BIOPLEX 2200 EBV IGG PANEL FOR BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM

K062211 · Bio-Rad Laboratories, Inc. · LSE · Dec 8, 2006 · Microbiology

Device Facts

Record IDK062211
Device NameBIOPLEX 2200 EBV IGG PANEL FOR BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM
ApplicantBio-Rad Laboratories, Inc.
Product CodeLSE · Microbiology
Decision DateDec 8, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3235
Device ClassClass 1

Indications for Use

The BioPlex™ 2200 EBV IgG kit is a multiplex flow immunoassay intended for the qualitative detection of IgG antibodies to three (3) separate EBV antigens; Epstein-Barr Virus Nuclear Antigen-1 (EBV NA-1), Viral Capsid Antigen (EBV VCA), and Early Antigen diffuse (EBV EA-D) in human serum. The test system can be used in conjunction with the BioPlex 2200 EBV IgM kit as an aid in the laboratory diagnosis of infectious mononucleosis (IM). The EBV IgG kit is intended for use with the Bio-Rad BioPlex 2200 System. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens, or infants. Assay performance characteristics have not been established for the diagnosis of nasopharyngeal carcinoma, Burkitt’s lymphoma, and other EBV-associated lymphomas.

Device Story

Multiplex flow immunoassay system; uses dyed beads coated with recombinant EBV antigens (NA-1, VCA p18, EA-D). Patient sample, diluent, and bead reagent incubated in reaction vessel; anti-human IgG/phycoerythrin conjugate added. Instrument detects bead identity via dye fluorescence and antibody binding via phycoerythrin fluorescence. Includes Internal Standard, Serum Verification, and Reagent Blank beads for quality control. Used in clinical laboratories by technicians. Output is relative fluorescence intensity (RFI) converted to antibody index (AI). Results aid clinicians in determining EBV serological status (e.g., primary acute, past infection) when used with IgM kit; supports infectious mononucleosis diagnosis.

Clinical Evidence

Bench-only study. Evaluated 621 banked serum samples from patients with ordered EBV tests across 3 U.S. sites. Compared BioPlex 2200 EBV IgG results against commercially available microplate EIAs. Overall percent agreement for NA-1 IgG was 95.6% positive/97.4% negative; VCA IgG was 96.2% positive/99.4% negative; EA-D IgG was 88.1% positive/84.1% negative. Precision and reproducibility studies performed per CLSI EP5-A2.

Technological Characteristics

Multiplex flow immunoassay; dyed beads coated with E. coli-derived recombinant EBV proteins. Conjugate: murine monoclonal anti-human IgG/phycoerythrin. Buffer: MOPS, triethanolamine, PBS. Preservatives: ProClin 300, sodium azide. Instrument: BioPlex 2200 System. Connectivity: Integrated system. Software: Automated data processing and RFI calculation.

Indications for Use

Indicated for qualitative detection of IgG antibodies to EBV NA-1, EBV VCA, and EBV EA-D in human serum to aid in laboratory diagnosis of infectious mononucleosis. Not for use in immunocompromised/immunosuppressed patients, neonates, infants, or diagnosis of EBV-associated lymphomas/carcinomas.

