K062543 · Focus Diagnostics, Inc. · MXJ · Feb 1, 2007 · Microbiology
Device Facts
Record ID
K062543
Device Name
HERPESELECT 1 AND 2 PLEXUS IGG, MODEL MP0900G
Applicant
Focus Diagnostics, Inc.
Product Code
MXJ · Microbiology
Decision Date
Feb 1, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3305
Device Class
Class 2
Indications for Use
Focus Diagnostics’ Plexus™ HerpeSelect®1 and 2 IgG is intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2 in human sera. The test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV-1 and HSV-2 infection. The predictive value of a positive or negative result depends on the population's prevalence and the pretest likelihood of HSV-1 and HSV-2 infection. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, for testing of immunocompromised patients, for use by a point of care facility or for use with automated equipment.
Device Story
Plexus HerpeSelect 1 and 2 IgG is a chemiluminescent-based fluorescent bead assay for qualitative detection of human IgG antibodies to HSV-1 and HSV-2 in serum. The device utilizes an antigen bead suspension containing distinct gG1 and gG2 antigen bead types. Procedure: 1) Patient serum incubation with antigen beads; 2) Addition of phycoerythrin-conjugated goat anti-human IgG to form a sandwich; 3) Measurement of fluorescence from each bead type compared against a cut-off calibrator. Requires Luminex xMAP or Bio-rad BioPlex 2200 instrumentation. Used in clinical laboratories to aid in HSV infection diagnosis. Output provides qualitative antibody status, assisting clinicians in patient management and diagnostic decision-making.
Clinical Evidence
Clinical performance evaluated in 300 expectant mothers and 300 sexually active adults. Compared against HerpeSelect 1/2 Immunoblot IgG. Expectant mothers: HSV-1 agreement 96.5% (pos)/92.2% (neg); HSV-2 agreement 94.3% (pos)/95.5% (neg). Sexually active adults: HSV-1 agreement 91.0% (pos)/96.5% (neg); HSV-2 agreement 96.3% (pos)/97.4% (neg). CDC panel (n=100) showed 100% agreement. Reproducibility studies showed %CV ≤ 10%. Cross-reactivity with CMV, EBV, VZV was 0-5%.
Technological Characteristics
Multiplexed immunoassay using antigen bead suspension. Beads fluoresce at distinct wavelengths/intensities. Three-step procedure: incubation with patient sera, addition of phycoerythrin-conjugated goat anti-human IgG, and fluorescence measurement. Standalone laboratory assay; non-automated.
Indications for Use
Indicated for sexually active adults or expectant mothers to aid in the presumptive diagnosis of HSV-1 and HSV-2 infection. Not for pediatric, neonatal, or immunocompromised populations.
Regulatory Classification
Identification
Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.
Special Controls
*Classification.* Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).
Predicate Devices
HerpeSelect 1 and 2 Immunoblot IgG
HerpeSelect-1 ELISA IgG
HerpeSelect-2 ELISA IgG
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number: K062543
B. Purpose for Submission: New device clearance
C. Measurand: Herpes Simplex Virus
D. Type of Test: Chemiluminescent technology for antibody technology
E. Applicant: Focus Diagnostics
F. Proprietary and Established Names: Plexus™ HerpeSelect®1 and 2 IgG
## G. Regulatory Information:
1. Regulation section: 866.3305
2. Classification: II
3. Product code: MXJ/MYF
4. Panel: Microbiology
## H. Intended Use:
Focus Diagnostics’ Plexus™ HerpeSelect®1 and 2 IgG is intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2 in human sera. The test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV-1 and HSV-2 infection. The predictive value of a positive or negative result depends on the population's prevalence and the pretest likelihood of HSV-1 and HSV-2 infection. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, for testing of immunocompromised patients, for use by a point of care facility or for use with automated equipment.
2. Indication(s) for use: NA
3. Special conditions for use statement(s): NA
4. Special instrument requirements: Luminex xMAP® or Bio-rad BioPlex® 2200
{2}
# I. Device Description:
The Focus Diagnostics Plexus HerpeSelect®1 and 2 IgG uses an Antigen Bead suspension that contains two distinct HSV antigen bead types that fluoresce at different wavelengths and/or intensities: gG1 beads and gG2 beads.
The Focus Diagnostics Plexus HerpeSelect®1 and 2 IgG is a three step procedure.
1. Patient sera are diluted, and the diluted sera are incubated with Antigen Beads. If HSV antibodies are present, then the antibodies bind to the corresponding antigen beads.
2. Phycoerythrin—conjugated goat anti-human IgG, (Conjugate) is added, and the Conjugate binds to the bound HSV antibody (if present), and forms a Conjugate-HSV antibody-antigen bead sandwich.
