CALIBRATOR FOR AUTOMATED SYSTEMS (C.F.A.S.), MODEL 10759350 360

K062319 · Roche Diagnostics Corp. · JIX · Sep 13, 2006 · Clinical Chemistry

Device Facts

Record IDK062319
Device NameCALIBRATOR FOR AUTOMATED SYSTEMS (C.F.A.S.), MODEL 10759350 360
ApplicantRoche Diagnostics Corp.
Product CodeJIX · Clinical Chemistry
Decision DateSep 13, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Intended Use

Calibrator for Automated Systems (C.f.a.s.) is for use in the calibration of quantitative Roche methods on Roche chemistry analyzers as specified in the enclosed value sheet.

Device Story

Modified calibrator for clinical chemistry assays; reagent composition updated via addition of pyridoxal-phosphate; stability profile after reconstitution altered; traceability updated for select analytes. Device used in clinical laboratory settings to calibrate diagnostic assays. Modification maintains fundamental scientific technology of previously cleared devices.

Clinical Evidence

Bench testing only; verification and validation activities performed per design control procedures to confirm impact of reagent and stability modifications.

Technological Characteristics

Lyophilized human serum-based calibrator containing chemical additives. Includes pyridoxal phosphate. Stability: 8 hours at 15-25°C, 2 days at 2-8°C, 4 weeks frozen at -15 to -25°C. Traceability established through standards or reference methods. Used on Roche clinical chemistry analyzers.

Indications for Use

Indicated for the calibration of quantitative Roche methods on Roche clinical chemistry analyzers. Intended for professional use in clinical laboratory settings.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SEP 1 3 2006 ## 510(k) Summary - Calibrator for Automated Systems (C.f.a.s) Introduction Roche Diagnostics Corporation hereby submits this Special 510(k): Device Modification to provide notification of modifications to our Calibrator for Automated Systems (C.f.a.s.). This calibrator was originally cleared for use as K990460 and modified via K033501. Modifications to the calibrator include: Modification to reagent composition: addition of pyridoxal phosphate . Change of stability claim after reconstitution . Modifications to value assignment procedure: including use of internal ● laboratory . Changes to traceability for select analytes Roche Diagnostics Submitter name, address, 9115 Hague Rd contact Indianapolis IN 46250 (317) 521-3723 Contact person: Corina Harper Date prepared: May 12, 2006 Device Name Proprietary name: Calibrator for Automated Systems (C.f.a.s.) Common name: Calibrator for Automated Systems (C.f.a.s.) Classification name: Calibrator, Multi-analyte mixture Device Calibrator for Automated Systems (C.f.a.s.) is for use in the calibration of Description quantitative Roche methods on Roche chemistry analyzers as specified in the enclosed value sheet. Calibrator for Automated Systems (C.f.a.s.) is a lyophilized calibrator based on human serum. Intended use Calibrator for Automated Systems (C.f.a.s.) is for use in the calibration of quantitative Roche methods on Roche chemistry analyzers as specified in the enclosed value sheet. {1}------------------------------------------------ | Predicate | We claim substantial equivalence to the Calibrator for Automated Systems | |-----------|--------------------------------------------------------------------------| | Device | (C.f.a.s.) cleared as K033501. | Substantial The table below indicates the similarities and differences between the equivalency modified Calibrator for Automated Systems (C.f.a.s.) and its predicate device Similarities/Diff (Calibrator for Automated Systems (C.f.a.s.), K033501). erences | Feature | Predicate device: C.f.a.s.<br>(K033501) | Modified device: C.f.a.s. | |-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use/<br>Indications for<br>Use | Calibrator for Automated Systems<br>(C.f.a.s.) is for use in the calibration<br>of quantitative Roche methods on<br>Roche clinical chemistry analyzers<br>as specified in the lot specific value<br>sheet. | Same | | Reagent<br>composition | Lyophilized human serum with<br>chemical additives and material of<br>biological origin | Same<br>Pyridoxal-phosphate addition | | Stability - shelf<br>life and on-board | 2-8 °C until expiration date<br>Reconstituted:<br>8 hours at 15 to 25° C<br>2 days at 2 to 8° C<br>8 hours at 15 to 25° C<br>2 weeks (frozen once) at (-15) to (-<br>25)° C | Same<br>Reconstituted:<br>8 hours at 15 to 25° C<br>2 days at 2 to 8° C<br>8 hours at 15 to 25° C<br>4 weeks (frozen once) at (-15) to (-<br>25)° C | {2}------------------------------------------------ | Traceability | Traceability through Master Lot<br>(ML) to standards or reference<br>methods. | Traceability through standards or<br>reference methods analyzed at all<br>set point validation runs. | |---------------------|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | Traceability changed for the<br>following analytes:<br>Direct Bilirubin<br>Calcium<br>Creatinine, Jaffe/RB/Compensated<br>Iron | | | | Magnesium, Xylidyl Blue<br>/Chlorophosphonazo<br>UIBC<br>Uric Acid | | Value<br>Assignment | Values to new lots are assigned by<br>running them as samples after<br>calibrating the system with the<br>Master Lot of C.f.a.s. | Values to new lots are assigned by<br>running them as samples after<br>calibrating the system with<br>previously assigned C.f.a.s lot.<br>Values are verified by using<br>reference material, Master lot<br>C.f.a.s. and previously assigned lots<br>of C.f.a.s. | Proposed Proposed labeling sufficient to describe the device, its intended use, and the Labeling directions for use can be found in Section V. We believe the proposed version of the device labeling presented contains all of the technical information required per 21 CFR 809.10. のアイデントです。 Validation and Development activities were conducted under appropriate design control Design Control procedures and the overall product specifications were met. The Declaration of Conformity with Design Controls and Results of Risk Analysis are provided in Section 5.1. Analytical Performance. #### Confidentiality Roche Diagnostics Corporation requests that the FDA not disclose the nature or existence of this submission until the substantial equivalence decision has been reached. {3}------------------------------------------------ Modification of the Calibrator for Automated Systems (C.f.a.s.) does not Closing affect the intended use or indications for use of the device as described in the labeling, nor does it alter the fundamental scientific technology of the device. Therefore, we trust the information provided in this Special 510(k) will support a decision of substantial equivalence of the Calibrator for Automated Systems (C.f.a.s.) to the predicate. > If you have any questions or require further information, please do not hesitate to contact this office. - · Phone: (317) 521-3831 - · FAX: (317) 521-2324 - · email: corina.harper@roche.com {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 SEP 1 3 2006 Ms. Corina Harper Roche Diagnostics Corporation 9115 Hague Road Indianapolis, IN 46250-0457 Re: k062319 Trade/Device Name: Calibrator for Automated System C.f.s.a. Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIX Dated: August 8, 2006 Received: August 9, 2006 Dear Ms. Harper: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {5}------------------------------------------------ ### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Alberto Garcia Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K062319 Calibrator for automated systems: C.f.a.s. Indications For Use: Calibrator for automated systems (C.f.a.s.) is for use in the calibration of quantitative Canorator for advinated bytwiss (chemistry analyzers as specified in the enclosed value sheet. Prescription Use XXX AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Page 1 of Carol Benson Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety K062319
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...