Regulatory Classification

Identification

Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k062211 B. Purpose for Submission: This is an original bundled application for a new multiplexed flow immunoassay for the qualitative detection of IgG antibodies of 3 separate analytes; Epstein-Barr Virus Nuclear Antigen (EBV NA-1), Viral Capsid Antigen (EBV VCA), and Early Antigen diffuse (EBV EA-D) in human serum to aid in the diagnosis of infectious mononucleosis. C. Measurand: Epstein-Barr Virus Nuclear Antigen (EBV NA-1), Viral Capsid Antigen (EBV VCA), and Early Antigen diffuse (EBV EA-D) D. Type of Test: Multiplexed flow immunoassay E. Applicant: BIO-RAD LABORATORIES, INC. F. Proprietary and Established Names: BioPlex 2200 EBV IgG Panel on the BioPlex 2200 Multi-Analyte Detection System BioPlex 2200 EBV IgG Control Set BioPlex 2200 EBV IgG Calibrator Set G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | LSE | I | 866.3235 - EPSTEIN-BARR VIRUS SEROLOGICAL REAGENTS | Microbiology (83) | {1} 2 H. Intended Use: 1. Intended use(s): The BioPlex™ 2200 EBV IgG kit is a multiplex flow immunoassay intended for the qualitative detection of IgG antibodies to three (3) separate EBV antigens; Epstein-Barr Virus Nuclear Antigen-1 (EBV NA-1), Viral Capsid Antigen (EBV VCA), and Early Antigen diffuse (EBV EA-D) in human serum. The test system can be used in conjunction with the BioPlex 2200 EBV IgM kit as an aid in the laboratory diagnosis of infectious mononucleosis (IM). The EBV IgG kit is intended for use with the Bio-Rad BioPlex 2200 System. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens, or infants. Assay performance characteristics have not been established for the diagnosis of nasopharyngeal carcinoma, Burkitt’s lymphoma, and other EBV-associated lymphomas. 2. Indication(s) for use: See above 3. Special conditions for use statement(s): Prescription use only 4. Special instrument requirements: The BioPlex 2200 EBV IgG kit is intended for use with the BioPlex 2200 System instrument and software. I. Device Description: The EBV IgG kit is a multiplex flow immunoassay for the qualitative detection of IgG antibodies to three (3) separate EBV antigens; Epstein-Barr Virus Nuclear Antigen-1 (EBV NA-1), Viral Capsid Antigen (EBV VCA), and Early Antigen diffuse (EBV EA-D) in human serum. See test principle below for more details. J. Substantial Equivalence Information: {2} 3 Predicate Device 1: EBNA IGG EIA TEST SYSTEM by Immuno Probe, Inc. 510k number: K951549 | Component | Similarities | | | --- | --- | --- | | | Device | Predicate | | Measurand | EBNA-1 IgG | EBNA-1 IgG | | Matrices | Serum | Serum | | Intended Use | Aid in diagnosis of infectious mononucleosis | Aid in diagnosis of infectious mononucleosis | | | Differences | | | | Device | Predicate | | Technology | Multiplexed flow immunoassay | Traditional ELISA | | Intended Use | Qualitative detection | Semi-quantitative detection | Predicate Device 2: EBV VCA IGG EIA TEST SYSTEM by Immuno Probe, Inc. 510k number: K980912 | Component | Similarities | | | --- | --- | --- | | | Device | Predicate | | Measurand | EBV VCA IgG | EBV VCA IgG | | Matrices | Serum | Serum | | Intended Use | Qualitative detection of EBV VCA IgG to aid in diagnosis of infectious mononucleosis | Qualitative detection of EBV VCA IgG to aid in diagnosis of infectious mononucleosis | | | Differences | | | | Device | Predicate | | Technology | Multiplexed flow immunoassay | Traditional ELISA | Predicate Device 3: EBV EA-D IGG EIA TEST SYSTEM by Immuno Probe, Inc. 510k number: K973123 | Component | Similarities | | | --- | --- | --- | | | Device | Predicate | | Measurand | EBV EA-D IgG | EBV EA-D IgG | | Matrices | Serum | Serum | | Intended Use | Qualitative detection of EBV EA-D IgG to aid in diagnosis of infectious mononucleosis | Qualitative detection of EBV EA-D IgG to aid in diagnosis of infectious mononucleosis | | | Differences | | | | Device | Predicate | {3} | Technology | Multiplexed flow immunoassay | Traditional ELISA | | --- | --- | --- | K. Standard/Guidance Document Referenced (if applicable): | STANDARDS | | --- | | Title and Reference Number | | Interference Testing in Clinical Chemistry; Approved Guideline (EP 7-A) | | Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition (EP5-A2) | | Other Standards | | | | | --- | --- | --- | --- | | GUIDANCE | | | | | Document Title | Office | Division | Web Page | | Review Criteria for In Vitro Diagnostic Devices for Detection of IGM Antibodies to Viral Agents | OIVD | DIHD | http://www.fda.gov/cdrh/ode/527.pdf | | Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable - Guidance for Sponsors, Institutional Review Boards, Clinical Investigators and FDA Staff | CBER/OIVD | | http://www.fda.gov/cdrh/oivd/guidance/1588.html | L. Test Principle: The EBV IgG kit uses multiplex flow immunoassay, a methodology that greatly resembles traditional EIA, but permits simultaneous detection and identification of many antibodies in a single tube. Three (3) different populations of beads are coated with $E.