3. Fluorescence from each distinct HSV antigen bead type is measured and compared against a Cut-off Calibrator.
# J. Substantial Equivalence Information:
1. Predicate device name(s): HerpeSelect 1 and 2 Immunoblot IgG, HerpeSelect 1 ELISA IgG, HerpeSelect 2 ELISA IgG
2. Predicate 510(k) number(s): K000238, K02129, K021486
3. Comparison with predicate:
| Item | Device | Predicate |
| --- | --- | --- |
| | Focus Plexus HerpeSelect 1 and 2 IgG | Focus HerpeSelect 1 and 2 Immunoblot IgG |
| Similarities | | |
| Same intended use | Qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2 in human sera. | Qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2 in human sera. |
| Same indications for use | The test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV-1 and HSV-2 infection. | The test is indicated for testing sexually active adults or expectant mothers for aiding in the presumptive diagnosis of HSV-1 and HSV-2 infection. |
| Same type specific HSV-1 antigen | Recombinant gG1 antigen | Recombinant gG1 antigen |
| Same type specific HSV-2 antigen | Recombinant gG2 antigen | Recombinant gG2 antigen |
| Same immunoglobulin type | IgG | IgG |
| Same sample matrix | Serum | Serum |
| Differences | | |
| Methodology | Luminex-based fluorescent bead assay | Immunoblot assay |
| CLIA complexity | High | Moderate |
{3}
K. Standard/Guidance Document Referenced (if applicable): Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays (http://www.fda.gov/cdrh/oivd/guidance/1305.pdf)
L. Test Principle:
The Focus Diagnostics Plexus HerpeSelect®1 and 2 IgG uses an Antigen Bead suspension that contains two distinct HSV antigen bead types that fluoresce at different wavelengths and/or intensities: gG1 beads and gG2 beads.
The Focus Diagnostics Plexus HerpeSelect®1 and 2 IgG is a three step procedure.
1. Patient sera are diluted, and the diluted sera are incubated with Antigen Beads. If HSV antibodies are present, then the antibodies bind to the corresponding antigen beads.
2. Phycoerythrin—conjugated goat anti-human IgG, (Conjugate) is added, and the Conjugate binds to the bound HSV antibody (if present), and forms a Conjugate-HSV antibody-antigen bead sandwich.
4. Fluorescence from each distinct HSV antigen bead type is measured and compared against a Cut-off Calibrator.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility
Inter-laboratory Reproducibility and Inter/Intra-assay Reproducibility Focus, a clinical laboratory located in Southern California, and a university laboratory located in Northern California assessed the device's inter-laboratory reproducibility and inter/intra-assay reproducibility. Each of the three laboratories tested eleven samples in triplicate on five different days.
Inter-laboratory Reproducibility and Inter/Intra-assay Reproducibility
| Sample | HerpeSelect 1 Plexus IgG Results | | | | | HerpeSelect 2 Plexus IgG Results | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Intra- and Inter-assay | | | Inter-Lab | | Intra- and Inter-assay | | | Inter-Lab | |
| | Mean Index | Intra-assay % CV | Inter-assay % CV | Mean Index | % CV | Mean Index | Intra-assay % CV | Inter-assay % CV | Mean Index | % CV |
| 9 | 4.93 | 3.6 | 10.3 | 4.94 | 3.7 | 3.88 | 3.4 | 10 | 3.87 | 2.5 |
| 6 | 4.24 | 3.8 | 8.7 | 4.22 | 3.3 | 4.9 | 2.6 | 8.5 | 4.89 | 2.1 |
| 2 | 3.87 | 4.8 | 7.9 | 3.86 | 1.3 | 3.36 | 4.3 | 7.7 | 3.35 | 2 |
| 8 | 3.27 | 4.9 | 9.1 | 3.25 | 3 | 4.56 | 3.1 | 8.3 | 4.55 | 1.5 |
| 4 | 3.24 | 4.9 | 7.4 | 3.22 | 2.1 | 2.55 | 4.5 | 8.9 | 2.54 | 5.8 |
{4}
4
| 1 | 3.04 | 4.3 | 8.9 | 3.02 | 2.3 | 2.71 | 3.8 | 9.3 | 2.7 | 2.1 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 12^{b} | 2.13 | 7.9 | 8.7 | 2.13 | 4.1 | 1.87 | 7.2 | 8.8 | 1.87 | 3.4 |
| 3 | 0.34 | 9.1 | 14.9 | 0.34 | 6.8 | 0.06 | 8.7 | 28.3 | 0.06 | 22.6 |
| 10^{c} | 0.19 | 9.9 | 213.1 | 0.19 | 59 | 0.12 | 11.4 | 334.2 | 0.4 | 103.8 |
| 10^{d} | 0.13 | 10 | 15.8 | 0.12 | 1.9 | 0.06 | 11.5 | 41.7 | 0.06 | 38.3 |
| 7 | 0.18 | 8.3 | 16.3 | 0.17 | 9.4 | 0.06 | 8.1 | 23.7 | 0.06 | 17.3 |
| 5 | 0.14 | 9 | 16 | 0.14 | 2.7 | 0.06 | 8.3 | 39.8 | 0.06 | 38.1 |
a. Excludes two runs at one site that were invalid because the Negative Control index was beyond the acceptable QC criteria (it appears that the Positive Control was run in those wells since the indices were about 1.9 for both gG1 and gG2)
b. Samples 12 (inter-lab reproducibility) and 14 (inter-lot reproducibility below) were separate samples, but they were made with the same sera Samples 11 did not have sufficient volume to be sent to investigators.
c. This line includes all data for Sample 10, including one run at Lab 2, where it appears that Sample I may have been run instead since the indices were about 2.7 for both gG1 and gG 2.
d. This line includes all data for Sample 10. except for one run at Lab 2, where it appears that Sample I may have been run instead since the indices were about 2.7 for both gGI and gG2
## Inter-Lot Reproducibility
Focus assessed the device's Inter-lot Reproducibility by testing eleven samples with three separate lots. The samples were run in triplicate. Each lot had a different set of gG-I and gG2 beads, a different lot of conjugate (made from 2 different stock conjugates), and a different lot of calibrator (made from 2 different combinations of positive and negative sera). The results of the studies are summarized in the tables below:
| Sample | HSV-1 | | HSV-2 | |
| --- | --- | --- | --- | --- |
| | Mean | Inter-Lot | Mean | Inter-Lot |
| | Index | %CV | Index | %CV |
| 9 | 5.2 | 7.4 | 3.9 | 12.8 |
| 6 | 4.36 | 8.5 | 4.76 | 9.2 |
| 2 | 3.6 | 7.3 | 3.19 | 9.3 |
| 4 | 3.29 | 7.9 | 2.54 | 6.4 |
| 8 | 3.23 | 11.3 | 4.45 | 9.4 |
| 1 | 3.14 | 5.8 | 2.73 | 4.8 |
| 12/14* | 2.22 | 10.5 | 1.86 | 7.8 |
| 3 | 0.31 | 17 | 0.11 | 50.9 |
| 7 | 0.15 | 31.3 | 0.08 | 21.8 |
| 5 | 0.1 | 45.6 | 0.06 | 24.9 |
| 10 | 0.09 | 50.6 | 0.06 | 26.9 |
*Samples 12 (inter-lab reproducibility above) and 14 (inter-lot reproducibility) were separate samples, but they were made with the same sera. Samples II and 13 did not have sufficient volume to be sent to investigators.
{5}
b. Linearity/assay reportable range: not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods): not applicable
d. Detection limit: not applicable
e. Analytical specificity
i. Cross-reactivity
**Cross-reactivity (n = 51)**
Focus assessed cross-reactivity with two groups of samples: a "HSV ELISA dual negative" group (n=37), and a "HSV ELISA mixed seroreactivity" group (n=14).
The HSV ELISA dual negative group (n=37) included samples that were sero-negative with both the HerpeSelect-1 ELISA IgG and HerpeSelect 2 ELISA IgG, and were sero-positive by at least one of a FDA cleared CMV ELISA IgG (n = 18), a home brew VZV ACIF (n=32), a FDA cleared EBV VCA IgG (n=31). The HerpeSelect 1 and 2 Plexus IgG was HSV-1 negative with all but one of the HSV ELISA dual negatives, and equivocal with one sample (the one sample was CMV+ VZV+ and EBV+). The HerpeSelect 1 and 2 Plexus IgG was HSV-2 negative with all but one of the HSV ELISA dual negatives, and equivocal with one sample (the one sample was CMV- VZV+ and EBV+).
The HSV ELISA mixed reactivity group (n=14) included samples that were sero-positive with either the HerpeSelect-1 ELISA IgG or HerpeSelect 2 ELISA IgG, and were sero-positive by at least one of a FDA cleared CMV ELISA IgG (HSV-1 neg n = 2, HSV-2 neg n = 9), a home brew VZV ACIF (HSV-1 neg n = 1, HSV-2 neg n = 2), a FDA cleared EBV VCA IgG (HSV-1 neg n = 1, HSV-2 neg n = 0). The HerpeSelect 1 and 2 Plexus IgG was HSV-1 negative with all of the HSV-1 ELISA negatives in the mixed reactivity group.
