← Product Code [CAF](/productcode/CAF) · K062689

# SIDESTREAM PLUS (K062689)

_Respironics , Ltd. · CAF · Feb 16, 2007 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K062689

## Device Facts

- **Applicant:** Respironics , Ltd.
- **Product Code:** [CAF](/productcode/CAF.md)
- **Decision Date:** Feb 16, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5630
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric, 3rd-Party Reviewed

## Indications for Use

The Sidestream Plus is a handheld nebulizer designed to aerosolize medication approved for nebulization and prescribed by a physician. The Sidestream Plus is intended for adult and pediatric patients consistent with the indications for the aerosol medication.

## Device Story

Sidestream Plus is a breath-enhanced reusable nebulizer; operates with external compressor providing air pressure/flow. Device aerosolizes physician-prescribed liquid medications for inhalation. Used in clinical or home settings by patients or caregivers. Enhances aerosol delivery during patient inhalation; improves medication administration efficiency. No complex electronics or software; purely mechanical pneumatic operation.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on design and operating principle comparison to predicate devices.

## Technological Characteristics

Breath-enhanced reusable nebulizer; pneumatic operation via external compressor. Class II device. No software, electronics, or connectivity. Materials and sterilization methods not specified in provided text.

## Regulatory Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

## Predicate Devices

- Pari LC+ ([K935540](/device/K935540.md))

## Reference Devices

- Ventstream ([K933535](/device/K933535.md))
- Pari LC Star ([K963924](/device/K963924.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K062689

## Section 5

FEB 1 6 2007

## 510(K) SUMMARY

Prepared: August 1, 2006

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.

The assigned 510(k) number is:

#### Submitter's Identification: 1.

Respironics Ltd. Heath Place West Sussex, PO 22 9SL UK Phone: +44 870 770 3548 Fax: +44 870 770 2002 Contact Name: Henry Marshall, Manager, OA/RA e-mail: henry.marshall(@respironics.com

Establishment Registration Number :0681154

## Foreign Manufacturer:

Respironics Medical Products (Shenzhen) Co., Ltd. Tong Xi Chong Village, Boa on Districts Shenzhen, People's Republic of China

· Establishment Registration Number: 3003001596

## Official Correspondent:

Lauren Ziegler Sr. Manager, QA, RA, and Clinical Affairs 41 Canfield Road Cedar Grove, NJ 07009 Phone: 973-571-2608 Fax: 973-857-9521 e-mail: Lauren.ziegler@respironics.com

{1}------------------------------------------------

### 2. · Name of the Device:

Sidestream Plus Nebulizer Common Name or Classification Name (21 CFR Part 807.87) of Device: Nebulizer, 21 CFR Part 868.5630

## 3. Predicate Device Information:

Identification of legally marketed device which we claim substantial equivalence to: Pari LC+ K935540 Pari Respiratory Equipment 2943 Oak Lake Boulevard Mindlothian, Va 23112-3998

#### 4. Device Description:

Sidestream Plus is a Class II device. It is a breath-enhanced reusable nebulizer to be used with a compressor that is capable of providing appropriate air pressure and flow to operate the device according to specification.

The Sidestream Plus is a breath-enhanced nebulizer and works utilizing the same operating principles as the Ventstream (K933535), the Pari LC Star (K963924) and the Pari LC + (K935540).

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The logo is black and white and appears to be a simple, clean design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Respironics, Limited C/O Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services 2307 East Aurora Road Twinsburg, Ohio 44087

FEB ] & 2007

Re: K062689

Trade/Device Name: Sidestream Plus Nebulizer Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: February 5, 2007 Received: February 6, 2007

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

## Page 2 - Mr. Lehtonen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu-Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: Sidestream Plus Nebulizer

Indications for Use:

The Sidestream Plus is a handheld nebulizer designed to aerosolize medication approved for nebulization and prescribed by a physician. The Sidestream Plus is intended for adult and pediatric patients consistent with the indications for the aerosol medication.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mhhl Vhhl

K062689

---

**Source:** [https://fda.innolitics.com/device/K062689](https://fda.innolitics.com/device/K062689)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
