← Product Code [MVM](/productcode/MVM) · K062787

# ELIA CELIKEY IGA WELL, MODEL 14-5517-01; ELIA CELIAC CONTROL, MODEL 83-1011-01 (K062787)

_Phadia US, Inc. · MVM · Dec 29, 2006 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K062787

## Device Facts

- **Applicant:** Phadia US, Inc.
- **Product Code:** [MVM](/productcode/MVM.md)
- **Decision Date:** Dec 29, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5660
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

EliA™ Celikey IgA is intended for the in vitro semi-quantitative measurement of IgA antibodies directed to tissue transglutaminase (tTG) in human serum and plasma (EDTA, citrate). EliA Celikey IgA is based on recombinant human tissue transglutaminase as antigen and is useful as an aid in the clinical diagnosis of patients with celiac disease in conjunction with other laboratory and clinical findings. EliA Celikey IgA uses the EliA IgA method on the instrument ImmunoCAP 100 and ImmunoCAP 250.

## Device Story

EliA Celikey IgA is a semi-quantitative fluoroenzyme immunoassay for detecting IgA antibodies to tissue transglutaminase (tTG) in human serum or plasma. The device uses microwells coated with recombinant human tTG. Patient samples are processed on ImmunoCAP 100 or ImmunoCAP 250 automated analyzers; samples are diluted 1:100. If anti-tTG IgA is present, it binds to the tTG antigen; bound antibodies are detected using an enzyme-labeled anti-IgA conjugate (β-galactosidase). A development solution (4-Methylumbelliferyl-β-D-galactoside) is added, producing fluorescence proportional to the amount of bound IgA. Results are compared to a calibration curve. The device is intended for use in clinical laboratories by trained personnel to assist physicians in diagnosing celiac disease. Heparin plasma is contraindicated due to interference.

## Clinical Evidence

Clinical performance evaluated in 194 subjects (93 celiac disease, 101 non-celiac normal biopsy). Clinical sensitivity was 95.7% (89/93; 95% CI 89.4-98.8%) and clinical specificity was 95.0% (96/101; 95% CI 88.8-98.4%). Method comparison against the predicate (n=244) showed 100% positive and negative percent agreement. Analytical precision demonstrated intra-assay CV ≤10.4% and inter-assay CV ≤13.3%. No hook effect observed up to 654 μg/L.

## Technological Characteristics

Fluoroenzyme immunoassay; solid phase: microwells coated with human recombinant tTG; energy source: automated analyzer (ImmunoCAP 100/250); detection: fluorescence via β-galactosidase conjugate and 4-Methylumbelliferyl-β-D-galactoside substrate; sample matrices: serum, EDTA plasma, citrate plasma; reaction temperature: 37°C controlled; calibration: 6-level IgA calibrators traceable to WHO IRP 67/86.

## Regulatory Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

## Predicate Devices

- Celikey® Tissue Transglutaminase IgA Antibody Assay (k041174)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k062787

B. Purpose for Submission:
New Device

C. Measurand:
Anti-tissue transglutaminase (tTG) IgA Antibodies

D. Type of Test:
Semi-quantitative fluoroenzyme immunoassay

E. Applicant:
Phadia US, Inc.

F. Proprietary and Established Names:
EliA™ Celikey IgA Immunoassay

G. Regulatory Information:
1. Regulation section:
21 CFR §866.5660 Multiple autoantibodies immunological test system
2. Classification:
Class II
3. Product code:
MVM, Autoantibodies, endomysial (tissue transglutaminase)
4. Panel:
Immunology 82

H. Intended Use:
1. Intended use(s):
EliA™ Celikey IgA is intended for the in vitro semi-quantitative measurement of IgA antibodies directed to tissue transglutaminase (tTG) in human serum and plasma (EDTA, citrate). EliA Celikey IgA is based on recombinant human tissue transglutaminase as antigen and is useful as an aid in the clinical diagnosis of patients with celiac disease in conjunction with other laboratory and clinical findings. EliA Celikey IgA uses the EliA IgA method on the instrument ImmunoCAP 100 and ImmunoCAP 250.
2. Indication(s) for use:
Same as intended use.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
ImmunoCAP 100 and ImmunoCAP 250 (k061165)

I. Device Description:
The EliA reagents are available as modular packages, each purchased separately. The EliA Celikey IgA wells are coated with human recombinant tissue transglutaminase (tTG). These are packed in carriers which are stored in sealed aluminum foil bags containing a desiccant. The EliA Method-Specific reagents consists of: six levels of ready to use EliA IgA calibrators (0, 0.3, 1.5, 15, 80 µg/L); IgA calibrator well (coated with mouse monoclonal antibodies); ready to use positive and negative

{1}

controls; ready to use IgA curve control (5 μg/L); IgA conjugate (β-Galactosidase anti-IgA mouse monoclonal antibodies) in PBS; sample diluent concentrate (PBS with BSA); ready to use development solution (0.1% 4-Methylumbelliferyl-β-D-galactoside); and stop solution (4% Sodium Carbonate).