The HerpeSelect 1 and 2 Plexus IgG was HSV-2 negative with all of the HSV-2 ELISA negatives in the mixed reactivity group.
| Cross-reactant | HSV ELISAs | HerpeSelect-1 Plexus | | | | | HerpeSelect-2 Plexus | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | n | Neg | Eqv* | Pos | % POS | n | Neg | Eqv* | Pos | % POS |
| CMV IgG + | Dual Neg | 18 | 17 | 1 | 0 | 5.6% (1/18)
95%CI 0.1-27.3% | 18 | 18 | 0 | 0 | 0.0% (0/18) 95%CI
0.0-18.5% |
| | +/- or -/+ | 2 | 2 | 0 | 0 | 0.0% (0/2) 95%CI
0.0-84.2% | 9 | 9 | 0 | 0 | 0.0% (0/9) 95%CI
0.0-33.6% |
| | Total | 20 | 19 | 1 | 0 | 5.0% (1/20)
95%CI 0.1-24.9% | 27 | 27 | 0 | 0 | 0.0% (0/27) 95%CI
0.0-12.8% |
{6}
6
iv. Interfering substances: not applicable
f. Assay cut-off
2. Comparison studies:
a. Method comparison with predicate device: Immunoblot IgG as the reference method, see clinical studies
b. Matrix comparison: not applicable
3. Clinical studies:
PERFORMANCE CHARACTERISTICS
Summary of Studies (see details below)
| Study | HerpeSelect 1 Plexus IgG Results | HerpeSelect 2 Plexus IgG Results |
| --- | --- | --- |
| Expectant Mothers (Indicated Population) | Agreement with positives | 96.5% |
| | Agreement with negatives | 92.2% |
| Sexually Active Adults (Indicated Population) | Agreement with positives | 91.0% |
| | Agreement with negatives | 96.5% |
| CDC HSV/CMV Panel | Agreement with positives | 100% |
| | Agreement with negatives | 100% |
| Low Prevalence Population | Agreement with negatives | 97.9% |
| Cross-reactivity with CMV, EBV and VZV. | Cross-reactivity | 0-5% |
| Reproducibility | %CV of positives | ≤10% |
Reactivity with Expectant Mothers (n = 300)
{7}
Focus (n = 150) and an external investigator (n = 150) assessed the device's reactivity with sera from Expectant Mothers. The sera were sequentially submitted to the laboratory, archived, and masked. The external investigator was an University laboratory located in Northern California, and the sera were collected in the Pacific Northwestern United States. The HerpeSelect Plexus results were compared to the HerpeSelect 1 ELISA IgG and the HerpeSelect 2 ELISA IgG, using the Focus HerpeSelect 1 and 2 Immunoblot IgG as the reference method.
## HSV-1 Reactivity
The Focus HerpeSelect 1 Immunoblot IgG was: HSV-1 positive for 170 samples, HSV-1 negative with 128 samples, and HSV Common Antigen band positive for two samples. The Plexus HerpeSelect 1 agreed with: 96.5% (164/170) of Immunoblot positives, and 92.2% (118/128) of Immunoblot negatives. The two Immunoblot HSV Common Antigen band positives were both negative in the Plexus.
## HSV-2 Reactivity
The Focus HerpeSelect 2 Immunoblot IgG was: HSV-2 positive for 122 samples, HSV-2 negative with 176 samples, and HSV Common Antigen band positive for two samples. The Plexus HerpeSelect 2 agreed with: 94.3% (115/122) of Immunoblot positives, and 95.5% (168/176) of Immunoblot negatives. The two Immunoblot HSV Common Antigen band positives were both negative in the Plexus.
