← Product Code [DSH](/productcode/DSH) · K062007

# DMS 300-7 (K062007)

_Diagnostic Monitoring Software · DSH · Aug 16, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K062007

## Device Facts

- **Applicant:** Diagnostic Monitoring Software
- **Product Code:** [DSH](/productcode/DSH.md)
- **Decision Date:** Aug 16, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2800
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The DMS 300-7 Holter ECG recorder is intended to acquire, record, and store up to 24-hours of three (3) channel ECG data for patients undergoing continuous ambulatory electrocardiography. The DMS 300-7 performs no cardiac analysis by itself and is intended to be used with the DMS Premier Holter system. In turn, the cardiac data and analysis provided by the Premier Holter system will be used by trained medical personnel in the diagnosis of patients with various rhythm patterns. The DMS 300-7 is indicated for use in a clinical setting, by qualified medical professionals only, for recording 3-channel ECG data of patients requiring ambulatory ECG monitoring for up to 24-hours. Holter ECG is appropriate for the indications below: Evaluation of adult patients with symptoms suggesting arrhythmia. Evaluation of adult patients with pacemakers. Evaluation of patients with suspected abnormal heart rate changes. Evaluation of patients with suspected abnormal QRS interval changes with drug therapy treatment.

## Device Story

DMS 300-7 is a 3-channel ambulatory Holter ECG recorder; captures continuous ECG data over 24 hours via 5 or 7-electrode cables; stores data on removable non-volatile compact flash card. Patient-activated event button flags symptomatic cardiac events. Device is used in clinical settings by qualified medical professionals. Data is downloaded via PCMCIA reader to DMS Premier Holter system for analysis. Device performs no onboard cardiac analysis; provides raw digitized ECG data to external system. Clinicians use Premier Holter system output to diagnose rhythm patterns, evaluate pacemakers, and monitor drug therapy response. Benefits include continuous cardiac monitoring for ambulatory patients.

## Clinical Evidence

Bench testing only. Comparison of final test results between the predicate 300 ECG Recorder and the modified DMS 300-7 using identical input signals demonstrated equivalent data processing performance.

## Technological Characteristics

3-channel ECG recorder; 8-bit resolution; 0.05 to 100 Hz bandwidth; >60 dB common mode rejection. Power: single AA alkaline battery. Storage: removable non-volatile compact flash card. Connectivity: PCMCIA reader interface to PC. Dimensions: 4.94 x 2.75 x 0.94 inches. Weight: 4 oz (w/o battery).

## Regulatory Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

## Predicate Devices

- Diagnostic Monitoring Software (Scole) 300 ECG Recorder ([K923664](/device/K923664.md))
- Datrix DR512 Holter Recorder ([K982975](/device/K982975.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### Special 510(k) Notification: Device Summary

July 14, 2006

#### Submitter:

William Parsons, Official Correspondent Diagnostic Monitoring Software 292 Kingsbury, 32B P.O. Box 3109 Stateline, NV 89449 Fax: 714 953 0260 Tel: 714 953 2653 e-mail: holterdms@hotmail.com Contact: William Parsons

AUG 16 2006

Trade Name: DMS 300-7 Common Name: Holter ECG Recorder Requlation Number: 21 CFR 870.2800 Requlation Name: Medical Magnetic Tape Recorder Regulatory Class: Il (two) Product Code: MAAS 1854 Establishment Registration Number: 2028190

#### Legally marketed device to which S.E. is claimed:

The DMS 300-7 Holter ECG recorder is an addition to its predecessor device, Model 300 ECG Recorder, which includes a reduction in size and a battery change from 9V to AA battery power source. Its intended use, indications for use, functions, and recording output ECG data are the same as the predicate device, Model 300 ECG Recorder. The Final Test results of the predicate 300 ECG Recorder (K923664) and the modified submission device DMS 300-7, show that testing of the device with the same input signals produces the same results in the processing of the data with the same Holter playback device. This modified DMS 300-7 Holter recorder is substantially equivalent (SE) to the following legally marketed predicate devices that were cleared 510(k)s under 21 CFR 870.2800, Class II:

- Diagnostic Monitoring Software (Scole) 300 ECG Recorder (K923664). .
- Datrix DR512 Holter Recorder (K982975) ●

#### Description

The modified DMS 300-7 Holter ECG recorder is intended for use as a part of a Holter analysis system and is designed to be used with the DMS Premier Holter system. The DMS 300-7 provides three (3) channels of continuous ECG recorded over a 24-hour time period. The DMS 300-7 acquires, digitizes, and stores data to be later analyzed by the Premier Holter system. The Premier Holter system processes pre-recorded patient 3-channel ECG data that has been stored in the DMS 300-7. The cardiac data provided by the DMS 300-7 recorder and the Premier 

{1}------------------------------------------------

Holter system is used by trained medical personnel to assist in the diagnosis of patients with various ECG rhythm patterns.

The DMS 300-7 is a 3-channel ECG recorder that uses either a five (5) or seven (7) electrode ECG cable, as is the same with the predicate Model 300 ECG Recorder. The five (5) electrode ECG cable records a modified V1, V3, Lead II ECG recording. The seven (7) electrode ECG cable records three (3) bipolar leads, and can be used for an XYZ recording. A patient activated ECG Event button can be pressed by the patient to flag a symptomatic cardiac event ECG, as is the same with the predicate Model 300 ECG Recorder.