# J. Substantial Equivalence Information:

1. Predicate device name(s):

Celikey® Tissue Transglutaminase IgA Antibody Assay

2. Predicate 510(k) number(s):

k041174

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | New Device | Predicate Devices  |
|   | EliATM Celikey IgA | Celikey® Tissue Transglutaminase IgA Antibody Assay  |
|  Antigen | Human recombinant tTG | Same  |
|  Assay Type | ELISA | Same  |
|  Solid phase | Microwells | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended use | For the semi-quantitative measurement of IgA antibodies directed to tissue transglutaminase (tTG) in human serum and plasma (EDTA, citrate). EliA Celikey IgA is based on recombinant human tissue transglutaminase as antigen and is useful as an aid in the clinical diagnosis of patients with celiac disease in conjunction with other laboratory and clinical findings. | For the semi-quantitative and qualitative measurement of IgA antibodies directed to tissue transglutaminase (tTG) in human serum and plasma. Celikey is based on recombinant human tissue transglutaminase as antigen and is useful as an aid in the clinical diagnosis of patients with celiac disease.  |
|  Assay Format | Semi-quantitative | Semi-quantitative and Qualitative  |
|  Assay Platform | 48 well determinations | 96 well determinations  |
|  Instrumentation | ImmunoCAP 100 and ImmunoCap 250(fully automated analyzers) | Microplate reader with 620 nm filter  |
|  Signal | Fluorescence | Optical Density  |

{2}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Reagent concept | Modular reagent concept (test-method specific and general reagents) | All reagents in a single kit  |
|  Internal Controls | Positive and negative control sera available separately | Positive and negative control sera included in the kit  |
|  Calibration | Total IgA calibrator (0, 0.3, 1.5, 15, 80 μg/L)
Option to store curve up to 28 days and run curve controls (provided in kit) in each assay for calibration | Analyte specific (tTG antibody concentrations of 0, 3, 7, 16, 40, 100 U/mL) in each assay.  |
|  Conjugate | Anti-human IgA β-galactosidase (mouse monoclonal antibodies) | Anti-human IgA horseradish peroxidase (goat)  |
|  Sample type and dilution | Serum or plasma (EDTA, citrate) at 1:100 by the instrument | Serum or plasma at 1:101 dilution manually  |
|  Development solution | 4-Methylumbelliferyl-β-D-galactoside | TMB Chromogen  |
|  Stop solution | 4% Sodium Carbonate | 0.5 M H₂SO₄ ready to use  |
|  Reaction temperature | 37°C controlled | Room temperature, 18-25 °C  |
|  Cut-off | Semi-quantitative: 10.0 U/mL | Semi-quantitative: 8.0 U/mL;
Qualitative: Ratio 1.4  |

K. Standard/Guidance Document Referenced (if applicable):
None referenced.

L. Test Principle:
The EliA Celikey IgA wells are coated with human recombinant tTG. If present in the patient’s specimen, antibodies to tTG bind to their specific antigen in the wells. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgA (EliA IgA conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away and the bound complex is incubated with a development solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The higher the response value, the more specific IgA is present in the specimen. To evaluate test results, the response for the patient samples is compared directly to the response for IgA calibrators.

M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:

{3}

The precision of the assay on ImmunoCAP 100 instrument was determined by testing four serum samples in duplicate on seven instruments in 23 runs on one EliA Celikey IgA well batch, together with one set of system reagent. Results showed that 2 samples positive with anti-tTG IgA (11.2-27.3U/mL) had ≤4.1 %CV and ≤6.8% CV for the intra-assay and inter-assay studies respectively; and 2 negative samples (1.0-7.9 U/mL) had ≤7.2 %CV and ≤13.3% CV for the intra-assay and inter-assay studies respectively (see below).

ImmunoCAP 100

|  Sample | Mean (U/mL) | Intra-assay (%CV) | Inter-assay (%CV)  |
| --- | --- | --- | --- |
|  1 | 1.0 | 7.2 | 13.3  |
|  2 | 7.9 | 4.4 | 4.6  |
|  3 | 11.2 | 4.1 | 6.0  |
|  4 | 27.3 | 2.9 | 6.8  |

The precision of the assay on ImmunoCAP 250 instrument was determined by testing four serum samples in duplicate on two instruments in 14 runs on one EliA Celikey IgA well batch, together with one set of system reagent. Results showed that 3 samples positive with anti-tTG IgA (11.0-126.1 U/mL) had ≤5.1 %CV and ≤4.4% CV for the intra-assay and inter-assay studies respectively; and one negative sample (0.4 U/mL) had 10.4 %CV and 8.1% CV for the intra-assay and inter-assay studies respectively (see below).

