MEDIC4ALL TELEMEDICINE SYSTEM, MODEL VMS-01 (NIBP)

K062127 · Medic4all (Israel) , Ltd. · DRG · Nov 8, 2006 · Cardiovascular

Device Facts

Record IDK062127
Device NameMEDIC4ALL TELEMEDICINE SYSTEM, MODEL VMS-01 (NIBP)
ApplicantMedic4all (Israel) , Ltd.
Product CodeDRG · Cardiovascular
Decision DateNov 8, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2910
Device ClassClass 2

Indications for Use

The Medic4All Telemedicine System Model VMS-01 is intended to collect and transmit physiological information, such as non-invasive blood pressure, via standard telephone lines or the internet to local or remote computer systems for use by the patient or a healthcare professional.

Device Story

Medic4All VMS-01 is a wireless telephonic/internet-based physiologic monitoring system; collects physiological data (e.g., non-invasive blood pressure); transmits data via standard telephone lines or internet to local/remote computer systems. System enables interactive remote measurement, processing, storage, and display of data; supports optional verbal and video conferencing between patient and caregiver. Used in home or clinical settings; operated by patients or healthcare professionals. Output allows providers to review patient data for clinical decision-making; facilitates remote care management and periodic monitoring.

Clinical Evidence

Bench testing only. Verification and validation activities included EMC/electrical safety, unit testing, integration testing, data integrity, usability, environmental, compatibility, performance, system, and load/performance testing. All tests confirmed the system functioned as intended.

Technological Characteristics

Wireless telephonic/internet-based physiologic monitoring system. Supports data measurement, transmission, processing, storage, and display. Includes optional hardware for verbal and video conferencing. Connectivity via standard telephone lines or internet.

Indications for Use

Indicated for patients requiring periodic monitoring of physiological information, including non-invasive blood pressure, for transmission to healthcare professionals or for patient use.

Regulatory Classification

Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 5 510(K) SUMMARY Image /page/0/Picture/1 description: The image shows the text "K062127 pg 1 of 2" written in black ink on a white background. The text appears to be handwritten. The text "K062127" is on the top line, and the text "pg 1 of 2" is on the bottom line. NOV - 8 2005 This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. § 807.92. ## Submitter: #### Medic4All (Israel) LTD. 10 Hamefalsim Street, POB 4222 Petach Tikva, Israel 49000 Phone number: +972-3-9226610 Fax number: +972-3-9226615 #### Contact Person: Jonathan S. Kahan Hogan & Hartson L.L.P. Columbia Square 555 Thirteenth Street, NW Washington, DC 20004-1109 Phone: (202) 637-5794 Date Prepared: July 21, 2006 #### Name of the Device: Medic4All Telemedicine System VMS-01 #### Predicate Devices: The VMS-01 is substantially equivalent to: - The Philips Medical Systems M3810A TeleMonitoring System with . M3812B TeleStation (K023749). - The BL-Healthcare, Inc. Remote Care Management System (K051470). . {1}------------------------------------------------ K062127 pg 2 of 2 #### Intended Use / Indications for Use The Medic4All Telemedicine System Model VMS-01 is intended to collect and transmit physiological information, such as non-invasive blood pressure, via standard telephone lines or the internet to local or remote computer systems for use by the patient or a healthcare professional. ## Technological Characteristics Medic4All's VMS-01 telemedicine system (VMS-01) is a wireless telephonic/internet-based physiologic monitoring system, designed for care of patients requiring periodic monitoring. It is designed for interactive, remote, physiological data measurement, transmission, processing, storage and display of data. Additionally, the system allows optional verbal and video conferencing between a care giver and a patient through common telephone lines or the internet and optional hardware. #### Performance Data Verification and validation testing activities were conducted to establish the performance and reliability characteristics of the telemedicine system. Testing involved EMC/Electrical safety; unit testing; integration testing; data integrity testing; usability testing; environmental testing; compatibility testing; performance testing, system testing, and load/performance testing for the whole telemedicine system. In all instances, the VMS-01 functioned as intended. #### Substantial Equivalence The VMS-01 system is substantially equivalent to the Philips Medical Systems M3810A TeleMonitoring System with M3812B TeleStation (K023749) and the BL-Healthcare, Inc. Remote Care Management System (K051470). The VMS-on has the same intended uses and substantially similar indications, technological characteristics, and principles of operation as these predicate devices. The minor differences in indications for use and technology between the VMS-01 and the predicate devices raise no new issues of safety or effectiveness, as demonstrated by performance test data. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV - 8 2006 Medic4All Ltd. c/o Mr. Jonathan S. Kahan Partner Hogan & Hartson 555 Thirteenth St. NW Washington D.C. 20004 Re: K062127 Trade Name: Wireless Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: Class II Product Code: DRG Dated: October 24, 2006 Received: October 24, 2006 Dear Mr. Kahan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bram D/Zuckerman, M.D. Bram D /Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### INDICATIONS FOR USE STATEMENT 4 510(k) Number (if known): K002 \27 # Device Name: Telemedicine System Model VMS-01 Indications for Use: The Medic4All Telemedicine System Model VMS-01 is intended to collect and transmit physiological information, such as noninvasive blood pressure, via standard telephone lines or the internet to local or remote computer systems for use by the patient or a healthcare professional. Prescription Use AND/OR Over-The-Counter Use (21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of Division Sign-Off Division of Cardio 510(k) Number K062127
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...