$ coli derived recombinant proteins, EBV NA-1 (28kD and 45kD), EBV VCA p18 (40kD), and EBV EA-D (28kD) associated with infectious mononucleosis. The BioPlex 2200 System combines an aliquot of patient sample, sample diluent, and bead reagent into a reaction vessel. The mixture is incubated at $37^{\circ}\mathrm{C}$ . After a wash cycle, antihuman IgG antibody, conjugated to phycoerythrin (PE), is added to the dyed beads and this mixture is incubated at $37^{\circ}\mathrm{C}$ . The excess conjugate is removed in another wash cycle, and the beads are re-suspended in wash buffer. The bead mixture then passes through the detector. The identity of the dyed beads is determined by the fluorescence of the dyes, and the amount of antibody captured by the antigen is determined by the fluorescence of the attached PE. Raw data is calculated in relative fluorescence intensity (RFI). {4} Three additional dyed beads, an Internal Standard Bead (ISB), a Serum Verification Bead (SVB) and a Reagent Blank Bead (RBB) are present in each reaction mixture to verify detector response, the addition of serum to the reaction vessel and the absence of significant non-specific binding in serum. Refer to the BioPlex 2200 System Operation Manual for more information. The instrument is calibrated using a set of seven (7) distinct calibrator vials, supplied separately by Bio-Rad Laboratories. A combination of four (4) vials representing four (4) different antibody concentrations are used for calibration. The result for each of these antibodies is expressed as an antibody index (AI). ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: **Precision Studies:** A precision panel, consisting of six (6) panel members was prepared by Bio-Rad Laboratories. Two (2) of the six (6) panel members had high levels of the antibodies contained in the BioPlex 2200 EBV IgG kit (EBV NA-1 IgG, EBV VCA IgG, and EBV EA-D IgG) and two (2) of the six (6) panel members had antibody levels near the cutoff, both prepared from positive patient samples. Two (2) of the six (6) panel members were negative (one high negative and one low negative) for both of the analytes. Precision testing was performed at Bio-Rad Laboratories on one lot of the EBV IgG kit, one lot of the EBV IgG Calibrator Set and one lot of the EBV IgG Control Set. Each of the six (6) panel members was tested in duplicate (x2) on two (2) runs per day for ten (10) days using one (1) lot of EBV IgG kit, one (1) lot of EBV IgG Calibrator Set and one (1) lot of EBV IgG Control Set (2 times x 2 runs x 10 days = 40 replicates per panel member). The data were analyzed for intra-assay and inter-assay precision according to the principles described in the Clinical Laboratory Standards Institute guidance EP5-A2, revised November 2004 and ISO/TR 22971:2205. The standard deviation (SD) and percent coefficient of variation (%CV) were calculated. Results can be found in Tables S - U. 5 {5} Table S. Precision Results; BioPlex 2200 EBV NA-1 IgG | EBV NA-1 IgG Panel Members | Sample N* | Al Mean | Within-Run | | Between-Day | | Between-Run | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | High Positive 1 | 42 | 4.6 | 0.2 | 4.2% | 0.3 | 6.3% | 0.3 | 5.9% | 0.4 | 9.6% | | High Positive 2 | 42 | 4.6 | 0.2 | 4.0% | 0.2 | 5.4% | 0.4 | 8.6% | 0.5 | 10.9% | | Low Positive 1 | 42 | 1.9 | 0.1 | 5.5% | 0.0 | 0.0% | 0.2 | 11.9% | 0.2 | 13.1% | | Low Positive 2 | 42 | 2.2 | 0.1 | 5.0% | 0.0 | 0.0% | 0.2 | 10.4% | 0.3 | 11.5% | | High Negative | 43 | 0.7 | 0.1 | 7.1% | 0.0 | 3.8% | 0.1 | 12.1% | 0.1 | 14.5% | | Low Negative | 44 | 0.0 | 0.0 | 0.0% | 0.0 | 0.0% | 0.0 | 0.0% | 0.0 | 0.0% | *Additional samples were run. Table T. Precision Results; BioPlex 2200 EBV VCA IgG | EBV VCA IgG Panel Members | Sample N* | Al Mean | Within-Run | | Between-Day | | Between-Run | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | High Positive 1 | 42 | 3.5 | 0.1 | 3.9% | 0.1 | 3.9% | 0.2 | 6.8% | 0.3 | 8.7% | | High Positive 2 | 42 | 3.4 | 0.2 | 5.0% | 0.1 | 3.7% | 0.3 | 9.1% | 0.4 | 11.0% | | Low Positive 1 | 42 | 1.6 | 0.1 | 8.4% | 0.0 | 2.6% | 0.1 | 8.1% | 0.2 | 11.9% | | Low Positive 2 | 42 | 1.3 | 0.1 | 7.4% | 0.1 | 4.1% | 0.1 | 10.2% | 0.2 | 13.3% | | High Negative | 43 | 0.6 | 0.1 | 10.8% | 0.0 | 0.0% | 0.1 | 11.4% | 0.1 | 15.7% | | Low Negative | 44 | 0.2 | 0.0 | 16.3% | 0.0 | 6.8% | 0.0 | 5.1% | 0.0 | 18.4% | *Additional samples were run. Table U. Precision Results; BioPlex 2200 EBV EA-D IgG | EBV EA-D IgG Panel Members | Sample N* | Al Mean | Within-Run | | Between-Day | | Between-Run | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | High Positive 1 | 42 | 4.3 | 0.2 | 4.6% | 0.0 | 0.0% | 0.3 | 6.9% | 0.4 | 8.3% | | High Positive 2 | 42 | 4.2 | 0.3 | 6.0% | 0.3 | 6.3% | 0.3 | 7.2% | 0.5 | 11.2% | | Low Positive 1 | 42 | 2.3 | 0.2 | 10.3% | 0.1 | 4.5% | 0.2 | 9.8% | 0.3 | 14.9% | | Low Positive 2 | 42 | 2.3 | 0.2 | 7.0% | 0.2 | 6.8% | 0.2 | 9.1% | 0.3 | 13.4% | | High Negative | 43 | 0.7 | 0.1 | 13.4% | 0.0 | 0.0% | 0.1 | 12.9% | 0.1 | 18.6% | | Low Negative | 44 | 0.2 | 0.1 | 29.8% | 0.0 | 5.3% | 0.0 | 0.0% | 0.1 | 30.3% | *Additional samples were run. ## Reproducibility Studies: A reproducibility panel, consisting of nine (9) panel members was prepared by Bio-Rad Laboratories. Two (2) of the nine (9) panel members had high levels of EBV NA-1 and EBV VCA; two (2) of the nine (9) panel members had high levels {6} of EBV EA-D; two (2) of the nine (9) panel