← Product Code [GGC](/productcode/GGC) · K062306

# PLASMACON N, PLASMACON L-1, PLASMACON L-2 (K062306)

_R2 Diagnostics, Inc. · GGC · Mar 16, 2007 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K062306

## Device Facts

- **Applicant:** R2 Diagnostics, Inc.
- **Product Code:** [GGC](/productcode/GGC.md)
- **Decision Date:** Mar 16, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5425
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

PlasmaCon N is a human lyophilized plasma control intended for use as a normal control with citrated plasma to monitor the performance of the Prothrombin Time (PT) and Activated Partial Thromboplastin Time (aPTT) tests. PlasmaCon L-1 is a human lyophilized plasma control intended for use as a mid-level abnormal control with citrated plasma to monitor the performance of the Prothrombin Time (PT) and Activated Partial Thromboplastin Time (aPTT) tests. PlasmaCon L-2 is a human lyophilized plasma control intended for use as a high level abnormal control with citrated plasma to monitor the performance of the Prothrombin Time (PT) and Activated Partial Thromboplastin Time (aPTT) tests.

## Device Story

Lyophilized citrated human plasma controls; used to verify system performance for PT and aPTT coagulation assays. Operated by qualified laboratory professionals in clinical settings; used with manual, semi-automated, or automated coagulation analyzers. Healthcare providers use control results to monitor assay performance and ensure accuracy of patient screening tests. Benefits include standardized quality control for coagulation testing.

## Clinical Evidence

Bench testing only. Precision/reproducibility evaluated per CLSI EP15-A guidelines; within-run and between-run CVs <15% across multiple instrument/reagent combinations (ACL3000+, MLA1000c, ACL Advance, Dade BCS, Stago STA). Stability validated via real-time open/closed vial studies.

## Technological Characteristics

Lyophilized citrated human plasma. Formulated for use as normal, mid-level abnormal, and high-level abnormal controls. Compatible with manual, semi-automated, and automated coagulation analyzers. No specific hardware or software components; device is a chemical control reagent.

## Regulatory Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

## Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

## Predicate Devices

- Trinity Biotech Normal Coagulation Control Plasma ([K895262](/device/K895262.md))
- Trinity Biotech Abnormal Coagulation Control Plasma Level-1 ([K895260](/device/K895260.md))
- Trinity Biotech Abnormal Coagulation Control Plasma Level-2 ([K895261](/device/K895261.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:
K062306

B. Purpose for Submission:
To obtain clearance for the PlasmaCon N, PlasmaCon L-1, and PlasmaCon L-2 coagulation control plasmas.

C. Measurand:
Prothrombin Time and Activated Partial Thromboplastin Time

D. Type of Test:
Assayed controls

E. Applicant:
R² Diagnostics, Inc.

F. Proprietary and Established Names:
PlasmaCon N, PlasmaCon L-1, and PlasmaCon L-2

G. Regulatory Information:
1. Regulation section:
21 CFR 864.5425
2. Classification:
Class II
3. Product code:
GIZ, GGC
4. Panel:

{1}

81 Hematology

H. Intended Use:

1. Intended use(s):

PlasmaCon N is a human lyophilized plasma control intended for use as a normal control with citrated plasma to monitor the performance of the Prothrombin Time (PT) and Activated Partial Thromboplastin Time (aPTT) tests.

PlasmaCon L-1 is a human lyophilized plasma control intended for use as a mid-level abnormal control with citrated plasma to monitor the performance of the Prothrombin Time (PT) and Activated Partial Thromboplastin Time (aPTT) tests.

PlasmaCon L-2 is a human lyophilized plasma control intended for use as a high level abnormal control with citrated plasma to monitor the performance of the Prothrombin Time (PT) and Activated Partial Thromboplastin Time (aPTT) tests.

