The Escort M8 Vital Signs Patient Monitor Model 3810 is intended for use as a vital signs monitor for adult and pediatric patients in a hospital setting.
Device Story
Escort M8 Model 3810 is a multiparameter patient monitor for hospital use. It collects physiological signals via sensors for ECG, non-invasive/invasive blood pressure, temperature, respiration, SpO2, and capnometry. The device processes these signals to calculate heart rate, respiration rate, and detect arrhythmias. Data is displayed on an 8.4-inch color TFT LCD. Clinicians use the monitor to observe real-time patient status and alarms. The device aids in clinical decision-making by providing continuous monitoring and automated arrhythmia analysis (e.g., asystole, ventricular fibrillation, PVCs). It benefits patients by enabling timely intervention through continuous vital sign tracking and alarm notification.
Clinical Evidence
No clinical data. Evidence consists of bench testing and validation using patient simulators under simulated use conditions. Performance requirements for accuracy and precision were verified against standards including IEC 60601-2-34 (invasive pressure), ISO 21647 (capnometry), and AAMI/ANSI EC 57 (arrhythmia detection). Electrical safety, EMC, and biocompatibility were confirmed via UL 60601-1, IEC 60601-1-2, and ISO 10993-1.
Technological Characteristics
Multiparameter monitor with 8.4-inch color TFT LCD (640x480). Parameters: ECG, NIBP, IBP, Temp, Resp, SpO2, EtCO2. Connectivity: 902-928 MHz spread spectrum wireless transmitter/receiver. Standards: UL 60601-1, IEC 60601-1-2, ISO 10993-1, ISO 21647, AAMI/ANSI EC 57. Arrhythmia detection algorithm compliant with AAMI/ANSI EC 57.
Indications for Use
Indicated for adult and pediatric patients in hospital settings for monitoring vital signs, including heart rate, ECG, non-invasive/invasive blood pressure, temperature, respiration, SpO2, and capnometry (eTCO2/FiCO2), with arrhythmia detection.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Zoe Medical Nightingale Monitoring System (K001775)
Datascope Passport 2 with View 12 ECG Analysis Module (K020550)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510K SUMMARY OF SAFETY AND EFFECTIVENESS FOR ESCORT M8 VITAL SIGNS PATIENT MONITOR MODEL 3810
## 1. Submitter Information:
Invivo Corporation 12601 Research Parkway Orlando, Fl 32826
0C7 1 2 2006
| Contact Person: | Rusty Kelly |
|-----------------|-------------------------|
| Title: | Quality Control Manager |
| Telephone No.: | (407) 275-3220 Ext: 166 |
| Fax No.: | (407) 206-9568 |
| Date Prepared: | July 5, 2006 |
## 2. Device Name:
Classification Name: Monitor, cardiac (incl. Cardiotachometer and rate alarm)
Propietary Name: Escort M8 Vital Signs Patient Monitor Model 3810
| Device | Product Code | CFR Reference | Class |
|-----------------------------|--------------|---------------|-------|
| ECG | DRT | 870.2300 | II |
| non invasive blood pressure | DXN | 870.1130 | II |
| invasive blood pressure | DSK | 870.1110 | II |
| temperature | FLL | 880.2910 | II |
| respiration | BZQ | 868.2375 | II |
| SPO2 (Oximeter) | DQA | 870.2700 | II |
| ETCO2 (Capnometer) | CCK | 868.1400 | II |
| arrhythmia detection | MHX | 870.1025 | II |
| transmitter and receiver | DRG | 870.2910 | II |
| recorder | DSF | 870.2810 | II |
| display | DXJ | 870.2450 | II |
Product Codes:
# 3. Device Description:
The Escort M8 Vital Signs Patient Monitor Model 3810 is a multiparameter patient monitor intended to monitor adult and pediatric patient vital signs for general hospital use. The Escort M8 contains the hardware and software needed to perform complex data gathering and signal processing tasks that allow it to produce accurate and reliable measurements of physiological parameters such as heart rate, ECG, non invasive blood pressure, temperature, respiration, SPO2, and capnometry (eTCO2 and FiCO2), provides arrhythmia detection and displays this information on an LCD display.
