CELLSEARCH CIRCULATING TUMOR CELL KIT

K062013 · Veridex, LLC · NQI · Dec 14, 2006 · Immunology

Device Facts

Record IDK062013
Device NameCELLSEARCH CIRCULATING TUMOR CELL KIT
ApplicantVeridex, LLC
Product CodeNQI · Immunology
Decision DateDec 14, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.6020
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K062013 · Dec 14, 2006CELLSEARCH CIRCULATING TUMOR CELL KITVeridex, LLCPatient medical records; Radiological imaging studies; Clinical monitoring dataThe submission uses retrospective clinical data to compare CTC enumeration results with standard-of-care radiological imaging (WHO criteria) to support the clinical utility of the device as an aid in monitoring metastatic breast cancer patients.Metastatic breast cancer; Radiological imaging; Progression-free survival; Overall survival; Clinical monitoring

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective cohort analysis138 patients with measurable metastatic breast cancer; Sample Size: 138 patients (231 imaging studies)Not applicable for this studyConcordance between CTC counts (>=5 vs <5) and radiological imaging response (Progressive vs Non-Progressive Disease)

Indications for Use

The CellSearch™ Circulating Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood. The presence of CTC in the peripheral blood, as detected by the CellSearch™ Circulating Tumor Cell Kit, is associated with decreased progression free survival and decreased overall survival in patients treated for metastatic breast cancer. The test is to be used as an aid in the monitoring of patients with metastatic breast cancer. Serial testing for CTC should be used in conjunction with other clinical methods for monitoring breast cancer. A CTC count of 5 or more per 7.5 mL of blood at any time during the course of the disease is predictive of shorter progression free survival and overall survival.

Device Story

Device uses ferrofluid-based capture reagent with magnetic nanoparticles coated with anti-EpCAM antibodies to isolate circulating tumor cells (CTC) from 7.5 mL whole blood samples. Following immunomagnetic enrichment, samples are labeled with fluorescent reagents: anti-CK-PE (cytokeratin), DAPI (nucleus), and anti-CD45-APC (leukocytes). CellTracks AutoPrep System dispenses mixture into cartridge; MagNest device uses magnetic field to concentrate cells. CellTracks Analyzer II or CellSpotter Analyzer scans cartridge surface, acquires images, and displays events with co-located CK-PE and DAPI fluorescence. User performs final classification in gallery format; tumor cells defined as EpCAM+, CK+, DAPI+, and CD45-. Used in clinical laboratory settings by trained personnel. Output provides CTC count; count ≥5 per 7.5 mL indicates shorter progression-free and overall survival. Results used as adjunct to clinical monitoring and imaging to inform treatment decisions.

Clinical Evidence

Prospective multi-center clinical trial (N=177) of metastatic breast cancer patients starting new therapy. Primary endpoints: progression-free survival (PFS) and overall survival (OS). Patients with ≥5 CTCs/7.5 mL at baseline had significantly shorter median PFS (2.7 vs 7.0 months, p=0.0001) and OS (10.9 vs 21.9 months, p<0.0001). Results confirmed at follow-up intervals. Concordance with radiological imaging (WHO criteria) showed 78% overall agreement; CTCs identified as independent prognostic factor.

Technological Characteristics

Immunomagnetic enrichment system. Components: ferrofluid nanoparticles (magnetic core, polymeric coating, anti-EpCAM antibodies), immunofluorescent reagents (anti-CK-PE, DAPI, anti-CD45-APC). Hardware: CellTracks AutoPrep System, MagNest cell presentation device, CellTracks Analyzer II/CellSpotter Analyzer. Connectivity: DICOM-compliant imaging. Software: Automated image acquisition and cell identification algorithm.

Indications for Use

Indicated for patients with metastatic breast cancer to aid in monitoring disease status. Contraindications: None stated.

Regulatory Classification

Identification

An immunomagnetic circulating cancer cell selection and enumeration system is a device that consists of biological probes, fluorochromes, and other reagents; preservation and preparation devices; and a semiautomated analytical instrument to select and count circulating cancer cells in a prepared sample of whole blood. This device is intended for adjunctive use in monitoring or predicting cancer disease progression, response to therapy, and for the detection of recurrent disease.

