CARDIAC MARKERS CONTROL AND CALIBRATION VERIFICATION CONTROL SET

K062593 · Compass Bioscience · JJY · Nov 22, 2006 · Clinical Chemistry

Device Facts

Record IDK062593
Device NameCARDIAC MARKERS CONTROL AND CALIBRATION VERIFICATION CONTROL SET
ApplicantCompass Bioscience
Product CodeJJY · Clinical Chemistry
Decision DateNov 22, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

The Compass Bioscience Cardiac Markers Controls are to be used as a quality control material to assess the accuracy and precision of laboratory test methods used to measure the antigens and enzymes contained in the control. It is intended to validate the measurement of these analytes in patient samples. Three levels of control are provided to allow the performance of the analyte test methods to be monitored within the clinically significant range. The Compass Bioscience Cardiac Marker Calibration Verification Control Set is used to verify the calibration of various test methods over the measurable range of the test.

Device Story

Compass Bioscience Cardiac Markers Controls and Calibration Verification Control Set are frozen, human EDTA plasma-based materials containing preservatives, stabilizers, and sodium azide. Used in clinical laboratory settings to assess accuracy, precision, and calibration of test methods for CK-MB, Troponin I, Myoglobin, and BNP. Controls are provided in three levels; calibration verification sets in four levels (including zero). Healthcare providers use these materials to monitor assay performance within clinically significant ranges. Donor units are tested for HBsAg, HCV, HIV-1/2, and syphilis. Stability is maintained at -20°C.

Clinical Evidence

No clinical data. Bench testing only. Stability studies conducted using Kennon prediction models on Biosite Triage MeterPlus, Abbott AxSYM, and Bayer ADVIA Centaur platforms to establish open and closed vial stability durations.

Technological Characteristics

Human EDTA plasma matrix; fortified with human source antigens/enzymes and reagent-grade chemicals; sodium azide (<0.1%) preservative. Liquid form; frozen storage (-20°C) or refrigerated (2-8°C). No electronic or software components.

Indications for Use

Indicated for use as quality control material to assess accuracy and precision of laboratory test methods for measuring antigens and enzymes in patient samples, and to verify calibration of test methods over the measurable range. No specific patient population age or gender restrictions.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K062593 Prepared: 8/29/2006 Submitted by: Establishment Address: Evy K. Johnson Compass Bioscience 1850 Evergreen Street Duarte CA 91010 Phone: 626.359.96.45 FAX: 626.359.9645 NOV 2 2 2006 ### Establishment Registration Number: 2032652 Contact Person: Dr. Carter Grandjean, Director R&D Proprietary Name: Cardiac Markers Control and Calibration Verification Set Common Name: Cardiac Markers Control and Calibration Verifier Controls Classification Name: Multi Analyte Controls, All Kinds (assayed and unassayed) 75 JJY Substantial Equivalence: The Compass Bioscience Cardiac Markers Control and Calibration Set are supplied liguid in three or 4 levels and consist of a human plasma matrix containing preservatives to which reagent grade chemicals and human source antigens and enzymes have been added at different concentrations to achieve the levels. Assayed values are determined from in house and interlaboratory data. The Compass Bioscience controls and calibration verifiers are substantially equivalent to other such controls and verifiers in general use, such as the Triage® Profiler S.O.B.™ (Shortness of Breath) Control and the Triage® Profiler S.O.B.™ (Shortness of Breath) Calibration Verification Controls which are supplied liquid in a two level control set and five level calibration set as a human EDTA plasma matrix with human source constituents and pure chemicals added by the manufacturer. ### Description: Compass Bioscience Cardiac Markers Controls are supplied in three levels, 2 x 1 mL each level per box and as a four Level Calibration Verification set, 2 x 1 mL each level per box. The controls are supplied as a ready-to-use frozen liquid, requiring no reconstitution or dilution. They are prepared in a human EDTA plasma matrix fortified to target levels with human source material and reagent grade chemicals added at different concentrations to achieve the levels. Sodium Azide <0.1% has been added as preservative to inhibit microbial growth. ### Intended Use: The Compass Bioscience Cardiac Markers Controls are intended for use as quality control materials to assess the accuracy and precision of assay procedures for the analytes included in the control. The Compass Bioscience Cardiac Marker Calibration Set is used to verify the calibration of various test methods over the measurable range of the test. # Technological Characteristics Compared to Predicate Devices: The Compass Bioscience control products employ a similar matrix and constituent formulation to the equivalent predicate device listed above: human plasma matrix fortified with human source material, {1}------------------------------------------------ reagent grade chemicals and azide as a preservative. The Compass Bioscience Control and Calibration Verification Set also have similar storage and stability requirements as the equivalent devices. # Performance Characteristics: The closed vial stability claim made for this product is 1 year when stored at -20° C, based on accelerated stability studies.2 The Cardiac Markers control was stored at 2-8°C to simulate 1 year storage -20°C. An increase or decrease of >10% of analyte recovery compared to the initial test value ± the highest allowable instrument/reagent imprecision was used as the analyte failure criterion for determining shelf-life. Real time stability testing is ongoing on multiple lots of product. The closed and opened vial stability claim for this product when stored at 2-8° C is 30 days for CK-MB and Myoglobin and 14 days for Troponin I and BNP. Real time testing was used to determine the closed and open vial refrigerated shelf life. The Cardiac Markers Control at 2 - 8° C and the recovery (vs. day 0) was measured in intervals of 7 days. The product passed stability if its recovery was within ±(10% + % CVmethod) after 44 days of storage at 2- 8°C. Multiple lots of product were tested with no significant difference in performance. The equivalent predicate device, Triage® Profiler S.O.B.™ Control claims a 2 year shelf life for storage at -20°C and a single use. The Bayer Healthcare LLC Liquid Cardiac Markers Plus Control claims a 20 day opened/closed vial stability at refrigerated (2-8°C) storage. ## Assayed Values Assay values for the package inserts were established from assays performed in the Compass Bioscience laboratory using three Triage® MeterPlus readers with the Triage® Profiler S.O.B.™ Panel tests, the Triage® CardioProfiler Panel, and the Triage® Cardiac Panel, and from interlaboratory data using instrument manufacturers' reagents. Mean values for the listed lots of controls were calculated from multiple instruments and reagent lots available at the time of assay. Ranges were determined based on 2 SD - 3SD of the overall mean values from stability and interlaboratory data. <sup>2</sup> L. Kennon, Stability Prediction Model, Journal of Pharmaceutical Sciences 53:7, 815-818, 1964. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is positioned inside a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 NOV 2 2 2006 Ms. Evy K. Johnson Compass Bioscience 1850 Evergreen Street Duarte, CA 91010 Re: k062593 Trade/Device Name: Cardiac Markers Control Cardiac Markers Calibration Verification Control Set Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJY Dated: August 30, 2006 Received: September 6, 2006 Dear Ms. Johnson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Albert Gutt Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K K062593 Device Name: Cardiac Markers Control ## Cardiac Markers Calibration Verification Control Set Indications For Use: The Compass Bioscience Cardiac Markers Controls are to be used as a quality control material to assess the accuracy and precision of laboratory test methods used to measure the antigens and enzymes contained in the control. It is intended to validate the measurement of these analytes in patient samples. Three levels of control are provided to allow the performance of the analyte test methods to be monitored within the clinically significant range. The Compass Bioscience Cardiac Marker Calibration Control Set is used to verify the calibration of various test methods over the measurable range of the test. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Carole Benam Division Sign Off Page 1 of 1 sion Sion-Of Office of In Vitro Diagnostic Device Evaluation and Safety K062593
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...