← Product Code [FPL](/productcode/FPL) · K062052

# LMT NOMAG IC 3.0 (K062052)

_Lmt Lammers Medical Technology GmbH · FPL · Jul 31, 2006 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K062052

## Device Facts

- **Applicant:** Lmt Lammers Medical Technology GmbH
- **Product Code:** [FPL](/productcode/FPL.md)
- **Decision Date:** Jul 31, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5410
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric, 3rd-Party Reviewed

## Indications for Use

The LMT nomag IC 3.0 MR-incubator is an infant incubator system for temporary use during MR-imaging in clinical environments provided with a suitable trolley for intra hospital transport. The LMT nomag IC 3.0 MR-incubator provides a controlled environment of warmth and humidity for premature babies and sick infants up to a body weight of 4.5 kg (10 lbs) and a maximum body length of 55 cm (21.7 inches). The LMT nomag IC 3.0 is suitable for use with MR scanners of field strengths up to 3.0T.

## Device Story

Infant incubator system designed for use within MR imaging environments and intra-hospital transport. Provides controlled microclimate (air temperature and humidity) for preterm/term infants requiring warming therapy. Operates in air control mode using closed-loop feedback for temperature and humidity regulation; humidification via distilled water vaporization. Includes trolley for transport and power supply. Used in clinical settings by healthcare professionals. MR compatibility allows infant to remain in incubator during scanning; device does not interfere with scanner performance, and scanner does not affect incubator function. Benefits include continuous thermal support for vulnerable infants during diagnostic imaging.

## Clinical Evidence

Bench testing only. Compatibility with 1.5T and 3.0T MR scanners verified using phantom testing protocols. No clinical data provided.

## Technological Characteristics

Neonatal transport incubator; closed-loop air temperature and humidity control; humidification via distilled water vaporization. MR-compatible design for field strengths up to 3.0T. Compliant with IEC 60601-2-20, IEC 60601-1, and ISO 10993 standards.

## Regulatory Identification

A neonatal transport incubator is a device consisting of a portable rigid boxlike enclosure with insulated walls in which an infant may be kept in a controlled environment while being transported for medical care. The device may include straps to secure the infant, a battery-operated heater, an AC-powered battery charger, a fan to circulate the warmed air, a container for water to add humidity, and provision for a portable oxygen bottle.

## Predicate Devices

- LMT nomag IC 1.5 ([K033565](/device/K033565.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Sec. 5 510(k) Summary LMT nomag IC 3.0

Image /page/0/Picture/1 description: The image shows the text "K062952 (p. 1 of 3)" written in black ink at the top. Below this text is a logo consisting of five black dots followed by the letters "LMT" in a bold, sans-serif font. The letters are also black, and the overall image has a high contrast.

JUL 3 1 2006

Preparation Date: Feb, 17th 2006

Applicant LMT Lammers Medical Technology GmbH Osterweide 8 23568 Luebeck Germany

Establishment Registration No. 3004083324

| Contact Person: | Thomas Michael Bohnen<br>Tel: +49-451-58098-0<br>Fax: +49-451-58098-15<br>Email: bohnen@lammersmedical.com |
|-----------------|------------------------------------------------------------------------------------------------------------|
|-----------------|------------------------------------------------------------------------------------------------------------|

Common Device Name Neonatal Transport Incubator

Device Panel: General Hospital

Product Code FPL (Incubator, Neonatal Transport) Regulation No. 880.5410

Class

Propietary Name LMT nomag IC 3.0 REF: LMT-0100004-A

- Performance Standard: IEC 60601-2-20: 1996 IEC 60601-1 (Ed.2): 1995 ISO 10993: 1997
2

- Legally marketed device LMT nomag IC 1.5 K#033565
Reason for 510(k): extended indications for use

{1}------------------------------------------------

Sec. 5 510(k) Summary LMT nomag IC 3.0

Image /page/1/Picture/1 description: The image shows the text "K462452 (0.2043)" at the top. Below this text are four black dots in a horizontal row, followed by the letters "LMT" in a bold, sans-serif font. The letters are aligned horizontally and appear to be part of a logo or identifier.