members had antibody levels near the cutoff for EBV NA-1 and EBV VCA; two (2) of the nine (9) panel members had antibody levels near the cutoff for EBV EA-D. All were prepared from positive patient samples. One (1) of the nine (9) panel members was negative for all three (3) analytes contained in the BioPlex 2200 EBV IgG kit. In addition, three (3) lots of the EBV IgG Control Set [1 positive control (antibody positive) and a negative control (antibody negative)] were also tested. Reproducibility testing was performed at each of three (3) US testing facilities on a total of three (3) lots of the EBV IgG kit, three (3) lots of the EBV IgG Calibrator Set and three (3) lots of the EBV IgG Control Set. Each testing facility evaluated reproducibility using one (1) kit lot of EBV IgG with matched calibrators and controls. The panels were provided to each of the testing sites. Each of the nine (9) panel members and positive and negative controls was tested in quadruple (x4) on each day for three (3) days at each of three (3) US testing facilities using one (1) lot of EBV IgG reagent pack, one (1) lot of EBV IgG Calibrator Set and one (1) lot of EBV IgG Control Set (4 times x 3 days x 3 sites = 36 replicates per panel member and controls). The data were analyzed for intra-assay and inter-assay reproducibility according to the principles described in the Clinical Laboratory Standards Institute guidance EP5-A2, revised November 2004 and ISO/TR 22971:2205. The standard deviation (SD) and percent coefficient of variation (%CV) were calculated. Positive results can be found in Tables P-R. 7 {7} Table P. Reproducibility; BioPlex 2200 EBV NA-1 IgG | EBV NA-1 IgG Panel Members | Sample N | Grand Mean AI | Within-Run | | Between-Day | | Between-Run | | Between-Site* | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | High Positive 1 | 36 | 4.2 | 0.1 | 3.2% | 0.0 | 0.0% | 0.1 | 2.9% | 0.3 | 7.4% | 0.4 | 8.5% | | High Positive 2 | 36 | 4.3 | 0.1 | 3.4% | 0.1 | 2.5% | 0.1 | 2.8% | 0.3 | 7.5% | 0.4 | 9.0% | | Low Positive 1 | 36 | 1.5 | 0.1 | 4.8% | 0.0 | 0.0% | 0.1 | 5.7% | 0.2 | 11.6% | 0.2 | 13.8% | | Low Positive 2 | 36 | 2.1 | 0.1 | 3.2% | 0.0 | 1.1% | 0.1 | 3.7% | 0.2 | 9.4% | 0.2 | 10.7% | | Positive Control | 36 | 2.9 | 0.1 | 1.9% | 0.0 | 1.1% | 0.1 | 2.4% | 0.5 | 17.2% | 0.5 | 17.5% | *Between site variance includes between lot variance. Table Q. Reproducibility; BioPlex 2200 EBV VCA IgG | EBV VCA IgG Panel Members | Sample N | Grand Mean AI | Within-Run | | Between-Day | | Between-Run | | Between-Site* | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | High Positive 1 | 36 | 3.3 | 0.1 | 3.5% | 0.0 | 0.0% | 0.1 | 3.3% | 0.4 | 12.7% | 0.4 | 13.5% | | High Positive 2 | 36 | 3.2 | 0.1 | 3.4% | 0.0 | 0.0% | 0.1 | 3.7% | 0.2 | 7.3% | 0.3 | 8.9% | | Low Positive 1 | 36 | 1.5 | 0.1 | 5.0% | 0.0 | 0.0% | 0.1 | 5.5% | 0.2 | 16.2% | 0.3 | 17.8% | | Low Positive 2 | 36 | 1.3 | 0.1 | 5.8% | 0.0 | 0.0% | 0.1 | 5.0% | 0.1 | 7.6% | 0.1 | 10.8% | | Positive Control | 36 | 2.3 | 0.1 | 2.8% | 0.0 | 0.0% | 0.1 | 2.7% | 0.1 | 5.6% | 0.2 | 6.9% | *Between site variance includes between lot variance. Table R. Reproducibility; BioPlex 2200 EBV EA-D IgG | EBV EA-D IgG Panel Members | Sample N | Grand Mean AI | Within-Run | | Between-Day | | Between-Run | | Between-Site* | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | High Positive 1 | 36 | 4.1 | 0.2 | 4.4% | 0.0 | 0.8% | 0.0 | 0.0% | 0.4 | 8.7% | 0.4 | 9.8% | | High Positive 2 | 36 | 4.0 | 0.1 | 3.7% | 0.0 | 0.7% | 0.1 | 3.5% | 0.2 | 6.2% | 0.3 | 8.1% | | Low Positive 1 | 36 | 2.3 | 0.2 | 8.8% | 0.1 | 2.8% | 0.0 | 0.0% | 0.1 | 5.7% | 0.2 | 10.9% | | Low Positive 2 | 36 | 2.2 | 0.1 | 4.6% | 0.0 | 0.0% | 0.1 | 3.5% | 0.1 | 4.2% | 0.2 | 7.2% | | Positive Control | 36 | 3.0 | 0.1 | 3.1% | 0.0 | 0.0% | 0.1 | 2.5% | 0.6 | 18.8% | 0.6 | 19.2% | *Between site variance includes between lot variance. b. Linearity/assay reportable range: Not Applicable. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Not Applicable. d. Detection limit: Not applicable. {8} # e. Analytical specificity: A cross-reactivity study was performed to determine if samples from various disease states and other potentially interfering factors interfere with test results when tested with the BioPlex 2200 EBV IgG kit. A panel of ten (10) specimens* positive for each cross reactant were evaluated for possible cross reactivity with the BioPlex 2200 EBV IgG kit for each of the three EBV IgG antibody assays. Due to the high prevalence of EBV IgG antibodies in the normal population, the test specimens were also evaluated on corresponding commercially available microplate EIAs. Most of the samples evaluated were high positive for each disease state. The majority of all samples that elicited a positive result were also confirmed positive by the corresponding commercially available microplate EIA, indicating reactivity to EBV IgG antibodies rather than cross reactivity with a potentially interfering factor. Results can be found in Table S. *Due to limited availability of samples, only four E. coli specimens were evaluated. Table S. Cross-Reactivity | Cross Reactives | N | Method | BioPlex 2200 EBV IgG | | | | --- | --- | --- | --- | --- | --- | | | | | EBV NA-1 IgG | EBV VCA