Plexus HerpeSelect 1 IgG Reactivity with Expectant Mothers (n = 300)
| Lab | Herpe - Select Immunoblot | Plexus HerpeSelect-1 | | | | | HerpeSelect-1 ELISA | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | n | Ne g | Eqv | Pos | % Agreement | n | Ne g | Eqv | Pos | % Agreement |
| Site 1 | Pos | 84 | 6 | 0 | 78 | 92.9% (78/84) 95%CI 85.1-97.3% | 84 | 5 | 1 | 78 | 92.9% (78/84) 95%CI 85.1-97.3% |
| Focus | Pos | 86 | 0 | 0 | 86 | 100% (86/86) 95%CI 95.8-100% | 86 | 0 | 0 | 86 | 100% (86/86) 95%CI 95.8-100% |
| Combined Labs | Pos | 170 | 6 | 0 | 164 | 96.5% (164/170) 95%CI 92.5-98.7% | 170 | 5 | 1 | 164 | 96.5% (164/170) 95%CI 92.5-98.7% |
| Site 1 | Neg | 66 | 61 | 1 | 4 | 92.4% (61/66) 95%CI 83.2-97.5% | 66 | 59 | 2 | 5 | 89.4% (59/66) 95%CI 79.4-95.6% |
| Focus | Neg | 62 | 57 | 2 | 3 | 91.9% (57/62) 95%CI 82.2-97.3% | 62 | 59 | 1 | 2 | 95.2% (59/62) 95%CI 86.5-99.0% |
| Combined Labs | Neg | 128 | 118 | 3 | 7 | 92.2% (118/128) 95%CI 86.1-96.2% | 128 | 118 | 3 | 7 | 92.2% (118/128) 95%CI 86.1-96.2% |
| Site 1 | Com | 0 | 0 | 0 | 0 | Na | 0 | 0 | 0 | 0 | Na |
| Focus | Com | 2 | 2 | 0 | 0 | Na | 2 | 2 | 0 | 0 | Na |
{8}
8
| Combined Labs | Com | 2 | 2 | 0 | 0 | Na | 2 | 2 | 0 | 0 | Na |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
Plexus HerpeSelect 2 IgG Reactivity with Expectant Mothers (n = 300)
| Lab | Herpe - Select Immunoblot | Plexus HerpeSelect-2 | | | | | HerpeSelect-2 ELISA | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | n | Ne g | Eqv | Pos | % Agreement | n | Ne g | Eqv | Pos | % Agreement |
| Site 1 | Pos | 60 | 3 | 1 | 56 | 93.3% (56/60)
95%CI 83.8-98.2% | 60 | 2 | 0 | 58 | 96.7% (58/60)
95%CI 88.5-99.6% |
| Focus | Pos | 62 | 2 | 1 | 59 | 95.2% (59/62)
95%CI 86.5-99.0% | 62 | 1 | 0 | 61 | 98.4% (61/62)
95%CI 91.3-100% |
| Combined Labs | Pos | 122 | 5 | 2 | 115 | 94.3% (115/122)
95%CI 88.5-97.7% | 122 | 3 | 0 | 119 | 97.5% (119/122)
95%CI 93.0-99.5% |
| Site 1 | Neg | 90 | 88 | 0 | 2 | 97.8% (88/90)
95%CI 92.2-99.7% | 90 | 86 | 0 | 4 | 95.6% (86/90)
95%CI 89.0-98.8% |
| Focus | Neg | 86 | 80 | 3 | 3 | 93.0% (80/86)
95%CI 85.4-97.4% | 86 | 80 | 1 | 5 | 93.0% (80/86)
95%CI 85.4-97.4% |
| Combined Labs | Neg | 176 | 168 | 3 | 5 | 95.5% (168/176)
95%CI 91.2-98.0% | 176 | 166 | 1 | 9 | 94.3% (166/176)
95%CI 89.8-97.2% |
| Site 1 | Com | 0 | 0 | 0 | 0 | na | 0 | 0 | 0 | 0 | na |
| Focus | Com | 2 | 2 | 0 | 0 | Na | 2 | 2 | 0 | 0 | Na |
| Combined Labs | Com | 2 | 2 | 0 | 0 | Na | 2 | 2 | 0 | 0 | Na |
## Reactivity with Sexually Active Adults (n = 300)
Focus (n = 150) and an external investigator (n = 150) assessed the device's reactivity with sera from sexually active adults. The sera were sequentially submitted to the laboratory, archived, and masked. The external investigator was a clinical laboratory located in Southern California, and the sera were collected in the Pacific Northwestern United States. The HerpeSelect Plexus results were compared to the HerpeSelect 1 ELISA IgG and the HerpeSelect 2 ELISA IgG, using the Focus HerpeSelect 1 and 2 Immunoblot IgG as the reference method.
## HSV-1 Reactivity
The Focus HerpeSelect 1 Immunoblot IgG was: HSV-1 positive for 157 samples, HSV-1 negative with 142 samples, and HSV Common Antigen band positive for one sample.
The Plexus HerpeSelect 1 agreed with: 91.0% (142/156) of Immunoblot positives (one sample was not run on the Plexus device), and
96.5% (137/142) of Immunoblot negatives.
## HSV-2 Reactivity
{9}
The Focus HerpeSelect 2 Immunoblot IgG was: HSV-2 positive for 109 samples, HSV-2 negative with 190 samples, and HSV Common Antigen band positive for one sample. The Plexus HerpeSelect 2 agreed with: $96.3\%$ (105/109) of Immunoblot positives, and $97.4\%$ (184/189) of Immunoblot negatives (one sample was not run on the Plexus device).