2. Indication(s) for use:

Same

3. Special conditions for use statement(s):

4. Special instrument requirements:

I. Device Description:

The PlasmaCon Control Plasma devices contain lyophilized citrated human plasma for use in the verification of system performance for PT and aPTT assays. The PlasmaCon Control plasmas are packages in 10 vials each containing 1 ml of reagent. Controls plasmas should only be used in an appropriate clinical laboratory by qualified laboratory professionals. The tests may be performed manually or using semi-automated and automated coagulation analyzers.

PlasmaCon N is prepared from a pool of fresh citrated plasma from donors with normal coagulation parameters. The pooled plasma is buffered with Hepes and lyophilized to maximize stability of all plasma constituents.

PlasmaCon L-1 and PlasmaCon L-2 are prepared from a pool of fresh citrated plasma from donors with normal coagulation parameters. The pooled plasma is depleted of clotting factors II, VII, IX, and X. The control plasma is buffered with Hepes and lyophilized to maximize stability of all plasma constituents.

J. Substantial Equivalence Information:

{2}

3

1. Predicate device name(s):

a) Trinity Biotech Normal Coagulation Control Plasma
b) Trinity Biotech Abnormal Coagulation Control Plasma Level-1
c) Trinity Biotech Abnormal Coagulation Control Plasma Level-2

2. Predicate 510(k) number(s):

a) K895262
b) K895260
c) K895261

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | PlasmaCon N | Normal Coagulation Control Plasma  |
|  Intended Use | For use as a normal control with citrated plasma to monitor the performance of the PT and (aPTT) tests. | Use as a normal control in PT, aPTT, and fibrinogen determinations.  |
|  Matrix | Citrated human plasma | Citrated human plasma  |
|  Device compatibility | Manual, semi or fully automatic processing | Manual, semi or fully automatic processing  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | PlasmaCon N | Normal Coagulation Control Plasma  |
|  Stability (reconstituted) | Stable for 8 hours when stored at 2 – 8°C. | Stable for 8 hours stored at room temperature.  |
|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | PlasmaCon L-1 | Abnormal Coagulation Control Plasma Level 1  |
|  Intended Use | For use as an abnormal mid-level control with citrated plasma to monitor the performance of the PT and | Use as an abnormal control in PT, aPTT, and fibrinogen determinations.  |

{3}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | (aPTT) tests. |   |
|  Matrix | Citrated human plasma | Citrated human plasma  |
|  Device compatibility | Manual, semi- or fully automatic processing | Manual, semi- or fully automatic processing  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | PlasmaCon L-1 | Abnormal Coagulation Control Plasma Level 1  |
|  Stability (reconstituted) | Stable for 8 hours when stored at 2 – 8°C. | Stable for 8 hours stored at room temperature.  |
|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | PlasmaCon L-2 | Abnormal Coagulation Control Plasma Level 2  |
|  Intended Use | For use as an abnormal high control with citrated plasma to monitor the performance of the PT and (aPTT) tests. | Use as an abnormal control in PT, aPTT, and fibrinogen determinations.  |
|  Matrix | Citrated human plasma | Citrated human plasma  |
|  Device compatibility | Manual, semi- or fully automatic processing | Manual, semi- or fully automatic processing  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | PlasmaCon L-2 | Abnormal Coagulation Control Plasma Level 2  |
|  Stability (reconstituted) | Stable for 8 hours when stored at 2 – 8°C. | Stable for 8 hours stored at room temperature.  |

# K. Standard/Guidance Document Referenced (if applicable):

Points to Consider Guidance Document on Assayed and Unassayed Quality Control Material, February 3, 1999.

# L. Test Principle:

{4}

PlasmaCon Control reagents can be used in all testing in the same manner as may be citrated plasma sample. Tests may be performed manually or using semi-automated and automated coagulation analyzers.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

Data was collected for within run and between run precision using three levels of lyophilized controls according to procedures developed under CLSI EP15-A approved guidelines. Within run and between run precision studies obtained CV values of less than 15%.

