NEUVIZ DUAL MULTI-SLICE CT SCANNER SYSTEM

K062451 · Neusoft Medical Systems Co., Ltd. · JAK · Sep 28, 2006 · Radiology

Device Facts

Record IDK062451
Device NameNEUVIZ DUAL MULTI-SLICE CT SCANNER SYSTEM
ApplicantNeusoft Medical Systems Co., Ltd.
Product CodeJAK · Radiology
Decision DateSep 28, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The NeuViz Dual systems are intended to produce cross-section images of head and whole body by computer reconstruction of X-ray transmission data taken at different angles.

Device Story

NeuViz Dual is a whole-body X-ray computed tomography (CT) scanner featuring a continuously rotating tube-detector system capable of simultaneous 2-slice acquisition. Available in two variants (H and L) with differing tube capacities. Device utilizes an on-board high-frequency high-voltage generator to produce X-radiation; solid-state detectors capture transmission data. System software provides interactive control for scanning, image reconstruction, and image archive/evaluation. Operated by clinical staff in a radiology setting. Produces 2D and 3D images for electronic storage or filming. Output assists clinicians in diagnostic evaluation of head and body anatomy.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Multi-slice CT scanner (2-slice capability); high-frequency high-voltage generator; solid-state detector array; continuously rotating gantry; 2D/3D image reconstruction software; complies with 21 CFR Subchapter J (Federal Diagnostic X-ray Equipment Standard).

Indications for Use

Indicated for head and whole body cross-sectional imaging via computer reconstruction of X-ray transmission data.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) ## Attachment 1 Summary of Safety and Effectiveness 24 Page 1 of 2 This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h) ## General Information: | Device Name: | NeuViz Dual Multi-slice CT Scanner System | |-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------| | Classification Name: | 21 CFR Part 892.1750<br>Computed Tomography X-ray System | | Model Name | NeuViz Dual (H); NeuViz Dual (L) | | Classification: | Class II | | Performance Standard: | 21 CFR Subchapter J, Federal Diagnostic X-ray Equipment<br>Standard | | Manufacture: | PHILIPS AND NEUSOFT MEDICAL SYSTEMS CO., LTD.<br>Neusoft Park, Hun Nan Industrial Area, Shenyang 110179,<br>P.R.China | | Distributor: | NEUSOFT MEDICAL SYSTEMS CO., LTD.<br>No.3-11,Wenhua Road, Heping District,<br>Shenyang, P.R.China<br>Post Code : 110004 | | Submitter: | Contact : Tianyanfang<br>Title : Manager of Q&R Department<br>Tel : 86-24-83660649<br>Fax : 86-24-83780480<br>E-Mail : Tianyanfang@neusoft.com | Summary prepared : September 10th, 2006 {1}------------------------------------------------ K062451 Safety and Effectiveness information Intended Uses: The NeuViz Dual systems are intended to produce cross-section images of head and whole body by computer reconstruction of X-ray transmission data taken at different angles. Device Description: The NeuViz Dual systems are whole body X-ray computed tomography scanner which features a continuously rotating tube-detector system and multi-slice capability of 2 slices simultaneously. There are two variants with different tube capacity: NeuViz Dual (H) and NeuViz Dual (L). The system software is an interactive program used for X-ray scan control, image reconstruction, and image archive/evaluation. Predicated Device: CT-C3000DUAL and CT-C2800DUAL Family of Dual-slice CT Scanner Systems (K041542) Statement of Substantial Equivalence: The NeuViz Dual systems are of comparable type and substantially equivalent to the CT-C300DUAL and CT-C2800DUAL Family of Dual-slice CT Scanner Systems (K041542) that comply with the same or equivalent standards and has the same intended uses. Both of these system use on-board high frequency High-Voltage generator to generate X-radiation from X-ray tube. The X-ray transmission data is detected by the solid-state detector and is reconstructed by the computer which has an interactive user interface. Both of these devices produce two dimensional image and 3D image that can be filmed or electronically stored for future review. Page 2 of 2 {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it, representing medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the caduceus symbol. Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 SEP 2 8 2006 Neusoft Medical Systems Co., Ltd. c/o Mr. Tamas Borsai TÜV Rheinland of North America 12 Commerce Road NEWTOWN-CT- 06470 Re: K062451 Trade/Device Name: NeuViz Dual Multi-slice CT Scanner System Regulation Number: 21 CFR §892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: September 12, 2006 Received: September 13, 2006 Dear Mr. Borsai: We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 70(x) promatice is substantially equivalent (for the indications for referenced above and have determined the devices marketed in interstate commence use stated in the enclosure) to legally marketed predicate and anyants and adouices that use stated in the encrosule/ to regally manuced processor in the Medical Devices that prior io May 26, 1970, the chacallent date of the Federal Food, Drug, and Cosmetic have been reclassified in accordance with the provisions of the Federal Food, MAA). You mov have been reclassified in accordance will and proven application (PMA). You may, Act (Act) that do not require approval of a premarket approval application (PMA). The parent therefore, market the device, subject to the general controls provisions of the general therefore, market the device, subject to the general vonn, and registration, list in the fulters on controls provisions of the riot morade requiritions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Prematke If your device is classince (see above) into entrols. Existing major regulations affecting your Approval), it may be subject to such adentonal vontons, Title 21, Parts 800 to 898. In addition, FDA device can be found in the Odde of I concerning your device in the Federal Register. Image /page/2/Picture/10 description: The image shows a circular logo with the text "1906 - 2006" at the top and "Centennial" at the bottom. The letters "FDA" are prominently displayed in the center of the logo. There are three stars below the word "Centennial". The logo is surrounded by a dotted circle. Protecting and Promoting Public Health {3}------------------------------------------------ Page 2 - Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow your o begin marketing your device in your-Scation 510(k) rms retter notification. The FDA finding of substantial equivalence of your device to a legally promated predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |----------------|----------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Attachment 2 Indications for Use Page 1 of 1 510(k) Number (if known): K062451 Device Name: NeuViz Dual Multi-slice CT Scanner System Indications for Use: The NeuViz Dual Multi-slice CT Scanner System -NeuViz Dual (H) and NeuViz Dual (L) - are intended to produce cross-section images of head and body by computer reconstruction of X-ray transmission data taken at different angles. | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|--| |--------------------------------------------------------|--| Prescription Use Nancy Brogdon (Division Sign-Off Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K062451
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%