ImmunoCAP 250

|  Sample | Mean (U/mL) | Intra-assay (%CV) | Inter-assay (%CV)  |
| --- | --- | --- | --- |
|  1 | 0.4 | 10.4 | 8.1  |
|  2 | 11.0 | 4.8 | 2.8  |
|  3 | 15.6 | 4.1 | 1.6  |
|  4 | 126.1 | 5.1 | 4.4  |

b. Linearity/assay reportable range:

Linearity by dilution was tested using 5 anti-tTG IgA positive samples (20.5-72.8 μg/L). These samples were tested in duplicates for dilutions 1/100, 1/200, 1/400, 1/800 and 1/1600. The results of the dilutions were compared with their expected value. The ratio observed/expected values were calculated.

Specifications for the ratio (O/E) were 0.8 and 1.2 within all sample dilutions. Samples 1 and 5 had results of 1.2-1.4, which were slightly above 1.2. The following sentence had been included in package insert: "Please note that due to differing binding characteristics of the antibodies in patient samples, not all sera can be diluted linearly within the measuring range".

Hook Effect:

The possibility of antigen excess occurring when using the device was evaluated with serum sample above the assay measuring range. The measuring range for EliA Celikey IgA is from 0.1 to ≥128 EliA U/mL. No

{4}

hook effect was observed for concentrations up to  $654~\mu \mathrm{g / L}$ , which is 8 fold above the measuring range.

c. Traceability, Stability, Expected values (controls, calibrators, or methods): The EliA IgA calibrators are traceable (via unbroken of calibrations) to the International Reference Preparation (IRP) 67/86 of the Human Serum Immunoglobulins A, G, M from WHO. New batches of IgA calibrators are compared to a secondary standard (standardized with the IRP) or the IRP directly and adjusted accordingly to meet the correct concentration (6 levels). There are no international standards for tTG antibodies. The positive and negative controls are prepared in house and arbitrary EliA Units/mL are assigned during the development process.

# d. Detection limit:

The lower limit of the measuring range was determined by measuring dilutions (1/2, 1/4, and 1/8) of Calibrator 0.3 (0.3 μg/L) in EliA IgA Calibrator wells. The results in Response Units (RU) were compared with the result of the sample diluent on EliA Celikey IgA wells. All the samples were measured in triplicates. The discrimination ability 'D' of the assay should be  $&gt;2$  for calibrator 0.3 diluted 1:4 (i.e. 0.075 μg/L IgA/L). The 1/8 diluted calibrator 0.3 (0.0375 μg/L) still can be discriminated from background given by the signal of the diluent on Celikey IgA wells (as shown in the table below). The lower limit of detection was set at 0.05 μg/L corresponding to 0.1 EliA U/mL.

|  Results on Calibrator Wells |   |   |   |
| --- | --- | --- | --- |
|  Sample ID | Mean Response Units RU | SD | D*  |
|  Calibrator 0.3 1/2 | 109 | 1.9 | 37.3  |
|  Calibrator 0.3 1/4 | 67 | 8.9 | 7.2  |
|  Calibrator 0.3 1/8 (0.0375μg/L) | 43 | 4.2 | 8.8  |
|  Results on Celikey IgA Wells |   |   |   |
|  Sample ID | Mean RU | SD |   |
|  Sample Diluent | 1 | 2.2 |   |

# e. Analytical specificity:

Interfering substances

The potential interferent and the corresponding blanks were added to two anti-tTG positive sera from patients with Celiac disease. The spiked samples were tested in triplicate. The ratio of the result of the sample spiked with the interfering substances and the sample spiked with a buffer blank was determined as shown in the table below:

{5}

|   |   | Positive Sample |   |   | Low Positive Sample  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Additive | blank/spiked sample | Conc. [μg/l] | CV % | Ratio | Conc. [μg/l] | CV % | Ratio  |
|  Bilirubin F | blank | 29.8 | 2.0 | 1.04 | 6.9 | 1.3 | 1.00  |
|   |  sample | 30.9 | 3.1 |   | 6.9 | 0.9  |   |
|  Bilirubin C | blank | 28.7 | 3.6 | 1.04 | 6.8 | 1.7 | 1.01  |
|   |  sample | 29.8 | 5.6 |   | 6.9 | 2.4  |   |
|  Haemoglobin | blank | 31.7 | 3.6 | 0.93 | 7.1 | 4.6 | 0.97  |
|   |  sample | 29.3 | 2.4 |   | 6.9 | 2.9  |   |
|  Chyle | blank | 30.4 | 6.3 | 1.03 | 7.0 | 4.5 | 1.00  |
|   |  sample | 31.4 | 5.9 |   | 7.0 | 0.4  |   |
|  Rheumatoid factor | blank | 33.5 | 4.6 | 0.98 | 7.1 | 3.2 | 1.01  |
|   |  sample | 32.7 | 3.9 |   | 7.2 | 4.7  |   |

