TRU-MR LARYNGOSCOPE SET

K062523 · Truphatek International, Ltd. · CCW · Nov 17, 2006 · Anesthesiology

Device Facts

Record IDK062523
Device NameTRU-MR LARYNGOSCOPE SET
ApplicantTruphatek International, Ltd.
Product CodeCCW · Anesthesiology
Decision DateNov 17, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5540
Device ClassClass 1
AttributesPediatric

Indications for Use

The Tru-MR ™ laryngoscope set is used to facilitate and aid in tracheal intubation in a Magnetic Resonance (MR) environment, not to exceed a 3.0 Tesla static magnetic field.

Device Story

Tru-MR™ is a rigid laryngoscope set consisting of handles and blades; designed for use in Magnetic Resonance (MR) environments up to 3.0 Tesla. Device facilitates tracheal intubation by providing visualization of the airway. Operated by clinicians in clinical settings where MR imaging is performed. Materials are specially treated to ensure MR compatibility; tested per ASTM F2052-02 to ensure safety in high-strength magnetic fields. Output is the physical visualization of the airway to assist the clinician in placing an endotracheal tube. Benefits include enabling airway management for patients requiring intubation while undergoing MR procedures.

Clinical Evidence

Bench testing only. Testing performed according to ASTM F2052-02 to verify MR compatibility in a 3.0 Tesla environment, demonstrating < 45-degree string deflection.

Technological Characteristics

Rigid laryngoscope set including handles and various blade configurations (e.g., Miller, Mac). Materials treated for MR compatibility. Tested per ASTM F2052-02 for magnetic field interaction in 3.0 Tesla environments. Cleaning via cold solution or autoclave.

Indications for Use

Indicated for tracheal intubation in patients of all ages within a Magnetic Resonance (MR) environment up to 3.0 Tesla.

Regulatory Classification

Identification

A rigid laryngoscope is a device used to examine and visualize a patient's upper airway and aid placement of a tracheal tube.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K062523 Premarket Notification 510(k) Section 5 – 510(k) Summary Truphatek Tru-MR™ ### 5 510(k) Summary | Non-Confidential Summary of Safety and Effectiveness | |------------------------------------------------------| | Page 1 of 2 | | | Page 1 of 2<br>25-Aug-06 | NOV 17 2006 | |----------------------------------------------------------------------------------------------|------------------------------------------------------|-------------| | Truphatck International Ltd.<br>Park Poleg-C, Sapir Industrial zone<br>Netanya, 42504 ISRAEL | Tel - 011-972-9-885-1155<br>Fax - 011-972-9-885-1212 | | | Official Contact: | David Grey, CEO | | | Proprietary or Trade Name: | Tru-MR™ Laryngoscope set - MR Conditional | | | Common/Usual Name: | Laryngoscope handle and blades - MR conditional | | | Classification Name: | Rigid Laryngoscope | | | Device: | Tru-MR™ | | | Predicate Devices: | Minrad MR tested laryngoscope - K041852 | | | Device Description: | | | The Truphatek Tru-MRTM MR conditional laryngoscope handle, battery, and blades are standard blades and handles that have been specially treated to be useable in a magnetic resonance (MR) environment. Testing has been performed according to ASTM F2052-02 in a 3.0 Tesla environment. There are many handle and blade configurations offered. | Indications for Use: | | |-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indicated Use -- | The Tru-MR ™ laryngoscope set is used to facilitate and<br>aid in tracheal intubation in a Magnetic Resonance (MR)<br>environment, not to exceed a 3.0 Tesla static magnetic<br>field. | | Patient Population -- | No limitations | | Environment of Use -- | Magnetic resonance (MR) environment up to 3.0 Tesla | {1}------------------------------------------------ ## Non-Confidential Summary of Safety and Effectiveness Page 2 of 2 25-Aug-06 | Device Attributes: | | |------------------------|----------------------------------------------| | Design | | | Configurations offered | Handles, many blades, i.e. Miller, Mac, etc. | | Cleaning | Cold solution, autoclave | | Testing | | | ASTM F2052-02<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>For Person Research Parcel of Children<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A . A | In 3.0 Tesla < 45 degree string deflection<br>while and the first and of the program and the comments of the comments of the first and the first and<br>Comments of Concession of Children Company of<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| # Differences Between Other Legally Marketed Predicate Devices The Tru-MR™ laryngoscope set viewed as substantially equivalent to the following predicate device -- Minrad -- K041852. There are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate devices. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, topped by a pair of wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Truphatek International, Limited C/O Mr. Paul Dryden President ProMedic, Incorporated 3460 Pointe Creck Court, #102 Bonita Springs, Florida 34134 NOV 1 7 2006 Re: K062523 Trade/Device Name: Tru MRTM Regulation Number: 21 CFR 868.5540 Regulation Name: Rigid Larvngoscope Regulatory Class: I Product Code: CCW Dated: August 25, 2006 Received: August 29, 2006 Dear Mr. Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Dryden Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Onete y.m., @lice Omd Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use Statement 4 Page 1 of 1 | 510(k) Number: | K 062523 (To be assigned) | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Tru MRTM | | Indications for Use: | The Tru-MR TM laryngoscope set is used to facilitate and<br>aid in tracheal intubation in a Magnetic Resonance (MR)<br>environment, not to exceed a 3.0 Tesla static magnetic<br>field. | Prescription Use XX (Part 21 CFR 801 Subpart D) 01 Over-the-counter use __ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Smite y. Michael ons ni Sign-Citi) on of Anesthesiology, General Hospital, don Control. Dental Number: K062523
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...