Special Controls

The device is classified as Class II under regulation 21 CFR 866.6020 with special controls. The special control guidance document "Immunomagnetic Circulating Cancer Cell Selection and Enumeration System" is available at https://www.fda.gov/cdrh/oivd/guidance/1531.pdf

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Immunomagnetic Circulating Cancer Cell Selection and Enumeration System.” See § 866.1(e) for availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k062013 B. Purpose for Submission: To expand the intended use of the CellSearch Assay as an aid in the monitoring of patients with metastatic breast cancer in conjunction with other clinical methods C. Measurand: EpCam, Cytokeratins 8, 18 and/or 19, and CD45 D. Type of Test: A semi-automated qualitative immunomagnetic-capture, immunofluorescent detection image analysis test. E. Applicant: Veridex, LLC, A Johnson and Johnson Company F. Proprietary and Established Names: CellSearch™ Circulating Tumor Cell Kit (Epithelial) G. Regulatory Information: 1. Regulation section: 21 CFR 866.6020-Immunomagnetic Circulating Cancer Cell Selection and Enumeration System 2. Classification: Class II 3. Product code: NQI system, immunomagnetic, circulating cancer cell, enumeration 4. Panel: Immunology 82 H. Intended Use: 1. Intended use(s): The CellSearch™ Circulating Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood. 2. Indication(s) for use: The presence of CTC in the peripheral blood, as detected by the CellSearch™ Circulating Tumor Cell Kit, is associated with decreased progression free survival and decreased overall survival in patients treated for metastatic breast cancer. The test is to be used as an aid in the monitoring of patients with metastatic breast cancer. Serial testing for CTC should be used in conjunction with other clinical methods for monitoring breast cancer. A CTC count of 5 or more per 7.5 mL of blood at any time during the course of the disease is predictive of shorter progression free survival and overall survival. 3. Special conditions for use statement(s): For prescription use only. 4. Special instrument requirements: The CellTracks® AutoPrep system (k040077) and the CellTracks® Analyzer II. {1} (k050145 and k060110) The CellTracks® Analyzer II is a semi-automated fluorescence microscope intended to enumerate fluorescently labeled cells that are immunomagnetically selected and distributed over a viewing surface I. Device Description: The CellSearch™ Circulating Tumor Cell Kit consists of anti-EpCAM Ferrofluid (mouse monoclonal antibody (mAB) to EpCAM conjugated-magnetic nanoparticles in buffer with BSA and ProClin 300), staining reagent (phycoerythrin (PE)-conjugated mouse anti-cytokeratins mAB to and allophycocyanin (APC)-conjugated mouse anti-CD45 mAB in buffer with BSA and sodium azide), nucleic acid dye (4', 6-diamidino-2-phenylindole, dihydrochloride and ProClin 300), capture enhancing reagent, permeabilization reagent, cell fixative, dilution buffer, conical tubes and caps and cartridges and cartridge plugs. J. Substantial Equivalence Information: 1. Predicate device name(s): UroVysion Bladder cancer Recurrence Kit and Vitros Immunodiagnostic Products CA15-3 Assay 2. Predicate 510(k) number(s): k013785 (UroVysion) and k983690 (Vitros) 3. Comparison with predicate: | | CellSearch CTC Kit (k062013) | UroVysion Bladder Cancer Recurrence Kit (k013785) | | --- | --- | --- | | Intended Use: Cancer Monitoring | The test is to be used as an aid in the monitoring of patients with metastatic breast cancer. Serial testing for CTC should be used in conjunction with other clinical methods for monitoring breast cancer. A CTC count of 5 or more per 7.5 mL of blood at any time during the course of the disease is predictive of shorter progression free survival and overall survival. | Results from the UroVysion Kit are intended for use as a noninvasive method for monitoring for tumor recurrence in conjunction with cystoscopy in patients previously diagnosed with bladder cancer. | | Primary Cancer Site | Breast | Bladder | | Measurand | CTC positive for EpCam, Cytokeratins 8, 18 and/or 19 | Aneuploidy for chromosomes 3, 7, 17, and loss of the 9p21 locus | | Assay Type | Cellular | Cellular | | Technology | Immunomagnetic capture, enrichment and detection | Fluorescent in situ hybridization (FISH) | | Fluorescent reagents | Anti-CK-PE, DAPI, and | Chromosome Enumeration Probe | {2} | | CellSearch CTC Kit (k062013) | UroVysion Bladder Cancer Recurrence Kit (k013785) | | --- | --- | --- | | | anti-CD45-APC | (CEP®) 3 SpectrumRed™, CEP 7 SpectrumGreen™, CEP 17 SpectrumAqua™, and Locus Specific Identifier (LSI®) 9p21 SpectrumGold™. | | Recognition of positive cells | Visual recognition through fluorescent signal | Visual recognition through fluorescent signal | | Means of measurement | Fluorescent cell counting compared to cutoff value | Fluorescent cell counting compared to cutoff values | | Cutoff for Progression/Survival | Crossing of one threshold value at any time point | Crossing of one of two threshold values at one time point | | | CellSearch CTC Kit (k062013) | Vitros Immunodiagnostic Products CA 15-3 Assay (k983690) | | --- | --- | --- | | Intended Use: Cancer Monitoring | The test is to be used as an aid in the monitoring of patients with metastatic breast cancer. Serial testing for CTC should be used in conjunction with other clinical methods for monitoring breast cancer. A CTC count of 5 or more per 7.5 mL of blood at any time during the course of the disease is predictive of shorter progression free survival and overall survival. | Serial test results obtained with the VITROS® CA 15-3 assay, in patients who are clinically free of disease, should be used in conjunction with all relevant information derived from diagnostic test, physical examination and full medical history in accordance with appropriate patient management procedures used for early detection of recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring progression or response to treatment. | | Primary Cancer Site | Breast | Breast | | Sample Type | Whole blood | Serum, plasma (heparin or EDTA) | | Technology | Immunomagnetic capture, enrichment and detection | Solid-phase immunoassay | | Cutoff for Progression/Survival | Crossing of one threshold value at any time point | Significant difference between test results from two different time points | # K. Standard/Guidance Document Referenced (if applicable): The CellSearch™ Circulating Tumor Cell kit was developed in conformance to the following standards and guidances. ISO 14971 Medical Devices- Application of Risk Management to Medical Devices Guidance for Industry and FDA Staff Class II Special Controls Guidance Document: Circulating Cancer Cell Selection and Enumeration System (May 11, {3} 2004) Guidance for Industry and FDA Staff: Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use (November 30, 2004) "Guidance Document for the Submission of Tumor Associated Antigens Premarket Notifications, [510(k)], to FDA to Guide Manufacturers L. Test Principle: Epithelial cells are immunomagnetically labeled by targeting the Epithelial Cell Adhesion Molecule (EpCAM) antigen. Anti-EpCAM monoclonal antibodies conjugated to ferrofluid particles are colloidal and, when mixed with a sample containing the target epithelial cells, bind to the EpCAM antigen associated with the epithelial cells. After immunomagnetic selection of epithelial cells from 7.5 mL of blood, fluorescent reagents are added at this time to discriminate between the immunomagnetically selected cells. Anti-Cytokeratin – Phycoerythrin (CK-PE) stains the intracellular cytoskeleton cytokeratin proteins expressed in cells of epithelial origin, anti-CD45-Allophycocyan (CD45-APC) stains leukocytes and DAPI stains DNA present in the cell nucleus. The processed reagent/sample mixture is dispensed by the CellTracks® AutoPrep System into a cartridge that is inserted into a MagNest® cell presentation device. The strong magnetic field of the MagNest® device causes the magnetically-labeled target cells to move to the surface of the cartridge. The cartridge is then placed on the CellTracks® Analyzer II for data acquisition and analysis. The CellTracks Analyzer II scans the entire surface of the cartridge with a series of fluorescence filters that are defined for a given assay and acquires images of PE, APC and DAPI fluorescence staining of the entire viewing surface. After data acquisition is completed, the images are analyzed for any event where cytokeratin-PE and DAPI are within a specified space in the cartridge, i.e. indicating the possible presence of a cell with a