K062154 · Vital Images, Inc. · LLZ · Aug 10, 2006 · Radiology
Device Facts
Record ID
K062154
Device Name
VITAL CONNECT 4.0
Applicant
Vital Images, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 10, 2006
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
ViTALConnect, Version 4.0 is a medical diagnostic software system intended to process, analyze, review, and distribute multi-dimensional digital images acquired from a variety of imaging devices including: CT, MR, CR/DR/DX, SC, US, NM, PET, XA, and RF, etc. In addition, the ViTALConnect system has the following specific intended uses: Advanced Vessel Analysis is an option within the ViTALConnect system intended to aid clinicians in the study and analysis of selected vessels, the inspection of circulatory anatomy, the quantification of stenosis, distance measurements and cross-sectional diameters, volume measurements, and assist in the planning and monitoring of vascular therapies. The feature operates with CT and MRI data. Pet Visualization is an option within the ViTALConnect system that provides for the overlay, inspection, and measurement of two different image modalities (primarily directed at PET/CT). The option also includes the ability to measure Standard Uptake Values on PET datasets. The overlay operation enables clinicians to obtain a better understanding of the joint information that would otherwise have to be compared separately. It is important to note that the clinician retains the ultimate responsibility for making pertinent diagnosis based on their standard procedures, including visual comparison of separate images. Collaboration Mode is a capability within the ViTALConnect system that allows multiple users to collaboratively interact with ViTALConnect system features and tools.
Device Story
ViTALConnect 4.0 is a medical diagnostic software system for processing, reviewing, and analyzing multi-dimensional digital images; inputs include DICOM data from CT, MR, CR/DR/DX, SC, US, NM, PET, XA, and RF modalities. System allows clinicians to retrieve, display, manipulate, annotate, and measure images; features include Advanced Vessel Analysis (stenosis quantification, volume/diameter measurements) and PET Visualization (modality overlay, Standard Uptake Value measurement). Collaboration Mode enables multi-user interaction. Used in clinical settings by physicians/clinicians to aid in anatomy/pathology analysis and therapy planning. Output provided via display, film/paper print, or report export. Clinicians retain ultimate diagnostic responsibility; device serves as an aid to standard clinical workflows.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, integration testing, verification testing, and software validation performed according to internal quality procedures. Risk management plan implemented to control potential hazards.
Technological Characteristics
Medical image processing software; DICOM-compliant; supports multi-dimensional visualization of CT, MR, CR/DR/DX, SC, US, NM, PET, XA, and RF data. Features include vessel analysis, PET/CT overlay, and multi-user collaboration. Software-based; operates on standard clinical hardware.
Indications for Use
Indicated for clinicians requiring processing, analysis, review, and distribution of multi-dimensional digital images (CT, MR, CR/DR/DX, SC, US, NM, PET, XA, RF) for diagnostic purposes, including vascular therapy planning/monitoring and multi-modality image overlay.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
VITALConnect System- formally The iConnection System (K040876)
Submission Summary (Full Text)
{1}------------------------------------------------
# 6.0 510(k) Summary
#### Submitter's Name / Contact Person
AUG 1 0 2006
Timothy J. Kappers, MBA, RAC Director, Quality Systems, Regulatory & Clinical Affairs Vital Images, Inc. 5850 Opus Parkway, Suite 300 Minnetonka, MN 55343
#### General Information
| Device Trade Name | VITALConnect™ 4.0 - Medical Image Processing Software | |
|----------------------------------------|-----------------------------------------------------------------------------------------|--|
| Common / Usual Name | System, Image Processing, Radiological | |
| Classification | 892.2050 Picture Archiving and Communications System<br>(LLZ; Class II) | |
| Identification of<br>Predicate Devices | VITALConnect System- formally The iConnection System<br>(K040876)<br>Vital Images, Inc. | |
#### Device Description
The ViTALConnect system is a medical diagnostic device that allows the processing, review. analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices.
The ViTALConnect system provides multi-dimensional visualization of digital images to aid clinicians in their analysis of anatomy and pathology. The ViTALConnect system user interface follows typical clinical workflow patterns to process, review, and analyze digital images, including:
- . Retrieve image data over the network via DICOM
- Display images that are automatically adapted to exam type via dedicated protocols .
- Select images for closer examination from a gallery of up to six 2D or 3D views .
- Interactively manipulate an image in real-time to visualize anatomy and pathology .
- Annotate, tag, measure, and record selected views .