Plexus HerpeSelect 1 IgG Reactivity with Sexually Active Adults (n = 300)
| Lab | Herpe - Select Immunoblo t | Plexus HerpeSelect-1 | | | | | HerpeSelect-1 ELISA | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | n | Ne g | Eqv | Pos | % Agreement | n | Ne g | Eqv | Pos | % Agreement |
| Site 2 | Pos | 71 | 4 | 2 | 65 | 91.5% (65/71) 95%CI 82.5-96.8% | 71 | 3 | 0 | 68 | 95.8% (68/71) 95%CI 88.1-99.1% |
| Focus | Pos | 85* | 5 | 3 | 77 | 90.6% (77/85) 95%CI 82.3-95.9% | 86 | 4 | 2 | 80 | 93.0% (80/86) 95%CI 85.4-97.4% |
| Combined Labs | Pos | 156 | 9 | 5 | 142 | 91.0% (142/156) 95%CI 85.4-95.0% | 157 | 7 | 2 | 147 | 93.6% (147/157) 95%CI 88.6-96.9% |
| Site 2 | Neg | 79 | 78 | 1 | 0 | 98.7% (78/79) 95%CI 93.1-100% | 79 | 77 | 1 | 1 | 97.5% (77/79) 95%CI 91.2-99.7% |
| Focus | Neg | 63 | 59 | 2 | 2 | 93.7% (59/63) 95%CI 84.5-98.2% | 63 | 60 | 0 | 3 | 95.2% (60/63) 95%CI 86.7-99.0% |
| Combined Labs | Neg | 142 | 137 | 3 | 2 | 96.5% (137/142) 95%CI 92.0-98.9% | 142 | 137 | 1 | 4 | 96.5% (137/142) 95%CI 92.0-98.9% |
| Site 2 | Com | 0 | 0 | 0 | 0 | na | 0 | 0 | 0 | 0 | na |
| Focus | Com | 1 | 1 | 0 | 0 | na | 1 | 1 | 0 | 0 | na |
| Combined Labs | Com | 1 | 1 | 0 | 0 | na | 1 | 1 | 0 | 0 | na |
*One of 300 samples was not run in the Plexus HerpeSelect, and that one sample was HSV-1 negative/HSV-2 positive in the Immunoblot.
Plexus HerpeSelect 2 IgG Reactivity with Sexually Active Adults (n = 300)
| Lab | Herpe - Select Immunoblo t | Plexus HerpeSelect-2 | | | | | HerpeSelect-2 ELISA | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | n | Ne g | Eqv | Pos | % Agreement | n | Ne g | Eqv | Pos | % Agreement |
| Site 2 | Pos | 47 | 3 | 1 | 43 | 91.5% (43/47) 95%CI 79.6-97.6% | 47 | 1 | 0 | 46 | 97.9% (46/47) 95%CI 88.7-99.9% |
| Focus | Pos | 62 | 0 | 0 | 62 | 100%(62/62) 95%CI 94.2-100% | 62 | 0 | 1 | 61 | 98.4%(61/62) 95%CI 91.3-100% |
{10}
| Combined Labs | Pos | 109 | 3 | 1 | 105 | 96.3% (105/109) 95%CI 90.9-99.0% | 109 | 1 | 1 | 107 | 98.2% (107/109) 95%CI 93.5-99.8% |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Site 2 | Neg | 103 | 100 | 1 | 2 | 97.1% (100/103) 95%CI 91.7-99.4% | 103 | 102 | 0 | 1 | 99.0% (102/103) 95%CI 94.7-100% |
| Focus | Neg | 86* | 84 | 0 | 2 | 97.7% (84/86) 95%CI 91.8-99.7% | 87 | 84 | 1 | 2 | 96.6% (84/87) 95%CI 90.3-99.3% |
| Combined Labs | Neg | 189 | 184 | 1 | 4 | 97.4% (184/189) 95%CI 93.9-99.1% | 190 | 186 | 1 | 3 | 97.9% (186/190) 95%CI 94.7-99.4% |
| Site 2 | Com | 0 | 0 | 0 | 0 | na | 0 | 0 | 0 | 0 | na |
| Focus | Com | 1 | 1 | 0 | 0 | na | 1 | 1 | 0 | 0 | na |
| Combined Labs | Com | 1 | 1 | 0 | 0 | na | 1 | 1 | 0 | 0 | na |
*One of 300 samples was not run in the Plexus HerpeSelect, and that one sample was HSV-1 negative/HSV-2 positive in the Immunoblot.
# Agreement with CDC Panel $(\mathbf{n} = 100)$
The following information is from a serum panel obtained from the CDC and tested by Focus Diagnostics. The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC.