For within run, 10 vials of each sample were pooled, tested in duplicate and recorded in duplicate or triplicate.

For between run, 2 vials of each sample were pooled, tested in duplicate and recorded each day for 5 days.

In comparison studies, normal and abnormal control plasma controls were tested using PT and aPTT reagents from multiple manufacturers on multiple instrument types to include:

- r² Diagnostics Phospholin ES on the ACL3000+, MLA1000c and ACL Advance
- r² Diagnostics Phosphoplastin RL on the ACL3000+, MLA1000c and ACL Advance
- Dade Behring Thromborel S on the Dade BCS
- Dade Behring Pathromtin SL on the Dade BCS
- Diagnostica Stago Neoplastine Cl+ on the Stago STA
- Diagnostica Stago Auto PTT on the Stago STA

The %CV ranges for all instrument and reagent combinations are:

|  Control | Within run PT | Between run PT | Within run APTT | Between run APTT  |
| --- | --- | --- | --- | --- |
|  PlasmaCon N | 0.5 – 1.05 | 0.4 – 2.5 | 0.51 – 4.1 | 0.7 – 2.9  |
|  PlasmaCon L-1 | 0.6 – 1.31 | 0.4 – 7.0 | 0.4 – 5.4 | 0.4 – 3.9  |
|  PlasmaCon L-2 | 0.44 – 5.3 | 0.8 – 13.5 | 0.4 – 8.0 | 0.2 – 4.0  |

#### b. Linearity/assay reportable range:

Not applicable.

{5}

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The value assignment protocol was determined by testing the PlasmaCon N reagent, the ACL3000+, and the Phosphoplastin RL (PT) reagent and Phospholin ES (APTT) reagent in the prothrombin time (PT) and activated partial thromboplastin time (APTT) assays. Ranges were determined based on a 97.5% confidence interval; thus 3SD ranges around the mean were calculated. The manufacturer recommends that each laboratory establish the mean values and expected control ranges for their particular laboratory's instrument-reagent system.

Open vial and closed vial real time stability tests were performed. Results are determined by pass/fail criteria.

Open vial stability studies were conducted PlasmaCon N, PlasmaCon L-1 and PlasmaCon L-2. Seven vials of test control plasma was stored at room temperature was run at the following intervals: 0, 4, 5, 6, 7, 8, and 9 hours (or until three consecutive failed tests) in a PT and APTT assay.

Closed vial stability studies were conducted PlasmaCon N, PlasmaCon L-1 and PlasmaCon L-2. One vial of test control plasma was stored at 2-8° C was run monthly for 3 months, once on months 6, 12, 18, 24, 30 and 36 (or until three consecutive failed tests) in a PT and APTT assay. Data was provided through month 7.

d. Detection limit:

Not applicable.

e. Analytical specificity:

Not applicable.

f. Assay cut-off:

Not applicable.

2. Comparison studies:

a. Method comparison with predicate device:

Not applicable.

b. Matrix comparison:

Not applicable.

{6}

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable.

b. Clinical specificity:

Not applicable.

c. Other clinical supportive data (when a. and b. are not applicable):

4. Clinical cut-off:

Not applicable.

5. Expected values/Reference range:

Expected values were determined PT and APTT using the Phosphoplastin RL reagent for PT and the Phospholin ES reagent for APTT. Results for the  $\mathbf{r}^2$  Diagnostics reagents on the ACL3000+ are:

|  Control | Expected Value  |   |
| --- | --- | --- |
|   |  PT | APTT  |
|  PlasmaCon N | 12.4 – 13.0 | 28.9 – 31.1  |
|  PlasmaCon L-1 | 17.7 – 18.9 | 43.3 – 45.7  |
|  PlasmaCon L-2 | 46.8 – 58.2 | 73.1 – 73.7  |

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K062306](https://fda.innolitics.com/device/K062306)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