nucleus that expresses cytokeratin. Images from each fluorescent color as well as a composite image of the cytokeratin staining (green) and the nuclear staining (purple) are presented to the user in a gallery for final cell classification. A cell is classified as a tumor cell when it its EpCAM+ (i.e., it is captured), CK+, DAPI+ and CD45-. A check mark placed by the operator next to the composite images classifies the event as a Circulating Tumor Cell (CTC) and the software tallies all the checked boxes to obtain the CTC count. M. Performance Characteristics (if/when applicable): 1. Analytical performance was presented originally in K031588: a. Precision/Reproducibility: i. System Reproducibility with CellSearch™ Circulating Tumor Cell Control Three separate CellSearch™ Circulating Tumor Cell Control samples were prepared and processed each day for over 30 days, per the long run method of NCCLS guideline EP5-A². Each single-use sample bottle contains a low and a 4 {4} high concentration of cells from a fixed cell line that have been pre-stained with two different fluorochromes. Summary statistics for the high and low control cells is presented below. Table 1. Summary of Precision Analyses | | Low | High | | --- | --- | --- | | N | 99 | 99 | | Mean cell count | 48 | 969 | | Total Precision Standard Deviation (ST) % CV | 18% | 5% | # ii. System Reproducibility with Patient Samples A total of 163 duplicate samples were collected from 47 patients over the course of the clinical study. These samples were processed at multiple sites to determine the reproducibility of CTC measurements. The regression equation for the comparison of these 163 duplicate samples was $y = 0.98x + 0.67$ , $R^2 = 0.99$ . Table 2 shows the summary of the data for replicates where the average of the two CTC results was $&lt; 5$ compared to those where the average (avg.) was $\geq 5$ . Table 2. Reproducibility of CTC Counts in Duplicate Samples (n=163) with an Average of $&lt; 5$ or $\geq 5$ CTC per $7.5\mathrm{mL}$ of blood. | | CTC <5 | CTC ≥5 | | --- | --- | --- | | Number of Duplicates | 123 | 40 | | Mean CTC Count of Duplicates | 0.7 | 210 | | Avg. Duplicate Standard Deviation | 0.5 | 12 | | Avg. %CV of Duplicates | 60% | 20% | # b. Accuracy/Recovery: Blood samples from a single healthy donor were pooled and five of six 7.5 mL aliquots were spiked with 5, 20, 81, 325 and 1300 cultured breast cancer cells (SK-Br-3). The sixth tube was unspiked pooled blood and served as a zero point. These samples were processed on the CellTracks® AutoPrep System with the CellSearch™ Circulating Tumor Cell Kit and CTC counts were determined on the CellTracks® Analyzer II. The experiment was repeated for four additional donors. The observed cell counts were plotted against the results of the expected cell count. The results are summarized in Table 3. Table 3. Percent Detection Estimates. | Expected Tumor Cell Count | Mean Observed Tumor Cell Count | Range of Percent Recovery | | --- | --- | --- | | 1300 | 1215 | 91 to 95% | | 325 | 308 | 82 to 101% | | 81 | 85 | 80 to 136% | | 20 | 22 | 95 to 140% | | 5 | 7 | 120 to 200% | {5} To determine the overall, or least squares fit, for the comparison of the observed and expected cell counts across all the data, linear regression analysis was performed. The regression equation for these 30 samples was $y = 0.93x + 3.87$, $R^2 = 0.999$. The results of this study indicate that on average over the tested CTC range the recovery, as derived from regression analysis, is 93%. Given the linear response of the tumor cell counts, one would expect the slope of the observed versus expected plot to be 1.0. However, the slope was 0.93. This is because the CellTracks® AutoPrep System with CellSearch™ CTC Kit involves the capture and fluorescent labeling of cells followed by their detection and enumeration by the CellTracks® Analyzer II. The loss of cells could therefore be attributed to one of the following possibilities; 1) the recovery of only 93% of the tumor cells spiked into 7.5 mL of blood by the CellTracks® AutoPrep System, 2) the detection of only 93% of the tumor cells present in the sample chamber by the CellTracks® Analyzer II or 3) a combination of both of these sources of error. **Linearity/assay reportable range:** Another way to examine the previous data is to analyze it