The interfering substances Bilirubin C, Bilirubin F, Chyle, Hemoglobin and Rheumatoid Factor did not adversely affect the results of the new device.

f. Assay cut-off:

The purpose of the normal sera studies was to evaluate expected values in the normal population and to confirm the defined cut-off. Samples from 400 apparently healthy Caucasian adult blood donors were measured. The individuals were equally distributed by age and gender.

The results were equally distributed and not dependent on age or gender. The 99th percentile (7.4 U/mL) lies below the lower limit of the equivocal range (7-10 U/mL).

2. Comparison studies:

a. Method comparison with predicate:

Testing was performed on 244 samples from clinically defined patients (93 celiac, 101 non-celiac normal biopsy, 10 Ulcerative Colitis, 10 Morbus Crohn, 20 Rheumatoid Arthritis, 10 Rheumatoid factor positive samples. Equivocals were excluded from the percent agreement calculations. The Positive Percent Agreement was 100% (89/89); the Negative Percent Agreement was 100% (144/144); and the Overall Agreement was 100% (233/233) (see table below).

|   | Celikey IgA on Varelisa™  |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Positive | Equivocal* | Negative | Total  |
|  EliA Celikey IgA | Positive | 89 | 6 | 0 | 95  |
|   |  Equivocal* | 0 | 3 | 1 | 4  |
|   |  Negative | 0 | 1 | 144 | 145  |
|   |  Total | 89 | 10 | 145 | 244  |

*Equivocal results were excluded from the analysis

Positive percent agreement = 100% (89/89)

Negative Percent Agreement = 100% (144/144)

Overall Agreement = 100% (233/233)

b. Matrix comparison:

{6}

The assay uses serum and plasma (EDTA, citrate) as sample matrices. Sixty sets of samples from different donors were tested. Each set contained serum, EDTA, heparin, and citrate plasma samples. Specifications were as follows: mean quota of plasma to serum concentration should be 0.8-1.2 for positive sera and negative samples should not switch to positive in all serum and plasma sample. Linear regression comparing the quota between serum and each type of plasma for the samples was performed as shown below:

![img-0.jpeg](img-0.jpeg)

The specification for the positive samples in this study was fulfilled for EDTA (1.04) and citrate (0.98), but not for heparin (1.37). All negative samples remain negative in all serum and plasma samples. The following sentence was included in package insert Specimen Collection section: "the use of plasma preparations with heparin is not recommended because heparin interferes with the measurement of tTG antibodies".

# c. Instrument Platform comparison:

For this comparison study, a total of 36 samples distributed over the measuring range were assayed as follows: 4 negative samples, 2 equivocal samples and 30 positive samples. All samples were run on three ImmunoCAP250 instruments and three ImmunoCAP100 instruments in two runs and in single replicates. Specification for correlation was fulfilled  $(r &gt; 0.9)$ . Results of the conformance study are summarized in the figure below:

{7}

![img-1.jpeg](img-1.jpeg)

3. Clinical studies:

a. Clinical Sensitivity and Specificity:

Ninety three (93) patients with known diagnosis of Celiac Disease (CD) and 101 non-CD patients with normal biopsy were tested for this study. Results showed clinical sensitivity of 95.7% (89/93); (95% CI 89.4-98.8%) and clinical specificity of 95% (96/101); (95% CI 88.8-98.4%) as shown in the table below:

|  n=194 | Diagnosis  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive (Celiac disease) | Negative (Disease Controls) | Total  |
|  EliA Celikey IgA | Positive | 89 | 5 | 94  |
|   |  Negative | 4 | 96 | 100  |
|   |  Total | 93 | 101 | 194  |

Sensitivity: 95.7% (89/93); 95% CI 89.4-98.8%

Specificity: 95.0% (96/101); (95% CI 88.8-98.4%)

b. Other clinical supportive data (when a. is not applicable):

Not applicable.

4. Clinical cut-off:

Same as assay cut-off.

5. Expected values/Reference range:

Expected values in the normal population should be negative.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

{8}

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

9

---

**Source:** [https://fda.innolitics.com/device/K062787](https://fda.innolitics.com/device/K062787)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