IgG | EBV EA-D IgG | | ANA | 10 | BioPlex 2200 | 9 | 10 | 7 | | | | EIA | 9 | 10 | 7 | | | | Discrepants | 0 | 0 | 0 | | Rheumatoid Factor | 10 | BioPlex 2200 | 10 | 10 | 1 | | | | EIA | 10 | 10 | 1+ | | | | Discrepants | 0 | 0 | 0 | | Toxo IgG | 10 | BioPlex 2200 | 9 | 9 | 2 | | | | EIA | 9 | 9+ | 2+ | | | | Discrepants | 0 | 0 | 0 | | Rubella IgG | 10 | BioPlex 2200 | 10 | 10 | 2 | | | | EIA | 10 | 10 | 1+ | | | | Discrepants | 0 | 0 | 1 | | CMV IgG | 10 | BioPlex 2200 | 10 | 10 | 2 | | | | EIA | 10 | 10 | 2** | | | | Discrepants | 0 | 0 | 1 | | VZV IgG | 10 | BioPlex 2200 | 8 | 8 | 1 | | | | EIA | 9 | 8 | 1 | | | | Discrepants | 1 | 0 | 0 | | HSV-1 IgG | 10 | BioPlex 2200 | 10 | 10 | 2 | | | | EIA | 10 | 10 | 3 | | | | Discrepants | 0 | 0 | 1 | | HSV-2 IgG | 10 | BioPlex 2200 | 10 | 10 | 3 | | | | EIA | 10 | 10 | 4 | | | | Discrepants | 0 | 0 | 1 | | HIV | 10 | BioPlex 2200 | 10 | 9 | 1 | | | | EIA | 10 | 10 | 2** | | | | Discrepants | 0 | 1 | 1 | | E. coli | 4 | BioPlex 2200 | 4 | 4 | 0 | | | | EIA | 4 | 4 | 0+ | | | | Discrepants | 0 | 0 | 0 | | Pregnant women | 10 | BioPlex 2200 | 9 | 9 | 3 | | | | EIA | 9 | 10 | 3+ | | | | Discrepants | 0 | 1 | 0 | *One Equivocal Sample; **Two Equivocal Samples {9} Interfering Substances: Testing for interfering substances was conducted according to CLSI Protocol EP7-A (Vol. 22, No. 27). No significant interference was observed in any of the substances tested. The following substances, listed in Table T, were tested $(N = 10)$ at maximum levels on one reagent lot. Table T. Interfering Substances | Substance | Concentration | | --- | --- | | Hemoglobin | ≤ 500 mg/dL | | Bilirubin (unconjugated) | ≤ 20 mg/dL | | Bilirubin (conjugated) | ≤ 20 mg/dL | | Triglycerides | ≤ 3000 mg/dL | | Protein (total) | ≤ 12 g/dL | | Cholesterol | ≤ 500 mg/dL | | Red Blood Cells | ≤ 0.4% Concentrate | | Gamma-globulin | ≤ 2.5 g/dL | | Ascorbic Acid | ≤ 3 mg/dL | # f. Assay cut-off: A final cut-off of 1 AI was established for all BioPlex 2200 EBV IgG kit assays based on an evaluation of 808 serum samples with the BioPlex 2200 EBV IgG kit assays and corresponding commercially available microplate EIA tests. ROC analysis was performed for each BioPlex 2200 EBV IgG assay using this population of samples. Samples that were equivocal on the commercially available microplate EIA tests were excluded from the analysis. This analysis was used to optimize sensitivity and specificity for the BioPlex 2200 EBV IgG kit assays. A final cut-off of 1 AI was established for all BioPlex 2200 EBV IgG kit assays based on the data collected. # 2. Comparison studies: # a. Method comparison with predicate device: Performance of the BioPlex 2200 EBV IgG kit was tested against corresponding commercially available microplate EIAs. A total of 621 banked serum samples from patients for which an EBV test were ordered were tested at 3 U.S. clinical testing sites. The BioPlex 2200 EBV IgG kit was run in conjunction with the BioPlex 2200 EBV IgM kit to allow for a complete antibody response profile. The characterization by antibody response was not compared with clinical data regarding presence, absence or status of disease. Two (2) samples were excluded due to RBB analysis error messages during BioPlex 2200 EBV IgM testing. One (1) sample was excluded due to RBB {10} analysis error messages during BioPlex 2200 EBV IgG testing. Using Table A as a guideline, results were analyzed by BioPlex 2200 EBV IgG analytes and corresponding EBV IgG reference assays according to serological characterization based on reference assay results. For the purpose of percent agreement calculations, BioPlex 2200 EBV IgG equivocal results were assigned to the opposite clinical interpretation than that of the corresponding reference assay result. Likewise, the reference IgG assay equivocal results were assigned to the opposite clinical interpretation than that of the corresponding BioPlex 2200 EBV IgG result. Results from all sites are shown and summarized in Tables H-M. Table H. BioPlex 2200 EBV NA-1 IgG vs. EIA: Comparison by Serological Pattern Characterization | EBV Serological Status | Reference EBV NA-1 IgG Interpretation | | | | | | | | | Total | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Positive | | | Equivocal | | | Negative | | | | | | BioPlex 2200 EBV NA-1 IgG | | | BioPlex 2200 EBV NA-1 IgG | | | BioPlex 2200 EBV NA-1 IgG | | | | | | Pos | Ind | Neg | Pos | Ind | Neg | Pos | Ind | Neg | | | | N | N | N | N | N | N | N | N | N | N | | Primary Acute | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 31 | 31 | | Late Acute | 104 | 0 | 4 | 0 | 0 | 0 | 0 | 0 | 2 | 110 | | Recovering | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 4 | 4 | | Previous Infection | 285 | 1 | 4 | 0 | 0 | 0 | 4 | 0 | 11 | 305 | | Susceptible | 0 | 0 | 0 | 0 | 0 | 1 | 1 | 0 | 125 | 127 | | Inconclusive | 20 | 0 | 9 | 0 | 0 | 0 | 0 | 0 | 12 | 41 | | Overall | 409 | 1 | 17 | 0 | 0 | 1 | 5 | 0 | 185 | 618 | Table I. BioPlex 2200 EBV NA-1 IgG vs. EIA: Percent Agreement &amp; Confidence Intervals by Serological Pattern Characterization | EBV Serological Status | Positive Agreement | | 95% CI | Negative Agreement | | 95% CI | | --- | --- | --- | --- | --- | --- | --- | | Primary Acute | (0/0) | N/A* | N/A* | (31/31) | 100% | 89.0 -100% | | Late Acute | (104/108) | 96.3% | 90.9 - 98.6% | (2/2) | 100% | 34.2 - 100% | | Recovering | (0/0) | N/A* | N/A* | (4/4) | 100% | 51.0 -100% | | Previous Infection | (285/290) | 98.3% | 96.0 - 99.3% | (11/15) | 73.3% | 48.0 - 98.1% | | Susceptible | (0/1) | 0.0% | N/A* | (125/126) | 99.2% | 95.6 - 99.9% | | Inconclusive | (20/29) | 69.0% | 50.8 - 82.7% | (12/12) | 100% | 75.8 - 100% | | Overall | (409/428) | 95.6% | 93.2 - 97.1% | (185/190) | 97.4% | 94.0 - 98.9% | *In cases where agreement resulted in a numerator of zero (0), 95% confidence interval could not be calculated; in cases where agreement resulted in (0/0) samples, percent agreement and 95% confidence interval could not be calculated. {11} Table J. BioPlex 2200 EBV VCA IgG vs. EIA: Comparison by Serological Pattern Characterization | EBV Serological Status | Reference EBV VCA IgG Interpretation | | | | | | | | | Total | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Positive | | | Equivocal | | | Negative | | | | | | BioPlex 2200 EBV VCA IgG | | | BioPlex 2200 EBV VCA IgG | | | BioPlex 2200 EBV VCA IgG | | | | | | Pos | Ind | Neg | Pos | Ind | Neg | Pos | Ind | Neg | | | | N | N | N | N | N | N | N | N | N | N | | Primary Acute | 4 | 0 | 3 | 0 | 0 | 1 | 0 | 0 | 23 | 31 | | Late Acute | 106 | 0 | 4 | 0 | 0 | 0 | 0 | 0 | 0 | 110 | | Recovering | 4 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 4 | | Previous Infection | 296 | 1 | 8 | 0 | 0 | 0 | 0 | 0 | 0 | 305 | | Susceptible | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 127 | 127 | | Inconclusive | 16 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 24 | 41 | | Overall | 426 | 1 | 15 | 0 | 0 | 1 | 1 | 0 | 174 | 618 | Table K. BioPlex 2200 EBV VCA IgG vs. EIA: Percent Agreement &amp; Confidence Intervals by Serological Pattern Characterization | EBV Serological Status | Positive Agreement | | 95% CI | Negative Agreement | | 95% CI | | --- | --- | --- | --- | --- | --- | --- | | Primary Acute | (4/8) | 50.0% | 21.5 - 78.5% | (23/23) | 100% | 85.7 - 100% | | Late Acute | (106/110) | 96.4% | 91.0 - 98.6% | (0/0) | N/A* | NA* | | Recovering | (4/4) | 100% | 51.0 - 100% | (0/0) | N/A* | NA* | | Previous Infection | (296/305) | 97.0% | 94.5 - 98.4% | (0/0) | N/A* | NA* | | Susceptible | (0/0) | N/A* | NA* | (127/127) | 100% | 97.1 - 100% | | Inconclusive | (16/16) | 100% | 80.6 - 100% | (24/25) | 96.0% | 80.5 - 99.3% | | Overall | (426/443) | 96.2% | 93.9 - 97.6% | (174/175) | 99.4% | 96.8 - 99.9% | *In cases where agreement resulted in (0/0) samples, percent agreement and 95% confidence interval could not be calculated. {12} Table L. BioPlex 2200 EBV EA-D IgG vs. EIA: Comparison by Serological Pattern Characterization | EBV Serological Status | Reference EBV EA-D IgG Interpretation | | | | | | | | | Total | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Positive | | | Equivocal | | | Negative | | | | | | BioPlex 2200 EBV EA-D IgG | | | BioPlex 2200 EBV EA-D IgG | | | BioPlex 2200 EBV EA-D IgG | | | | | | Pos | Ind | Neg | Pos | Ind | Neg | Pos | Ind | Neg | | | | N | N | N | N | N | N | N | N | N | N | | Primary Acute | 18 | 1 | 2 | 0 | 0 | 0 | 2 | 1 | 7 | 31 | | Late Acute | 72 | 1 | 3 | 4 | 0 | 0 | 6 | 3 | 21 | 110 | | Recovering | 4 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 4 | | Previous Infection | 0 | 0 | 0 | 5 | 3 | 3 | 23 | 22 | 249 | 305 | | Susceptible | 0 | 0 | 0 | 0 | 1 | 1 | 0 | 9 | 116 | 127 | | Inconclusive | 10 | 2 | 1 | 0 | 0 | 0 | 2 | 2 | 24 | 41 | | Overall | 104 | 4 | 6 | 9 | 4 | 4 | 33 | 37 | 417 | 618 | Table M. BioPlex 2200 EBV EA-D IgG vs. EIA: Percent Agreement &amp; Confidence Intervals by Serological Pattern Characterization | EBV Serological Status | Positive Agreement | | 95% CI | Negative Agreement | | 95% CI | | --- | --- | --- | --- | --- | --- | --- | | Primary Acute | (18/21) | 85.7% | 65.4 - 95.0% | (7/10) | 70.0% | 39.7 - 89.2% | | Late Acute | (72/76) | 94.7% | 87.2 - 97.9% | (21/34) | 61.8% | 45.0 - 76.1% | | Recovering | (4/4) | 100% | 51.0 - 100% | (0/0) | N/A* | N/A* | | Previous Infection | (0/3) | 0.0% | N/A* | (249/299) | 83.3% | 78.6 - 87.1% | | Susceptible | (0/1) | 0.0% | N/A* | (116/125) | 92.8% | 86.9 - 96.2% | | Inconclusive | (10/13) | 76.9% | 49.7 - 91.8% | (24/28) | 85.7% | 68.5 - 94.3% | | Overall | (104/118) | 88.1% | 81.1 - 92.8% | (417/496) | 84.1% | 80.6 - 87.0% | *In cases where agreement resulted in a numerator of zero (0), 95% confidence interval could not be calculated; in cases where agreement resulted in (0/0) samples, percent agreement and 95% confidence interval could not be calculated. Comparison of Characterization EBV Serological Status: Using Table A as a guideline, samples characterized into serological status associated with EBV disease, using the commercially available microplate EIA and agglutination tests, were compared with characterizations using BioPlex 2200 EBV IgG and IgM kits. The EBV IgG kit was run in conjunction with the EBV IgM kit to allow for a complete antibody response profile. The characterization by antibody response was not compared with clinical data regarding presence, absence or status of disease. Results from 618 serum samples tested at 3 U.S. clinical testing sites are