← Product Code [DXO](/productcode/DXO) · K062769

# PRESSUREWIRE CERTUS, MODELS 12006 AND 12306 (K062769)

_Radi Medical Systems AB · DXO · Dec 4, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K062769

## Device Facts

- **Applicant:** Radi Medical Systems AB
- **Product Code:** [DXO](/productcode/DXO.md)
- **Decision Date:** Dec 4, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2870
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

PressureWire® is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a vessel. The signal output from the senor is used for calculation and presentation of any physiological parameters, functions or indices based on pressure or temperature, e.g. Fractional Flow Reserve (FFR).

## Device Story

PressureWire® is a .014" guidewire featuring an integrated pressure and temperature sensor; includes a detachable cable for connection to a diagnostic computer. Used by physicians in clinical settings to navigate percutaneous catheters through vessels. The device measures intravascular pressure and temperature; signals are transmitted to a diagnostic computer for calculation and display of physiological indices, such as Fractional Flow Reserve (FFR). This data assists clinicians in assessing coronary and peripheral vessel physiology to inform clinical decision-making and patient management.

## Clinical Evidence

Bench testing only. Biocompatibility and performance testing were conducted to verify that the device meets specified safety and performance requirements.

## Technological Characteristics

.014" guidewire; integrated pressure and temperature sensor; hydrophilic distal coating; detachable cable for external computer connectivity.

## Regulatory Identification

A catheter tip pressure transducer is a device incorporated into the distal end of a catheter. When placed in the bloodstream, its mechanical or electrical properties change in relation to changes in blood pressure. These changes are transmitted to accessory equipment for processing.

## Predicate Devices

- PressureWire® Sensor ([K031662](/device/K031662.md))
- Safe-Cross Deflecting Catheter ([K040481](/device/K040481.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K062769

## 510(k) Summary

| Submitter:           | Radi Medical Systems AB<br>Palmbladsgatan 10<br>SE-754 50 Uppsala, Sweden<br>Phone:(+46) 18161000 | DEC - 4 2006 |
|----------------------|---------------------------------------------------------------------------------------------------|--------------|
| Contact:             | Mats Granlund                                                                                     |              |
| Date Prepared:       | September 13, 2006                                                                                |              |
| Proprietary Name:    | PressureWire®                                                                                     |              |
| Common Name:         | Pressure Guidewire                                                                                |              |
| Classification Name: | Catheter Tip Pressure Transducer (21 CFR 870.2870)<br>Product Code DXO                            |              |
| Predicate Devices:   | PressureWire® Sensor, K031662<br>Safe-Cross Deflecting Catheter, K040481                          |              |

#### Device Description

Pressure Wire® is a .014" guidewire with an integrated pressure and temperature sensor, together with a detachable cable for connection to a diagnostic computer.

#### Intended Use of the Device:

PressureWire® has the same intended use as the submitter's predicate device. Pressure Wire® is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a vessel. The signal output from the senor is used for calculation and presentation of any physiological parameters, functions or indices based on pressure or temperature, e.g. Fractional Flow Reserve (FFR).

#### Technological Characteristics:

PressureWire® is similar in basic material, design, construction and mechanical performance to the predicate device. The main modification is that the distal part has been coated with a hydrophilic coating.

#### Performance Data

Biocompatibility and performance testing indicate that PressureWire® satisfies safety and performance requirements of the device specifications and do not raise additional safety issues.

#### Conclusion

On the basis of the testing conducted, it may be concluded that PressureWire® satisfies specified safety and performance requirements. PressureWire® is substantially equivalent to the predicate device.

{1}------------------------------------------------

Public Health Service

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wing and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC - 4 2006

Radi Medical Systems c/o Mats Granlund Director, Quality & Regulatory Affairs Palmbladsgatan 10 Uppsala, Sweden SE-754 50

Re: K062769 Trade/Device Name: PressureWire® Regulation Number: 21 CFR 870.2870 Regulation Name: Catheter tip pressure transduccr Regulatory Class: II Product Code: DXO, DQX Dated: October 25, 2006 Received: October 26, 2006

Dear Mr. Granlund:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do net require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Mats Granlund

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdri/industry/support/index.html.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

510(k) Number (if known): K062769

Device Name:

Pressure Wire®

Indications for Use:

PressureWire® is indicated to direct a catheter through a blood vessel and to measure physiological parameters in the coronary and peripheral blood vessels.

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ision of Cardiovascular Devices

510(k) Number K062769

Page 1 of

---

**Source:** [https://fda.innolitics.com/device/K062769](https://fda.innolitics.com/device/K062769)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