## Intended Use of the LMT nomag IC 3.0

The LMT nomag IC 3.0 is an infant incubator system for temporary use during MR-imaging in clinical environments and provided with a suitable trolley for intra hospital transport.

The LMT nomag IC 3.0 MR-incubator provides a controlled environment of warmth and humidity for premature babies and sick infants up to a body weight of 4.5 kg (10 lbs) and a maximum body length of 55 cm (21.7 inches).

The LMT nomag IC 3.0 is suitable for use with MR scanners of field strengths up to 3.0T.

## Function of the Device

The LMT Nomag IC 3.0 is an incubator for preterm and term-born infants who are scheduled for Magnetic Resonance (MR) Imaging and are dependant on warming therapy. It provides a microclimate of air temperature and humidity suiting the infants needs. The incubator has a trolley for intra clinical transport providing power.

The incubator runs in air control mode and has closed loop control for both air temperature and humidity. Humidification works hygienically via vaporizing of distilled water.

#### Design and Specifications

The basic incubator performance of the LMT nomag IC 3.0 is defined by the IEC particular standard and the design changes over the nomag IC 1.5. do not affect these performance figures.

#### Testing under MR influence

The compatibility to MR scanners with 3T field strength has been proven in extensive bench testing.

lmage evaluation was performed to the same standards as with the nomag IC 1.5.

{2}------------------------------------------------

Sec. 5 510(k) Summary LMT nomaq IC 3.0

Image /page/2/Picture/1 description: The image shows the text "Kψ62φ52 (P.3ΩΔ3)" written in black ink at the top. Below this text is a row of three black dots followed by the letters "LMT" in a bold, sans-serif font. The letters are also black, and the overall image has a simple, clean appearance.

The compatibility with MR scanners of different manufacturers (Siemens, Philips, GE 1.5T and 3T) have been tested according to an established protocol using phantoms.

During this testing the behaviour of the incubator has shown no adverse effects in neither direction. The incubator performance was not influenced in any way by the scanner and the imaging has not shown significant artefacts caused by the incubator. Other functions of the scanner were also not influenced by the incubator.

## SE Statement

The LMT Nomag IC 3.0 incubator is substantially equivalent to the legally marketed device. It is compliant with international harmonized standards for incubators by the IEC.

The testing did not rise new risks or questions associated with the use of a MR compatible incubator.

The LMT Nomag IC 3.0 works as safe and effective as the legally marketed device.

V.N.BoC

Thomas Michael Bohnen RA LMT Lammers Medical Technology GmbH, Germany

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle.

JUL 3 1 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

LMT Lammers Medical Technology GmbH C/O Mr. Jeffrey D. Rongero Responsible Third Party Official Underwriters Laboratories, Incorporated 12 Laboratory Drive Research Triangle, North Carolina 27709

Re: K062052

Trade/Device Name: LMT Nomag IC 3.0 Regulation Number: 21 CFR 880.5410 Regulation Name: Neonatal Transport Incubator Regulatory Class: II Product Code: FPL Dated: June 19, 2006 Received: July 20, 2006

Dear Mr. Rongero:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affècting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

# Page 2 - Mr. Rongero

.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrlv/industrv/support/index.html.

Sincerely vours,

Chiu Lin, Ph.D.

Chiu Lin. Ph.D. Director Division of Anesthesiology. General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

______________________________________________________________________________________________________________________________________________________________________________ Device Name:

Indications for Use:

The LMT nomag IC 3.0 MR-incubator is an infant incubator system for temporary use during MR-imaging in clinical environments provided with a suitable trolley for intra hospital transport.

The LMT nomag IC 3.0 MR-incubator provides a controlled environment of warmth and humidity for premature babies and sick infants up to a body weight of 4.5 kg (10 lbs) and a maximum body length of 55 cm (21.7 inches).

The LMT nomag IC 3.0 is suitable for use with MR scanners of field strengths up to 3.0T.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

His C. Murphy & Elizabeth M. Petton (Widow of John)
1789

sthesiblogy, General Hospi

Page 1 of 1

umber Krizdz

---

**Source:** [https://fda.innolitics.com/device/K062052](https://fda.innolitics.com/device/K062052)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
