The Pressio® Ventricular Intracranial Pressure Monitoring Kit with Tunneling, Model PSO-VT, is indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage applications. The Pressio® Intracranial Parenchymal Pressure Monitoring Kit with Tunneling, Model PSO-PT, is indicated for use in subdural or parenchymal pressure monitoring. The Pressio® Intracranial Parenchymal Pressure Monitoring Kit with Bolt, Model PSO-PB, is indicated for use in parenchymal pressure monitoring.
Device Story
Pressio® ICP Monitoring System monitors intracranial pressure (ICP) via implantable catheters (ventricular or parenchymal). System includes monitor (PSO-3000), catheters (PSO-VT, PSO-PT, PSO-PB), transmitter, and interface unit. Catheters use strain gauge technology to sense pressure. Data transmitted to monitor for display of ICP curves. Monitor can connect to hospital patient monitors via interface cable for centralized display. Used in clinical settings by medical staff. Provides real-time pressure data to assist clinical decision-making regarding neurological status and CSF management. Benefits include accurate, continuous ICP monitoring to guide patient care.
Clinical Evidence
No clinical data. Evidence based on bench testing, including biocompatibility studies per ISO 10993 and performance testing of catheter designs. Safety testing conducted per IEC 60601-1 and UL 2601 standards.
Technological Characteristics
Strain gauge sensing technology. System components: monitor, implantable catheters, serial transmitter, interface control unit, cables. Connectivity: standalone or via connection to hospital patient monitors. Compliance with IEC 60601-1 and UL 2601 safety standards. Biocompatible materials per ISO 10993.
Indications for Use
Indicated for patients requiring intracranial pressure monitoring via intraventricular, subdural, or parenchymal access, and for cerebrospinal fluid drainage (PSO-VT model).
Regulatory Classification
Identification
An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.
{0}------------------------------------------------
K062584
JUL -5 2007
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### Name, Address, Phone and Fax Number of the Applicant 3.1
Sophysa SA 22 rue Jean Rostand Parc Club Orsay Université 91893 ORSAY Cedex, France Tel: 011-331-69 35 35 00 Fax: 011 331 69 35 36 90
#### Contact Person 3.2
Jean-Christophe Audras 22 rue Jean Rostand Parc Club Orsay Université 91893 ORSAY Cedex, France Tel: 011-331-69 35 35 06
#### Date Prepared 3.3
June 19, 2007
#### 3.4 Device Name
The Pressio® ICP Monitoring system is composed of the following elements:
- Pressio® ICP Monitor (PSO-3000) .
- Pressio® ICP Catheters (PSO-VT, PSO-PT, PSO-PB): there are four . available types of implantable catheters:
- Catheter implanted in ventricles via tunnelling (PSO-VT) .
- Catheter implanted in parenchyma via tunnelling (PSO-PT) .
- Catheter implanted in parenchyma via a bolt (PSO-PB) .
- Pressio® Serial Transmitter Model (PSO-TX00) .
- Pressio® Intracranial Pressure Interface Control Unit (PSO-IN00) .
- Catheter Extension Cable (PSO-EC20) .
- Monitor connexion cable (PSO-MCxx): "xx" depends on the type of . patient monitor available in the hospital, it exists 9 different references
- Power Supply Cable (PSO-AC) .
- Pressio® Pole Clamp (PSO-CL) .
- . Pressio® Disposable Hand Drill (PSO-DR)
Page 52
{1}------------------------------------------------
## 3.5 Device Description
The Pressio® ICP Monitoring System (PSO-3000) is an electromedical device designed for the monitoring of a patient's intracranial pressure (ICP) via a catheter implanted in the parenchyma (PSO-PB or PSO-PT) or in the ventricles (PSO-VT). The Pressio® ICP Monitoring System is sold as a kit containing a Pressio® ICP Monitor (PSO-3000), a Power Supply Cable (PSO-AC) and the Catheter Extension Cable (PSO-EC20).
The Pressio® ICP Monitor can be connected to a patient monitor via a compatible monitor connection cable (PSO-MCxx). This permits medical staff to display intracranial pressure curves on the patient monitor. This connection is not necessary for Pressio™ Intracranial Pressure Monitor functioning. The calibration of patient monitor is performed via a host monitor calibration key on Pressio™ Intracranial Pressure Monitor.
