Street<br>Stoughton, MA 02072<br>FDA Registration No. not yet assigned<br>OCT 13 2001 |
| Device Common<br>Name: | Endoscopic overtube |
| Classification: | According to Section 513 of the Federal Food, Drug, and<br>Cosmetic Act, the device classification is Class II. |
| Predicate Device: | K052084 - Spirus Medical, Inc. Endo-Ease Endoscopic Overtube |
### Product Description:
The device described in this 510(k) consists of a a modified sterile, single use, flexible over ube designed for use with currently marketed flexible endoscopes in the lower GI tract.
#### Indications for Use:
The Spirus Medical, Inc. Endo-Ease Endoscopic Overtube is indicated for use with a flexible endoscope to aid in endoscopic insertion and advancement during diagnostic and therapeutic lower GI flexible endoscopy.
#### Safety and Performance:
This submission is a Special 510(k): Device Modification as described in FDA's guidance document entitled "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications." In support of this 510(k), Spirus Medical has provided certification of compliance to 21 CFR 820.30 Design Control requirements, and a description of the internal Risk Analysis procedure. A description and summary of the verification and validation testing methods and results utilized to qualify the device modifications is included in this submission.
### Conclusion:
Based on the indications for use, technological characteristics, performance testing and comparison to predicate devices, the proposed modified Spirus Medical, Inc. Endo-Ease Endoscopic Overtube has been shown to be safe and effective for its intended use.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle faces to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
## 13 2006
Spirus Medical, Inc. c/o Pamela Papineau, R.A.C. Delphi Medical Device Consulting, Inc. 5 Whitcomb Avenue AYER MA 01432
Re: K062805
> Trade/Device Name: Endo-Ease Endoscopic Overtube Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED and FDF Dated: September 13, 2006 Received: September 19, 2006
Dear Ms. Papineau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/1/Picture/10 description: The image is a circular logo commemorating the centennial of the FDA from 1906-2006. The letters FDA are in bold, stacked on top of the word Centennial. There are three stars below the word Centennial. The logo is surrounded by a dotted circle.
Protecting and Promoting Public Health.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
1f you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Spirus Medical, Inc. September 13, 2006
Special 510(k) Premarket Notification: Device Modification Modified Endo-Ease Endoscopic Overtube
Page 1 of 1
510(k) Number (if known): 八
Device Name: Endo-Ease Endoscopic Overtube
Indications for Use:
The Spirus Medical, Inc. Endo-Ease Endoscopic Overtube is indicated for use with a flexible endoscope to aid endoscopic insertion and advancement during diagnostic and therapeutic lower GI flexible endoscopy.
Prescription Use X (Per 21 CFR 801 Subpart D)
OR
Over-the -Counter Use _ (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Lynn
---
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number