← Product Code [GWF](/productcode/GWF) · K062765

# MODIFICATION TO NUVASIVE NEUROVISION JJB SYSTEM (K062765)

_Nuvasive, Inc. · GWF · Jan 24, 2007 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K062765

## Device Facts

- **Applicant:** Nuvasive, Inc.
- **Product Code:** [GWF](/productcode/GWF.md)
- **Decision Date:** Jan 24, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.1870
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The NeuroVision JJB System is used for intraoperative monitoring and neurological status assessment by the administration of brief electrical stimulus pulses to neural tissues and the EMG monitoring of the associated muscle groups. The System is used in conjunction with other NuVasive devices to assist in gaining controlled access to, and visualization of, the spine.

## Device Story

NeuroVision JJB System provides intraoperative neurophysiologic monitoring during spine surgery; reduces risk of nerve root injury. System inputs: electrical stimulus pulses delivered to neural tissues; EMG signals captured from associated muscle groups via surface electrodes. System components: reusable Patient Module; Control Unit with embedded computer, touch screen, and interface card; various disposable/reusable conductive probes and leads. Operation: clinician delivers electrical pulses to neural tissues; system monitors resulting EMG responses in muscle groups. Output: real-time EMG monitoring data displayed on touch screen. Clinical utility: assists surgeons in gaining controlled access to and visualization of the spine; provides feedback on nerve status to guide surgical maneuvers and prevent nerve damage.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use.

## Technological Characteristics

System comprises a patient module, control unit with embedded computer, touch screen, and interface card. Uses conductive probes, electrodes, and leads for EMG monitoring and electrical nerve stimulation. Operates as an evoked response electrical stimulator (21 CFR 882.1870).

## Regulatory Identification

An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.

## Predicate Devices

- NeuroVision JJB System ([K062765](/device/K062765.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## VII. 510(k) Summary

# JAN 2 4 2007

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular §807.92, the following summary of information is provided:

#### Submitted by: A.

Laetitia Cousin Director of Regulatory Affairs and Quality Assurance NuVasive, Incorporated 4545 Towne Centre Court San Diego, California 92121 Telephone: (858) 909-1868 Fax: (858) 909-2068

#### B. Device Name

| Trade or Proprietary Name: | NuVasive NeuroVision JJB System           |
|----------------------------|-------------------------------------------|
| Common or Usual Name:      | Electromyography (EMG) monitor/stimulator |
| Classification Name:       | Surgical nerve stimulator/locator         |
| Device Class:              | Class II                                  |
| Classification:            | $§$ 874.1820, $§$ 882.1870                |
| Product Code:              | 77ETN, 84GWF                              |

#### C. Predicate Devices

The subject NeuroVision JJB System is substantially equivalent to the NeuroVision JJB System currently manufactured and distributed commercially in the U.S. by NuVasive.

#### D. Device Description

The NVJJB System utilizes conventional neurophysiologic monitoring to reduce the incidence of injury to nerve roots during instrumented spine surgery. In the procedure, stimulus evoked electromyography is used to determine changes in nerves. Corresponding muscle groups are monitored using surface electrodes, while stimulation is used to detect nerve responses.

The Neuro Vision JJB System consists of a reusable Patient Module, a Control Unit comprised of an embedded computer with touch screen controls and an interface card, and an assortment of disposable and reusable conductive probes, electrodes, and electrode leads.

{1}------------------------------------------------

#### Intended Use E.

The NeuroVision JJB System is used for intraoperative monitoring and neurological status The Neuro vision 952 byteen is as brief electrical stimulus pulses to neural tissues and the assessment by the associated muscle groups. The System is used in conjunction with other NuVasive devices to assist in gaining controlled access to, and visualization of, the spine.

### Comparison to Predicate Devices F.

The subject device has indications for use identical to those of its predicate, and employs the same principles of operation. Due to this equivalency, the device raises no new safety or effectiveness issues.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services, USA. The seal is circular and features an eagle with its wings spread. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" are written around the eagle in a circular fashion.

Public Health Service

JAN 2 4 2007

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NuVasive, Incorporated c/o Ms. Laetitia Cousin Director of Regulatory Affairs and Ouality Assurance 4545 Towne Centre Court San Diego, California, 92121

Re: K062765

Trade Name: NuVasive NeuroVision® JJB System Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked response electrical stimulator Regulatory Class: II Product Code: GWF Dated: December 26, 2006 Received: December 28, 2006

Dear Ms. Cousin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Laetitia Cousin

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### D. Indications for Use Statement

1062765 510(k) Number (if known):

Device Name: NeuroVision JJB System

Indications for Use:

The NeuroVision JJB System is used for intraoperative monitoring and neurological status assessment by the administration of brief electrical stimulus pulses to neural tissues and the EMG monitoring of the associated muscle groups. The System is used in conjunction with other NuVasive devices to assist in gaining controlled access to, and visualization of, the spine.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use

![image](image)

**(Division Sign-Off)**

**Division of General, Res**

Division of General, Rest an tive, and Neurological Devices

**510(k) Number:** Lo66 2165

Page - 13 -

---

**Source:** [https://fda.innolitics.com/device/K062765](https://fda.innolitics.com/device/K062765)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