The test panel consists of 100 samples. This panel contains duplicate samples of 50 test sera. The duplicates serve to test for reproducibility. There are 16 HSV-1 positive, 7 HSV-2 positive, 11 double-positive and 16 double-negative sera resulting in 54 HSV-1 positive and 36 HSV-2 positive specimens.
# Determination of positive and negative samples
Of the 54 HSV-1 positive samples, the HerpeSelect® Plexus IgG correctly identified $100\%$ (54/54).
Of the 36 HSV-2 positive samples, the HerpeSelect® Plexus IgG correctly identified $100\%$ (36/36).
Of the 22 double positive samples, the HerpeSelect® Plexus IgG correctly identified $100\%$ (22/22).
Of the 32 double negative samples, the HerpeSelect® Plexus IgG correctly identified $100\%$ (32/32).
Agreement with CDC Panel (n = 100)
| Sample Type | CDC Result | | n | HerpeSelect-1 Plexus Results | | | | HerpeSelect-2 Plexus Results | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | HSV1 | HSV2 | | Neg | Eqv | Pos | % Agreement | Neg | Eqv | Pos | % Agreement |
| HSV-1 Positive | Pos | Neg | 32 | 0 | 0 | 32 | 100% (32/32) 95%CI 89.1-100% | 32 | 0 | 0 | 100% (32/32) 95%CI 89.1-100% |
| HSV-2 Positive | Neg | Pos | 14 | 14 | 0 | 0 | 100% (14/14) 95%CI 76.8-100% | 0 | 0 | 14 | 100% (14/14) 95%CI 76.8-100% |
| Dual | Pos | Pos | 22 | 0 | 0 | 22 | 100% (22/22) | 0 | 0 | 22 | 100% (22/22) |
{11}
| Positive | | | | | | | 95%CI 84.6-100% | | | | 95%CI 84.6-100% |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Dual Negative | Neg | Neg | 32 | 32 | 0 | 0 | 100% (32/32)
95%CI 89.1-100% | 32 | 0 | 0 | 100% (32/32)
95%CI 89.1-100% |
## CDC Panel Reproducibility
All paired sera were correctly identified: The Focus Diagnostics HerpeSelect® 1 and 2 Plexus IgG identified 16 out of 16 paired HSV-1 positive and HSV-2 negative (100%), 7 out of 7 paired HSV-2 positive and HSV-1 negative (100%), 11 out of 11 paired double-positive (100%) and 16 out of 16 paired double-negative (100%) samples.
## Specificity with a Low Prevalence Population (n = 77)
Focus (n = 77) assessed the device's reactivity with sera from a low prevalence population. Focus selected sera from patients aged 18 and 19 years, and that had been submitted to a clinical laboratory in Southern California from states having a history of low sexually transmitted disease prevalence. Focus excluded sera that were submitted for sexually transmitted diseases, herpesvirus testing, and tests indicating the patient may be immunocompromised. The sera were sequentially selected, archived and masked. The HerpeSelect Plexus results were compared to the Focus HerpeSelect 1 and 2 Immunoblot IgG.
## HSV-1 Reactivity
The Focus HerpeSelect 1 Immunoblot IgG was: HSV-1 positive for 28 samples, HSV-1 negative with 47 samples, and HSV Common Antigen band positive for two samples.
The Plexus HerpeSelect 1 agreed with: 96.4% (27/28) of Immunoblot positives (one HSV-1 Immunoblot positive was equivocal in the Plexus device), and 97.9% (46/47) of Immunoblot negatives (one HSV-1 Immunoblot negative was positive in the Plexus device). One Immunoblot HSV Common Antigen band positive was equivocal in the Plexus, and the other sample was Plexus negative.
## HSV-2 Reactivity
The Focus HerpeSelect 2 Immunoblot IgG was: HSV-2 positive for four samples, HSV-2 negative with 71 samples, and HSV Common Antigen band positive for two samples. The Plexus HerpeSelect 2 agreed with: 75.0% (3/4) of Immunoblot positives (one HSV-1 Immunoblot positive was equivocal in the Plexus device), and 100% (71/71) of Immunoblot negatives (one HSV-1 Immunoblot negative was positive in the Plexus device). Both Immunoblot HSV Common Antigen band positives were negative in the Plexus.
