DATEX-OHMEDA S/5 E-PSM(P) MODULE (CONSISTING OF E-PSM, E-PSMP AND E-INTPSM MODULES) AND ACCESSORIES
Applicant
GE Healthcare
Product Code
MHX · Cardiovascular
Decision Date
Sep 29, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Datex-Ohmeda S/5 PSM module (consisting of E-PSM, E-PSMP and E-INTPSM Modules) and accessories is intended for monitoring hemodynamic parameters of hospitalized patients.
Device Story
Hemodynamic plug-in parameter module for modular monitoring systems; inputs include ECG leads, SpO2 sensors, NIBP cuffs, temperature probes, and invasive pressure transducers. Module processes physiological signals; transmits data to monitor for display, trending, and alarm management. Used in clinical settings (hospitals) by qualified medical personnel. E-INTPSM interface module enables compatibility with S/5 F-CU8 and Compact monitor frames; allows mounting on IV-poles or anesthesia machines. Output provides real-time waveforms and vital sign readings to assist clinicians in patient assessment and clinical decision-making.
Clinical Evidence
Bench testing only. Device underwent validation and verification of specifications against recognized standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, ISO 9919, and EN 12470-4.
Technological Characteristics
Modular hemodynamic monitor; includes NIBP, 12-lead ECG, Impedance Respiration, SpO2, invasive pressure, and temperature measurement boards. Connectivity via module bus data communication. Software-based processing for ECG (heartbeat/arrhythmia detection) and NIBP. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, IEC 60601-2-49, ISO 9919, and EN 12470-4.
Indications for Use
Indicated for monitoring hemodynamic parameters of all hospital patients. Includes ECG (ST-segment/arrhythmia), Impedance respiration (ages 3+), NIBP (weight 5kg/11 lbs+), Temperature, SpO2 (including motion), and invasive blood pressure. For use by qualified medical personnel only.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
# KOG2576
Page 1 of 4
SEP 2 9 2006
# Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5 E-PSM(P) Module (consisting of E-PSM, E-PSMP and E-INTPSM Modules) and accessories
# GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
# COMPANY NAME/ADDRESS/PHONE/FAX:
GE Healthcare 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
#### NAME OF CONTACT:
Mr. Joel Kent
DATE:
August 22, 2006
#### DEVICE NAME as required by 807.92(a)(2)
#### TRADE NAME:
Datex-Ohmeda S/5 E-PSM(P) Module (consisting of E-PSM, E-PSMP and E-INTPSM Modules) and accessories
### COMMON NAME:
Multi-parameter Hemodynamic Module
{1}------------------------------------------------
## CLASSIFICATION NAME:
## The following Class II classifications appear applicable:
| Product Code | Classification Name | CFR Section |
|--------------|-------------------------------------------------------------|-------------|
| MHX | Arrhythmia detector & alarm | 870.1025 |
| MLD | Monitor ST-segment & alarm | 870.1025 |
| DQA | Oximeter | 870.2700 |
| DPZ | Ear Oximeter | 870.2710 |
| DRT | Cardiac Monitor (including cardiotachometer and rate alarm) | 870.2300 |
| DPS | Electrocardiograph | 870.2340 |
| DXN | Non-invasive blood pressure measurement system | 870.1130 |
| DSK | Blood pressure computer | 870.1110 |
| DRQ | Transducer signal amplifier and conditioner | 870.2060 |
| FLL | Clinical Electronic Thermometer | 880.2910 |
## NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The Datex-Ohmeda E-PSM Module (consisting of E-PSM, E-PSMP and E-INTPSM modules) and accessories is substantially equivalent in safety and effectiveness to the legally marketed and decessories to adocument of the module (consisting E-PSM and E-PSMP Modules) and accessories (K043551).
## DEVICE DESCRIPTION as required by 807.92(a)(4)
E-PSMP is a hemodynamic plug-in parameter module including the NIBP measurement, 12-lead ECG with the Impedance Respiration measurement, SpO2 with the plethysmographic waveform, two invasive pressure measurements (P1 and P2) and two temperature measurements (T1 and T2). two firessere measurements ( module for a modular monitoring system. This module can E-t own is a nemouring. I modular monitor and in other monitors using the new F-CU5P frame. With the new mounting accessories for the PSM module, E-PSM(P) plug-in parameter module can be removed from the FM monitor frame or S/5 F-CU5P frame and used near the patient on an IV-pole or anesthesia machine. Via the new E-INTPSM module and the new paticin on an IV-poic of unebhe massis. (P) plug-in parameter module can be attached thounting accessories for the S/5 F-CU8 frame or with Compact Monitor frames or used near the uncenty to a monitor asuage as a machine. S/5 F-CU8 frames or Compact Monitor frames have patient on I v pole of anesticoln his software licer. The monitors display waveforms and measurement readings, and handle the trending and alarm management. The ECG (e.g. heart beat medium in readings, and hance money and expiration algorithms are in the monitor software. and arrifything detection) and send them to the monitor. The NIBP, SpO2, Temperature and I he modules measure organis are in the module. There are two available options of the module: E-Invasive Pressure agormes P1 and P2 and E-PSM without P1 and P2. There are three r Sivil Will Invasive pressures I v aE-PSMP module for processing the measurement signals. Each parameter creat boards made the 2 1 chith software. The new E-INTPSM module and the new processing board has a mechanical and electrical interfacing between the E-PSM(P) module and mounting accessories are neemanted. a -INTPSM module and the new mounting accessories include only connections for the power supply and module bus data communication. There are no microcontrollers or software in the module or mounting accessories.