The Pressio® ICP Monitoring System is composed of the following:
- Pressio® ICP Monitor (PSO-3000)
- Pressio® ICP Catheters: there are three available types of implantable . catheters:
- Catheter implanted in ventricles via tunnelling (PSO-VT) .
- Catheter implanted in parenchyma via tunnelling (PSO-PT) .
- Catheter implanted in parenchyma via a bolt (PSO-PB) .
- . Pressio® Serial Transmitter Model (PSO-TX00)
- . Pressio® Intracranial Pressure Interface Control Unit (PSO-IN00)
- Catheter Extension Cable (PSO-EC20) .
- Monitor Connexion Cable (PSO-MCxx): "xx" depends on the type of . patient monitor available in the hospital, it exists 9 different references
- Power Supply Cable (PSO-AC) .
- Pressio® Pole Clamp (PSO-CL) .
- . Pressio® Disposable Hand Drill (PSO-DR)
### 3.6 Device Intended Use
The Pressio® Ventricular Intracranial Pressure Monitoring Kit with Tunneling, Model PSO-VT, is indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage applications.
The Pressio® Intracranial Parenchymal Pressure Monitoring Kit with Tunneling, Model PSO-PT, is indicated for use in subdural or parenchymal pressure monitoring.
The Pressio® Intracranial Parenchymal Pressure Monitoring Kit with Bolt, Model PSO-PB, is indicated for use in parenchymal pressure monitoring.
{2}------------------------------------------------
# Substantial Equivalence Summary 3.7
The Pressio® ICP monitoring system is substantially equivalent to the Codman® ICP express monitoring system (K945585, K914479, K991222 and K974088) in terms of intended use, materials, design, function and operating characteristics. The two monitoring systems utilize the same technology (strain gauge) to monitor Intracranial pressure (ICP).
## Device Testing 3.8
Biocompatibility studies were conducted per ISO 10993 standard and have demonstrated that the materials used to manufacture the Pressio® ICP monitoring system are safe for its intended use.
In addition, the three Implantable catheters were subjected to extensive performance testing. Results of the testing showed that the catheter designs are safe for their intended uses.
The Pressio® Intracranial Pressure Monitor, the Pressio® Interface Control Unit and the Pressio® Serial Transmitter underwent numerous safety tests, including testing to IEC 60601-1 and UL 2601.
Finally, the manufacturing process of the Pressio® ICP Monitoring System complies with the United States Food and Drug Administration and European Standards for the manufacturing of medical devices.
{3}------------------------------------------------
Image /page/3/Picture/12 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features the department's seal, which includes an abstract caduceus symbol with three figures in profile, representing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the seal.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 5 2007
SOPHYSA % Mr. Jean-Christophe Audras Director, Regulatory Affairs 22 rue Jean Rostand Parc Club Orsay Universite 91893 ORSAY Cedes - France
Re: K062584/S2
Trade/Device Name: Pressio® ICP Monitoring System Regulation Number: 21 CFR 882.1620 Regulation Name: Intracranial pressure monitoring device Regulatory Class: II Product Code: GWM Dated: June 25, 2007 Received: June 26, 2007
Dear Mr. Audras:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
# Page 2 - Mr. Jean-Christophe Audras
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
Anastacia In Bile
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Appendix E. Indications for Use
510(k) Number (if known): 上の62584
Device Name: Pressio® Intracranial Pressure Monitoring System Indications for Use:
The Pressio® Ventricular Intracranial Pressure Monitoring Kit with Tunneling, Model PSO-VT, is indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage applications.
The Pressio® Intracranial Parenchymal Pressure Monitoring Kit with Tunneling, Model PSO-PT, is indicated for use in subdural or parenchymal pressure monitoring.
The Pressio® Intracranial Parenchymal Pressure Monitoring Kit with Bolt, Model PSO-PB, is indicated for use in parenchymal pressure monitoring.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Posted November 13, 2003)
Anastasia Mc Bile
(Division Sign-Off Division of General, Restorative. and Neurological Devices
**510(k) Number** K062584
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.