SERENITY PSF (PNEUMATIC SKIN FLATTENING) SYSTEM

K062589 · Inolase 2002, Ltd. · GEX · Nov 30, 2006 · General, Plastic Surgery

Device Facts

Record IDK062589
Device NameSERENITY PSF (PNEUMATIC SKIN FLATTENING) SYSTEM
ApplicantInolase 2002, Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateNov 30, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Serenity PSF™ (Pneumatic Skin Flattening) System is an accessory for a compatible legally marketed Laser or Intense Pulse Light System for use in hair removal. Its handpiece produces a negative pressure over the skin surface just prior to the administration of the treatment beam through it.

Device Story

The Serenity PSF™ is a pneumatic accessory for laser or IPL hair removal systems. It generates negative pressure (200-700 mmHg) via a handpiece to flatten the skin at the treatment site. The handpiece features a transparent, thermally conductive sapphire window through which the laser or IPL beam is delivered. By creating tight mechanical contact, the device expels blood from the treatment site, enhancing light penetration and facilitating heat removal from the skin. This process aims to reduce post-treatment erythema. The device is operated by a clinician in a clinical setting in conjunction with a primary light-based treatment system.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Pneumatic suction system (200-700 mmHg); handpiece with sapphire window; thermally conductive interface; mechanical skin flattening; standalone accessory for laser/IPL systems.

Indications for Use

Indicated for use as an accessory to legally marketed laser or intense pulsed light (IPL) systems for hair removal.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ## 510(k) Summarv K 062589 Pursuant to CFR 807.92, the following 510(k) Summary is provided: - 1. (a) Submitter's George J. Hattub Address: MedicSense, USA 291 Hillside Avenue Somerset, MA 02726 NOV 3 U 2006 1. (b) Manufacturer Inolase 2002, Ltd. Address: 13 Giborei Israel St., PO Box 8497 Netanya (New Industrial Area) Israel, 42504 Mfg. Phone: 972-9-865-6750 Raphi Shavit, CEO Contact Person: Date: Revised- November 12, 2006 2. Device & Laser Handpiece Accessory System (Class 2), Product Code GEX, Classification Name: 21 CFR 878.4810 - Tradename of device: Serenity PSF™ (Pneumatic Skin Flattening) System Aesthera AlP™ Intense Pulsed Light System (K041554) Predicate Device: 3. The Serenity PSF™ (Pneumatic Skin Flattening) System is a device which has the 4. Description: ability to produce a vacuum at a level of 200-700 mmHg. Suction can be delivered through its handpiece, which has a sapphire window at its distal end, to be positioned over the skin for laser or IPL treatment. When the PFS system is activated, the negative pressure results in the flattening of the skin of the treatment site. During the time duration of suction, the treatment beam of the laser or IPL is administered through the sapphire window of the handpiece, which is transparent and thermally conductive. The resultant tight mechanical contact produced by the PSF expels blood from the treatment site which enhances the light penetration of the laser or IPL in tissue as well as the removal of its heat from the skin. This can reduce the possibility of post treatment erythema. The Serenity PSF™ (Pneumatic Skin Flattening) System is an accessory for a 5. Intended Use: compatible legally market Laser or Intense Pulse Light System for use in hair removal. Its handpiece produces a negative pressure over the skin surface just prior to the administration of the treatment beam through it. 6. Comparison of Technological With respect to technology, the Serenity PSF™ (Pneumatic Skin Flattening) System Characteristics: is substantially equivalent to one of its predicate device in that it produces a pneumatic connection of the treatment site for the delivery of photonic treatment. When the PSF™ System is used in combination with a legally marketed device such as an IPL system, it also has the same basic intended use as this predicate device. ్ {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes representing the department's commitment to health, services, and people. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Inolase 2002, Ltd. % MedicSense, USA Mr. George J. Hattub, RAC & CQE Senior Staff Consultant 291 Hillside Avenue Somerset, Massachusetts 02726 NOV 3 0 2006 Re: K062589 Trade/Device Name: Serenity PSF™ (Pneumatic Skin Flattening) System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: September 21, 2006 Received: September 25, 2006 Dear Mr. Hattub: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. George J. Hattub, RAC & CQE forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality of the (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will and w yours of substantial equivalence of your device of your device to a legally prematics notification. The stassification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your acc (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K 06 25 8-9 Device Name: Serenity PSF™ (Pneumatic Skin Flattening) System Indications For Use: The Serenity PSF™ (Pneumatic Skin Flattening) System is an accessory for a compatible legally marketed Laser or Intense Pulse Light System for use in hair removal. Its handpiece produces a negative pressure over the skin surface, just prior to the administration of the treatment beam through it. (revised 11/12/06) Prescription Use ﻨﮯ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) **(Division Sign-Off)** Page 1 of 1 Division of General, Restorative, and Neurological Devices 16062559 510(k) Number_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...