The DMS 300-7 uses a single AA alkaline battery as its power source, and stores acquired ECG data on a removable, non-volatile memory compact flash card. Stored ECG data is then downloaded to the hard disk drive of the Premier Holter system via a standard PCMCIA reader interface. The recorded data will remain in memory until it has been cleared by the clinician.

## Comparison to the Sponsor's Predicate Devices

| Specifications        | DMS 300-7             | 300 ECG Recorder      | DR512                 |
|-----------------------|-----------------------|-----------------------|-----------------------|
| Predicate Device      | No                    | Yes                   | Yes                   |
| Owner                 | DMS                   | DMS                   | Datrix                |
| 510(k) number         |                       | K923664               | K982975               |
| ECG Channels          | 3-channel ECG         | 3-channel ECG         | 3-channel ECG         |
| Resolution            | 8-Bit                 | 8-Bit                 | 8-Bit                 |
| Recording Duration    | 24-hours              | 24-hours              | 24-hours              |
| Bandwidth             | 0.05 to 100 Hz        | 0.05 to 100 Hz        | 0.05 to 50 Hz         |
| Common Mode Rej       | > 60 db               | > 60 db               | > 60 db               |
| Power Source          | AA alkaline           | 9V                    | 9V or AA alkaline     |
| Average Current Drain | 5 mA                  | 9 mA                  | 5 mA                  |
| Event ECG Button      | Yes                   | Yes                   | Yes                   |
| Operating Temp        | 0 to 60 C             | 0 to 50 C             | 0 to 60 C             |
| 1-MV Input =          | 1-CM square wave      | 1-CM square wave      | 1-CM square wave      |
| Dimensions            | 4.94x2.75x0.94 in.    | 5.20x2.90x1.16 in.    | 4.94x2.75x0.94 in.    |
| Weight                | 4 oz. w/o battery     | 8 oz. w/o battery     | 4 oz. w/o battery     |
| Processing System     | Premier Holter system | Premier Holter system | Premier Holter system |
| ECG Data Transfer     | PC's hard disk        | PC's hard disk        | PC's hard disk        |

The DMS 300-7 is substantially equivalent (SE) to the predicate below devices.

{2}------------------------------------------------

### Intended Uses and Indications for Use of the Modified Device:

The "Intended Uses" of the modified DMS 300-7 Holter recorder are exactly the same as the predicate devices (300 ECG Recorder and DR512). Its intended use is to acquire, record, and store up to 24-hours of three (3) channel ECG data for patients undergoing continuous ambulatory electrocardiography. The DMS 300-7 performs no cardiac analysis by itself and is intended to be used with the DMS Premier Holter system. The predicate devices (300 ECG Recorder and DR512) also are processed by the DMS Premier Holter system.

The "Indications for Use" of the modified DMS 300-7 Holter recorder is indicated for use in a clinical setting, by qualified medical professionals only, for recording 3channel ECG data of patients during an ambulatory 24-hour time period. It is not a life-supporting system, and is not connected to an AC power source. Ambulatory 24-hour electrocardiography is used for the below indications:

- Evaluation of patients with symptoms suggesting arrhythmia. .
- Evaluation of patients with pacemakers. .
- Evaluation of patient heart rate changes. .
- Evaluation of patient QRS interval changes. .
- Evaluation of patient response to drug therapy treatment. .

The "Indications for Use" for the modified DMS 300-7 are exactly the same as the predicate devices (300 ECG Recorder and DR512).

{3}------------------------------------------------

Public Health Service

Image /page/3/Picture/11 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is depicted in black, and the text is also in black against a white background. The seal is simple and iconic, representing the department's role in promoting health and well-being in the United States.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 6 2006

Diagnostic Monitoring Software c/o William Parsons 292 Kingsbury, 32B P.O. Box 3109 Stateline, NV 89449

Re: K062007

Trade Name: DMS 300-7 Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II (two) Product Code: DSH Dated: August 8, 2006 Received: August 8, 2006

Dear Mr. Parsons:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. William Parsons

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bram D. Zuckerman, M.D.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): K062007

DMS Holter Recorder: Model DMS 300-7 Device Name:

Indications For Use:

The DMS 300-7 Holter ECG recorder is intended to acquire, record, and store up to Ine Dro 900 - Horee- (3) channel ECG data for patients undergoing continuous 24-nours of thece (e) chambly. The DMS 300-7 performs no cardiac analysis by amoulatory exceeded to be used with the DMS Premier Holter system. In turn, the cardiac data and analysis provided by the Premier Holter system will be che cardiac and and and and by trained medical personnel in the diagnosis of patients with various rhythm patterns.

The DMS 300-7 is indicated for use in a clinical setting, by qualified medical Ine DMD 500 7 15 Indroaded Seding ECG data of patients requiring ambulatory professionals only) 2011 / 2019 to 24-hours. Holter ECG is appropriate for the indications below:

.Evaluation of adult patients with symptons suggesting arrhythmia .Evaluation of adult patients with pacemakers .Evaluation of patients with suspected abnormal heart rate changes .Evaluation of patients with suspected abnormal QRS interval changes with drug therapy treatment Evaluation of patien Over-The-Counter Use Prescription Use (21 CFR 801 Subpart C)

(Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

vision Sign-Off)
vision of Cardiovascu
K0(k) Number K062007

Page 1 of I

---

**Source:** [https://fda.innolitics.com/device/K062007](https://fda.innolitics.com/device/K062007)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