{2}------------------------------------------------
#### INTENDED USE as required by 807.92(a)(5)
Intended Use:
The Datex-Ohmeda S/5 PSM module (consisting of E-PSM, E-PSMP and E-INTPSM Modules) and accessories is intended for monitoring hemodynamic parameters of hospitalized patients.
Indications for use:
The Datex-Ohmeda S/5 PSM module (consisting of E-PSM, E-PSMP and E-INTPSM Modules) and accessories are indicated for monitoring of hemodynamic parameters of all hospital patients. The hemodynamic parameters of the module comprise ECG (including STsegment and arrhythmia), Impedance respiration, NIBP, Temperature, SpO2 (including monitoring during conditions of clinical patient motion), and invasive blood pressure. Impedance Respiration measurement is indicated for patients ages 3 and up. The NIBP measurement is indicated for patients who weigh 5kg (11 lbs.) and up. The device is indicated for use by qualified medical personnel only.
#### SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The Datex-Ohmeda E-PSM Module (consisting of E-PSM, E-PSMP and E-INTPSM modules) and accessories is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda E-PSM module (consisting E-PSM and E-PSMP Modules) and accessories (K043551).
The E-PSMP module (consisting E-PSM, E-PSMP and E-INTPSM Modules) and accessories have the following similarities compared to the predicate E-PSMP (K043551):
- same intended use and indications for use (only difference is to add E-INTPSM to the . name)
- identical fundamental scientific technology .
- . use the same operating principle
- the Customer and parameter specifications are the same except the Venous stasis values ◆ of the NIBP measurement
- have the same safety and effectiveness .
- have the same user interface at the monitor and alarms .
- are manufactured using the same processes .
The main differences between the new E-PSMP and the predicate E-PSMP (K043551) is primarily due to fact that the new E-PSM module has the following changes:
- added a new interface module, E-INTPSM, on the intended use and indications for use .
- with E-INTPSM the E-PSM(P) module can now used with S/5 F-CU8 monitor frame . and Compact monitor
- added new mounting accessories on the accessory list of the PSM module .
- added three reusable temperature probes and twelve disposable temperature probes .
- added an invasive pressure sensor on the accessory list of the PSM module .
- added new temperature and invasive pressure Care cables in the E-PSM(P) module . accessory list
- changed core material of the NIBP choke that is used in the power supply circuit of the . NIBP measurement board
- modified the software of the NIBP board to improve the Venous Stasis pressure used .
{3}------------------------------------------------
# SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The Datex-Ohmeda S/5 PSM module (consisting of E-PSM, E-PSMP and E-INTPSM Modules) and I IN Dates Onlined OF I DN I assist the standards below. The device has been thoroughly tested through validation and verification of specifications.
- FDA regulation 21 CFR 898.12 .
- IEC 60601-1:1988 + Amendments: A1:1991, ,A2:1995, .
- . IEC 60601-1-2:2001
- IEC 60601-1-4:1996 + A1 1999 .
- ANSI/AAMI ES1 (1993) .
- CAN/CSA C22.2 No. 601-1-M90 + SI (1994)+Amdt2:1998 .
- IEC 60601-2-27 (1994) .
- IEC 60601-2-30 (1999) .
- IEC 60601-2-34 (2000) .
- IEC 60601-2-49:2001 .
- EN 12470-4:2000 .
- ISO 9919 (1994) / EN 865:1997 .
- UL 2601-1 : 1997 .
#### CONCLUSION:
The summary above shows that there are no new questions of safety and effectiveness for the The Datex-Ohmeda S/5 PSM module (consisting of E-PSM, E-PSMP and E-INTPSM Modules) and accessories compared to the legally marketed (predicate) Datex-Ohmeda F-PSM module (consisting E-PSM and E-PSMP Modules) and accessories (K043551).
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a stylized logo of the United States Department of Health and Human Services (HHS). The logo features a stylized caduceus, which is a symbol of medicine, with a bird-like figure in the center. The bird-like figure is composed of three curved lines that resemble wings. The text "HHS" is located below the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
525 11 8 2006
GE Healthcare c/o Mr. Joel C. Kent Manager, Quality and Regulatory Affairs 86 Pilgrim Road Needham, MA 02492
Re: K062576
Trade Name: Datex-Ohmeda S/5 E-PSM(P) Module and accessories Regulation Number: 21 CFR 870.1025 Regulation Name: Physiological Patient Monitor (With Arrhythmia Detection or Alarm) Regulatory Class: Class II (two) Product Code: MHX, MLD, DSK, DRQ, DQA, DPZ, DRT, DPS, DXN, FLL Dated: August 28, 2006 Received: August 31, 2006
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass stated in to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, arorely initions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rr your de rice is existing (tional controls. Existing major regulations affecting your device can may oe subject to back as baseral Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Joel C. Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240)276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bermuda for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): __KCG ೭১76
# Device Name: Datex-Ohmeda S/5 PSM Module, (consisting of E-PSM, E-PSMP and E-INTPSM Modules) and accessories
Indications for Use:
The Datex-Ohmeda S/5 PSM module (consisting of E-PSM, E-PSMP and E-INTPSM Modules) and accessories are indicated for monitoring of hemodynamic parameters of all hospital patients. The hemodynamic parameters of the module comprise ECG (including ST-segment and arrhythmia), Impedance respiration, NIBP, Temperature, SpO2 (including monitoring during conditions of clinical patient motion), and invasive blood pressure.
Impedance Respiration measurement is indicated for patients ages 3 and up.
The NIBP measurement is indicated for patients who weigh 5kg (11 lbs.) and up.
The device is indicated for use by qualified medical personnel only.
Over-The-Counter Use _ Prescription Use AND/OR -(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ahumma
Division Sign-Off
Page _/ of _/
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.