INNERSCAN BODY COMPOSITION MONITOR, INNERSCAN SEGMENTAL BODY COMPOSITION MONITOR
K062652 · Tanita Corporation of America · MNW · Dec 28, 2006 · Cardiovascular
Device Facts
Record ID
K062652
Device Name
INNERSCAN BODY COMPOSITION MONITOR, INNERSCAN SEGMENTAL BODY COMPOSITION MONITOR
Applicant
Tanita Corporation of America
Product Code
MNW · Cardiovascular
Decision Date
Dec 28, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The TANITA Ironman Innerscan Body Composition Monitor measures body weight, impedance and estimates total and segmental body fat percentage, body water, total and segmental muscle mass, physique rating, bone mass, visceral fat rating with healthy range, basal metabolic rate(BMR), daily caloric intake(DCI) and metabolic age using BIA (Bioelectrical Impedance Analysis). The device is intended to be used by generally healthy children 7-17 years old and generally healthy adults with active, moderately active, to inactive lifestyles for body composition assessment in the home environment. Actual measurements made by the BC-558 include weight and impedance. Based upon these measured values, values are calculated for: total and segmental body fat percentage, body water, total and segmental muscle mass, physique rating, bone mass, visceral fat rating with healthy range, basal metabolic rate (BMR), daily caloric intake (DCI) and metabolic age.
Device Story
Computer-operated body composition analyzer; utilizes 8-electrode segmental Bioelectrical Impedance Analysis (BIA) to measure weight and impedance. Device calculates total and segmental body fat percentage, body water, muscle mass, bone mass, visceral fat, BMR, DCI, and metabolic age. Used in home environment by patients for self-assessment. Output displayed to user to monitor body composition trends over time; aids in tracking health and fitness status.
Clinical Evidence
Bench testing only. Performance validated by comparing BIA methodology against internal BIA and DXA reference standards.
Technological Characteristics
Segmental BIA technology; 8 electrodes; 50kHz measurement frequency; max 500μA current. Measures weight and impedance. Battery-powered (AA). Dimensions: 150kg capacity, 100g/50g increments. Data memory for 31 days/52 weeks/36 months.
Indications for Use
Indicated for body composition assessment in generally healthy children (7-17 years) and adults with active to inactive lifestyles. Used in home environment.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
TANITA Body Fat Analyzer Professional and Consumer Models (K040778)
TANITA Body Fat Analyzer Professional and Consumer Models (K033157)
Submission Summary (Full Text)
{0}------------------------------------------------
#### SUMMARY
This summary of 510(k) safety and effectiveness information is accordance with the requirements of SMDA 1990 and 21 CFR 8807.92
Trade Name: Common Name:
TANITA Ironman Innerscan Body Composition Monitor: Model BC-558 Segmental Body Composition Monitor Classification Name: ANALYZER, BODY COMPOSITION 21 CFR §807.92
DEC 2 8 2006
Description of Applicant Device:
The TANTA Ironman Innerscan Body Composition Monitor BC-558 are a computer-operated body composition analyzer that utilizes BlA (bigledrical impedance analysis) to determine body fat percent, muscle mass, bone weight, visceral fat rating, Basal Metabolic Rate (BMR), metabolism age, physique rating, daily calorie intake (DCI), body water percent and segmental fat percent.
Intended Uses of Applicant Device:
The TANTA Ironnan Innerscan Body Composition Monitor measures body weight, impedance and estimates total and segmental body fat percentage, body water, total and segmental music news , bone mass, viseral fat bailed on bothere on bother on be becember and intake(OC) and metabolic age using Black on a natists). The device is ntended to be used by generally helthy hillfer ?-17 year old and generally healthy adults with active, to inactive lifestyles for body composition assessment in the home environment
Aculal measurements made by the BC-558 include weight and impedance. Based upon these measured values, values are calculated for: total and segmental body fat percentage, body water, total and segmental muscle mass, visce ratios be valoc baceratif frating with healthy rang, basal metabolic rate (BMR), daily caloric intake (DCI) and metabolic age.
