← Product Code [LYZ](/productcode/LYZ) · K062522

# POWDER-FREE VINYL PATIENT EXAMINATION GLOVES (K062522)

_Jiangsu Jaysun Glove Co., Ltd. · LYZ · Sep 8, 2006 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K062522

## Device Facts

- **Applicant:** Jiangsu Jaysun Glove Co., Ltd.
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** Sep 8, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Disposable vinyl patient examination glove; worn on examiner's hand or finger; acts as physical barrier to prevent cross-contamination between patient and examiner. Used in clinical settings by healthcare providers. Device is non-sterile, powder-free, and clear. Performance verified against ASTM D5250-00e4 standards for dimensions, physical properties, and powder residuals; biocompatibility confirmed via primary skin irritation and dermal sensitization testing.

## Clinical Evidence

No clinical data. Substantial equivalence is based on non-clinical bench testing, including physical property testing, water leak testing for pinholes, and biocompatibility assessments (primary skin irritation and dermal sensitization).

## Technological Characteristics

Vinyl patient examination glove; non-sterile; powder-free. Meets ASTM D5250-00e4 standards for dimensions and physical properties. Biocompatibility per ISO 10993-10. Pinhole testing per 21 CFR 800.20. Powder residual <2mg/glove per ASTM D6124-01.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves ([K032908](/device/K032908.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: K OG 252 2(applicant leave blank)

#### Premarket Notification |510(k)| Summary

[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

| Submitter's name :             | JIANGSU JAYSUN GLOVE CO., LTD                                                                              |
|--------------------------------|------------------------------------------------------------------------------------------------------------|
| Submitter's address :          | NO. 199 JIANLING AVENUE, SUQIAN ECONOMIC DEVELOPMENT<br>ZONE, SUQIAN CITY, JIANGSU PROVINCE, 223800, CHINA |
| Phone number :                 | (86)527-4568809                                                                                            |
| Fax number :                   | (86)527-4568806                                                                                            |
| Name of contact person:        | Ms. Sea XU                                                                                                 |
| Date the summary was prepared: | Aug. 20,2006                                                                                               |

## ((a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

| Device Name:            | Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) |  |
|-------------------------|-------------------------------------------------------------------|--|
| Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves                      |  |
|                         | Other clients private labeling                                    |  |
| Common Name:            | Patient examination glove                                         |  |
| Classification Name:    | Patient examination glove                                         |  |
| Device Classification:  | I                                                                 |  |
| Regulation Number:      | 21 CFR 880.6250                                                   |  |
| Panel:                  | General Hospital (80)                                             |  |
| Product Code:           | LYZ                                                               |  |

## ((a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* powder free vinyl patient examination gloves , Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-00-4

Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 .

[(a)(4)| A description of the device

: 10 : 10 : 10 : 10 : 10 : 1

{1}------------------------------------------------

Device Description : powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-00-4.

### [(a)(5)] The summary describes the intended use of the device

Device Intended Use: powder free vinyl patient examination glove, Clear(non-colored) is a Device intended Ester powate not wall purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

### [(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

| Characteristics       | Standard                                              | Device performance                                    |
|-----------------------|-------------------------------------------------------|-------------------------------------------------------|
| Dimension             | ASTM standard D 5250-00e4                             | Meets                                                 |
| Physical Properties   | ASTM standard D 5250-00e4                             | Meets                                                 |
| Freedom from pinholes | 21 CFR 800.20                                         | Meets                                                 |
| Powder Residual       | ASTM standard D 5250-00e4                             | Meets                                                 |
| Biocompatability      | and D6124-01<br>Primary Skin Irritation in<br>rabbits | <2mg/glove<br>Passes<br>Not a Primary Skin Irritation |
|                       | Dermal sensitization in the<br>guinea pig             | Passes<br>Not a Dermal sensitization                  |

# ((b)(1) A brief discussion of the nonclinical submitted, reference, or relied on in the ((0)(1) A brief diseason ission for a determination of substantial equivalence .

Powder free vinyl patient examination gloves , Clear(non-colored) meet requirements per ASTM D5250-0064, per ASTM D6124-01, per 21 CFR 800.20 and ISO10993-10.

# {(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket ((0)(2)) { }) { } }) { } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } }

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

#### ((b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a){3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims .

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is simple and recognizable, conveying the department's mission and purpose.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Jiangsu Javsun Glove Company, Limited C/O Ms. Chu Xiaoan Beijing Easy-Link Company Room 1606 Building 1, Jianxiang Yuan No. 209 Bei Si Huan Zhong Road Beijing, China 100083

SEP - 8 2006

Re: K062522

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-Colored) Regulation Number: 880.6250 Regulation Name: Patient Examination Regulatory Class: I Product Code: LYZ Dated: August 20, 2006 Received: August 28, 2006

Dear Ms. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{3}------------------------------------------------

#### Page 2 - Ms. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Devicc Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### INDICATIONS FOR USE

Applicant: JIANGSU JAYSUN GLOVE CO., LTD

510(k) Number (if known): *

Device Name:_Powder Free Vinyl Patient Examination Gloves, Clear(Non-colored)

Indications For Use:

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Shule H. Murphy, KS 9/8/16

nesthesiology, General Hospil.al, ontrol, Dental Devices

) Number: K012522

---

**Source:** [https://fda.innolitics.com/device/K062522](https://fda.innolitics.com/device/K062522)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).