as a dilution series to evaluate test linearity. The confounding variable of percent recovery was removed by using the observed value of the original sample divided by the dilution factors to determine the expected values for the dilution series for each patient sample. Regression of all of these numbers of observed tumor cells versus the numbers of expected tumor cells yielded a slope of 1.007, an intercept of 3.0, an $r^2 = 0.99$ and $r = 0.995$. Therefore, once the percent recovery (cell loss) was factored out of the CTC values of each of the original samples, this analysis of the data demonstrated that the detection of CTC was linear over the reportable range of 0 to 1238 tumor cells. c. Traceability, Stability, Expected values (controls, calibrators, or methods): No recognized reference material or method. d. Detection limit: One CTC per 7.5 mL can be detected by the CellTracks® Analyzer II resulting in a limit of detection of 1 CTC in a cartridge. Linear regression shows that on average, 93% of CTC present in a 7.5 mL blood sample are recovered using the CellTracks® AutoPrep System (see Recovery section). The loss of approximately 7% of the CTC in the sample is not sufficient to reduce the limit of detection of 1 CTC. e. Analytical specificity: **Interfering Substances:** SK-BR-3 cells spiked into blood samples were exposed to potential interfering substances and compared to untreated controls. Toxic levels (5 times therapeutic index) of the following cancer drugs, over-the-counter drugs, and other exogenous substances were tested: cyclophosphamide, Mitomycin C®, Procrit®, biotin, 5-fluorouracil, methotrexate, tamoxifen citrate, paclitaxel, 6 {6} Arimidex®, acetaminophen, acetylsalicylic acid, caffeine, dextromethorphan, Aredia®, Human Anti-Mouse Antibody (HAMA) type 1, HAMA type 2, Herceptin®, and ibuprofen. No significant differences in SK-BR-3 cell numbers were detected, indicating that these substances do not interfere with the CellSearch™ kit. Samples spiked with toxic levels of doxorubicin resulted in aberrant staining of leukocytes as cytokeratin and CD45 dual positive cells, due to the doxorubicin being a fluorescent compound that is incorporated into nucleated cells. If seen, the staining pattern of all cells being CD45 positive and cytokeratin positive is obvious and easily identified by the operator as a known interference staining profile. If blood is drawn after the recommended 7-day washout period, following doxorubicin infusion, this interference is unlikely to be observed in clinical practice given controlled therapeutic levels and rapid drug clearance. Potential interference from lipemia was studied by adding Intralipid to samples to a concentration of 2.6%, which corresponds to greater than 1000 mg/dL triglyceride. Samples were lysed to simulate total hemolysis. Bilirubin at 7.4 mg/dL, HAMA 1/HAMA 2 and hematocrit from 18-60% were studied. Lipemia, hemolysis, icterus and a broad range of hematocrit values do not interfere with the CellSearch™ test. HAMA 1 and HAMA 2 also do not interfere, indicating that individuals receiving mouse Ig by parenteral routes can be tested successfully with the CellSearch™ test. f. Assay cut-off: Results are reported as the number of CTC/7.5 mL of blood. A CTC count of 5 or more per 7.5 mL of blood is predictive of shorter progression free survival and overall survival. This cut-off was established in the 510(k) of the previous version of this assay (k031588). 2. Comparison studies: a. Method comparison with predicate device: Direct comparison to the UroVysion Bladder cancer Recurrence Kit is not feasible since it is different measurands in a different intended population. As for the Vitros CA15-3 assay, the measurand is a serum tumor marker and not for circulating tumor cells. b. Matrix comparison: Since there is only one matrix for this test, i.e. whole blood, no matrix comparison studies were performed. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): USE OF CTC TO MONITOR CLINICAL STATUS {7} 8 # Relationship between survival, CTCs, and disease assessment by imaging Radiological imaging is one of the primary means of determining disease status and response to therapy in metastatic breast cancer patients. To establish the relationship of CTCs, measured at two different time points, to clinical status as determined by imaging, CTCs and imaging results were compared 1) to the true clinical endpoint overall survival and 2) to each other. # CTC Previous data has shown that patients with $\geq 5$ CTCs / $7.5\mathrm{mL}$ of blood at any succeeding follow-up visit after the initiation of therapy had a higher likelihood of progressive disease and decreased overall survival compared to patients with $&lt; 5$ CTCs / $7.5\mathrm{mL}$ of blood. # Imaging All imaging sites were in compliance with Digital Imaging and Communications in Medicine (DICOM) standards. Using standardized digital images, two expert radiologists (readers), working individually and blinded to clinical information, classified each follow-up disease assessment (total of 231 imaging studies) from 138 patients with measurable disease as indeterminate (I), stable disease (S), partial response (PR), or progressive disease (PD) according to World Health Organization (WHO) bi-dimensional criteria. Measurable disease was defined as the presence of at least one lesion $\geq 2\mathrm{cm}$ in its longest dimension. Readers identified up to eight lesions per patient per time point by describing the longest dimension of the lesion and the longest perpendicular dimension. These two dimensions were multiplied and the "cross product" was reported. Summed measurements for the cross-products were calculated, and percent change from the previous time point was determined. Although all patients had measurable disease, non-measurable lesions (still detectable by radiology) were included in the determination of patient status as described in the WHO guidelines. Progressive disease was defined as a $&gt;25\%$ increase in the sum of all lesions or appearance of a new measurable or non-measurable lesion. Partial response was defined as a decrease in the sum of all lesions of $\geq 50\%$ and no new lesions. - Radiology interpretations from the two expert radiologists were classified as followed: - S and PR were considered to both reflect non-progressive disease (NPD) - PD was considered to reflect progressive disease - In situations where one of the radiologists rendered a classification of Indeterminate (I) but the other radiologist rendered a classification of S, PR or PD, the classification of the latter radiologist was used for comparison to CTCs (n=11) - When both radiologists rendered a classification of Indeterminate (I), then the data was not used in the comparison to CTCs (n=3) - A third independent radiologist adjudicated disagreements between the two primary readers regarding PD and NPD (n=27) {8} - In situations where the third independent radiologist rendered a classification of Indeterminate (I), the data was not used in the comparison to CTCs $(n = 2)$ - In serial imaging studies, radiology results that were less than one month from a previous tabulated observation were not used $(n = 1)$ . - The CTC results obtained within $\pm$ one month of the imaging study were classified as $&lt; 5$ CTC and $\geq 5$ CTC. If more than one CTC value was obtained within $\pm$ one month of the imaging study, the CTC result obtained closest to the date of the imaging study was used. # Concordances between CTC and Radiological Monitoring Imaging studies are a major component of the current standard of care for determining disease progression and response to treatment in the metastatic breast cancer setting. To support the effectiveness of CTCs in making these clinical assessments, two-by-two tabulations of concordant and discordant observations between CTCs and radiological imaging were constructed using the previously described criteria. Using only the $1^{\text{st}}$ follow-up imaging study, the radiological response at this visit was compared with the CTC results obtained within $\pm$ one month of this imaging study. A total of 134 of the 138 patients $(97\%)$ had CTC results that met this criterion. The result of this "patient-wise" comparison between CTCs and imaging is shown in Table 4. Table 4. Patient-Wise Comparison of CTC and Imaging | Response at 1stFollow-Up Imaging Study | CTCs within ± 1 Month of Imaging <5 CTCs/7.5mL | ≥5 CTCs/7.5mL | Total | | --- | --- | --- | --- | | Non-Progressive Disease | 85 | 9 | 94 | | Progressive Disease | 20 | 20 | 40 | | Total | 105 | 29 | 134 | | Measurement | Estimate | Lower 95% CI | Upper 