← Product Code [MQP](/productcode/MQP) · K062132

# AVS PL PEEK SPACER SYSTEM (K062132)

_Stryker Spine · MQP · Aug 16, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K062132

## Device Facts

- **Applicant:** Stryker Spine
- **Product Code:** [MQP](/productcode/MQP.md)
- **Decision Date:** Aug 16, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Stryker Spine AVS™ PL PEEK Spacer is a vertebral body replacement indicated for use in the thoraco-lumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body resected or excised during partial and total vertebrectomy procedures due to tumor or trauma, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. It is recommended to pack bone graft material inside the implant. The Stryker Spine AVS™ PL PEEK Spacer is intended for use with supplemental fixation. The supplemental fixation systems that may be used with the Stryker Spine AVS™ PL PEEK Spacer include, but are not limited to, Stryker Spine plate or rod systems (XIA, Spiral Radius 90D, and Trio).

## Device Story

Rectangular, hollow frame spinal implant; machined from medical-grade PEEK OPTIMA LT1; incorporates Tantalum marker pins for radiographic visualization. Used in thoraco-lumbar spine (T1-L5) to replace collapsed/damaged vertebral bodies following vertebrectomy. Implant packed with bone graft material; requires supplemental fixation (e.g., Stryker XIA, Spiral Radius 90D, or Trio systems). Operates as a structural spacer to restore vertebral height and decompress neural tissues. Used by surgeons in clinical/OR settings.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing and compliance with FDA's Guidance for Spinal System 510(k)'s (May 3, 2004).

## Technological Characteristics

Material: PEEK OPTIMA LT1; Tantalum marker pins. Form factor: Rectangular, hollow frame with lateral fenestrations. Dimensions: Various footprints, heights, and lordotic angles. Connectivity: None. Sterilization: Not specified.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Stryker Spine AVS™ PL PEEK Spacer ([K050624](/device/K050624.md))

## Reference Devices

- AVS™ PL Peek System ([K040731](/device/K040731.md))
- XIA (Stryker Spine)
- Spiral Radius 90D (Stryker Spine)
- Trio (Stryker Spine)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Line Extension to the AVS™ PL PEEK Spacer System

Special 510(k) Premarket Notificatio

## Special 510(k) Summary of Safety and Effectiveness: Special STo(R) Summary of PEEK Spacer System

| Proprietary Name:                  | AVSTM PL PEEK Spacer System<br>AUG 1 6 2006                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:                       | Spinal Fixation Appliances                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Proposed Regulatory Class:         | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                    | Spinal Vertebral Body Replacement Device,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                    | 21 CFR 888.3060                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Product Code:               | MQP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| For Information contact:           | Simona Voic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                    | Regulatory Affairs Project Manager                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                    | 2 Pearl Court                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                    | Allendale, NJ 07401                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                    | Telephone: (201) 760-8145                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                    | Fax: (201) 760-8345                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                    | Email: Simona. Voic@stryker.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Date Summary Prepared:             | July 25, 2006                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Predicate Device                   | Stryker Spine AVSTM PL PEEK Spacer (K050624)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Predicate Device Information       | The AVSTM PL Peek Spacer is a rectangular shaped,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                    | hollow frame implant with lateral fenestrations,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                    | machined from medical grade PEEK OPTIMA LT1.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                    | The spacer incorporates two (2) Tantalum marker pins                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                    | to aid in radiographic visualization. The AVSTM PL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                    | Peek Spacer System is comprised of six (6) different                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Description of Device Modification | sized footprints with a variety of heights and lordotic<br>angles. The AVSTM PL Peek System was determined<br>substantial equivalent via K040731 and K050624.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Intended Use                       | This Special 510(k) premarket notification is intended<br>to introduce additional sizes of the AVSTM PL Peek<br>Spacer.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                    | The Stryker Spine AVSTM PL PEEK Spacer is a<br>vertebral body replacement indicated for use in the<br>thoraco-lumbar spine (T1-L5) to replace a collapsed,<br>damaged, or unstable vertebral body resected or<br>excised during partial and total vertebrectomy<br>procedures due to tumor or trauma, to achieve anterior<br>decompression of the spinal cord and neural tissues,<br>and to restore the height of a collapsed vertebral body.<br>It is recommended to pack bone graft material inside<br>the implant.<br>The Stryker Spine AVSTM PL PEEK Spacer is intended<br>for use with supplemental fixation. The supplemental<br>fixation systems that may be used with the Stryker<br>Spine AVSTM PL PEEK Spacer include, but are not<br>limited to, Stryker Spine plate or rod systems (XIA,<br>Spiral Radius 90D, and Trio). |

{1}------------------------------------------------

{2}------------------------------------------------

Summary of the Technological Characteristics

Documentation is provided which demonstrates the additional sizes of the Stryker Spine AVS™ PL PEEK Spacer to be substantially equivalent to their predicate device in terms of its material, design, and indications for use. Compliance with FDA's Guidance for Spinal System 510(k)'s May 3, 2004 was also demonstrated for the subject spacers.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" arranged in a circular pattern around the symbol. The caduceus is a common symbol associated with healthcare and medicine, and its presence in the logo signifies the department's role in promoting and protecting the health of the nation.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 6 2006

Stryker Spine % Ms. Simona Voic Regulatory Affairs Project Manager 2 Pearl Court Allendale, New Jersey 07401

Re: K062132

Trade/Device Name: AVS™ PL PEEK Spacer System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: MQP Dated: July 25, 2006 Received: July 26, 2006

Dear Ms. Voic:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This

{4}------------------------------------------------

Page 2 - Ms. Simona Voic

letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and this, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may ontain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/suport/index.html.

Sincerely yours.

Per Sera D. Primo no
DEO V.S.C.R.R

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K062132

Line Extension to the AVS™ PL PEEK Spacer System

Special 510(k) Premarket Notification

## Indications for Use

510(k) Number (if known):

Device Name: AVS™ PL PEEK Spacer System

Indications for Use:

The Stryker Spine AVS™ PL PEEK Spacer is a vertebral body replacement indicated for use in the thoraco-lumbar spine (T1-L5) to replace a collapsed, damaged, or unstable verteral body resected or excised during partial and total vertebrectory procedures due to tumor or trauma, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body.

It is recommended to pack bone graft material inside the implant.

The Stryker Spine AVS™ PL PEEK Spacer is intended for use with supplemental fixation. The Supplemental fixation systems that may be used with the Stryker Spine AVS™ PL PEEK Spacer include, but are not limited to, Stryker Spine plate or rod systems (XIA, Spiral Radius 90D, and Trio).

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

P.O. hma

(Division Sign-Of Division of General, Restorative, and Neurological Devices

Page 1 of 1

**510(k) Number** K062132

---

**Source:** [https://fda.innolitics.com/device/K062132](https://fda.innolitics.com/device/K062132)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