#### Predicate Devices
TANITA Body Fat Analyzer Professional and Consumer Models K040778 and K033157
Scientific Concepts and Significant Performance Characteristics: wn in SECTION 9 and APPENDIX 1 1St പ്പി
| *Same as shown in SECTION 9. and APPENDIX 1. (Substantial Equivalence Matrix) | | | | |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|---|
| | New | Predicate Device#: K040778 | Predicate Device#: K033157 | |
| | BC-558 Specification | BC-53X Specification | BC-418 Specification | |
| Product Description | Body composition monitor/scale that<br>utilizes a segmental BIA technology to<br>determine internal body composition. | Body composition monitor/scale that<br>utilizes a "foot-to-foot" BIA technology<br>to determine internal body composition. | Body composition monitor/scale that<br>utilizes a segmental BIA technology to<br>determine internal body composition | |
| Analytical Method /<br>Measurement | Segmental BIA In house BIA and<br>DXEA reference | Foot-to Foot BIA In house BIA and<br>DXEA/Deuterium Dilution reference | Segmental BIA In house BIA and<br>DXEA reference | |
| Measurement Frequency | 50kHz | 50kHz | 50kHz | |
| Measurement Current | Max. 500μA | Max. 500μA | Max. 90 μ A | |
| Number of Electrodes | 8 | 4 | 8 | |
| Specifications | | | | |
| Capacity | 150 kg / 330 lb | 150 kg / 330 lb | 200 kg / 440 lb | |
| Increments | 100 g / 0.2 lb or 50 g / 0.1 lb | 100 g / 0.2 lb | 100 g / 0.2 lb | |
| Body Fat % | 0.1% | 0.1% | 0.1% | |
| User Memory | 4 | 4 | | |
| Input Age | 7-99<br>7-17: Child<br>18-99: Adult | 7-99<br>7-17: Child<br>18-99: Adult | 7-99<br>7-17: Child<br>18-99: Adult | |
| Input Height | 90cm~220cm | 100 cm - 220 cm | 90 cm - 249 cm | |
| Input Activity Level | 1 - 3 | 1 - 3 | | |
| Input Body Type | Standard / Athlete | Standard / Athlete | Standard / Athlete | |
| Data Memory | 31days / 1day average<br>52weeks / 1week average<br>36months / 1month average | 4weeks / 1week average<br>2months / 1month average | | |
| Power Supply | AA Batteries | AA Batteries | AC Adapter / DC5V 3.5A | |
| Printer Function | | | ✓ | |
| Computer Interface | - | | RS-232 | |
| Functions | | | | |
| Weight | ✓ | ✓ | ✓ | |
| | Full body | ✓ | ✓ | ✓ |
| | Right arm | ✓ | - | ✓ |
| Body Fat % | Left arm | ✓ | - | ✓ |
| | Right feet | ✓ | - | ✓ |
| | Left feet | ✓ | - | ✓ |
| | Trunk | ✓ | | ✓ |
| % Fat Judge<br>Body Water% | | ✓ | ✓ | ✓ |
| | Full body | ✓ | ✓ | ✓ |
| | Right arm | ✓ | | ✓ |
| Muscle Mass | Left arm | ✓ | | ✓ |
| | Right feet | ✓ | | ✓ |
| | Left feet | ✓ | - | ✓ |
| | Trunk | ✓ | - | ✓ |
| Physique Rating | | ✓ | ✓ | - |
| Bone mass | | ✓ | ✓ | - |
| Visceral Fat | | ✓ | ✓ | ✓ |
| Visceral Fat Judge | | ✓ | ✓ | - |
| BMR | | ✓ | ✓ | ✓ |
| DCI | | ✓ | ✓ | ✓ |
| Metabolic Age | | ✓ | ✓ | - |
{1}------------------------------------------------
### 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information in accordance with the requirements of SMDA 1990 and 21 CFR 8807.92
Side by side comparison of the TANITA Ironman Innerscan Body Composition Monitor BC-558 to the predicate that the applicant devices are substantially equivalent to those legally marketed devices.
Based on the results of using the previously approved BlA methodobogy with our whole body BlA, it was concluded that the TANITA i ronman Innerstan Body Composition Moritor BC-558 performs as well as the predicate devices and therefore have proven its safety and efficacy.
Beth Mackey
TANITA Corporation of America Marketing Director
Tel: (847)434-3981
Fax: (847)640-7978
July 7, 2006 Revised in red. December 14. 2006
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure embracing a globe, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
## DEC 2 8 2006
Ms. Beth Mackey Marketing Director TANITA Corporation of America, Inc. 2625 South Clearbrook Drive ARLINGTON HEIGHTS IL 60005
Re: K062652
> Trade/Device Name: TANITA Ironman Innerscan Body Composition Monitor Model BC-558 Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: December 14, 2006 Received: December 15, 2006
Dear Ms. Mackey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image is a circular logo or emblem. The central element is the letters "FDA" in a stylized font, with the word "Centennial" written below it in a cursive font. Above the letters "FDA" are the numbers "1906 - 2006". The entire design is surrounded by a circular border with text, and there are three stars below the word "Centennial".
ing and Promoting Public
{3}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.
Sincerely yours,
Nancy Cbrogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K 062652
Device Name: TANITA Ironman Innerscan Body Composition Monitor Model BC-558 Indications For Use: The TANITA Ironman Innerscan Body Composition Monitor measures body weight, impedance and estimates total and segmental body fat percentage, body water, total and segmental muscle mass, physique rating, bone mass, visceral fat rating with healthy range, basal metabolic rate(BMR), daily caloric intake(DCI) and metabolic age using BIA (Bioelectrical Impedance Analysis). The device is intended to be used by generally healthy children 7-17 years old and generally healthy adults with active, moderately active, to inactive lifestyles for body composition assessment in the home environment. Actual measurements made by the BC-558 include weight and impedance. Based upon these measured values, values are calculated for: total and segmental body fat percentage, body water, total and segmental muscle mass, physique rating, bone mass, visceral fat rating with healthy range, basal metabolic rate (BMR), daily caloric intake (DCI) and metabolic age.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancyc Brogdon
(vers 6/25/05)(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K062652
Page 1 of
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.