shown in Table N. {13} Table N. Comparison of EBV Serological Status | EBV Serological status | BioPlex 2200 EBV IgG & IgM Profile | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Primary Acute | Late Acute | Recovering | Previous Infection | Susceptible | Inconclusive | Total | % Serological Agreement | 95% Confidence Interval | | Commercially Available Assays | Primary Acute | 30 | 0 | 0 | 0 | 0 | 1 | 31 | 96.8% | 83.8 - 99.4% | | | Late Acute | 5 | 90 | 1 | 13 | 0 | 1 | 110 | 81.8% | 73.6 - 87.9% | | | Recovering | 1 | 0 | 3 | 0 | 0 | 0 | 4 | 75.0% | 30.0 - 95.4% | | | Previous Infection | 0 | 31 | 2 | 263 | 4 | 5 | 305 | 86.2% | 81.9 - 89.7% | | | Susceptible | 4 | 0 | 0 | 0 | 122 | 1 | 127 | 96.1% | 91.1 - 98.3% | | | Inconclusive | 6 | 10 | 0 | 7 | 11 | 7 | 41 | 17.1% | 8.5 - 31.3% | | | Overall | 46 | 131 | 6 | 283 | 137 | 15 | 618 | 83.3% | 80.2 - 86.1% | Note: Calculations are performed for unshaded areas only. b. Matrix comparison: Not Applicable. 3. Clinical studies: Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: Expected values for the EBV IgG kit are presented by age and gender in the following tables for serum samples from unselected hospitalized pediatric and adult patients $(N = 303)$ and patients for which an EBV test was ordered $(N = 620)$ . A total of 621 serum samples from patients for which an EBV test was ordered were tested. One (1) sample from the patients for which an EBV test was ordered population was excluded due to RBB analysis error messages during BioPlex 2200 EBV IgG testing. For all analytes, results of $\leq 0.8$ AI are negative, 0.9 and $1.0$ AI are indeterminate, and $\geq 1.1$ AI are reported as positive. {14} Table B. Hospitalized Patient Samples: EBV NA-1 IgG | Age | Gender | BioPlex 2200 EBV NA-1 IgG | | | | | | Total | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Positive | | Indeterminate | | Negative | | | | | | N | % | N | % | N | % | N | | < 5 years of age | F | 11 | 41% | 0 | 0% | 16 | 59% | 27 | | | M | 6 | 30% | 0 | 0% | 14 | 70% | 20 | | 5-12 years of age | F | 13 | 59% | 0 | 0% | 9 | 41% | 22 | | | M | 15 | 44% | 0 | 0% | 19 | 56% | 34 | | 13-20 years of age | F | 28 | 80% | 0 | 0% | 7 | 20% | 35 | | | M | 10 | 67% | 0 | 0% | 5 | 33% | 15 | | 21-30 years of age | F | 5 | 83% | 0 | 0% | 1 | 17% | 6 | | | M | 1 | 50% | 0 | 0% | 1 | 50% | 2 | | 31-40 years of age | F | 10 | 100% | 0 | 0% | 0 | 0% | 10 | | | M | 11 | 100% | 0 | 0% | 0 | 0% | 11 | | 41-50 years of age | F | 13 | 100% | 0 | 0% | 0 | 0% | 13 | | | M | 7 | 100% | 0 | 0% | 0 | 0% | 7 | | 51-60 years of age | F | 22 | 96% | 0 | 0% | 1 | 4% | 23 | | | M | 18 | 95% | 0 | 0% | 1 | 5% | 19 | | 61-70 years of age | F | 11 | 100% | 0 | 0% | 0 | 0% | 11 | | | M | 12 | 100% | 0 | 0% | 0 | 0% | 12 | | 71-80 years of age | F | 11 | 100% | 0 | 0% | 0 | 0% | 11 | | | M | 6 | 100% | 0 | 0% | 0 | 0% | 6 | | 81-90 years of age | F | 11 | 100% | 0 | 0% | 0 | 0% | 11 | | | M | 5 | 83% | 0 | 0% | 1 | 17% | 6 | | 91-100 years of age | F | 0 | 0% | 0 | 0% | 0 | 0% | 0 | | | M | 2 | 100% | 0 | 0% | 0 | 0% | 2 | | Total | | 228 | 75% | 0 | 0% | 75 | 25% | 303 | {15} Table C. Hospitalized Patient Samples: EBV VCA IgG | Age | Gender | BioPlex 2200 EBV VCA IgG | | | | | | Total | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Positive | | Indeterminate | | Negative | | | | | | N | % | N | % | N | % | N | | < 5 years of age | F | 11 | 41% | 0 | 0% | 16 | 59% | 27 | | | M | 5 | 25% | 0 | 0% | 15 | 75% | 20 | | 5-12 years of age | F | 14 | 64% | 0 | 0% | 8 | 36% | 22 | | | M | 15 | 44% | 0 | 0% | 19 | 56% | 34 | | 13-20 years of age | F | 29 | 83% | 0 | 0% | 6 | 17% | 35 | | | M | 9 | 60% | 0 | 0% | 6 | 40% | 15 | | 21-30 years of age | F | 6 | 100% | 0 | 0% | 0 | 0% | 6 | | | M | 1 | 50% | 0 | 0% | 1 | 50% | 2 | | 31-40 years of age | F | 10 | 100% | 0 | 0% | 0 | 0% | 10 | | | M | 11 | 100% | 0 | 0% | 0 | 0% | 11 | | 41-50 years of age | F | 13 | 100% | 0 | 0% | 0 | 0% | 13 | | | M | 7 | 100% | 0 | 0% | 0 | 0% | 7 | | 51-60 years of age | F | 22 | 96% | 0 | 0% | 1 | 4% | 23 | | | M | 18 | 95% | 0 | 0% | 1 | 5% | 19 | | 61-70 years of age | F | 10 | 91% | 1 | 9% | 0 | 0% | 11 | | | M | 11 | 92% | 1 | 8% | 0 | 0% | 12 | | 71-80 years of age | F | 10 | 91% | 1 | 9% | 0 | 0% | 11 | | | M | 6 | 100% | 0 | 0% | 0 | 0% | 6 | | 81-90 years of age | F | 11 | 100% | 0 | 0% | 0 | 0% | 11 | | | M | 5 | 83% | 0 | 0% | 1 | 17% | 6 | | 91-100 years of age | F | 0 | 0% | 0 | 0% | 0 | 0% | 0 | | | M | 2 | 100% | 0 | 0% | 0 | 0% | 2 | | Total | | 226 | 75% | 3 | 1% | 74 | 24% | 303 | Table D. Hospitalized Patient Samples: EBV EA-D IgG | Age | Gender | BioPlex 2200 EA-D IgG | | | | | | Total | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Positive | | Indeterminate | | Negative | | | | | | N | % | N | % | N | % | N | | < 5 years of age | F | 2 | 7% | 0 | 0% | 25 | 93% | 27 | | | M | 1 | 5% | 1 | 5% | 18 | 90% | 20 | | 5-12 years of age | F | 4 | 18% | 0 | 0% | 18 | 82% | 22 | | | M | 5 | 15% | 0 | 0% | 29 | 85% | 34 | | 13-20 years of age | F | 9 | 26% | 2 | 6% | 24 | 69% | 35 | | | M | 3 | 20% | 3 | 20% | 9 | 60% | 15 | | 21-30 years of age | F | 2 | 33% | 0 | 0% | 4 | 67% | 6 | | | M | 1 | 50% | 1 | 50% | 0 | 0% | 2 | | 31-40 years of age | F | 4 | 40% | 2 | 20% | 4 | 40% | 10 | | | M | 6 | 55% | 0 | 0% | 5 | 45% | 11 | | 41-50 years of age | F | 5 | 38% | 0 | 0% | 8 | 62% | 13 | | | M | 2 | 29% | 0 | 0% | 5 | 71% | 7 | | 51-60 years of age | F | 8 | 35% | 5 | 22% | 10 | 43% | 23 | | | M | 10 | 53% | 1 | 5% | 8 | 42% | 19 | | 61-70 years of age | F | 3 | 27% | 0 | 0% | 8 | 73% | 11 | | | M | 5 | 42% | 1 | 8% | 6 | 50% | 12 | | 71-80 years of age | F | 4 | 36% | 2 | 18% | 5 | 45% | 11 | | | M | 1 | 17% | 0 | 0% | 5 | 83% | 6 | | 81-90 years of age | F | 7 | 64% | 1 | 9% | 3 | 27% | 11 | | | M | 5 | 83% | 0 | 0% | 1 | 17% | 6 | | 91-100 years of age | F | 0 | 0% | 0 | 0% | 0 | 0% | 0 | | | M | 1 | 50% | 0 | 0% | 1 | 50% | 2 | | Total | | 88 | 29% | 19 | 6% | 196 | 65% | 303 | {16} Table E. Samples from Patients for which an EBV Test was Ordered: EBV NA-1 IgG | Age | Gender | BioPlex 2200 EBV NA-1 IgG | | | | | | Total | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Positive | | Indeterminate | | Negative | | | | | | N | % | N | % | N | % | N | | < 5 years of age | F | 6 | 20% | 0 | 0% | 24 | 80% | 30 | | | M | 9 | 26% | 0 | 0% | 25 | 74% | 34 | | 5-12 years of age | F | 22 | 35% | 0 | 0% | 40 | 65% | 62 | | | M | 24 | 39% | 0 | 0% | 38 | 61% | 62 | | 13-20 years of age | F | 46 | 59% | 1 | 1% | 31 | 40% | 78 | | | M | 19 | 49% | 0 | 0% | 20 | 51% | 39 | | 21-30 years of age | F | 42 | 91% | 0 | 0% | 4 | 9% | 46 | | | M | 25 | 76% | 0 | 0% | 8 | 24% | 33 | | 31-40 years of age | F | 50 | 96% | 0 | 0% | 2 | 4% | 52 | | | M | 22 | 92% | 0 | 0% | 2 | 8% | 24 | | 41-50 years of age | F | 33 | 100% | 0 | 0% | 0 | 0% | 33 | | | M | 30 | 97% | 0 | 0% | 1 | 3% | 31 | | 51-60 years of age | F | 26 | 96% | 0 | 0% | 1 | 4% | 27 | | | M | 21 | 81% | 0 | 0% | 5 | 19% | 26 | | 61-70 years of age | F | 11 | 85% | 0 | 0% | 2 | 15% | 13 | | | M | 19 | 90% | 0 | 0% | 2 | 10% | 21 | | 71-80 years of age | F | 2 | 100% | 0 | 0% | 0 | 0% | 2 | | | M | 3 | 100% | 0 | 0% | 0 | 0% | 3 | | 81-90 years of age | F | 2 | 100% | 0 | 0% | 0 | 0% | 2 | | | M | 2 | 100% | 0 | 0% | 0 | 0% | 2 | | 91-100 years of age | F | 0 | 0% | 0 | 0% | 0 | 0% | 0 | | | M | 0 | 0% | 0 | 0% | 0 | 0% | 0 | | Total | | 414 | 67% | 1 | 0% | 205 | 33% | 620 | Table F. Samples from Patients for which an EBV Test was Ordered: EBV VCA IgG | Age | Gender | BioPlex 2200 EBV VCA IgG | | | | | | Total | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Positive | | Indeterminate | | Negative | | | | | | N | % | N | % | N | % | N | | < 5 years of age | F | 6 | 20% | 0 | 0% | 24 | 80% | 30 | | | M | 11 | 32% | 0 | 0% | 23 | 68% | 34 | | 5-12 years of age | F | 21 | 34% | 0 | 0% | 41 | 66% | 62 | | | M | 28 | 45% | 0 | 0% | 34 | 55% | 62 | | 13-20 years of age | F | 47 | 60% | 0 | 0% | 31 | 40% | 78 | | | M | 20 | 51% | 0 | 0% | 19 | 49% | 39 | | 21-30 years of age | F | 43 | 93% | 1 | 2% | 2 | 4% | 46 | | | M | 26 | 79% | 0 | 0% | 7 | 21% | 33 | | 31-40 years of age | F | 52 | 100% | 0 | 0% | 0 | 0% | 52 | | | M | 22 | 92% | 0 | 0% | 2 | 8% | 24 | | 41-50 years of age | F | 32 | 97% | 0 | 0% | 1 | 3% | 33 | | | M | 31 | 100% | 0 | 0% | 0 | 0% | 31 | | 51-60 years of age | F | 27 | 100% | 0 | 0% | 0 | 0% | 27 | | | M | 24 | 92% | 0 | 0% | 2 | 8% | 26 | | 61-70 years of age | F | 12 | 92% | 0 | 0% | 1 | 8% | 13 | | | M | 18 | 86% | 0 | 0% | 3 | 14% | 21 | | 71-80 years of age | F | 2 | 100% | 0 | 0% | 0 | 0% | 2 | | | M | 3 | 100% | 0 | 0% | 0 | 0% | 3 | | 81-90 years of age | F | 2 | 100% | 0 | 0% | 0 | 0% | 2 | | | M | 1 | 50% | 0 | 0% | 1 | 50% | 2 | | 91-100 years of age | F | 0 | 0% | 0 | 0% | 0 | 0% | 0 | | | M | 0 | 0% | 0 | 0% | 0 | 0% | 0 | | Total | | 428 | 69% | 1 | 0% | 191 | 31% | 620 | {17} Table G. Samples from Patients for which an EBV Test was Ordered: EBV EA-D IgG | Age | Gender | BioPlex 2200 EA-D IgG | | | | | | Total | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Positive | | Indeterminate | | Negative | | | | | | N | % | N | % | N | % | N | | < 5 years of age | F | 3 | 10% | 3 | 10% | 24 | 80% | 30 | | | M | 6 | 18% | 2 | 6% | 26 | 76% | 34 | | 5-12 years of age | F | 7 | 11% | 3 | 5% | 52 | 84% | 62 | | | M | 6 | 10% | 4 | 6% | 52 | 84% | 62 | | 13-20 years of age | F | 16 | 21% | 8 | 10% | 54 | 69% | 78 | | | M | 12 | 31% | 2 | 5% | 25 | 64% | 39 | | 21-30 years of age | F | 16 | 35% | 2 | 4% | 28 | 61% | 46 | | | M | 9 | 27% | 3 | 9% | 21 | 64% | 33 | | 31-40 years of age | F | 15 | 29% | 5 | 10% | 32 | 62% | 52 | | | M | 7 | 29% | 1 | 4% | 16 | 67% | 24 | | 41-50 years of age | F | 10 | 30% | 2 | 6% | 21 | 64% | 33 | | | M | 4 | 13% | 1 | 3% | 26 | 84% | 31 | | 51-60 years of age | F | 13 | 48% | 3 | 11% | 11 | 41% | 27 | | | M | 10 | 38% | 1 | 4% | 15 | 58% | 26 | | 61-70 years of age | F | 6 | 46% | 1 | 8% | 6 | 46% | 13 | | | M | 5 | 24% | 3 | 14% | 13 | 62% | 21 | | 71-80 years of age | F | 0 | 0% | 1 | 50% | 1 | 50% | 2 | | | M | 1 | 33% | 0 | 0% | 2 | 67% | 3 | | 81-90 years of age | F | 0 | 0% | 0 | 0% | 2 | 100% | 2 | | | M | 1 | 50% | 0 | 0% | 1 | 50% | 2 | | 91-100 years of age | F | 0 | 0% | 0 | 0% | 0 | 0% | 0 | | | M | 0 | 0% | 0 | 0% | 0 | 0% | 0 | | Total | | 147 | 24% | 45 | 7% | 428 | 69% | 620 | # N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. # O. Conclusion: 1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...