95% CI | | --- | --- | --- | --- | | Positive % Agreement | 50% | 34% | 66% | | Negative % Agreement | 90% | 83% | 96% | | Positive Predictive Value | 69% | 49% | 85% | | Negative Predictive Value | 81% | 72% | 88% | | Overall Agreement | 78% | 70% | 85% | | Odds Ratio | 9.4 | 3.4 | 26.8 | Using all of the follow-up imaging studies performed after the initiation of therapy on the 138 patients that rendered useable radiological response results $(n = 225)$ , these results were then compared to CTC results obtained within $\pm$ one month of the imaging study. A total of 219 of the 225 $(97\%)$ imaging studies had CTC results meeting this criterion. The result of this "observation-wise" comparison between CTCs and imaging is shown in Table 5. {9} Table 5. Observation-Wise Comparison of CTC and Imaging | Response at All Follow-Up Imaging Studies | CTCs within ± 1 Month of Imaging <5 CTCs/7.5mL | ≥5 CTCs/7.5mL | Total | | --- | --- | --- | --- | | Non-Progressive Disease | 151 | 16 | 167 | | Progressive Disease | 30 | 22 | 52 | | Total | 181 | 38 | 219 | | Measurement | Estimate | Lower 95% CI | Upper 95% CI | | --- | --- | --- | --- | | Positive % Agreement | 42% | 29% | 57% | | Negative % Agreement | 90% | 85% | 94% | | Positive Predictive Value | 58% | 41% | 74% | | Negative Predictive Value | 83% | 77% | 89% | | Overall Agreement | 79% | 73% | 84% | | Odds Ratio | 6.9 | 3.0 | 15.8 | In serial observations, only a minority of the transitions for imaging results between non-progressive disease and progressive disease coincided with a matching transition of CTC counts between $&lt; 5$ and $\geq 5$ CTCs/7.5 mL. Because the prognostic value of the CTC results at an earlier time-point were equivalent to that of the CTC results at the time of imaging, a patient-wise comparison using results from only the $1^{\text{st}}$ follow-up imaging study, performed approximately 9 weeks after the initiation of therapy, and the CTC results obtained at the $1^{\text{st}}$ follow-up, performed approximately 4 weeks after initiation of therapy, was constructed. All 138 patients had CTC results meeting this criterion. The result of this "patient-wise" comparison between CTCs at an earlier time point and imaging is shown in Table 6. Table 6. Patient-Wise Comparison of CTC and Imaging | Response at 1stFollow-Up Imaging Study | CTCs at 1stFollow-Up <5 CTCs/7.5mL | ≥5 CTCs/7.5mL | Total | | --- | --- | --- | --- | | Non-Progressive Disease | 84 | 12 | 96 | | Progressive Disease | 20 | 22 | 42 | | Total | 104 | 34 | 138 | | Measurement | Estimate | Lower 95% CI | Upper 95% CI | | --- | --- | --- | --- | | Positive % Agreement | 52% | 36% | 68% | | Negative % Agreement | 88% | 79% | 93% | | Positive Predictive Value | 65% | 46% | 80% | | Negative Predictive Value | 81% | 72% | 88% | | Overall Agreement | 77% | 69% | 84% | | Odds Ratio | 7.7 | 3.0 | 19.9 | {10} 11 4. Clinical cut-off: Results are reported as the number of CTC / 7.5 mL of blood. A CTC count of 5 or more per 7.5 mL of blood is predictive of shorter progression free survival and overall survival. This cut-off was established in the 510(k) of the previous version of this assay (k031588). 5. Expected values/Reference range: The EXPECTED VALUES were determined with a previous version of this test, as presented in k031588. The present version of the test was demonstrated to be substantially equivalent to that in k031588 in k050245. Healthy volunteers, non-malignant breast disease, non-malignant other disease Single point CTC analyses were performed on control groups of 145 healthy volunteers, 101 women with non-malignant breast disease, and 99 women with other non-malignant diseases. Epithelial cells are not expected to be present in the peripheral blood of healthy individuals. Of the 345 total samples from healthy volunteers and women with non-malignant disease, only one subject had more than 5 CTC/7.5 mL. The results are presented in Table 7. Table 7. Control Subjects | Category | N | Mean # CTC | SD | # Patients with ≥ 5 CTC | Min.* | Max.* | | --- | --- | --- | --- | --- | --- | --- | | Healthy | 145 | 0.1 | 0.2 | 0 | 0 | 1 | | Non-malignant breast disease | 101 | 0.2 | 1.2 | 1 | 0 | 12 | | Non-malignant other disease | 99 | 0.1 | 0.4 | 0 | 0 | 3 | * NCCLS Guideline C28-A2³ N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...