{2}------------------------------------------------
- Output selected views to standard film or paper printers, or post a report to an Intranet . Web server or export views to another DICOM device
- . Retrieve reports that are archived on a Web server
## Intended Use
ViTALConnect, Version 4.0 is a medical diagnostic software system intended to process, analyze, review, and distribute multi-dimensional digital images acquired from a variety of imaging devices including: CT, MR, CR/DR/DX, SC, US, NM, PET, XA, and RF, etc. In addition, the ViTALConnect system has the following specific intended uses:
Advanced Vessel Analysis is an option within the ViTALConnect system intended to aid clinicians in the study and analysis of selected vessels, the inspection of circulatory anatomy, the quantification of stenosis, distance measurements and cross-sectional diameters, volume measurements, and assist in the planning and monitoring of vascular therapies. The feature operates with CT and MRI data.
Pet Visualization is an option within the ViTALConnect system that provides for the overlay, inspection, and measurement of two different image modalities (primarily directed at PET/CT). The option also includes the ability to measure Standard Uptake Values on PET datasets.
The overlay operation enables clinicians to obtain a better understanding of the joint information that would otherwise have to be compared separately. It is important to note that the clinician retains the ultimate responsibility for making pertinent diagnosis based on their standard procedures, including visual comparison of separate images.
Collaboration Mode is a capability within the ViTALConnect system that allows multiple users to collaboratively interact with ViTALConnect system features and tools.
## Predicate Device Comparison
The ViTALConnect 4.0 system and its predicate device allow for the analysis, communication and media interchange of digital images acquired from a variety of acquisition devices. All devices support the DICOM protocol for communication of images with other medical imaging devices.
## Summary of Studies
The software utilized was designed, developed, tested, and validated according to written procedures. These procedures specify individuals within the organization responsible for developing and approving product specifications, coding, testing, validating and maintenance.
{3}------------------------------------------------
The ViTALConnect 4.0 system will successfully complete integration testing/verification testing prior to Beta validation. Software Beta testing/validation will be successfully completed prior to release. In addition, potential hazards have been studied and controlled by a Risk Management Plan.
# Conclusion
The ViTALConnect 4.0 system has the same intended uses as the predicate device and has very similar technological characteristics. Minor technological differences do not raise any new questions regarding safety or effectiveness of the device. Thus, the ViTALConnect 4.0 system is substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the edge. In the center of the seal is a stylized image of an eagle with its wings spread. The eagle is facing to the right and has three stripes on its wings.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
# AUG 1 0 2006
Vital Images, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services 1394 25th Street NW BUFFALO MN 55313
Re: K062154
Trade/Device Name: ViTAL Connect™ 4.0 Medical Image Processing Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 24, 2006 Received: July 28, 2006
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a black and white circular logo. The logo commemorates the centennial anniversary from 1906-2006. The letters "FDA" are in the center of the logo. The word "Centennial" is written below the letters "FDA".
*Protecting and Promoting Public Health*
{5}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlvindustry/support/index.html.
Sincerely yours, .
Nancy Cbrogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## 3.0 Intended Use Statement
# 510(k) Number (if known): __ K o L 2 15 1 Device Name: ViTALConnect™ 4.0 Medical Image Processing Software
ViTALConnect, Version 4.0 is a medical diagnostic software system intended to process, analyze, review, and distribute multi-dimensional digital images acquired from a variety of imaging devices including: CT, MR, CR/DR/DX, SC, US, NM, PET, XA, and RF, etc. In addition, the ViTALConnect system has the following specific intended uses:
Advanced Vessel Analysis is an option within the ViTALConnect system intended to aid clinicians in the study and analysis of selected vessels, the inspection of circulatory anatomy, the quantification of stenosis, distance measurements and cross-sectional diameters, volume measurements, and assist in the planning and monitoring of vascular therapies. The feature operates with CT and MRI data.
Pet Visualization is an option within the ViTALConnect system that provides for the overlay, inspection, and measurement of two different image modalities (primarily directed at PET/CT). The option also includes the ability to measure Standard Uptake Values on PET datasets.
The overlay operation enables clinicians to obtain a better understanding of the joint information that would otherwise have to be compared separately. It is important to note that the clinician retains the ultimate responsibility for making pertinent diagnosis based on their standard procedures, including visual comparison of separate images.
Collaboration Mode is a capability within the ViTALConnect system that allows multiple users to collaboratively interact with ViTALConnect system features and tools.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C Hodgson
Page of
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K062154
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.