{1}------------------------------------------------
# 4. Substantial Equivalence:
The Escort M8 Vital Signs Patient Monitor is substantially equivalent to the following devices:
| Zoe Medical Nightingale Monitoring System | 510(k) K001775 |
|-------------------------------------------------------|----------------|
| Escort II + 400 Series (Escort Prism) | 510(k) K014294 |
| Polaris 2004 Capnograph | 510(k) K040011 |
| Masimo SET Radical Pulse Oximeter | 510(k) K992340 |
| Datascope Passport 2 with View 12 ECG Analysis Module | 510(k) K020550 |
# 5. Indications for Use
The Escort M8 Vital Signs Patient Monitor is intended for use as a vital signs monitor for adult and pediatric patients in hospital settings.
# 6. Comparison to Predicate Devices
The Escort M8 Vital Signs Patient Monitor is a modification of the Zoe Medical Nightingale Monitoring System. This modification adds a capnometer, invasive blood pressure monitor, arrhythmia detector and alarm, strip chart recorder, replaces the existing oximeter, adds lead selectivity to ECG and increases the size of the display.
# 7. Technological Characteristics
A comparison of the technological characteristics of the Escort M8 Vital Signs Patient Monitor and the predicate devices has been performed. The results of this comparison demonstrate that the Escort M8 Vital Signs Patient Monitor is equivalent to the marketed predicate devices in technological characteristics.
## 8. Environmental and non-clinical testing
Applicable environmental and non-clinical testing was performed per UL60601 and IEC 60601-1-2 as well as other applicable standards and procedures. The Escort M8 Vital Signs Patient Monitor passed all tests.
## 9. Performance data
The performance data included in this submission to compare equivalency of the Escort M8 Vital Signs Patient Monitor with its predicate devices met the performance requirements for accuracy and precision and indicates substantial equivalence to the predicate devices. Equivalent performance in meeting user requirements was also determined.
{2}------------------------------------------------
## 10. Summary of performance Testing
Validation and Verification Testing confirmed that this device operates as designed and intended. The specifications for this device were verified in bench testing under simulated use conditions. Compliance to performance standards was confirmed by bench testing.
This device was validated using patient simulators under simulated use conditions. The functional requirements and user needs of the Escort M8 Vital Signs Patient Monitor were verified to have been met.
#### General
| Parameter | Requirement | Pass/Fail |
|-------------------|-------------------------------------|-----------|
| Electrical Safety | UL 60601-1:2003, IEC 60601-1-1:2000 | Pass |
| EMC | IEC 60601-1-2 | Pass |
| Biocompatibility | ISO 10993-1 | Pass |
| Shock | IEC 60068-2-27 | Pass |
| Vibration | IEC 60068-2-64 | Pass |
## Invasive Pressure Monitoring
| Parameter | Requirement | Pass/Fail |
|----------------------|----------------------|-----------|
| Measurement range | -10 to 300 mmHg | Pass |
| Measurement accuracy | ± 2 mmHg or ± 2% | Pass |
| Heart Rate Range | 30 to 240 bpm. | Pass |
| Heart Rate Accuracy | ± 4 bpm | Pass |
| Sensitivity | 5 uV/V/mmHg | Pass |
| Standard | IEC 60601-2-34: 2000 | Pass |
#### Capnometer
| Parameter | Requirement | Pass/Fail |
|------------------------|-----------------------------------------------|-----------|
| Accuracy 0 to 20 min | 0 to 38 mmHg: $\pm$ 4 mmHg | Pass |
| | 39 to 99 mmHg: $\pm$ 12% | |
| Accuracy 20 min and up | 0 to 38 mmHg: $\pm$ 2 mmHg | Pass |
| | 39 to 99 mmHg: $\pm$ 5% of reading +0.08% for | |
| | every 1mmHg above 38 mmHg | |
| Measurement Range | 0 to 99 mmHg ; 0 to 13.0 Vol%; 0 to 13.0 kPa | Pass |
| Flow Rate | 50ml per min. | Pass |
| Respiration Rate | 0 to 150 bpm | Pass |
| Respiration Accuracy | 0 to 70 bpm : $\pm$ 1 bpm | Pass |
| | 71 to 120 bpm $\pm$ 2 bpm | |
| | 121 to150 bpm $\pm$ 3 bpm | |
| Standard | ISO 21647:2004 | Pass |
#### Transmitter and Receiver
| Parameter | Requirement | Pass/Fail |
|---------------|----------------------------------------|-----------|
| Frequency | Spread spectrum 902 to 928 MHz | Pass |
| Transmission | Bidirectional with seamless auto-retry | Pass |
| FCC Licensing | 47 CFR Part 15 | Pass |
{3}------------------------------------------------
## Multiple Lead Selection on Cardiac Monitor
| Parameter | The consisted to the research to recommendent of the more of the more of the many of the comments of the comments of the comments of the comments of the comments of the comme<br>Requirement | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| eads | | |
| "LEALE MELLETE KENDY SECTIOn And Claim of California Collection of Call Collection of Collection of FREE BER FLEFF & FLORE A FRIENDEN OF FUNLES A<br>Vector | TEMPERATION CONSULTIONAL LE CLUMBAR LE COMMENT COMMENT COMMENT COMMENT COMMENT COMMENT COMMENT COMMENT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CO | |
| State of the secure to the manage with a many of the comments of the comments of the contribution of the | | |
The testing of AAMI/ANSI EC 13 will be completed prior to product release, testing still ongoing.