{12}
Specificity with Low Prevalence Population (n = 77)
| HerpeSelect Immunoblot | Plexus HerpeSelect-1 | | | | | Plexus HerpeSelect-2 | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | n | Ne g | Eq v | Pos | % Agreement | n | Ne g | Eq v | Pos | % Agreement |
| Pos | 28 | 0 | 1 | 27 | 96.4% (27/28) 95%CI 81.6-99.9% | 4 | 0 | 1 | 3 | 75.0% (3/4) 95%CI 19.4-99.4% |
| Neg | 47 | 46 | 0 | 1 | 97.9% (46/47) 95%CI 88.7-99.9% | 71 | 71 | 0 | 0 | 100% (71/71) 95%CI 94.9-100% |
| Com | 2 | 1 | 1 | 0 | Na | 2 | 2 | 0 | 0 | na |
4. Clinical cut-off:
5. Expected values/Reference range:
EXPECTED VALUES
An outside investigator assessed the device with masked, archived and unselected sera from 1) sexually active adults (n = 300), and 2) from expectant mothers (n = 300). The reference methods were Focus Diagnostics HerpeSelect 1 ELISA IgG and HerpeSelect 2 ELISA IgG. The observed prevalences and the hypothetical predictive values for the two populations are shown below. The positive predictive value will decrease proportionally to the prevalence of HSV infection as reflected in the table below. The calculations are based on HerpeSelect 1 and 2 Plexus IgG having
1) HSV-1 sensitivity of 96.5% and a HSV-1 specificity of 92.2% (expectant mothers), and
2) HSV-2 sensitivity of 94.3% and a HSV-2 specificity of 95.5% (expectant mothers).
3) HSV-1 sensitivity of 91.0% and a HSV-1 specificity of 96.5% (Sexually Active Adults), and
4) HSV-2 sensitivity of 96.3% and a HSV-2 specificity of 97.4% (Sexually Active Adults).
Observed Prevalence with Sexually Active Adults
| Sero-status | Observed Prevalence | | |
| --- | --- | --- | --- |
| | HerpeSelect ELISA | HerpeSelect Immunoblot | HerpeSelect Plexus |
| HSV-1 neg | 48.3% (145/300) | 47.3% (142/300) | 49.2% (147/299) |
| HSV-1 + | 50.3% (151/300) | 52.0% (156/300) | 48.2% (144/299) |
| HSV-2 neg | 62.7% (188/300) | 63.0% (189/300) | 62.9% (188/299) |
| HSV-2 + | 36.7% (110/300) | 36.3% (109/300) | 36.4% 9109/299) |
{13}
Observed Prevalence with Expectant Mothers
| Sero-status | Observed Prevalence | | |
| --- | --- | --- | --- |
| | HerpeSelect ELISA | HerpeSelect Immunoblot | HerpeSelect Plexus |
| HSV-1 neg | 41.7% (125/300) | 42.6% (128/300) | 42.0% (126/300) |
| HSV-1 + | 57.0% (171/300) | 56.7% (170/300) | 57.0% (171/300) |
| HSV-2 neg | 57.0% (171/300) | 58.7% (176/300) | 58.3% (175/300) |
| HSV-2 + | 42.7% (128/300) | 40.7% (122/300) | 40.0% (120/300) |
Expectant Mothers Prevalence vs. Hypothetical Predictive Values
| Prevalence | HSV-1 | | HSV-2 | |
| --- | --- | --- | --- | --- |
| | PPV | NPV | PPV | NPV |
| 50% | 92.5 | 92.2 | 95.4 | 95.5 |
| 40% | 89.2 | 94.7 | 93.3 | 97.0 |
| 30% | 84.1 | 96.5 | 90.0 | 98.0 |
| 25% | 80.5 | 97.3 | 87.5 | 98.5 |
| 20% | 75.6 | 97.9 | 84.0 | 98.8 |
| 15% | 68.6 | 98.5 | 78.7 | 99.2 |
| 10% | 57.9 | 99.1 | 70.0 | 99.5 |
| 5% | 39.4 | 99.6 | 52.4 | 99.8 |
Sexually Active Adults Prevalence vs. Hypothetical Predictive Values
| Prevalence | HSV-1 | | HSV-2 | |
| --- | --- | --- | --- | --- |
| | PPV | NPV | PPV | NPV |
| 50% | 96.3 | 96.5 | 97.4 | 97.4 |
| 40% | 94.5 | 97.6 | 96.1 | 98.3 |
| 30% | 91.8 | 98.5 | 94.1 | 98.9 |
| 25% | 89.7 | 98.8 | 92.5 | 99.1 |
{14}
| 20% | 86.7 | 99.1 | 90.3 | 99.3 |
| --- | --- | --- | --- | --- |
| 15% | 82.1 | 99.4 | 86.7 | 99.5 |
| 10% | 74.3 | 99.6 | 80.5 | 99.7 |
| 5% | 57.8 | 99.8 | 66.1 | 99.9 |
Note: Sexually active adult and expectant mother populations in different geographic areas may produce different frequency distributions from the table above. Each laboratory should establish frequency distributions for their specific patient populations.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.