## Arrhythmia Detector and Alarm
| Parameter | Requirement | Pass/Fail |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Heart Rate Range | 15 to 300 bpm. | Pass |
| Heart range accuracy | $\pm$ 2 bpm or $\pm$ 1%, whichever is greater | Pass |
| Pacer rejection | Rejects all pulses of amplitude $\pm$ 2mV to $\pm$ 700mV and duration 0.1 to 2 ms with no tail | Pass |
| Tall T-Wave rejection | Without Arrhythmia Option: Rejects all T-Waves less than or equal to 120% of 1mV QRS. With Arrhythmia Option: Rejects T-Waves less than or equal to 60% of 1mv QRS. | Pass |
| Arrhythmia analysis | Asystole, Couplets, Bigeminy, Trigminy, Ventricular Tachycardia, Ventricular Fibrillation, Ventricular Rhythm, Ventricular runs, Multi-focal PVC's, and R-on-T. | Pass |
| Standard | AAMI/ANSI EC 57: 1998 | Pass |
#### Oximeter
| Parameter | Requirement | Pass/Fail |
|-----------------------|----------------------------------|-----------|
| Saturation Range | 1% to 100% | Pass |
| Saturation Accuracy | 70 to 100% ± 2 digits -no motion | Pass |
| | 70 to 100% ± 3 digits -motion | Pass |
| | 0 to 69% unspecified | Pass |
| Saturation Resolution | 1% | Pass |
| Pulse rate range | 25 to 240 Bpm | Pass |
| Pulse rate accuracy | 25 to 240 ± 3 digits -no motion | Pass |
| | 25 to 240 ± 5 digits -motion | Pass |
| Pulse rate resolution | 1 BPM | Pass |
### Display
| Parameter | Requirement | Pass/Fail |
|------------------|-----------------------------|-----------|
| Type | Color Active Matrix TFT LCD | Pass |
| Size | 8.4 in./21.3 cm diagonal | Pass |
| Resolution | 640 x 480 pixels | Pass |
| Number of Traces | 4 channels | Pass |
# 11. Quality System
This device is being designed and manufactured in a quality system that has been certified to EN 13485:2003 requirements, and conforms with the Medical Device Directive requirements found in the European Council Directive 93/42/EEC.
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 2 2006
Invivo Corporation c/o Mr. Rusty Kelly Quality Control Manager 12601 Research Parkway Orlando, FL 32826
Re: K062144
Trade Name: Escort M8 Vital signs Patient Monitor, Model 3810 Regulation Number: 21 CFR 870.1025 Regulation Name: Physiological Patient Monitor (With Arrhythmia Detection or Alarm) Regulatory Class: Class II (two) Product Code: MHX, CCK, DSK, DRT, DQA Dated: July 26, 2006 Received: July 27, 2006
Dear Mr. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 – Mr. Rusty Kelly
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be advisou that i bromination that your device complies with other requirements of the Act that I DA has made a acterimant administered by other Federal agencies. You must or any I cueral statutes and registements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OF N Fart 807); moother (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality by systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product raction of to begin marketing your device as described in your Section 510(k) I mis letter will and in your e FDA finding of substantial equivalence of your device to a legally prematication of the results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rryou desire of Compliance at (240)276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
B. Ramimmar for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE
510K Number: K062144
Device Name: Escort M8 Vital Signs Patient Monitor Model 3810
Indications for Use: The Escort M8 Vital Signs Patient Monitor Model 3810 is intended for use as a vital signs monitor for adult and pediatric patients in a hospital setting.
Prescription Use J (21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Condurrence of CDRH, Office of Device Evaluation (ODE)
*Blummemman*
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K212144
Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.